MSI-Viking Gage, LLC.
7044 Fairfield Business Center, 45014, Fairfield, OH, United States
Visit website- Scope lines
- 17
- A2LA certificate
- 1
- Sites in our directory
- 3
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 17 scope lines.
A2LA certificate 1387.04 · 17 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 1 of 17
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
MSI-VIKING GAGE, LLC.
7044 Fairfield Business Center
Fairfield, OH 45014
Brian Bunnell Phone: 513 829 0900
CALIBRATION
Valid To: August 31, 2026 Certificate Number: 1387.04
In recognition of the successful completion of the A2LA evaluation process (including an assessment of the
organization’s compliance with R205 - A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory at the location listed below to perform the following calibrations
1, 9:
I. Dimensional
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Bore Micrometers
Up to 8 in
(16 + 4.9D) μin
Master rings
Dial Bore Gages
(Bore Gage w/ Indicator)
Up to 4 in
(27 + 1.6L) μin
(17 + 4.5L) μin
ULM
Master rings
Calipers3
50 μin Resolution:
100 μin Resolution
500 μin Resolution
Up to 6 in
(6 to 12) in
(12 to 24) in
(24 to 48) in
Up to 6 in
(6 to 12) in
(12 to 24) in
(24 to 48) in
Up to 6 in
(6 to 12) in
(12 to 24) in
(24 to 48) in
(48 to 80) in
(35 + 0.58L) μin
(42 + 1.9L) μin
(42 + 2.6L) μin
(34 + 3.2L) μin
(61 + 0.34L) μin
(64 + 1.4L) μin
(62 + 2.2L) μin
(49 + 3.1L) μin
(290 + 0.073L) μin
(290 + 1.9L) in
(290 + 0.75L) μin
(270 + 1.5L) μin
(250 + 2.0L) μin
Gage blocks | to 80 in |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 2 of 17
Parameter/Equipment
Range CMC2, 4 (±) Comments
Calipers3 (cont)
0.001 in Resolution
50uin Res
100 uin Res
500 uin Res
0.001 in Res
Up to 6 in
(6 to 12) in
(12 to 24) in
(24 to 48) in
(48 to 80) in
Up to 12 in
Up to 12 in
Up to 12 in
Up to 12 in
(580 + 0.037L) μin
(580 + 0.19L) μin
(580 + 0.39L) μin
(570 + 0.81L) μin
(550 + 1.2L) μin
(240 + 1.5L) μin
(250 + 0.15L) μin
(370 + 0.98L) μin
(620 + 0.059L) μin
Gage blocks
Caliper checker
Caliper Checkers
Up to 12 in
(90 + 0.92L) μin
Electronic indicator
amplifier, gage blocks
Caliper Gage3 – Internal,
External
Up to 8.0 in
(18 + 1.9L) μin
Gage blocks
Depth Step Gages
(0.5 to 11.5) in
(13 + 2.1L) μin
Electronic indicator
amplifier, gage blocks
Cylindrical Ring Gages3
(0.425 to 18) in
(6.1 + 4.1D) μin
Universal length
machine, master rings
Disc, Plug & Pin Gages3
Up to 20 in
(12 + 1.0D) μin
Universal length
machine
Electronic Indicator
Amplifier3
Up to 0.02 in
9.3 μin
Gage blocks, optical
flat | ||
| 3 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 01/31/2025 Page 3 of 17
Parameter/Equipment Range CMC2, 4 (±) Comments
Attribute Gage
Up to 12 in
Up to 27 in
Ext. Measurement
Up to 11.5 in
Int. Measurement
Up to 6 in
(130 + 5.4L) μin
(12 + 1.0D) μin
(12 + 2.1L) μin
0.000 30 in
Vision system
Universal length machine
Electronic indicator
amplifier, gage blocks
Rings, plugs, hand tools
Gage Blocks – Length Only
Up to 4 in
(4.1 + 1.8L) μin
