Guildline Instruments
Smiths Falls, Canada
Visit website- Scope lines
- 6
- A2LA certificate
- 1
- Sites in our directory
- 3
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 6 scope lines.
A2LA certificate 2906.01 · 6 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017 | - | - |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2906.01) 3/24/2026 Page 2 of 5
Parameter/Equipment
Range
CMC2, 3, 6 (±)
Comments
DC Resistance5 –
Generate and Measure
Resistance Ranges
Fixed Decade Points
(1 to 10) µΩ
(10 to 100) µΩ
100 µΩ to 1 mΩ
(1 to 10) mΩ
(10 to 100) mΩ
100 mΩ to 1 Ω
(1 to 10) µΩ
(10 to 100) µΩ
100 µΩ to 1 mΩ
(1 to 10) mΩ
(10 to 100) mΩ
100 mΩ to 1 Ω
1 Ω to 10 kΩ
(10 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
100 MΩ to 1 GΩ
1 µΩ
10 µΩ
100 µΩ
1 mΩ
10 mΩ
100 mΩ
1 µΩ
10 µΩ
100 µΩ
1 mΩ
10 mΩ
100 mΩ
1 Ω
10 Ω
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
1 GΩ
100 µΩ/Ω
80 µΩ/Ω
50 µΩ/Ω
12 µΩ/Ω
12 µΩ/Ω
12 µΩ/Ω
25 µΩ/Ω
15 µΩ/Ω
2.0 µΩ/Ω
1.0 µΩ/Ω
0.80 µΩ/Ω
0.70 µΩ/Ω
0.25 µΩ/Ω
0.50 µΩ/Ω
1.3 µΩ/Ω
3.0 µΩ/Ω
7.0 µΩ/Ω
11 µΩ/Ω
100 µΩ/Ω
80 µΩ/Ω
50 µΩ/Ω
10 µΩ/Ω
10 µΩ/Ω
10 µΩ/Ω
25 µΩ/Ω
15 µΩ/Ω
2.0 µΩ/Ω
1.0 µΩ/Ω
0.80 µΩ/Ω
0.70 µΩ/Ω
0.25 µΩ/Ω
0.25 µΩ/Ω
0.25 µΩ/Ω
0.25 µΩ/Ω
0.25 µΩ/Ω
1.0 µΩ/Ω
1.5 µΩ/Ω
4.5 µΩ/Ω
7.5 µΩ/Ω
12 µΩ/Ω
Ratio metric comparison against
reference high power resistance
standards and shunts, using
Guildline 6622A current
comparator bridges and
associated range extenders up to
1000 A
Ratio metric comparison against
reference resistance standards,
using Guildline 6622A current
comparator bridges up to 1000
V and associated range
extenders up to 100 A
Ratio metric comparison against
reference high powers resistance
standards and shunts, using
Guildline 6622A current
comparator bridges and range
extenders up to 1000 A
Ratio metric comparison against
reference resistance standards,
using interchange where 1:1
transfers exist on Guildline
6622A current comparator
bridges (up to 1000 V) and
associated range extenders up to
300 A | ||
| 3 | - | electrical resistance standardsShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2906.01) 3/24/2026 Page 3 of 5
Parameter/Equipment
Range
CMC2, 3, 4, 6 (±)
Comments
High Resistance –
Generate and Measure
Resistance Ranges
Fixed Decade Points
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
100 MΩ to 1 GΩ
(1 to 10) GΩ
(10 to 100) GΩ
100 GΩ to 1 TΩ
(1 to 10) TΩ
(10 to 100) TΩ
100 TΩ to 1 PΩ
(1 to 10) PΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
1 GΩ
10 GΩ
100 GΩ
1 TΩ
10 TΩ
100 TΩ
1 PΩ
10 PΩ
50 µΩ/Ω
12 µΩ/Ω
12 µΩ/Ω
15 µΩ/Ω
50 µΩ/Ω
70 µΩ/Ω
100 µΩ/Ω
190 µΩ/Ω
800 µΩ/Ω
0.80 %
8.0 %
50 µΩ/Ω
12 µΩ/Ω
12 µΩ/Ω
12 µΩ/Ω
15 µΩ/Ω
50 µΩ/Ω
70 µΩ/Ω
100 µΩ/Ω
190 µΩ/Ω
800 µΩ/Ω
0.80 %
8.0 %
Substitution of electrical
resistance standards, using
Guildline Teraohmmeter
6520 or 6530-XPR/C
The uncertainties are based
on the typical applied
voltage is 1000 V for
values 100 GΩ to 10 PΩ,
100 V for 10 GΩ, 10 V for
1 GΩ and 1 V for 100 MΩ
down to 100 kΩ.
