Lyri Calibrations
#4 Jack Smith Plaza, Eight Mile Rock, Freeport, Grand Bahama, Bahamas, Freeport, Grand Bahama, Bahamas
Visit website- Scope lines
- 6
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 6 scope lines.
A2LA certificate 7608.01 · 6 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017 | - | - |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7608.01) 06/09/2026 Page 2 of 5
Parameter/Equipment
Range CMC2, 5 (±) Comments
DC Current – Generate &
Measure3
Up to 24 mA
0.02 % of rdg + 2
counts
Fluke 725 process
calibrator
Electrical Simulation of
Thermocouple3
Type J
Type K
Type T
(-200 to 0) °C
(0 to 1200) °C
(-200 to 0) °C
(0 to 1370) °C
(-200 to 0) °C
(0 to 400) °C
1.0 °C
0.70 °C
1.2 °C
0.80 °C
1.0 °C
0.80 °C
Fluke 725 process
calibrator
II. Fluid Quantities
Parameter/Equipment
Range CMC2, 4 (±) Comments
Air Velocity – Measuring
Equipment3
(0.40 to 10.00) m/s
(10.0 to 25.00) m/s
(80 to 1965) ft/min
(1965 to 4900) ft/min
0.5 meters/sec
0.75 meters/sec
100 ft/min
150 ft/min
Anemometer | ||
| 3 | - | Cardiac Functions; Ventilators; Anesthesia Machines; DefibrillatorsShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7608.01) 06/09/2026 Page 3 of 5
III. Mechanical
Parameter/Equipment
Range CMC2 (±) Comments
Cardiac Functions –
Measure3
ECG Tests
Pacemaker Signals
Arrhythmia Function
Blood Pressure Function
Cardiac Output
Fetal/Maternal Function
Pulse Oximeter
0.05 mV to 5 V
0.1 mV to 10 V / (30-240) bpm
(30 to 300) bpm
Up to 300 mmHg
Up to 20 L/min
(30 to 240) bpm
(70 – 100) % SpO2
0.15 % rdg + 5 µV
0.2 % rdg + 10 µV
0.1 bpm
0.5 mmHg
2 % rdg
0.1 bpm
1 % SpO2
Fluke PROSIM 3
PROSIM SPOT
SP02
Ventilators3
Flow
Volume
Pressure
Temperature
Humidity
Up to 200 slpm
Up to 100 L
(-0.8 to 10) bar
Up to 50 °C
Up to 100 % RH
2 % rdg
2 % rdg
1 % rdg
0.5 °C
3 % RH
VT900A/Vapor
Anesthesia Machines3
Flow
Volume
Pressure
O2 Concentration
Up to 200 slpm
Up to 100 L
(-0.8 to 10) bar
Up to 100 %
2 % rdg
2 % rdg
1 % rdg
1 % rdg
VT900A/Vapor
Defibrillators3
Energy
Voltage
Current
Pulse Width
(1 to 360) J
(20 to 5000) V
(0.4 to 100) A
(1 to 50) ms
1.5 % rdg + 0.3 J
1 % rdg + 2 V
2 % rdg + 0.1 A
0.1 ms
Impulse 7000DP | See scope document |
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7608.01) 06/09/2026 Page 4 of 5
IV. Thermodynamics
Parameter/Equipment
Range CMC2 (±) Comments
Thermometers3
(-200 to 200) °C
(100 to 1100) °C
1.1 °C
2.0 °C
Type J, K, T, bath
V. Time & Frequency
Parameter/Equipment
Range CMC2, 4 (±) Comments
Frequency – Measure3
(2 to 1000) CPM
(1 to 1100) Hz
(1 to 10) kHz
Sensitivity
0.05 % rdg + 1 count
0.05 % rdg + 1 count
0.05 % rdg + 1 count
1 V peak-to-peak
minimum
Fluke 725 process
calibrator
________________________________________
1 This laboratory offers commercial calibration service and field calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the la boratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the Calibration and Measurement
Capability Uncertainty (CMC) found on the A2LA Scope. Allowance must be made for aspects such as the
environment at the place of calibration and for other possible adverse effects such as those caused by
transportation of the calibration equipment. The usual allowance for the uncertainty introduced by the item
being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual
measurement uncertainty achievable on a customer’s site being larger than the CMC.
4 The instrument or material being calibrated is defined by the parameter. This indicates the laboratory is
capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed
measurement parameter. | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 7608.01) 06/09/2026 Page 5 of 5
5 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.
6 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 6 | ISO/IEC 17025:2017 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.