Gulf Equipment Metrology Services (GEMS)
PO Box 8239 Bldg. 4148 cross road 118, Support Industrial Area 1, Al-Jubail Industrial City, 31961, Saudi Arabia, Al-Jubail Industrial City, Saudi Arabia
- Scope lines
- 7
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 7 scope lines.
A2LA certificate 3097.01 · 7 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3097.01) 05/30/2025 Page 1 of 6
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
GULF EQUIPMENT METROLOGY SERVICES (GEMS)
P.O Box 8239, Building No 4148, Cross Road 118, Support Industrial Area 1
Al-Jubail Industrial City, Saudi Arabia 31961
Jowell Liwanag Phone: +966 13 341 9451
CALIBRATION
Valid To: January 31, 2027 Certificate Number: 3097.01
In recognition of the successful co mpletion of the A2LA evaluation process (including an assessment of
the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory to perform the following calibrations
1, 6:
I. Acoustical
Parameter/Equipment
Range
CMC
2 (±)
Comments
Sound Level Calibrators,
Fixed Points
94 dB, 104 dB, & 114 dB
0.61 dB
3M DL-1
characterized w/ B&K
model 4226
Parameter/Range
Frequency
CMC2 (±)
Comments
Acoustical Levels –
Sound Level Meters, Noise
Dosimeters
94 dB, 104 dB, &
114 dB
31.5 Hz to 4 kHz, Octave
Step
8 kHz
(12.5 to 16) kHz
0.42 dB
0.5 dB
0.7 dB
B&K model 4226
multifunction acoustic
calibrator | ||
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3097.01) 05/30/2025 Page 2 of 6
II. Electrical – DC/Low Frequency
Parameter/Equipment
Range
CMC2, 5 (±)
Comments
Electrical Simulation
Calibration of
Thermocouple Devices
3
– Generate & Measure
Type B
Type C
Type E
Type J & L
Type K
Type N
Type R
Type S
Type T & U
(600 to 1820) °C
(0 to 2316) °C
(-250 to 1000) °C
(-210 to 1200) °C
(-200 to 1372) °C
(-200 to 1300) °C
(0 to 1767) °C
(0 to 1767) °C
(-250 to 400) °C
0.0063mV
0.0056 mV
0.0059 mV
0.0055 mV
0.0061 mV
0.0057 mV
0.0064 mV
0.0064 mV
0.0056 mV
Fluke 5522A
III. Fluid Quantities
Parameter/Equipment
Range
CMC
2, 4, 5 (±)
Comments
Density (Specific
Gravity)
(0 to 3) g/cm
3
0.0002 g/cm3
Mettler Toledo model DE45
density meter
Air Velocity –
Anemometers, Thermal,
Pitot, Vane Style &
Similar Equipment
(100 to 1000) fpm
(1000 to 9000) fpm
2.4 % rdg
1.2 % rdg
Omega Engineering model
WT4001-D-110V wind
tunnel w/ electronics box
WT 4401-D | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3097.01) 05/30/2025 Page 3 of 6
Parameter/Equipment
Range
CMC24, 5 (±)
Comments
Flow (Air)3 – Measure
(0 to 5) L/min
(0.23 to 0.6) CFM
(0.6 to 1.2) CFM
(1.2 to 1.8) CFM
(1.8 to 2.33) CFM
(0.85 to 2.0) CFM
(2.0 to 4.0) CFM
(4.0 to 6.0) CFM
(6.0 to 8.5) CFM
(3.79 to 10.0) CFM
(10.0 to 20.0) CFM
(20.0 to 30.0) CFM
(30.0 to 37.9) CFM
0.31 L/min
0.019 CFM
0.019 CFM
0.015 CFM
0.019 CFM
0.077 CFM
0.57 CFM
0.58 CFM
0.11 CFM
0.37 CFM
0.37 CFM
0.78 CFM
0.88 CFM
Aalborg XFM17
Brook Instruments
flowmeter models:
1110CJ61CUHAA
1110CK61CUHAA
1110CM61CUHAA
IV. Mechanical
Parameter/Equipment
Range
CMC2, 5 (±)
Comments
Vibration General
Purpose – Sensitivity
Magnitude
Accelerometer
Vibration Meters
(20 to 10 000) Hz
(20 to 10 000) Hz
7.3 % rdg
2 % rdg
3120A accelerometer &
vibration controller.