Voltages outside the typical
are available for most
resistance values note that
the uncertainties outside the
typical voltages at high
values may also depend on
applied voltage (1 V to
1000 V)
The above conditions are
applied in both fixed
decade points and
resistance ranges.
DC Voltage –
Generate
(10 to 220) mV
200 mV to 2.2 V
(2.2 to 11) V
(11 to 22) V
(22 to 220) V
(220 to 1100) V
15 µV/V + 0.51 µV
8.8 µV/V + 0.90 µV
6.3 µV/V + 4.0 µV
6.3 µV/V + 6.0 µV
9.0 µV/V + 60 µV
13 µV/V + 600 µV
Generate voltage using
calibrator: Fluke 5720A
DC Voltage – Measure
(10 to 200) mV
200 mV to 2 V
(2 to 20) V
(20 to 200) V
(200 to 1000) V
7.6 µV/V + 0.12 µV
5.0 µV/V + 0.51 µV
5.0 µV/V + 5.1 µV
7.5 µV/V + 51 µV
7.6 µV/V + 610 µV
Direct measurement of
voltage using DMM: Fluke
8508A | - |
| 4 | - | DC CurrentShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2906.01) 3/24/2026 Page 4 of 5
Parameter/Equipment
Range
CMC2, 3, 4, 6 (±)
Comments
DC Current – Generate
100 fA
1 pA
10 pA
100 pA
1 nA
10 nA
100 nA
1 µA
10 µA
2.0 %
0.20 %
0.20 %
500 µA/A
300 µA/A
300 µA/A
150 µA/A
150 µA/A
150 µA/A
Generate low current
through calibrator
(Fluke 5720A) DC
voltage stimulus over
reference resistance
value
Uncertainties are based
on applied voltage and
resistance value used
DC Current – Measure
(0.1 to 1) pA
(1 to 10) pA
(10 to 100) pA
100 pA to 1 nA
(1 to 10) nA
(10 to 100) nA
100 nA to 1 µA
(1 to 10) µA
10 µA to 100 nA
100 nA to 1 µA
(1 to 10) µA
(10 to 100) µA
100 µA to 1 mA
(1 to 10) mA
(10 to 100) mA
100 mA to 1 A
3.0 %
0.50 %
0.10 %
650 µA/A
400 µA/A
250 µA/A
200 µA/A
150 µA/A
39 µA/A + 50 pA
29 µA/A + 50 pA
29 µA/A + 110 pA
28 µA/A + 900 pA
28 µA/A + 6 nA
28 µA/A + 60 nA
44 µA/A + 600 nA
130 µA/A + 11 µA
Direct measurement of
low DC current using
Teraohmeter Guildline
6520 or 6530-XPR/C
Direct measurement of
DC current using
DMM:
Keysight/Agilent 3458A
_______________
1 This laboratory offers commercial calibration service.
2 Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a
laboratory can achieve within its scope of accreditation when performing more or less routine calibrations
of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded
uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of
k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be
greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Where ranges are not specified, CMC stated is for the cardinal points only.
4 In the statement of CMC, the value is defined as the percentage of reading. | - |
| 5 | - | ResistorsShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2906.01) 3/24/2026 Page 5 of 5
5 The stated uncertainties refer to negligible power dissipation. Resistors with significant power dissipation
can be measured at voltages up to 1000 V and currents up to 1000 A with uncertainties in the range (3 to
300) x 10-6. Resistors of modest dimensions suitable for oil immersion can be measured in the range from
(20 to 28) °C. Resistors which are not oil immersible, depending on dimensions, can be measured over a
temperature range (20 to 30) °C. Specific values are those which fall within ±0.5 % of the stated values.
6 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification.
7 This scope meets A2LA’s P112 Flexible Scope Policy. | - |
| 6 | ISO/IEC 17025:2017 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Other Locations
2 other sites in this organization
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.