Los Alamos National Laboratory Metrology Program and
P.O. Box 1663, Mail Stop D478, 87545-0001, Los Alamos, NM, United States
Visit website- Scope lines
- 5
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 5 scope lines.
A2LA certificate 7277.01 · 5 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert No. 7277.01) 9/16/2025 Page 1 of 4
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
LOS ALAMOS NATIONAL LABORATORY
METROLOGY PROGRAM AND CALIBRATION LABORATORY
P.O. Box 1663, Mail Stop D478
Los Alamos, NM 87545-0001
Christina Martinez Phone: 505 665 2744
CALIBRATION
Valid To: September 30, 2027 Certificate Number: 7277.01
In recognition of the successful completion of the A2LA evaluation process (including an assessment of
the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory to perform the following calibrations1, 6:
I. Dimensional
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Steel Gage Blocks
(0.01 to 5.0) in
(≥ 5 to 20) in
Up to 100 mm
(2.8 + 0.94L) µin
(5.1 + 0.71L) µin
(0.06 + 0.0014L) µm
Mechanical
comparison
Dial/Test Indicator
Up to 100 mm Analog
Up to 100 mm Digital
0.67 µm
0.73 µm
Gage calibrator
Micrometers, Outside7
Up to 1 in
120 µin
Gage blocks
Calipers8
Up to 6 in
(320 + 2.9L) µin
Gage blocks | ||
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert No. 7277.01) 9/16/2025 Page 2 of 4
II. Electrical DC/Low Frequency
Parameter/Equipment
Range
CMC2, 3 (±)
Comments
Resistance – Measure
1 Ω
10 Ω
100 Ω
1000 Ω
10 000 Ω
1.0 µΩ/Ω
1.6 µΩ/Ω
1.3 µΩ/Ω
1.0 µΩ/Ω
1.0 µΩ/Ω
MI 6242 bridge &
933 lR resistors
III. Mechanical
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Accelerometers
Sensitivity
Up to 100 mV/g:
(5 to 9) Hz
(10 to 99) Hz
100 Hz (ref)
(101 to 920) Hz
> 920 Hz to 5 kHz
(> 5 to 10) kHz
(> 10 to 15) kHz
(> 15 to 20) kHz
1.9 %
1.3 %
0.93 %
1.1 %
1.5 %
2.0 %
2.3 %
3.0 %
Comparison
method, PCB 9155
cal system
Torque Wrenches
(5 to 50) lbf·in
(30 to 400) lbf·in
(10 to 125) lbf·ft
(20 to 250) lbf·ft
0.71 %
0.68 %
0.67 %
0.81 %
CDI torque
calibrator | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert No. 7277.01) 9/16/2025 Page 3 of 4
IV. Thermodynamics
Parameter/Equipment
Range
CMC2, 5 (±)
Comments
Temperature – Measuring
Equipment
(-40 to 120) °C
(> 120 to 300) °C
(> 300 to 425) °C
(> 425 to 660) °C
(> 660 to 960) °C
(> 960 to 1200) °C
0.018 °C
0.019 °C
0.30 °C
0.47 °C
0.35 °C
1.7 °C
7340 bath with SPRT &
super-thermometer
6330 bath with SPRT &
super-thermometer
9173 well with SPRT &
super-thermometer
Additel 850 furnace with
high temp SPRT & super-
thermometer
Additel 850 furnace with
Type-S Thermocouple &
super-Daq
Infrared Temperature –
Measuring Equipment
(-15 to 0) °C
(> 0 to 50) °C
(> 50 to 100) °C
(> 100 to 200) °C
(> 200 to 350) °C
(> 350 to 500) °C
0.45 °C
0.39 °C
0.42 °C
0.54 °C
0.84 °C
1.1 °C
Fluke 4180 & 4181
calibrators, emissivity @
0.95
V. Time & Frequency
Parameter/Equipment
Range
CMC2 (±)
Comments
Frequency Offset
10 MHz
1.0 x 10-12 Hz/Hz
NIST FMAS
Stopwatches & Timers8
1 s to 24 hr
0.064 s/day
HK4500 timometer
________________________
1 This laboratory does not offer commercial calibration services. | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert No. 7277.01) 9/16/2025 Page 4 of 4
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a
percentage or fraction of the reading plus a fixed floor specification.
4 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches
and % is the percent of reading unless otherwise stated.
5 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
6 This scope meets A2LA’s P112 Flexible Scope Policy.
7 Resolution is not included in the CMC uncertainty budgets.
8 Human factor uncertainty component not included in the CMC uncertainty budgets. | ||
| 5 | ISO/IEC 17025:20 17; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.