Gage block comparator,
master gage blocks
Height Gages3
Up to 40 in
(290 + 0.94L) µin
Gage blocks
Indicators/LVDTs3
Up to 12 in
0.000 01 in
0.000 05 in
0.0001 in
0.0005 in
0.001 in
(17 + 3.1L) μin
(41 + 2.4L) μin
(59 + 1.3L) μin
(290 + 0.3L) μin
(580 + 0.85L) μin
ULM
Length Standards3
Up to 20 in
(14 + 1.4D) μin
Universal length machine
Levels3
Up to 12 in
140 μin
Surface plate, sine bar,
gage blocks | to 12 in |
| 4 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 4 of 16
Parameter/Equipment Range CMC2, 4 (±) Comments
Micrometers3 –
Groove
Depth
Inside
Specialty Micrometers
Up to 12 in
(12 to 24) in
(24 to 48) in
Up to 4 in
Up to 16 in
Up to 20 in
Up to 4 in
(7.9 + 3.3L) μin + 0.6R
(11 + 3.1L) μin + 0.6R
(3.7 – 3.1L) μin + 0.6R
(91 + 3.9L)µin + 0.6R
(580 + 3.3L) μin
(32 + 1.8D) μin + 0.6R
(0.96 + 16L) μin
Gage blocks
ULM
Indicator Calibrators3 (Mic
Head Type)
Up to 1 in
(22 + 4.2L) μin
ULM
Plain Pins Class ZZ3
Up to 2 in
37 μin
High accuracy
micrometer
Protractor –
Bevel
Digital3
Up to 180°
Up to 90°
11 arcsec or 0.003°
27 arcsec
Vision system
Sine plate, gage
blocks
Snap Gage3 –
Flatness of Anvils
Size
Up to 1 in
(Up to 20) in
3.3 μin
(57 + 1.1L) μin
Optical flat
Gage blocks
Steel Rules3
Up to 96 in
Up to 39 in
(300 + 0.27L) μin
0.0046 in
Vision system
Direct comparison | to 180 ° |
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 5 of 16
Parameter/Equipment Range CMC2, 4 (±) Comments
Steel Tapes3
Up to 100 ft
0.05 in
Direct comparison
Thickness & Feeler Gage3
Up to 2 in
(12 + 1.0D) μin
Universal length
machine (ULM)
Thread Measuring Wires3
(4 to 80) TPI
8.1 μin
Universal length
machine, gage blocks,
master pins
Angle Blocks
Up to 90°
11 arcsec or 0.003°
Vision system
Screw Thread Micrometer
Standards
Up to 12 in
(130 + 5.4L) μin
Vision system
Thread Plugs3 –
Major Diameter
Pitch Diameter
Tapered Thread Plug Gage –
Pitch Diameter
Step
Up to 8 in
Up to 80 TPI
Up to 4 in
Up to 1 in
(7.6 + 1.3D) μin
(68 + 1.2D) μin
(41 + 4.5D) μin
45 μin
Universal length
machine
Universal length
machine, thread
measuring wires
Universal length
machine, gage blocks
Gage blocks
Universal Measuring
Machines/Bench Micrometers3
Up to 12 in
(8 + 2.4L) μin
Gage blocks
Laser Micrometers3
Up to 1 in
(27 + 1.9D) µin
Class XX master pins
Radius Gages
Up to 12 in
(140 + 5.2L) μin
Vision system | ||
| 6 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 6 of 16
Parameter/Equipment Range CMC2, 4 (±) Comments
Surface Finish Testers3 – Ra,
Rz
(2 to 500) µin
2.5 µin
Master surface finish
patch
Surface Finish Standards3 –
Ra, Rz
(2 to 500) µin
6.2 µin
Direct comparison to
master surface patch
II. Dimensional Testing/Calibration5
Parameter/Equipment Range CMC2, 4, 8 (±) Comments
Surface Finish – Measure
(2 to 500) µin
5.2 µin
Master surface finish
patch
Flatness – Measure
Up to 20 in
Up to 2 in
31 μin
4.8 μin
Electronic indicator
Optical flat
Parallelism – Measure
Up to 20 in
14 μin
Electronic indicator
amplifier
Radius – Measure
Up to 12 in
Up to 304.8 mm
(140 + 5.2L) μin
