PCB Piezotronics, Inc., PCB Load & Torque Division
24350 Indoplex Circle, 48335, Farmington Hills, MI, United States
Visit website- Scope lines
- 3
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 3 scope lines.
A2LA certificate 1015.01 · 3 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1015.01) 05/28/2024 Page 1 of 2
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
PCB PIEZOTRONICS, INC., PCB LOAD & TORQUE DIVISION
24350 Indoplex Circle
Farmington Hills, MI 48335
Lance Pellens Phone: 716 684 0002 x 102222
CALIBRATION
Valid To: April 30, 2026 Certificate Number: 1015.01
In recognition of the successful completion of the A2LA evaluation process, (including an assessment of
the organization’s compliance with A2LA’s Calibration Program Requirements), accreditation is granted
to this laboratory to perform the following calibrations
1, 4:
I. Electrical – DC/Low Frequency
Parameter/Equipment
Range CMC2, 3 (±) Comments
Torque/Force Indicating
System
(0 to 2.5) mV/V
0.0041 mV/V
Precision bridge
simulator
II. Mechanical
Parameter/Equipment
Range CMC2 (±) Comments
Force – Tension and
Compression
Up to 250 lbf
(250 to 500) lbf
(500 to 2200) lbf
(2200 to 30 000) lbf
(30 000 to 100 000) lbf
0.021 % Full Scale
0.042 % Full Scale
0.025 % Full Scale
0.022 % Full Scale
0.036 % Full Scale
Deadweight
Reference load cell | ||
| 2 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1015.01) 05/28/2024 Page 2 of 2
Parameter/Equipment
Range CMC2 (±) Comments
Torque
Up to 2 lbf·in
(2 to 100) lbf·in
(100 to 3600) lbf·in
(3600 to 12 000) lbf·in
(12 000 to 144 000) lbf·in
(144 000 to 216 000) lbf·in
(12 000 to 64 800) lbf·in
(64 800 to 216 000) lbf·in
0.74 % Full Scale
0.046 % Full Scale
0.014 % Full Scale
0.017 % Full Scale
0.029 % Full Scale
0.019 % Full Scale
0.12 % Full Scale
0.047 % Full Scale
Torque arm with
weights
Reference torque
cell
___________________________
1 This laboratory offers commercial calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavi or of the customer’s device and to influences from the
circumstances of the specific calibration.
3 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
4 This scope meets A2LA’s P112 Flexible Scope Policy. | to 216000 lbf·in |
| 3 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.