Thermo Electron North America LLC operating as Unity Lab Services
4835 Decatur BLVD, Suite D, 46241, Indianapolis, IN, United States
Visit website- Scope lines
- 13
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 13 scope lines.
A2LA certificate 6843.01 · 13 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | - | CO₂ gas; Certified gas analyzer; Certified O₂ gasShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 1 of 12
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
THERMO ELECTRON NORTH AMERICA LLC OPERATING AS UNITY LAB SERVICES8
4835 Decatur Blvd, Suite D
Indianapolis, IN 46241
Martin Legault Phone: 613-796-1394
CALIBRATION
Valid to: December 31, 2027 Certificate Number: 6843.01
In recognition of the successful completion of the A2LA evaluation process (including an assessment of
the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory to perform the following calibrations1, 7, 8:
I. Chemical Quantities
Parameter/Equipment
Range
CMC2, 4, 9 (±)
Comments
CO₂ – Measuring
Equipment9
3 % CO2
7 % CO2
9 % CO2
0.025 % CO2
0.048 % CO2
0.053 % CO2
Certified CO₂ gas
CO₂ – Measuring
Equipment3
Up to 20 % CO₂
1.0 % CO₂
Certified gas analyzer
O₂ – Measuring
Equipment
1 % O2
6 % O2
20.8 % O2
0.017 % O2
0.066 % O2
0.034 % O2
Certified O₂ gas | See scope document |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 2 of 12
II. Electrical – DC/Low Frequency
Parameter/Range
Frequency
CMC2, 6 (±)
Comments
AC Volts – Generate
(1.0 to 32.999) mV
(10 to 45) Hz
(0.045 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.080 % + 6 µV
0.015 % + 6 µV
0.020 % + 6 µV
0.10 % + 6 µV
0.35 % + 12 µV
0.80 % + 50 µV
Fluke 5522A
AC Volts – Generate
(33 to 329.999) mV
(0.33 to 3.299 99) V
(3.3 to 32.9999) V
(33 to 329.999)
(300 to 1020) V
(10 to 45) Hz
(0.045 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
(0.045 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(10 to 45) Hz
(45 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(45 to 1000) Hz
(1 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(45 to 1000) Hz
(1 to 5) kHz
(5 to 10) kHz
0.030 % + 8 µV
0.015 % + 8 µV
0.016 % + 8 µV
0.035 % + 8 µV
0.080 % + 32 µV
0.20 % + 70 µV
0.030 % + 50 µV
0.015 % + 60 µV
0.019 % + 60 µV
0.030 % + 50 µV
0.070 % + 130 µV
0.030 % + 0.65 mV
0.015 % + 0.60 mV
0.024 % + 0.60 mV
0.035 % + 0.60 mV
0.090 % + 1.6 mV
0.019 % + 2.0 mV
0.020 % + 6.0 mV
0.025 % + 6.0 mV
0.030 % + 6.0 mV
0.20 % + 50 mV
0.030 % + 10 mV
0.025 % + 10 mV
0.030 % + 10 mV
Fluke 5522A | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 3 of 12
Parameter/Range
Frequency
CMC2, 6 (±)
Comments
AC Volts – Measure
(0 to 10) mV
(10 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
(0.001 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(0.001 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
(0.003 to 1) MHz
(1 to 2) MHz
(2 to 4) MHz
(4 to 8) MHz
