C12
Measurement accuracy and performance testing
Bench and, where required, clinical validation of the specific measurement claim against the particular standard for that device class.
If it makes a clinical blood-pressure accuracy claim, this may require measurement accuracy and performance testing.
Blood pressure accuracy is proven by a defined clinical investigation protocol, which is the single fact most worth knowing early.
- ISO 81060-2Not harmonised under the MDR
The clinical investigation protocol for automated non-invasive sphygmomanometers.
It is the clinical accuracy protocol, distinct from the equipment standard IEC 80601-2-30. Buyers confuse the two often.
Why? S059
Consolidated MDR harmonised-standards list, Implementing Decision (EU) 2021/1182 as amended, CELEX 02021D1182-20260617: no 81060 entry appears (absence re-verified 2026-08-17)
- IEC 80601-2-30Not harmonised under the MDR
The equipment requirements for automated non-invasive blood pressure measurement.
Why? S058
Consolidated MDR harmonised-standards list, Implementing Decision (EU) 2021/1182 as amended, CELEX 02021D1182-20260617: no 80601 entry appears (absence re-verified 2026-08-17)
Coverage in our register is thin: 2 laboratories reference this (counted 2026-08-18). Search the register
A laboratory can hold capability that its published scope text does not itemize. Absence from these counts is not evidence of absence.