VaLogic Bio
21 Byte Court, 21702, Frederick, MD, United States
Visit website- Scope lines
- 5
- A2LA certificates
- 2
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 5 scope lines.
A2LA certificate 7466.01 · 3 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7466.01) Revised 3/3/2026 Page 1 of 2
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
VALOGIC BIO LLC
21 Byte Court
Frederick, MD 21702
Ajay Bhale Phone: 240 529 1673
CALIBRATION
Valid To: June 30, 2027 Certificate Number: 7466.01
In recognition of the successful completion of the A2LA evaluation process (including an assessment of
the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory to perform the following calibrations
1, 5:
I. Mechanical
Parameter/Equipment
Range
CMC2, 5, 6 (±)
Comments
Pressure Measuring
Equipment3
Up to 100 psig
(100 to 150) psig
(150 to 300) psig
0.49 psig
0.70 psig
0.60 psig
Comparison to
Druck DPI 611-13G
RPM3
(2.5 to 600) RPM
(600 to 1720) RPM
(1720 to 3100) RPM
2.5 RPM
4.9 RPM
5.2 RPM
Comparison to
Omega HHT13
II. Thermodynamic
Parameter/Equipment
Range
CMC2, 5, 6 (±)
Comments
Temperature –
Measuring Equipment
(-196 to 200.0) °C
0.38 °C
Comparison to Fluke
1523 w/ 5628
Relative Humidity –
Measure3
(20 – 85) %RH
2.8 % RH
Comparison to
Vaisala HMI41 w/
HMP46
__________________________________________
1 This laboratory offers commercial calibration service and field calibration service. | ||
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7466.01) Revised 3/3/2026 Page 2 of 2
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards of nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the uncertainty introduced by the item being calibrated, (e.g. resolution) must also be
considered and this, on its own, could result in the actual measurement uncertainty achievable on a
customer’s site being larger than the CMC.
4 This scope meets A2LA’s P112 Flexible Scope Policy.
5 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter
6 The CMC is dependent upon the resolution of the unit under test. The CMC presented does not include
the resolution of the unit under test. The resolution will be included in the reported measurement
uncertainty at the time of calibration. | ||
| 3 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
A2LA certificate 7466.02 · 2 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | USP–1116; USP-1231; ISO-14644; USP–1116; USP–643; USP-645 | water; gas; air; Settle plates/surface plates; Compressed Gas | - |
| 2 | ISO/IEC 17025:2017 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.