(8 to 10) MHz
(0.001 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
(0.003 to 1) MHz
(1 to 2) MHz
(2 to 4) MHz
(4 to 8) MHz
(8 to 10) MHz
(0.001 to 1) kHz
(1 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
(0.003 to 1) MHz
(0.001 to 1) kHz
(1 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
0.032 % + 1.4 µV
0.043 % + 1.4 µV
0.044 % + 1.4 µV
0.39 % + 1.0 µV
1.3 % + 5.0 µV
2.6 % + 5.0 µV
0.009 % + 0.65 µV
0.014 % + 0.65 µV
0.027 % + 1.3 µV
0.066 % + 6.5 µV
0.26 % + 40 µV
1.3 % + 0.13 mV
1.9 % + 0.65 mV
5.2 % + 1.3 mV
10 % + 1.3 mV
19 % + 1.3 mV
0.008 % + 0.065 mV
0.014 % + 0.065 mV
0.027 % + 0.13 mV
0.066 % + 0.65 mV
0.26 % + 4.0 mV
1.3 % + 13 mV
1.9 % + 65 mV
5.2 % + 130 mV
10 % + 130 mV
19 % + 130 mV
0.030 % + 0.65 mV
0.13 % + 0.65 mV
2.6 % + 1.3 mV
0.027 % + 1.3 mV
0.066 % + 6.5 mV
0.45 % + 60 mV
1.3 % + 650 mV
0.03 % + 13 mV
0.45 % + 0.6 V
1.3 % + 6.5 V
0.027 % + 0.032 V
0.066 % + 0.13 V
Fluke 8588A | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 4 of 12
Parameter/Equipment
Range
CMC2, 6 (±)
Comments
DC Voltage – Generate
(0 to 329.9999) mV
(0.329 to 3.299 999) V
(3.29 to 32.999 99) V
(32.9 to 329.9999) V
(329 to 1020.000) V
0.0023 % + 2.0 µV
0.0013 % + 3.5 µV
0.0014 % + 35 µV
0.0021 % + 250 µV
0.0021 % + 1.8 mV
Fluke 5522A
DC Volts – Measure
(0 to 100) mV
(0.001 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
0.0007 % + 0.26 µV
0.0004 % + 0.39 µV
0.0004 % + 0.60 µV
0.0006 % + 39 µV
0.0006 % + 0.69 mV
Fluke 8588A
Resistance – Generate
(0 to 10.9999) Ω
(11 to 32.9999) Ω
(33 to 109.9999) Ω
(110 to 329.9999) Ω
(0.33 to 1.099 999) kΩ
(1.1 to 3.299 999) kΩ
(3.3 to 10.999 99) kΩ
(11 to 32.999 99) kΩ
(33 to 109.9999) kΩ
(110 to 329.999 99) kΩ
(0.33 to 1.099 999) MΩ
(1.1 to 3.299 99) MΩ
(3.3 to 10.999 99) MΩ
(11 to 32.999 99) MΩ
(33 to 109.9999) MΩ
(110 to 329.9999) MΩ
(330 to 1100) MΩ
0.0040 % + 0.01 Ω
0.0030 % + 0.015 Ω
0.0028 % + 0.015 Ω
0.0028 % + 0.02 Ω
0.0028 % + 0.02 Ω
0.0028 % + 0.20 Ω
0.0028 % + 0.1 Ω
0.0028 % + 1.0 Ω
0.0028 % + 1.0 Ω
0.0032 % + 10 Ω
0.0032 % + 10 Ω
0.0060 % + 150 Ω
0.013 % + 250 Ω
0.025 % + 2.5 kΩ
0.050 % + 3.0 kΩ
0.30 % + 100 kΩ
1.5 % + 500 kΩ
Fluke 5522A
Resistance – Measure
(0 to 1) Ω
(1 to 10) Ω
(10 to 100) Ω
(0.1 to 1) kΩ
(1 to 10) kΩ
(10 to 100) kΩ
(0.1 to 1) MΩ
(1 to 10) MΩ
(10 to 100) MΩ
(0.1 to 1) GΩ
0.0014 % + 5.2 µΩ
0.0010 % + 18 µΩ
0.0009 % + 60 µΩ
0.0009 % + 600 µΩ
0.0009 % + 0.0060 Ω
0.0009 % + 0.060 Ω
0.0011 % + 1.3 Ω
0.0014 % + 0.13 kΩ
0.0050 % + 13 kΩ
0.065 % + 1.3 MΩ
Fluke 8846A | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 5 of 12
Parameter/Range
Frequency
CMC2, 6 (±)
Comments
AC Current – Generate
(29.00 to 329.99) µA
(0.33 to 3.299 99) mA
(3.3 to 32.9999) mA
(33 to 329.999) mA
(0.33 to 1.099 99) A
(1.1 to 2.999 99) A
(3 to 10.9999) A
(11 to 20.5) A
(10 to 20) Hz
(20 to 45) Hz
(0.045 to 1) kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 20) Hz
