Cameron Instruments
Guelph, Canada
Visit website- Scope lines
- 6
- A2LA certificate
- 1
- Sites in our directory
- 4
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 6 scope lines.
A2LA certificate 4171.01 · 6 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017 | - | See scope document |
| 2 | - | GasShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 4171.01) Revised 04/09/2026 Page 2 of 5
Parameter/Equipment
Range
CMC2, 9 (±)
Comments
DC Current – Measure
(0 to 0.1) mA
(0.1 to 1) mA
(1 to 4) mA
(4 to 8) mA
(8 to 16) mA
(16 to 20) mA
(20 to 24) mA
(24 to 30) mA
(30 to 40) mA
(40 to 50) mA
1.2 µA
1.3 µA
1.8 µA
2.3 µA
3.5 µA
4.0 µA
4.6 µA
5.5 µA
11 µA
12 µA
AMC910
Resistance – Generate
(5 to 10) Ω
(10 to 100) Ω
(100 to 200) Ω
(200 to 400) Ω
(400 to 1000) Ω
(1000 to 4000) Ω
25 mΩ
24 mΩ
27 mΩ
30 mΩ
0.35 Ω
0.36 Ω
AMC910
II. Mechanical
Parameter/Equipment
Range
CMC2, 3, 4, 5, 7 (±)
Comments
Gauge Pressure –
Measure/Measuring Equipment
Gas
Gas – Differential6
(5 to 350) kPa
(25 to 1750) kPa
(100 to 7000) kPa
(1.4 to 172) kPa
(12 to 700) kPa
(14 to 7000) kPa
(2000 to 70 000) kPa
(-31 to 10) kPa
(-99 to 240) kPa
(240 to 700) kPa
(-99 to 1750) kPa
(1750 to 7000) kPa
(0 to 5000) kPa
(5000 to 20 685) kPa
(0 to 11 667) kPa
(11 667 to 41 400) kPa
(0 to 70 000) kPa
(-99 to 259) kPa
0.0016 % + 0.19 Pa
0.0014 % + 0.43 Pa
0.0018 % + 3.1 Pa
0.0016 % + 0.13 Pa
0.0015 % + 0.08 Pa
0.0018 % + 0.22 Pa
0.0030 % + 19 Pa
0.010 % + 10 Pa
12 Pa
0.005 %
180 Pa
0.010 %
500 Pa
0.010 %
1400 Pa
0.012 %
9100 Pa
0.0014 % + 0.64 Pa
Fluke PG7601
Ruska 2465
Fluke PG7202
BA100K – RPM4
A700KP – RPM4
A7M – PM600
A20M – PM600
A40M – RPM4
G70M – PM230
Fluke PG7601 &
RPM4 | - |
| 3 | - | Oil; GasShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 4171.01) Revised 04/09/2026 Page 3 of 5
Parameter/Equipment
Range
CMC2, 3, 4, 5, 7 (±)
Comments
Gauge Pressure –
Measure/Measuring
Equipment
Oil
(2000 to 70 000) kPa
(5000 to 210 000) kPa
(0 to 2000) kPa
(2000 to 20 685) kPa
(0 to 20 000) kPa
(20 000 to 206 850) kPa
0.0029 % + 66 Pa
0.0035 % + 60 Pa
400 Pa
0.02 %
4000 Pa
0.02 %
Fluke PG7202
A20M RPM4 E-DWT
A200M RPM4 E-DWT
Absolute Pressure –
Measure/Measuring
Equipment
Gas
Oil
(5 to 350) kPa
(25 to 1750) kPa
(100 to 7000) kPa
(1.4 to 172) kPa
(12 to 700) kPa
(14 to 7000) kPa
(2000 to 70 000) kPa +
Atmospheric Pressure
(70 to 110) kPa
(5 to 240) kPa
(240 to 700) kPa
(5 to 1750) kPa
(1750 to 7000) kPa
(97 to 5000) kPa
(5000 to 20 685) kPa
(97 to 11 667) kPa
(11 667 to 41 400) kPa
(97 to 70 000) kPa +
Atmospheric Pressure
(2000 to 70 000) kPa +
Atmospheric Pressure
(5000 to 210 000) kPa +
Atmospheric Pressure
0.0014 % + 0.6 Pa
0.0013 % + 0.54 Pa
0.0018 % + 3.1 Pa
0.0012 % + 1.1 Pa
0.0014 % + 0.94 Pa
0.0018 % + 0.43 Pa
0.0030 % + 20 Pa
0.010 %
12 Pa
0.005 %
180 Pa
0.010 %
500 Pa
0.010 %
1400 Pa
0.012 %
9100 Pa
0.0029 % + 62 Pa
0.0035 % + 71 Pa
Fluke PG7601
Ruska 2465
Fluke PG7202 and
RPM4
BA100K – RPM4
A700KP – RPM4
A7M – PM600
A20M – PM600
A40M – RPM4
G70M – PM230 and
Fluke RPM4
Fluke PG7202 and
Fluke RPM4
Fluke PG7202 and
Fluke RPM4 | See scope document |
| 4 | - | TemperatureShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 4171.01) Revised 04/09/2026 Page 4 of 5
III. Thermodynamics
Parameter/Equipment
Range
CMC2 (±)
Comments
Temperature – Dry Well
(-100 to -40) °C
(-40 to 155) °C
0.038 °C
0.037 °C
RTC159
Fluke 5609 PRT
Temperature – Dry Well
(-45 to -35) °C
(-35 to 50) °C
(50 to 140) °C
0.031 °C
0.020 °C
0.023 °C
Fluke 9170
Fluke 5609 PRT
Temperature – Dry Well
(33 to 420) °C
(420 to 600) °C
(600 to 660) °C
0.043 °C
0.064 °C
0.077 °C
RTC700
Fluke 5609 PRT
Temperature – Liquid
Bath
(-22 to 155) °C
0.034 °C
RTC158
Fluke 5609 PRT
_________________________
1 This laboratory offers commercial calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 In the statement of CMC, percentages are to be read as percent of reading, unless otherwise noted.
4 The contributions from the “best existing device”, such as resolution and repeatability of the unit under
calibration, are not included in the CMC uncertainties claim.”
5 The QRPT (RPM, PPC, PM230, PM600) uncertainty budgets are prepared in accordance with Fluke
technical note 8050TN11.
6 The Gas – Differential mode uncertainty budgets are prepared in accordance with Fluke technical note
9940TN02.
7 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
8 This scope meets A2LA’s P112 Flexible Scope Policy. | See scope document |
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 4171.01) Revised 04/09/2026 Page 5 of 5
9 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification. | ||
| 6 | ISO/IEC 17025:2017 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.