PO Box 8239 Bldg. 4148 cross road 118, Support Industrial Area 1, Al-Jubail Industrial City, 31961, Saudi Arabia, Al-Jubail Industrial City, Saudi Arabia

A2LA 3097.01ISO/IEC 17025Register refreshed 2026-11-20

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free, no obligation.
Scope lines
7
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 7 scope lines.

A2LA certificate 3097.01 · 7 lines

LineMethodMatrixRange
1

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3097.01) 05/30/2025 Page 1 of 6 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 GULF EQUIPMENT METROLOGY SERVICES (GEMS) P.O Box 8239, Building No 4148, Cross Road 118, Support Industrial Area 1 Al-Jubail Industrial City, Saudi Arabia 31961 Jowell Liwanag Phone: +966 13 341 9451 CALIBRATION Valid To: January 31, 2027 Certificate Number: 3097.01 In recognition of the successful co mpletion of the A2LA evaluation process (including an assessment of the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is granted to this laboratory to perform the following calibrations 1, 6: I. Acoustical Parameter/Equipment Range CMC 2 (±) Comments Sound Level Calibrators, Fixed Points 94 dB, 104 dB, & 114 dB 0.61 dB 3M DL-1 characterized w/ B&K model 4226 Parameter/Range Frequency CMC2 (±) Comments Acoustical Levels – Sound Level Meters, Noise Dosimeters 94 dB, 104 dB, & 114 dB 31.5 Hz to 4 kHz, Octave Step 8 kHz (12.5 to 16) kHz 0.42 dB 0.5 dB 0.7 dB B&K model 4226 multifunction acoustic calibrator
2

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3097.01) 05/30/2025 Page 2 of 6 II. Electrical – DC/Low Frequency Parameter/Equipment Range CMC2, 5 (±) Comments Electrical Simulation Calibration of Thermocouple Devices 3 – Generate & Measure Type B Type C Type E Type J & L Type K Type N Type R Type S Type T & U (600 to 1820) °C (0 to 2316) °C (-250 to 1000) °C (-210 to 1200) °C (-200 to 1372) °C (-200 to 1300) °C (0 to 1767) °C (0 to 1767) °C (-250 to 400) °C 0.0063mV 0.0056 mV 0.0059 mV 0.0055 mV 0.0061 mV 0.0057 mV 0.0064 mV 0.0064 mV 0.0056 mV Fluke 5522A III. Fluid Quantities Parameter/Equipment Range CMC 2, 4, 5 (±) Comments Density (Specific Gravity) (0 to 3) g/cm 3 0.0002 g/cm3 Mettler Toledo model DE45 density meter Air Velocity – Anemometers, Thermal, Pitot, Vane Style & Similar Equipment (100 to 1000) fpm (1000 to 9000) fpm 2.4 % rdg 1.2 % rdg Omega Engineering model WT4001-D-110V wind tunnel w/ electronics box WT 4401-D
3

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3097.01) 05/30/2025 Page 3 of 6 Parameter/Equipment Range CMC24, 5 (±) Comments Flow (Air)3 – Measure (0 to 5) L/min (0.23 to 0.6) CFM (0.6 to 1.2) CFM (1.2 to 1.8) CFM (1.8 to 2.33) CFM (0.85 to 2.0) CFM (2.0 to 4.0) CFM (4.0 to 6.0) CFM (6.0 to 8.5) CFM (3.79 to 10.0) CFM (10.0 to 20.0) CFM (20.0 to 30.0) CFM (30.0 to 37.9) CFM 0.31 L/min 0.019 CFM 0.019 CFM 0.015 CFM 0.019 CFM 0.077 CFM 0.57 CFM 0.58 CFM 0.11 CFM 0.37 CFM 0.37 CFM 0.78 CFM 0.88 CFM Aalborg XFM17 Brook Instruments flowmeter models: 1110CJ61CUHAA 1110CK61CUHAA 1110CM61CUHAA IV. Mechanical Parameter/Equipment Range CMC2, 5 (±) Comments Vibration General Purpose – Sensitivity Magnitude Accelerometer Vibration Meters (20 to 10 000) Hz (20 to 10 000) Hz 7.3 % rdg 2 % rdg 3120A accelerometer & vibration controller. Agate AT-2050
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3097.01) 05/30/2025 Page 4 of 6 Parameter/Equipment Range CMC2, 4, 5 (±) Comments Volumetric (Pipettes, Flask, Burets, Cylinders) (2 to 10) μl (> 10 to 20) μl (> 20 to 100) μl (> 100 to 200) μl (> 200 to 500) μl (> 500 to 1000) μl (> 1000 to 2000) μl (> 2000 to 5000) μl (> 5000 to 10 000) μl (> 10 000 to 20 000) μl (20 to 50) mL (> 50 to 200) mL (> 200 to 500) mL > 500 mL to 1 L (> 1 to 2) L 0.033 μl 0.048 μl 0.30 μl 0.40 μl 0.6 μl 1.3 μl 2.4 μl 5.9 μl 13 μl 24 μl 3.7 μl 26 μl 40 μl 61 μl 83 μl Gravimetric measurement using balances Pressure – Measure & Measuring Equipment 3 Hydraulic Pneumatic Vacuum (6000 to 20 000) psig (60 to 20 000) psig (20 000 to 39 000) psig (-13 to 3000) psig (-15 to 15) InH 2O (0.2 to 25) psig (2 to 1000) psig (0 to –13) psia 0.015 % 0.0035 % 9.2 psi 0.27 psi 0.0087 InH 2O + 0.01 % 0.0011 % 0.0027 % 0.002 psi Fluke PPCH-140M A140Ms/A40Ms Ruska model 2485-930S deadweight tester High pressure pump w/ Additel reference gage PM500-BG20M w/ Fluke 2271A PM500-BG3.5K / Fluke 2271A Fluke model 2465 gas piston gage Wika CPG2500 digital pressure gauge
5

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3097.01) 05/30/2025 Page 5 of 6 V. Time & Frequency Parameter/Equipment Range CMC2, 7 (±) Comments Frequency – Measuring Equipment (5 & 10) MHz 0.1 Hz to 20 MHz 20 MHz to 50 MHz 50 MHz to 31.8 GHz 5.9 x 10 -10 Hz 5.8 x10 -6 Hz 6.0 x 10-6 Hz 8.9 x 10-6 Hz Symmetricom 5071A cesium beam frequency standard w/ Fluke 910R GPS HP 3325A/HP 5345A HP 3325A opt 001/ HS 20/ 54610, Agilent E8257D opt 007, 1E1,1EUM 532, U01, UNX Frequency – Measure (0 to 400) MHz (0.4 to 1.6) GHz (1.5 to 26) GHz 3.0 x 10 -7 Hz 17 x 10-7 Hz 14 x 10-7 Hz HP 5345A HP 5345A/ HP 5355A HP 5345A/HP 5355A/ HP 5356B Stop Watches (0.15 to 24) hr 0.33 s Agilent 33220A, 53132A w/ Symmetricom 5071A Timers3 (0.15 to 1) hr 13 ms Agilent 33220A, 53132A w/ Symmetricom 5071A ____________________________________________ 1 This laboratory offers commercial calibration and field calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. 3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC. 4 In the statement of CMC, the value is defined as the percentage of reading, unless otherwise noted.
6

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3097.01) 05/30/2025 Page 6 of 6 5 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter. 6 This scope meets A2LA’s P112 Flexible Scope Policy.
7ISO/IEC 17025:2017; R205--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Accreditation data refreshed from the A2LA register on .

This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.