| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No.7011.01) Revised 01/30/2026 Page 1 of 2
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
XTRA ENVIRONMENTAL LLC
17 John Alden Circle
Bellingham, MA 02019
Gregory Brown Phone: 508 958 8022
CALIBRATION
Valid To: March 31, 2026 Certificate Number: 7011.01
In recognition of the successful co mpletion of the A2LA evaluation process (including an assessment of
the organization’s compliance with A2LA’s Calibration Program Requirements), accreditation is granted
to this laboratory to perform the following calibrations
1, 4:
I. Electrical– DC/Low Frequency
Parameter/Equipment
Range
CMC2 (±)
Comments
DC Voltage – Generate
3
(0 to 11.00) V
0.01 V
Process calibrators, Fluke 753
DC Resistance – Generate
3
(0 to 500) Ω
5.1 Ω
Process calibrators, Fluke 753
DC Current – Generate
3
(0.1 to 22) mA
0.12 mA
Process calibrators, Fluke 753
Electrical Simulation of
RTD Indicating Systems
3–
Pt 385, 100 Ω
(-200 to 250) °C
0.7 °C
Process calibrators, Fluke 753
Electrical Simulation of
Thermocouple Indicating
Stystems
3–
Type T
(-200 to 400) °C
1.2 °C
Process calibrators, Fluke 753 | -200 to 400 °C |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No.7011.01) Revised 01/30/2026 Page 2 of 2
II. Thermodynamics
Parameter/Equipment
Range
CMC2 (±)
Comments
Relative Humidity– Measure
3
(10 to 95) % RH
3.7 % RH
Vaisala HMP77/M-170
Temperature– Measure
3
Type T
(-80 to 200) °C
1.5 °C
Process calibrator Fluke 753
with calibrated thermocouples
_________________________________________
1 This laboratory offers commercial calibration service and field calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the small est uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3
Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expect ed to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as th e environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 3 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.