(0.44 + 0.0011L) μm
Vision system
Straightness – Measure
Up to 12 in
22 μin
Electronic indicator
amplifier
Video Measurement5 –
X, Y Axis
Up to 12 in
Up to 304.8 mm
(140 + 5.2L) μin
(0.44 + 0.0011L) μm
Vision system
Angular Measurements5
Up to 360°
38 arcsec or 0.010°
Optical comparator | 2 to 500 µin |
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 7 of 16
III. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC2, 7 (±) Comments
DC Voltage – Generate3
(0 to 329.9999) mV
(0 to 3.2999) V
(0 to 32.9999) V
(33 to 329.9999) V
(100 to 1020) V
24 uV/V + 3 uV
63 uV/V + 5 uV
60 uV/V + 50 uV
66 uV/V + 500 µV
66 uV/V + 1500 µV
Fluke 5500A
DC Voltage – Measure3
(0 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
6.2 µV/V + 0.30 µV
3.2 µV/V + 0.30 µV
1.1 µV/V + 0.50 µV
3.4 µV/V + 30 µV
6.6 µV/V + 0.10 mV
Agilent 3458A
DC High Voltage –
Measure
(1 to 9) kV
0.000 54 V/V – 0.12 V
High voltage meter
DC Current – Generate3
(0 to 3.299 99) mA
(0 to 33) mA
(0 to 330) mA
(0 to 2.2 A
(0 to 11) A
0.016 % + 0.05 uA
0.013 % + 0.25 uA
0.015 % + 3.3 uA
0.036 % + 44 uA
0.072 % + 330 uA
Fluke 5500A
DC Current – Measure3
(10 to 100) µA
100 µA to 1 mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
25 µA/A + 0.8 nA
23 µA/A + 5.0 nA
23 µA/A + 50 nA
37 µA/A + 0.50 µA
0.013 % + 10 µA
Agilent 3458A | ||
| 8 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 8 of 16
Parameter/Range Frequency CMC2, 7 (±) Comments
AC Voltage –
Generate3
(1 to 32.999) mV
(33 to 329.999) mV
(0.33 to 3.29999) V
(3.3V to 32.9999) V
(33 to 329.999)
(330 to 1020) V
(10 to 45) Hz
45Hz to 10kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45Hz to 10kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45Hz to 10kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45Hz to 10kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
45Hz to 1kHz
(1 to 10) kHz
(10 to 20) kHz
45Hz to 1kHz
(1 to 5) kHz
(5 to 10) kHz
0.42 % + 20 µV
0.18 % + 20 µV
0.024 % + 20 µV
0.30 % + 20 µV
0.42 % + 33 µV
1.2 % + 60 µV
0.30 % + 50 µV
0.06 % + 20 µV
0.12 % + 20 µV
0.19 % + 40 µV
0.29 % + 170 µV
0.84 % + 330 µV
0.18 % + 250 µV
0.04 % + 60 µV
0.096 % + 60 µV
0.16 % + 300 µV
0.29 % + 1700 µV
0.60 % + 3300 µV
0.18 % + 2500 µV
0.048 % + 600 µV
0.096 % + 2600 µV
0.23 % + 5000 µV
0.29 % + 1.7 mV
0.06 % + 6.6 mV
0.096 % + 15 mV
0.10 % + 33 mV
0.06 % + 80 mV
0.24 % + 100 mV
0.24 % + 500 mV
Fluke 5500A | ||
| 9 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 9 of 16
Parameter/Range Frequency CMC2, 7 (±) Comments
AC Voltage – Measure3
Up to 10 mV
(10 to 100) mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 750) V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 1) MHz
(1 to 2) MHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(1 to 2) MHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 to 1 MHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.030 % + 3.0 µV
0.020 % + 1.1 µV
0.040 % + 1.1 µV
0.11 % + 1.1 µV
0.51 % + 1.1 µV
4.1 % + 2.0 µV
0.0080 % + 4.0 µV
0.0080 % + 2.0 µV
0.016 % + 2.0 µV
