GRUENTEC. CÍA. LTDA
Eloy Alfaro S7-157 y Belisario Quevedo, (Basement), Quito, 170157, Ecuador, Quito, Ecuador
Visit website- Accredited standards
- 1
- A2LA certificate
- 1
- Sites in our directory
- 2
What this lab is accredited for
Our reading of the A2LA scope: each standard once, with its official name where we have checked it.
Only this lab. Searches the 1 standard number and the 1 official name we hold. When no standard matches, it searches the method text of every scope line.
ISO 8655-6Piston-operated volumetric apparatus - Part 6: Gravimetric reference measurement procedure for the determination of volume
ISO 8655-6:2022
Read from the A2LA scope document we downloaded on 2026-08-05. Register refreshed 2026-11-20.
Scope as published by A2LA
8 lines
ISO/IEC 17025
Active7126.01
The scope document we hold says it is valid until 2028-06-30.
Accredited scope, certificate 7126.01
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017 | pH buffer solutions; Conductivity standard solutions | - |
| 2 | - | laboratory glasswareShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7126.01) 06/03/2026 Page 2 of 7
II. Fluid Quantities
Parameter/Equipment
Range
CMC2 (±)
Comments
Volume – Laboratory Glass Ware
Single Volume Pipettes
0.5 mL
1 mL
2 mL
5 mL
10 mL
20 mL
25 mL
50 mL
100 mL
0.000 66 mL
0.0012 mL
0.0019 mL
0.0019 mL
0.0024 mL
0.0029 mL
0.0029 mL
0.0032 mL
0.0043 mL
Gravimetric method
Volume – Laboratory Glass Ware
One-Mark Volumetric Flasks
1 mL
2 mL
5 mL
10 mL
20 mL
25 mL
50 mL
100 mL
200 mL
250 mL
500 mL
1000 mL
2000 mL
5000 mL
0.0045 mL
0.0045 mL
0.0045 mL
0.0045 mL
0.0074 mL
0.0074 mL
0.011 mL
0.016 mL
0.023 mL
0.024 mL
0.036 mL
0.057 mL
0.1 mL
0.22 mL
Gravimetric method
Volume – Laboratory Glass Ware
Graduated Pipettes
0.1 mL
0.2 mL
0.5 mL
1 mL
2 mL
5 mL
10 mL
20 mL
25 mL
0.0017 mL
0.0017 mL
0.0019 mL
0.0026 mL
0.0069 mL
0.011 mL
0.02 mL
0.02 mL
0.02 mL
Gravimetric method | - |
| 3 | ISO 8655-6:2022 | Laboratory Glass Ware; Piston Pipettes | - |
| 4 | ISO 8655-6:2022 | Piston-Operated Volumetric Apparatus; Mass | - |
| 5 | OIML R111-1; OIML R76-1; SIM MWG7/cg-01/v.00 Ed. 2009; CEM TH-007 | mass weights; scales and balances; thermo-hygrometers; dataloggers | See scope document |
| 6 | CEM TH-007; CEM TH-001 | humidity; temperature | See scope document |
| 7 | - | Ovens, Environment Chambers, Freezers, Stoves, Incubators, Mufflers, Refrigerators, AutoclavesShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7126.01) 06/03/2026 Page 7 of 7
Parameter/Equipment
Range CMC2 (±) Comments
Temperature Controlled
Enclosures3 –
Ovens, Environment
Chambers, Freezers,
Stoves, Incubators,
Muffles, Refrigerators, etc.
Autoclaves3
(-20 to 0) °C
(0 to 4) °C
(>4 to 20) °C
(>20 to 50) °C
(>50 to 70) °C
(>70 to 100) °C
(>100 to 120) °C
(>120 to 150) °C
(>150 to 180) °C
(>180 to 200) °C
(>200 to 250) °C
(>250 to 300) °C
(>300 to 350) °C
(>350 to 400) °C
(>400 to 450) °C
(>450 to 500) °C
(>500 to 550) °C
121 °C
0.16 °C
0.16 °C
0.16 °C
0.16 °C
0.16 °C
0.16 °C
0.16 °C
0.2 °C
0.2 °C
0.2 °C
1.3 °C
1.3 °C
1.6 °C
1.6 °C
1.9 °C
1.9 °C
2.3 °C
0.19 °C
Comparison method
using Type K
thermocouple data
loggers
Comparison method
using wireless data
loggers
_____________________________________
1 This laboratory offers commercial calibration service and field calibration services.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects su ch as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g., resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 This scope meets A2LA’s P112 Flexible Scope Policy. | See scope document |
| 8 | ISO/IEC 17025:2017 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Eloy Alfaro S7-157 y Belisario Quevedo, (Basement), Quito, 170157, Ecuador, Quito, EC
-0.22016, -78.51233
Open in MapsOther Locations
1 other site in this organization
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.