What this lab is accredited for
We found no published test standard numbers in this lab’s scope. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 3 scope lines.
A2LA certificate 3589.01 · 3 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3589.01) Revised 11/25/2024 Page 1 of 2
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
XSENSOR TECHNOLOGY CORP
133 12 Avenue SE
Calgary, Alberta T2G 0Z9
CANADA
Tim Gorjanc Phone: 403 266 6612 ext. 248
CALIBRATION
Valid To: November 30, 2026 Certificate Number: 3589.01
In recognition of the successful completion of the A2 LA evaluation process, (including an assessment of
the organization’s compliance with A2LA’s Calibration Program Requirements) accreditation is granted to
this laboratory to perform the following calibrations
1, 4:
I. Mechanical
Parameter/Equipment
Range
CMC
2, 3, 5 (±)
Comments
Pressure – Measuring
Equipment
(0.1 to 512.0) psi
(0.689 to 3530) kPa
0.5 % of FS
6
or
(0.016P + 0.034) psi
(0.016P + 0.23) kPa
Whichever is greater
Comparison using the
force measurement
machine and fixtures of
known area to apply a
specific load.
Pressure – Measuring
Equipment (Pneumatic)
(5.0 to 200.0) mmHg
(0.7 to 26.7) kPa
(0.1 to 15.0) psi
(0.7 to 103.4) kPa
(0.1 to 60.0) psi
(0.7 to 413.7) kPa
(1.0 to 128.0) psi
(6.9 to 822.5) kPa
3.9 mmHg
0.5 kPa
0.1 psi
0.7 kPa
0.3 psi
2.1 kPa
1.3 psi
9.2 kPa
Comparison using the air
chamber pressure
controlled by IP
regulator/pressure
transducer/ DAQ | 0.1 to 512 psi |
| 2 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3589.01) Revised 11/25/2024 Page 2 of 2
Parameter/Equipment
Range
CMC2, 3, 5 (±)
Comments
Pressure – Measuring
Equipment (Pneumatic)
(cont)
(1.0 to 128.0) psi
(6.9 to 822.5) kPa
0.24 psi
1.4 kPa
Comparison using the air
chamber pressure
controlled by pressure
controller
7
_________________________________________________________
1 This laboratory does not offer commercial calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 In the statement of CMC, P is the numerical value of the pressure in the applicable units listed.
4 This scope meets A2LA’s P112 Flexible Scope Policy.
5 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory
is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed
measurement parameter.
6 In the statement of CMC, FS is the Full Scale, the maximum of the range of the sensor (Example: 10 psi to
300 psi = FS 300 psi).
7 Pressure controller model ALICAT PCPD-100PSIG-D/5P Range 130PSIG. | 1 to 128 psi |
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.
Accredited Laboratory
A2LA has accredited
XSENSOR TECHNOLOGY CORP
Calgary, Alberta, CANADA
for technical competence in the field of
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General requirements
for the competence of testing and calibration laboratories. This laboratory also meets R205 – Specific Requirements: Calibration
Laboratory Accreditation Program. This accreditation demonstrates technical competence for a defined scope and the
operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017).
Presented this 25th day of October 2024.
__ _____________________
Mr. Trace McInturff, Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 3589.01
Valid to November 30, 2026 | ||
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.