0.031 % + 2.0 µV
0.081 % + 2.0 µV
0.34 % + 10 µV
1.1 % + 10 µV
1.6 % + 10 µV
0.0090 % + 40 µV
0.0070 % + 20 µV
0.017 % + 20 µV
0.036 % + 20 µV
0.82 % + 20 µV
0.31 % + 100 µV
1.1 % + 100 µV
1.6 % + 100 µV
0.0080 % + 0.40 mV
0.0080 % + 0.20 mV
0.015 % + 0.20 mV
0.031 % + 0.20 mV
0.080 % + 0.20 mV
0.30 % + 1.0 mV
1.0 % + 1.0 mV
0.030 % + 4.0 mV
0.030 % + 2.0 mV
0.030 % + 2.0 mV
0.040 % + 2.0 mV
0.13 % + 2.0 mV
0.040 % + 4.0 mV
0.040 % + 2.0 mV
0.060 % + 2.0 mV
0.14 % + 2.0 mV
0.40 % + 2.0 mV
Agilent 3458A opt
002 | ||
| 10 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 10 of 16
Parameter/Range Frequency CMC2, 7 (±) Comments
AC High Voltage –
Measure
(1 to 10) kV
60 Hz
0.0015 V/V + 0.34 V
High voltage meter
AC Current – Measure3
(5 to 100) μA
(1 to 100) mA
(0.1 to 1) A
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 10 kHz
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 1 kHz
(5 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
0.57 % rdg + 0.030 % rng
0.18 % rdg + 0.030 % rng
0.07 % rdg + 0.030 % rng
0.07 % rdg + 0.030 % rng
0.55 % rdg + 0.020 % rng
0.18 % + 20 μA
0.080 % + 20 μA
0.043 % rdg + 0.02 % rng
0.090 % rdg + 0.02 % rng
0.50 % + 40 μA
0.70 % + 150 μA
0.47 % rdg + 0.020 % rng
0.19 % + 200 μA
0.10 % + 200 μA
0.13 % rdg + 0.020 % rng
0.12 % rdg + 0.020 % rng
1.2 % + 40 μA
Agilent 3458A opt
002
AC Current – Generate3
(0.029 to 0.329 99) mA
(0.33 to 3.2999) mA
(3.3 to 32.999) mA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.29 % rdg + 0.15 μA
0.15 % rdg + 0.15 μA
0.15 % rdg + 0.25 μA
0.47 % rdg + 0.15 μA
0.29 % rdg + 0.15 μA
0.24 % rdg + 0.3 μA
0.12 % rdg + 0.3 μA
0.12 % rdg + 0.3 μA
0.24 % rdg + 0.3 μA
0.70 % rdg + 0.3μA
0.24 % rdg + 0.03 mA
0.12 % rdg + 0.03 mA
0.11 % rdg + 0.03 mA
0.24 % rdg + 0.03 mA
0.70 % rdg + 0.03 mA
Fluke 5500A | 1 to 10 kV |
| 11 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 11 of 16
Parameter/Range
Frequency CMC2, 7 (±)
Comments
AC Current – Generate3
(cont)
(33 to 329.99) mA
(0.33 to 2.19999) A
(2.2 to 11) A
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(45 to 65) Hz
(65 to 500) Hz
500 Hz to 1 kHz
0.24 % rdg + 0.03 mA
0.12 % rdg + 0.03 mA
0.11 % rdg + 0.03 mA
0.24 % rdg + 0.03 mA
0.70 % rdg + 0.03 mA
0.24 % rdg + 0.3 A
0.12 % rdg + 0.3 A
0.87 % rdg + 0.3 A
0.070 % rdg + 0.002 A
0.12 % rdg + 0.002 A
0.38 % rdg + 0.002 A
Fluke 5500A
Parameter/Equipment
Range CMC2, 7 (±)
Comments
Resistance – Generate3 (0 to 10.999) Ω
(11 to 32.999) Ω
(33 to 109.999) Ω
(110 to 329.999) Ω
(0.330 to 1.099 99) kΩ
(1.1 to 3.299 99) kΩ
(3.3 to 10.9999) kΩ
(11 to 32.9999) kΩ
(33 to 109.999) kΩ
(110 to 329.999) kΩ
(0.33 to 1.099 99) MΩ
(1.1 to 3.299 99) MΩ
(3.3 to 10.9999) MΩ
(11 to 32.9999) MΩ
(33 to 109.999) MΩ
0.014 % + 93 mΩ
0.014 % + 17 mΩ
0.14 % + 17 mΩ
0.11 % + 17 mΩ
0.12 % + 66 mΩ
0.46 % + 55 mΩ
0.12 % + 660 mΩ
0.11 % + 690 mΩ
0.13 % + 6.9 Ω
0.15 % + 6.7 Ω
0.16 % + 87 Ω
0.20 % + 50 Ω
0.70 % + 650 Ω
1.4 % + 550 Ω
5.8 % + 6.5 kΩ
Fluke 5500A, 4-
wire
Fluke 5500A, 2-
wire
Resistance – Measure3 (0 to 10) Ω