Agate AT-2050 | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3097.01) 05/30/2025 Page 4 of 6
Parameter/Equipment
Range
CMC2, 4, 5 (±)
Comments
Volumetric (Pipettes,
Flask, Burets, Cylinders)
(2 to 10) μl
(> 10 to 20) μl
(> 20 to 100) μl
(> 100 to 200) μl
(> 200 to 500) μl
(> 500 to 1000) μl
(> 1000 to 2000) μl
(> 2000 to 5000) μl
(> 5000 to 10 000) μl
(> 10 000 to 20 000) μl
(20 to 50) mL
(> 50 to 200) mL
(> 200 to 500) mL
> 500 mL to 1 L
(> 1 to 2) L
0.033 μl
0.048 μl
0.30 μl
0.40 μl
0.6 μl
1.3 μl
2.4 μl
5.9 μl
13 μl
24 μl
3.7 μl
26 μl
40 μl
61 μl
83 μl
Gravimetric measurement
using balances
Pressure – Measure &
Measuring Equipment
3
Hydraulic
Pneumatic
Vacuum
(6000 to 20 000) psig
(60 to 20 000) psig
(20 000 to 39 000) psig
(-13 to 3000) psig
(-15 to 15) InH
2O
(0.2 to 25) psig
(2 to 1000) psig
(0 to –13) psia
0.015 %
0.0035 %
9.2 psi
0.27 psi
0.0087 InH
2O +
0.01 %
0.0011 %
0.0027 %
0.002 psi
Fluke PPCH-140M
A140Ms/A40Ms Ruska
model 2485-930S
deadweight tester
High pressure pump w/
Additel reference gage
PM500-BG20M w/ Fluke
2271A
PM500-BG3.5K / Fluke
2271A
Fluke model 2465 gas
piston gage
Wika CPG2500 digital
pressure gauge | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3097.01) 05/30/2025 Page 5 of 6
V. Time & Frequency
Parameter/Equipment
Range
CMC2, 7 (±)
Comments
Frequency – Measuring
Equipment
(5 & 10) MHz
0.1 Hz to 20 MHz
20 MHz to 50 MHz
50 MHz to 31.8 GHz
5.9 x 10
-10 Hz
5.8 x10
-6 Hz
6.0 x 10-6 Hz
8.9 x 10-6 Hz
Symmetricom 5071A
cesium beam frequency
standard w/ Fluke 910R
GPS
HP 3325A/HP 5345A
HP 3325A opt 001/ HS 20/
54610, Agilent E8257D
opt 007, 1E1,1EUM 532,
U01, UNX
Frequency – Measure
(0 to 400) MHz
(0.4 to 1.6) GHz
(1.5 to 26) GHz
3.0 x 10
-7 Hz
17 x 10-7 Hz
14 x 10-7 Hz
HP 5345A
HP 5345A/ HP 5355A
HP 5345A/HP 5355A/ HP
5356B
Stop Watches
(0.15 to 24) hr
0.33 s
Agilent 33220A, 53132A
w/ Symmetricom 5071A
Timers3
(0.15 to 1) hr
13 ms
Agilent 33220A, 53132A
w/ Symmetricom 5071A
____________________________________________
1 This laboratory offers commercial calibration and field calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 In the statement of CMC, the value is defined as the percentage of reading, unless otherwise noted. | ||
| 6 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 3097.01) 05/30/2025 Page 6 of 6
5 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges
indicated for the listed measurement parameter.
6 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 7 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.