Eddyfi Canada
Quebec, Canada
Visit website- Scope lines
- 9
- A2LA certificate
- 1
- Sites in our directory
- 3
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 9 scope lines.
A2LA certificate 2734.03 · 9 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 1 of 8
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
EDDYFI CANADA INC.
3425 Pierre-Ardouin St
Quebec, QC, CANADA G1P 0B3
Ghislain Cournoyer, Quality Assurance Director Phone: 418 263 3686
CALIBRATION
Valid To: August 31, 2027 Certificate Number: 2734.03
In recognition of the successful completion of the A2LA evaluation process (including an assessment of
the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory to perform the following calibrations
1, 8:
I. Electrical – DC/Low Frequency
Parameter/Equipment
Range
CMC2, 6 (±)
Comments
AC Resistance –
Measure
3
(0 to 10) Ω
0.45 Ω
Agilent 33250A,
33610 Tektronix
TDS-3032 (A, B,
or C), MDO3032,
MDO32
II. Time & Frequency (Ultrasound Instrumentation – Dynaray Series)
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Receiver – Ultrasound
Instrumentation3
Frequency
Response:
Low Cutoff
High Cutoff
(0.225 to 20)
MHz
None, HP1M, HP2M, HP5M, HP10M,
LP2M < 2.2 % , All others < 0.90 %
None, BP1-5M, HP1M, LP2M, LP5M,
LP10M, LP15M < 6.5 % , All others <
1.9 %
Agilent 33250A,
Tektronix TDS-
3032
(A, B, or C),
MDO3032,
MDO32, ZETEC
attenuator
10040260,
ZETEC
ABUXE127A | 0 to 10 Ω |
| 2 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 2 of 8
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Receiver – Ultrasound
Instrumentation3 (cont)
Frequency
Response:
Bandwidth
Center
Linearity of Time
Delays
Gain Linearity
Channel Gain
Variation
Linearity of Vertical
Display
Equivalent Input
Noise
(0.225 to 20)
MHz
Up to 50 µs
(0 to 80) dB
< 1.5 dB
(Zetec)
(0 to 28) dB
rel.
< 80 nV/√Hz
None, HP1M, HP2M, HP10M, LP2M,
LP15M < 2.5 % , All others < 1.0 %
None, < 3.6 % , All others < 1.8 %
0.0063 %
0.47 dB
0.65 dB
0.53 %
7.0 nV/√Hz
Agilent 33250A,
Tektronix TDS-
3032
(A, B, or C),
MDO3032,
MDO32, ZETEC
attenuator
10040260,
ZETEC
ABUXE127A
Transmitter –
Ultrasound
Instrumentation3
Voltage
Rise Time
Duration
Linearity of Time
Delays
(25 to 250) V
Up to 50 ns
(25 to 1000) ns
Up to 50 µs
1.8 %
0.78 ns
0.46 ns
0.070 %
Agilent 33250A,
Tektronix TDS-
3032
(A, B, or C),
MDO3032,
MDO32, ZETEC
ABUXE127A,
50 ohms/10W
load | 0.225 to 20 MHz |
| 3 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 3 of 8
III. Time & Frequency (Ultrasound Instrumentation – Zircon-15, Zircon-18, Topaz16, Topaz32, Quartz)
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Receiver – Ultrasound
Instrumentation3
Frequency Response –
Zircon-15, Zircon-18,
Topaz32, Quartz:
Phased Array
Low Cutoff
High Cutoff
Bandwidth
Center
Conventional
Low Cutoff
High Cutoff
Bandwidth
Center –
Conventional
Zircon-15
(0.50 to 15) MHz
Zircon-18, Topaz32,
Quartz
(0.50 to 18) MHz
< 0.70 %
HP1M ≤ 3.2 %
None, LP5M, LP10M,
HP2M ≤ 2.4 % ,
All others ≤ 1.5 %
< 0.80 %
< 0.90 %
< 0.70 %
HP1M ≤ 8.0 % ,
None, HP2M ≤ 5.2 %
LP5M, LP10M, BP1-5M,
BP2-10M ≤ 1.9 % ,
All others ≤ 0.90 %
< 0.90 %
LP5M, HP1M ≤ 1.3 % ,
All others ≤ 1.0 %
Agilent 33250A,
Tektronix TDS-
3032 (A, B, or
C), MDO3032,
MDO32, ZETEC
attenuator
10040260,
ZETEC
10040371 (ZIF),
ZETEC
10059869 (IPEX) | 0.5 to 18 MHz |
| 4 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 4 of 8
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Receiver – Ultrasound
Instrumentation3 (cont)
Frequency Response –
Topaz16:
Phased Array
Low Cutoff
High Cutoff
Bandwidth
Center – Phased Array
Conventional
Low Cutoff
High Cutoff
Bandwidth
Center
(0.50 to 18) MHz
< 0.80 %
None, LP2M, LP5M,
LP10M, HP1M, HP2M
< 2.3 % , All others < 1.4 %
LP2M < 1.1 % ,
All others < 0.90 %
None, LP2M, LP5M < 1.4 %
, All others < 1.0 %
< 1.0 %
None, HP1M, HP2M < 6.5
% LP2M, LP5M, LP10M,
HP5M, BP1-5M, BP5-
18M < 3.0 % ,
All others < 1.7 %
LP2M, BP5-18M, BP10-
18M < 1.6 % ,
All others < 1.0 %
None, LP2M, BP5-18M <
1.9 % , All others < 1.0 %
Agilent 33250A,
Tektronix TDS-
3032 (A, B, or
C), MDO3032,
MDO32, ZETEC
attenuator
10040260,
ZETEC
10040371 (ZIF),
10059869 (IPEX) | 0.5 to 18 MHz |
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 5 of 8
Parameter/Equipment
Range5
CMC2, 4, 5 (±)
Comments
Receiver – Ultrasound
Instrumentation3 (cont)
Linearity of Time Delays
Gain Linearity
Phased Array
Conventional
Channel Gain Variation
Phased Array
Conventional
Linearity of Vertical
Display
Equivalent Input Noise
Up to 20 µs
(0 to 70) dB
(0 to 70) dB
< 1.5 dB (Zetec)
< 1.5 dB (Zetec)
28 dB rel.