(10 to 100) Ω
100 Ω to 1 kΩ
(1 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
15 µΩ/Ω + 5 µΩ
15 µΩ/Ω + 50 µΩ
19 µΩ/Ω + 50 µΩ
52 µΩ/Ω + 50 mΩ
50 µΩ/Ω + 60 µΩ
50 µΩ/Ω + 150 µΩ
0.070 % + 5.0 kΩ
Agilent 3458A opt
002 | to 109.999 MΩ |
| 12 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 12 of 16
Parameter/Equipment Range CMC2, 7 (±) Comments
Capacitance – Generate3
(0.33 to 0.499) nF
(0.5 to 1.0999) nF
(1.1 to 3.2999) nF
(3.3 to 10.999) nF
(11 to 32.999) nF
(33 to 109.99) nF
(110 to 329.99) nF
(0.33 to 1.0999) µF
(1.1 to 3.2999) µF
(3.3 to 10.999) µF
(11 to 32.999) µF
(33 to 109.99) µF
(110 to 329.99) µF
330 µF to 1.1 mF
0.58 % + 0.012 nF
0.58 % + 0.012 nF
0.61 % + 0.011 nF
0.58 % + 0.012 nF
0.33 % + 0.11 nF
0.29 % + 0.13 nF
0.56 % + 1.2 nF
0.29 % + 1.2 nF
0.41 % + 3.5 nF
0.41 % + 12 nF
0.47 % + 35 nF
0.58 % + 120 nF
0.94 % + 350 nF
1.2 % + 350 nF
Fluke 5500A
Electrical Calibration of
Thermocouples &
Thermocouple Indicating
Systems3 –
Type B
Type C
Type E
(600 to 800) °C
(800 to 1000) °C
(1000 to 1550) °C
(1550 to 1820) °C
(0 to 150) °C
(150 to 650) °C
(650 to 1000) °C
(1000 to 1800) °C
(1800 to 2316) °C
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
0.59 °C
0.45 °C
0.39 °C
0.39 °C
0.37 °C
0.32 °C
0.38 °C
0.59 °C
0.97 °C
0.57 °C
0.19 °C
0.16 °C
0.18 °C
0.24 °C
Fluke 5500A | 0.33 to 1.1 mF |
| 13 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 13 of 16
Parameter/Equipment Range CMC2 (±) Comments
Electrical Calibration of
Thermocouples &
Thermocouple Indicating
Systems3 – (cont)
Type J
Type K
Type L
Type N
Type R
Type S
Type T
Type U
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
(-200 to -100) °C
(-100 to 800) °C
(800 to 900) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 410) °C
(410 to 1300) °C
(0 to 250) °C
(250 to 400) °C
(400 to 1000) °C
(1000 to 1767) °C
(0 to 250) °C
(250 to 1000) °C
(1000 to 1400) °C
(1400 to 1767) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 120) °C
(120 to 400) °C
(-200 to 0) °C
(0 to 600) °C
0.31 °C
0.19 °C
0.16 °C
0.20 °C
0.27 °C
0.37 °C
0.21 °C
0.19 °C
0.30 °C
0.46 °C
0.43 °C
0.30 °C
0.20 °C
0.45 °C
0.27 °C
0.22 °C
0.21 °C
0.31 °C
0.70 °C
0.43 °C
0.40 °C
0.43 °C
0.58 °C
0.44 °C
0.44 °C
0.51 °C
0.70 °C
0.28 °C
0.19 °C
0.16 °C
0.65 °C
0.31 °C
Fluke 5500A | ||
| 14 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 14 of 16
Parameter/Equipment
Range
CMC2 (±)
Comments
Electrical Calibration of RTD
Indicating Systems3 –
Pt 395, 100 Ω
Pt 3926, 100 Ω
Pt 3916, 100 Ω
Pt 385, 200 Ω
Pt 385, 500 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 630) °C
(630 to 800) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 630) °C
(-200 to -190) °C
(-190 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
0.060 °C
0.060 °C
0.090 °C
0.11 °C
0.12 °C
0.14 °C
0.26 °C
0.060 °C
0.060 °C
0.090 °C
0.10 °C
0.12 °C
0.14 °C
0.29 °C
0.050 °C
0.060 °C
0.070 °C
0.090 °C
0.090 °C
0.10 °C
0.12 °C
0.26 °C
0.050 °C
0.050 °C
0.050 °C
0.060 °C
0.15 °C
0.17 °C
0.18 °C
0.19 °C
0.050 °C
0.070 °C