(20 to 45) Hz
(0.045 to 1) kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 20) Hz
(20 to 45) Hz
(0.045 to 1) kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 20) Hz
(20 to 45) Hz
(0.045 to 1) kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 45) Hz
(0.045 to 1) kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 45) Hz
(0.045 to 1) kHz
(1 to 5) kHz
(5 to 10) kHz
(45 to 100) Hz
(0.1 to 1) kHz
(1 to 5) kHz
(45 to 100) Hz
(0.1 to 1) kHz
(1 to 5) kHz
0.20 % + 0.10 µA
0.15 % + 0.10 µA
0.13 % + 0.10 µA
0.30 % + 0.15 µA
0.80 % + 0.20 µA
1.6 % + 0.40 µA
0.20 % + 0.15 µA
0.13 % + 0.15 µA
0.10 % + 0.15 µA
0.20 % + 0.20 µA
0.50 % + 0.30 µA
1.0 % + 0.60 µA
0.16 % + 2.0 µA
0.09 % + 2.0 µA
0.04 % + 2.0 µA
0.08 % + 2.0 µA
0.20 % + 3.0 µA
0.40 % + 4.0 µA
0.10 % + 20 µA
0.090 % + 20 µA
0.040 % + 20 µA
0.10 % + 50 µA
0.20 % + 100 µA
0.40 % + 200 µA
0.18 % + 0.10 mA
0.050 % + 0.10 mA
0.60 % + 1.0 mA
2.5 % + 5.0 mA
0.18 % + 0.10 mA
0.060 % + 0.10 mA
0.60 % + 1.0 mA
2.5 % + 5.0 mA
0.060 % + 2.0 mA
0.10 % + 2.0 mA
3.0 % + 2.0 mA
0.12 % + 5.0 mA
0.15 % + 5.0 mA
3.0 % + 5.0 mA
Fluke 5522A | ||
| 6 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 6 of 12
Parameter/Range
Frequency
CMC2, 4, 6 (±)
Comments
AC Current – Measure
(0 to 10) µA
(10 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
(1 to 10) A
(10 to 30) A
(0.001 to 30) kHz
(0.001 to 2) Hz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(0.001 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(0.001 to 2) kHz
(2 to 10) kHz
(10 to 30) kHz
(0.01 to 10) kHz
(0.01 to 2) kHz
(2 to 10) kHz
0.20 % + 0.0025 µA
0.026 % + 0.005 µA
0.051 % + 0.05 µA
0.072 % + 0.50 µA
0.40 % + 1.0 µA
0.026 % + 5.0 µA
0.050 % + 5.0 µA
0.071 % + 5.0 µA
0.026 % + 0.10 mA
0.051 % + 0.10 mA
0.071 % + 0.10 mA
0.080 % + 0.50 mA
0.080 % + 12 mA
0.12 % + 12 mA
Fluke 8588A
Parameter/Equipment
Range
CMC2, 6 (±)
Comments
DC Current – Generate
(0 to 329.999) µA
(0.3 to 3.299 99) mA
(3.3 to 32.9999) mA
(33 to 329.999) mA
(0.3 to 1.099 99) A
(1.1 to 2.999 99) A
(3 to 10.9999) A
(11 to 20.5) A
0.015 % + 0.020 µA
0.010 % + 0.05 µA
0.010 % + 0.25 µA
0.010 % + 2.5 µA
0.020 % + 40 µA
0.038 % + 40 µA
0.050 % + 500 µA
0.10 % + 750 µA
Fluke 5522A
DC Current – Measure
(0 to 10) µA
(10 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
(1 to 10) A
(10 to 30) A
0.0031 % + 0.52 nA
0.0011 % + 0.50 nA
0.0010 % + 0.005 µA
0.0011 % + 0.05 µA
0.0043 % + 1.3 µA
0.013 % + 130 µA
0.022 % + 0.052 mA
0.063 % + 0.56 mA
Fluke 8588A | ||
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 7 of 12
Parameter/Equipment
Range
CMC2, 6 (±)
Comments
Capacitance – Generate
(220 to 339.9) pF
(0.4 to 1.0999) nF
(1.1 to 3.2999) nF
(3.3 to 10.9999) nF
(11 to 32.9999) nF
(33 to 109.999) nF
(110 to 329.999) nF
(0.33 to 1.099 99) µF
(1.1 to 3.299 99) µF
(3.3 to 10.9999) µF
(11 to 32.9999) µF
(33 to 109.999) µF
