Registers refreshed 2026-11-08.See refresh dates
United States

X-Rite Incorporated

4300 44th St. SE, 49512, Grand Rapids, MI, United States

Visit website
A2LA 2108.01ISO/IEC 17025Register refreshed 2026-11-06

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free. No lab gets your details until you choose.
Scope lines
6
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 6 scope lines.

A2LA certificate 2108.01 · 6 lines

LineMethodMatrixRange
1

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2108.01) 12/10/2025 Page 1 of 5 SCOPE OF ACCREDITATION TO ISO 17025:2017 X-RITE INCORPORATED 4300 44 th Street SE Grand Rapids, MI 49512 Kevin Irwin Phone: 616 219 8681 CALIBRATION Valid To: December 31, 2027 Certificate Number: 2108.01 In recognition of the successful completion of the A2LA evaluation process (including an assessment of the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is granted to this laboratory to perform the following calibrations 1, 7: I. Optical Quantities Parameter/Equipment5 Range CMC2, 4 (±) Comments6 Spectrophotometers – Model 962, 964, (0/45 Optical Geometry) Model CI60, CI61, CI62, CI64, CI64 UV, CI4xxx Model CI7xxx 3 Model eXact (0/45 Optical Geometry) (400 to 420) nm (430 to 700) nm 400 nm (410 to 700) nm (360 to 400) nm (410 to 750) nm (400 to 410) nm (420 to 700) nm 1.7 % R 0.92 % R 0.95 % R 0.41 % R 0.83 % R 0.43 % R 1.8 % R 1.1 % R White reflection standards with: SPP X-it 962 SPP CI6x SPP CI7x SPP eXact
2

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2108.01) 12/10/2025 Page 2 of 5 Parameter/Equipment5 Range CMC2, 4 (±) Comments6 Spectrophotometers – (cont) Model eXact 2 (ETV) (0/45 Optical Geometry) Model MA9X (Aspecular Optical Geometry): -15 degree 15 degree 25 degree 45 degree 75 degree 110 degree (400 to 410) nm (420 to 700) nm (400) nm (410 to 700) nm (400 to 420) nm (430 to 700) nm (400) nm (410 to 700) nm (400) nm (410 to 700) nm (400) nm (410 to 700) nm (400) nm (410 to 700) nm 1.8 % R 1.1 % R 0.96 % R 0.99 % R 0.91 % R 0.78 % R 0.86 % R 0.75 % R 0.81 % R 0.70 % R 0.88 % R 0.80 % R 0.95 % R 0.90 % R White reflection standards with: SPP ETV SPP final 2 MA9x
3--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2108.01) 12/10/2025 Page 3 of 5 Parameter/Equipment5 Range CMC2, 4 (±) Comments6 Spectrophotometers – (cont) Model MA-x Series and OEM variants (TOP Spectro-photometer) (Aspecular Optical Geometry): 15 degree 25 degree 45 degree 75 degree 110 degree Model MA-Tx Series and OEM variants (KOH Spectro-photometer) (Aspecular Optical Geometry): -15 degree 15 degree 25 degree 45 degree 75 degree 110 degree (400 to 410) nm (420 to 700) nm 400 nm (410 to 700) nm 400 nm (410 to 700) nm (400 to 410) nm (420 to 700) nm (400 to 700) nm 400 nm (410 to 700) nm (400 to 420) nm (430 to 700) nm 400 nm (410 to 700) nm 400 nm (410 to 700) nm (400 to 420) nm (430 to 700) nm 400 nm (410 to 700) nm 1.2 % R 1.1 % R 1.2 % R 1.1 % R 1.1 % R 0.96 % R 1.2 % R 1.0 % R 1.2 % R 1.1 % R 1.0 % R 1.1 % R 0.85 % R 0.95 % R 0.80 % R 0.88 % R 0.73 % R 0.98 % R 0.78 % R 0.98 % R 0.92 % R White reflection standards with: Topaz service manual Kohinoor service manual
400 to 700 nm
4--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2108.01) 12/10/2025 Page 4 of 5 Parameter/Equipment5 Range CMC2, 4 (±) Comments6 Spectrophotometers – (cont) Model MA-x Series and OEM Variants (TOP II Spectro-photometer) (Aspecular Optical Geometry): -15 degree 15 degree 25 degree 45 degree 75 degree 110 degree (400 to 700) nm (400 to 700) nm 400 nm (410 to 700) nm 400 nm (410 to 700) nm (400 to 410) nm (420 to 700) nm (400 to 700) nm 1.6 % R 1.2 % R 1.1 % R 1.0 % R 1.1 % R 0.96 % R 1.1 % R 1.0 % R 1.2 % R White reflection standards with: Topaz II service manual Optical Radiation3 (Lightbooth) 2300 K (Horizon) 2856 K (Illuminate A) 20 K 22 K X-Rite LightSpex Pro radiometer _____________ 1 This laboratory offers commercial calibration service and field calibration services. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. 3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC. 4 In the statement of CMC, R is the reflectance.
400 to 700 nm
5

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2108.01) 12/10/2025 Page 5 of 5 5 The product families have the same optical geometries and calibration procedures. The features on the display and outputs are the only differences. 6 White reflection standards apply to the calibration of approximately 80 % reflectance on a neutral white ceramic. Color calibration is mainly based on inter-instrument agreement. 7 This scope meets A2LA’s P112 Flexible Scope Policy.
6ISO/IEC 17025:2017; R205--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

4300 44th St. SE, Grand Rapids, MI, 49512, US

42.96324, -85.66786

Open in Maps

Accreditation data refreshed from the A2LA register on .