0.070 °C
0.080 °C
0.10 °C
0.10 °C
0.10 °C
0.12 °C
Fluke 5500A | ||
| 15 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 15 of 16
Parameter/Equipment Range CMC2 (±) Comments
Electrical Calibration of RTD
Indicating Systems3 (cont) –
Pt 385, 1000 Ω
PtNi 385, 120 Ω (Nil20)
Cu 427, 10 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(-100 to 260) °C
0.040 °C
0.040 °C
0.050 °C
0.070 °C
0.070 °C
0.090 °C
0.090 °C
0.27 °C
0.10 °C
0.10 °C
0.17 °C
0.35 °C
Fluke 5500A
IV. Mechanical
Parameter/Equipment Range CMC2, 6, 8 (±) Comments
Torque Wrenches3
(4 to 40) oz-in
(4 to 50) lbf-in
(30 to 400) lbf-in
(80 to 1000) lbf-in
20 to 250 lbf-ft
60 to 600 lbf-ft
0.23 % + 0.28 oz-inf
0.28 % + 0.083 lbf-in
0.4 % + 0.28 lbf-in
0.44 % + 0.55 lbf-in
0.46 % + 0.39 lbf-ft
0.28 % + 0.76 lbf-ft
Torque transducers
V. Time & Frequency
Parameter/Equipment Range CMC2, 6, 8 (±) Comments
Stop Watches/Time
Measurement3
Up to 24 hours
0.12 s
Quartz watch-clock
analyzer
Frequency – Measure
Up to 10 MHz
2.6 Hz
Agilent 3458A
________________________________________________________________ | ||
| 16 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1387.04) Revised 3/5/2025 Page 16 of 16
1 This laboratory offers commercial dimensional testing/calibration and on-site calibration services.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the Calibration and
Measurement Capability Uncertainty (CMC) found on the A2LA Scope. Allowance must be made for
aspects such as the environment at the place of calibration and for other possible adverse effects such as
those caused by transportation of the calibration equipment. The usual allowance for the uncertainty
introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own,
could result in the actual measurement uncertainty achievable on a customer’s site being larger than the
CMC.
4 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches
or millimeters. DL is the diagonal length in inches or millimeters. R is the resolution of the unit under
test. D is the numerical value of the nominal diameter of the device measured in inches or millimeters.
Pitch diameter is measured by the three-wire method. Major diameter is calibrated by direct measurement.
In the statement of CMC, T indicates temperature in degrees Celsius. In the statement of CMC, H is the
Humidity of the reading.
5 This laboratory meets R205 – Specific Requirements: Calibration Laboratory Accreditation Program for
the types of dimensional tests listed above and is considered equivalent to that of a calibration.
6 In the statement of CMC, a percent refers to a percentage of reading unless otherwise noted and IV stands
for indicated value.
7 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification.
8 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
9 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 17 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Other Locations
2 other sites in this organization
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.