VaLogic Bio
21 Byte Court, 21702, Frederick, MD, United States
Free. We check which accredited labs can do it, starting with this one, and reply within two working days with who can do it, how fast, and roughly what it costs.
Only the laboratory you pick gets your contact details. We ask the others without them.
Contact this laboratory
Work at VaLogic Bio?
Claim this profile to keep it accurate.
Register record
- A2LARefreshed
- 7466.02Active
- 7466.01Active
Accredited standards
We found no published test standard numbers in this lab’s scope (labs often list in-house methods instead). The full scope as published is below.
Scope as published by A2LA
6 lines
ISO/IEC 17025
Active7466.02
The scope document we hold says it is valid until 2027-06-30.
Accredited scope, certificate 7466.02
| Line | Discipline | Method | Matrix | Range |
|---|---|---|---|---|
| 1 | mechanical; thermodynamic | - | - | -196 to 3100 psig; RPM; °C; %RH |
| 2 | - | (A2LA Cert. No. 7466.01) Revised 3/3/2026 Page 2 of 2
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards of nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the uncertainty introduced by the item being calibrated, (e.g. resolution) must also be
considered and this, on its own, could result in the actual measurement uncertainty achievable on a
customer’s site being larger than the CMC.
4 This scope meets A2LA’s P112 Flexible Scope Policy.
5 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter
6 The CMC is dependent upon the resolution of the unit under test. The CMC presented does not include
the resolution of the unit under test. The resolution will be included in the reported measurement
uncertainty at the time of calibration. | ||
| 3 | - | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
ISO/IEC 17025
Active7466.01
The scope document we hold says it is valid until 2027-06-30.
Accredited scope, certificate 7466.01
| Line | Discipline | Method | Matrix | Range |
|---|---|---|---|---|
| 1 | mechanical; thermodynamic | - | - | -196 to 3100 psig; RPM; °C; %RH |
| 2 | - | (A2LA Cert. No. 7466.01) Revised 3/3/2026 Page 2 of 2
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards of nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the uncertainty introduced by the item being calibrated, (e.g. resolution) must also be
considered and this, on its own, could result in the actual measurement uncertainty achievable on a
customer’s site being larger than the CMC.
4 This scope meets A2LA’s P112 Flexible Scope Policy.
5 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter
6 The CMC is dependent upon the resolution of the unit under test. The CMC presented does not include
the resolution of the unit under test. The resolution will be included in the reported measurement
uncertainty at the time of calibration. | ||
| 3 | - | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.