< 80 nV/√Hz
0.014 %
0.29 dB
0.27 dB
0.40 dB
0.38 dB
0.36 %
3.3 nV/√Hz
Agilent 33250A,
Tektronix TDS-3032
(A, B, or C),
MDO3032, MDO32,
ZETEC attenuator
10040260,
ZETEC 10040371
(ZIF), ZETEC
10059869 (IPEX)
Transmitter – Ultrasound
Instrumentation3
Voltage
Phased Array
Conventional
Rise Time
Phased Array
Conventional
Duration
Phased Array
Conventional
Linearity of Time Delays
(35 to75) V
(50 to 200) V
Up to 50 ns
Up to 50 ns
(25 to 500) ns
(25 to 500) ns
Up to 20 µs
2.0 %
2.6 %
0.57 ns
0.34 ns
0.51 ns
0.48 ns
0.051 %
Agilent 33250A,
Tektronix TDS-3032
(A, B, or C),
MDO3032, MDO32,
ZETEC 10040371
(ZIF), ZETEC
10059869 (IPEX)
50 ohms/10W load | ||
| 6 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 6 of 8
IV. Time & Frequency (Ultrasound Instrumentation – Topaz64, Emerald64)
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Receiver – Ultrasound
Instrumentation3
Frequency Response:
Phased Array
Low Cutoff
High Cutoff
Bandwidth
Center
Conventional
Low Cutoff
High Cutoff
Bandwidth
Center
(0.50 to 18) MHz
HP0.6M, LP2M, LP5M, LP10M,
HP1M, BP1-5M: < ±6.5 %
HP2M, BP1.75-7M, BP2-10M:
< ±3.4 %
All others: < ±1.7 %
HP0.6M, LP2M, LP5M, LP10M,
BP1-5M:< ±2.5 %
All others: < ±1.1 %
HP0.6M, LP2M, LP5M, BP1-
5M, BP5-10M: < ±3.6 %
All others: < ±1.9 %
HP0.6M, LP2M, LP5M, LP10M,
BP1-5M, BP1.75-7M: < ±4.3 %
All others: < ±1.9 %
LP2M, BP1-4M, BP1.125-4.5M,
BP1.875-7.5M, BP5-20M,
BP0.55-6M: < ±4.2 %
All others: < ±3.0 %
LP2M, BP1-4M, HP3M:
< ±2.1 %
All others: < ±1.2 %
LP2M, BP1-4M, BP2-8M, BP5-
20M, HP10M: < ±3.1 %
All others: < ±2.0 %
LP2M, BP1-4M, BP1.125-4.5M,
BP2-8M, HP10M, HP3M,
BP0.55-6M: < ±3.7 %
All others: < ±2.0 %
Agilent 33250A,
Tektronix TDS-
3032 (A, B, or C),
MDO3032,
MDO32,
ZETEC attenuator
10040260,
ZETEC
10040371 (ZIF)
ZETEC
10059869 (IPEX) | 0.5 to 18 MHz |
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 7 of 8
Parameter/Equipment
Range5
CMC2, 4, 5 (±)
Comments
Receiver – Ultrasound
Instrumentation3 (cont)
Linearity of Time Delays
Gain Linearity
Phased Array
Conventional
Channel Gain Variation
Phased Array
Conventional
Linearity of Vertical Display
Equivalent Input Noise
Phased Array
Conventional
Up to 20 µs
(0 to 70) dB
(0 to 70) dB
< 1.5 dB (Zetec)
< 1.5 dB (Zetec)
28 dB rel.
< 80 nV/√Hz
0.088 %
0.29 dB
0.27 dB
0.41 dB
0.37 dB
1.2 %
9.0 nV/√Hz
28 nV/√Hz
Agilent 33250A,
Tektronix TDS-3032 (A,
B, or C), MDO3032,
MDO32, ZETEC
attenuator 10040260,
ZETEC 10040371 (ZIF),
ZETEC 10059869
(IPEX)
Transmitter – Ultrasound
Instrumentation3
Voltage
Phased Array
Conventional
Rise Time
Phased Array
Conventional
Duration
Phased Array
Conventional
Linearity of Time Delays
(10 to 75) V
(20 to 200) V
Up to 50 ns
Up to 50 ns
(25 to 500) ns
(25 to 500) ns
Up to 20 µs
2.2 %
2.7 %
0.57 ns
0.34 ns
0.49 ns
0.57 ns
0.056 %
Agilent 33250A,
Tektronix TDS-3032
(A, B, or C),
MDO3032, MDO32,
ZETEC
10040371 (ZIF)
ZETEC
10059869 (IPEX)
50 ohms/10W load
____________________________________________ | ||
| 8 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2734.03) 05/23/2025 Page 8 of 8
1 This laboratory offers commercial and field calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g., resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 In the statement of CMC, percentages represent percent of reading unless otherwise noted.
5 References to decibels (dB) refer to dB in voltage.
6 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification.
7 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
8 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 9 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.