(110 to 329.999) µF
(0.33 to 1.099 99) mF
(1.1 to 3.299 99) mF
(3.3 to 10.9999) mF
(11 to 32.9999) mF
(33 to 110) mF
0.50 % + 10 pF
0.50 % + 0.01 nF
0.50 % + 0.01 nF
0.25 % + 0.01 nF
0.25 % + 0.01 nF
0.25 % + 0.01 nF
0.25 % + 0.03 nF
0.25 % + 1.0 nF
0.25 % + 3.0 nF
0.25 % + 10 nF
0.40 % + 30 nF
0.45 % + 100 nF
0.45 % + 300 nF
0.45 % + 1.0 µF
0.45 % + 3.0 µF
0.45 % + 10 µF
0.75 % + 30 µF
1.1 % + 100 µF
Fluke 5522A
Capacitance – Measure
(0 to 1) nF
(1 to 10) nF
(10 to 100) nF
(0.1 to 1) µF
(1 to 10) µF
(10 to 100) µF
(0.1 to 1) mF
(1 to 10) mF
(10 to 100) mF
0.10 % + 0.001 nF
0.06 % + 0.002 nF
0.04 % + 0.01 nF
0.04 % + 0.1 nF
0.04 % + 1.0 nF
0.06 % + 10 nF
0.06 % + 0.1 µF
0.007 % + 1.0 µF
0.007 % + 10 µF
Fluke 8588A
Thermocouple –
Generate
Type E
Type J
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 750) °C
(760 to 1200) °C
0.50 °C
0.16 °C
0.14 °C
0.16 °C
0.21 °C
0.27 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
Fluke 5522A | ||
| 8 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 8 of 12
Parameter/Equipment
Range
CMC2 (±)
Comments
Thermocouple – Generate
(cont)
Type K
Type N
Type R
Type S
Type T
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120 °C
(120 to 1000) °C
(1000 to 1372) °C
(-200 to -100) °C
(-100 to 25) °C
(-25 to 120) °C
(120 to 410) °C
(410 to 1300) °C
(0 to 250) °C
(250 to 400) °C
(400 to 1000) °C
(1000 to 1767) °C
(0 to 250) °C
(250 to 1000) °C
(1000 to 1400) °C
(1400 to 1767) °C
(120 to 400) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 120) °C
(120 to 400) °C
0.33 °C
0.18 °C
0.16 °C
0.26 °C
0.40 °C
0.40 °C
0.22 °C
0.19 °C
0.18 °C
0.27 °C
0.57 °C
0.35 °C
0.33 °C
0.40 °C
0.47 °C
0.36 °C
0.37 °C
0.46 °C
0.14 °C
0.63 °C
0.37 °C
0.16 °C
0.14 °C
Fluke 5522A
RTD – Generate
Pt 385 – 100 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 640) °C
(640 to 800) °C
0.050 °C
0.050 °C
0.070 °C
0.090 °C
0.10 °C
0.12 °C
0.23 °C
Fluke 5522A | ||
| 9 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 9 of 12
Parameter/Equipment
Range
CMC2 (±)
Comments
RTD – Generate (cont)
Pt 3926 – 100 Ω
Pt 3916 – 100 Ω
Pt 3985 – 200 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 640) °C
(-200 to -190) °C
(-190 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
0.050 °C
0.050 °C
0.070 °C
0.090 °C
0.10 °C
0.12 °C
0.25 °C
0.040 °C
0.050 °C
0.060 °C
0.070 °C
0.080 °C
0.090 °C
0.10 °C
0.23 °C
0.040 °C
0.040 °C
0.040 °C
0.050 °C
0.12 °C
0.13 °C
0.14 °C
0.16 °C
Fluke 5522A
III. Mechanical
Parameter/Equipment
Range
CMC2 (±)
Comments
Centrifuges3
(1000 to 15 000) rpm
(>15 000 to 80 000) rpm
10 rpm
43 rpm
Optical tachometer | ||
| 10 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 10 of 12
IV. Thermodynamics
Parameter/Equipment
Range
CMC2, 9 (±)
Comments
Temperature – Measure3
(-80 to 105) °C
(>105 to 150) °C
(>150 to 200) °C
(>200 to 250) °C
(>250 to 300) °C
(>300 to 350 °C
(>350 to 450) °C
(>450 to 1000) °C
0.13 °C
1 °C
2 °C
5 °C
7 °C
10 °C
12 °C
17 °C
Certified thermometers
Temperature –
IR Systems
(-15 to 0) °C
(>0 to 50) °C
(>50 to 120) °C
(>50 to 200) °C
(>200 to 350) °C
(>350 to 500) °C
0.31 °C
0.22 °C
0.25 °C
0.32 °C
0.56 °C
0.96 °C
Fluke 4180
ε = 0.95
λ = (8 to 14) µm
Fluke 4181
ε = 0.95
λ = (8 to 14) µm
Thermocouple
Thermometer System
-196 °C
(-80 to 160) °C
(160 to 200) °C
(200 to 300) °C
(300 to 425) °C
(425 to 700) °C
(700 to 1200) °C
0.0073 °C
0.018 °C
0.021 °C
0.039 °C
0.30 °C
0.35 °C
0.44 °C
Liquid bath or dry
block calibrator
Liquid-In-Glass
Thermometers (Partial &
Total Immersion)
(-80 to 100) °C
(100 to 160) °C
0.016 °C
0.029 °C
Liquid temperature
bath
Platinum Resistance &
Thermistor Thermometer
System
-196 °C
(-80 to 100) °C
(100 to 160) °C
(160 to 200) °C
(200 to 300) °C
(300 to 425) °C
(425 to 600) °C
0.0073 °C
0.013 °C
0.019 °C
0.028 °C
0.038 °C
0.29 °C
0.47 °C
Liquid bath or dry
block calibrator | ||
| 11 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 11 of 12
Parameter/Equipment
Range
CMC2, 4, 9 (±)
Comments
Relative Humidity –
Hygrometer Systems
(10 to 35) % RH
(35 to 65) % RH
(65 to 95) % RH
95 % RH
0.23 % RH
0.43 % RH
0.55 % RH
0.76 % RH
Rotronic HyrgoGen
H2-S
Relative Humidity –
Measure3
Up to 95 % RH
1.0 % RH
Reference
hygrometer
V. Time & Frequency
Parameter/Equipment
Range
CMC2, 9 (±)
Comments
Frequency – Generate
(Up to 3 V)
(0.01 to 45) Hz
(0.045 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
0.014 Hz
0.0017 kHz
0.0042 kHz
0.016 kHz
0.074 kHz
1.3 kHz
Fluke 5522A internal
source
Frequency – Measure3
(Up to 1000 V)
(0.000 000 1 to 10) MHz
0.000 13 %
Fluke 8588A
Timers/Stopwatch
Up to 24 hr
0.013 s/day
Timometer 4500
Optical Tachometers
Up to 1000 RPM
(1000 to 10 000) RPM
(10 000 to 25 000) RPM
(25 000 to 100 000) RPM
0.15 RPM
0.20 RPM
0.65 RPM
1.5 RPM
Function generator &
lamp
__________________________
1 This laboratory offers commercial calibration services and field calibration services.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration. | ||
| 12 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 6843.01) 12/24/2025 Page 12 of 12
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 In the statement of CMC, percentages are absolute, unless otherwise indicated.
5 In the statement if CMC, percentages are percentages of reading, unless otherwise indicated.
6 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMCs are expressed as eithe r a specific value that covers the full range or as a
fraction/percentage of the reading plus a fixed floor specification.
7 This scope meets A2LA’s P112 Flexible Scope Policy.
8 Offices for administration of field technicians are located in 275 Aiken Road, Asheville, NC, USA 28804.
9 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter. | ||
| 13 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.