Anritsu Americas Sales Company
490 Jarvis Drive, 95037, Morgan Hill, CA, United States
Visit website- Scope lines
- 6
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 6 scope lines.
A2LA certificate 2160.05 · 6 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; IEC 17025; ANSI/NCSL Z540-1-1994 ANRITSU AMERICAS SALES COMPANY 490 | - | to 0.1 in |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2160.05) 07/31/2024 Page 2 of 5
Parameter/Range
Frequency
CMC2, 5 (±)
Comments
Power Level – Absolute3, 8
Measure & Generate, Type
N & K Connector,
Spectrum Analyzer,
Measuring Receiver
(-35 to 20) dBm
(-60 to 20) dBm
(-90 to -60)
(-35 to 20) dBm
(-60 to 20) dBm
(-90 to -60)
(0.01 to 100) MHz
(0.1 to 2) GHz
(2 to 12) GHz
(12 to 18) GHz
(0.01 to 2) GHz
(2 to 9) GHz
(9 to 13) GHz
(13 to 18) GHz
(0.01 to 18) GHz
(0.01 to 100) MHz
(0.1 to 7) GHz
(7 to 12) GHz
(12 to 17) GHz
(17 to 40) GHz
(0.01 to 3) GHz
(3 to 8) GHz
(8 to 12) GHz
(12 to 27) GHz
(27 to 40) GHz
(0.05 to 40) GHz
0.025 dB
0.030 dB
0.045 dB
0.055 dB
0.057 dB
0.063 dB
0.071 dB
0.088 dB
0.17 dB
0.035 dB
0.05 dB
0.06 dB
0.08 dB
0.085 dB
0.057 dB
0.063 dB
0.071 dB
0.088 dB
0.097 dB
0.18 dB
Type N power sensor
Spectrum analyzer, power
sensor with attenuator
Type K power sensor
Spectrum analyzer, power
sensor with attenuator
Power Level – Absolute6
Measure & Generate, Type
V & Waveguide Connector
(-35 to 20) dBm
(0.01 to 100) MHz
(0.1 to 2) GHz
(2 to 8) GHz
(8 to 12) GHz
(12 to 26) GHz
(26 to 40) GHz
(40 to 50) GHz
(50 to 67) GHz
0.04 dB
0.05 dB
0.06 dB
0.065 dB
0.08 dB
0.09 dB
0.12 dB
0.17 dB
Type V power sensor | ||
| 3 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2160.05) 07/31/2024 Page 3 of 5
Parameter/Range
Frequency
CMC2, 5, 7 (±)
Comments
Power Level – Absolute6
Measure& Generate, Type V
& Waveguide Connector
(cont)
(-35 to 20) dBm
(50 to 90) GHz
0.21 dB
Waveguide power
sensor
Frequency Modulation –
Measure
Modulation 10 %
Carrier
1 GHz
5 GHz
10 GHz
20 GHz
30 GHz
40 GHz
0.02 %
0.08 %
0.15 %
0.30 %
0.45 %
0.60 %
Function generator
measuring receiver
LO synthesizer
Amplitude Modulation –
Measure
(10 to 100) %
(1 to 20) GHz
(20 to 40) GHz
2.6 %
3.9 %
Function generator &
measuring receiver
Pulse Modulation – Measure
Power
Rise, Fall & Compression
High Level Noise
Magnitude
High Level Noise Phase
(0.01 to 40) GHz
< 500 kHz
(0.005 to 2.5) GHz
(2.5 to 5) GHz
(5 to 20) GHz
(20 to 40) GHz
< 500KHz
(0.005 to 2.5) GHz
(2.5 to 5) GHz
(5 to 20) GHz
(20 to 40) GHz
0.09 dB
0.011 ns
0.000 59 dB-rms
0.000 33 dB-rms
0.000 25 dB-rms
0.000 28 dB-rms
0.000 32 dB-rms
0.0055 deg-rms
0.0025 deg-rms
0.0029 deg-rms
0.0018 deg-rms
0.0028 deg-rms
Wide band oscilloscope
& function generator
Spectrum & network
analyzers | -35 to 20 dBm |
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2160.05) 07/31/2024 Page 4 of 5
III. Mechanical
Parameter/Equipment
Range
CMC2, 7 (±)
Comments
Torque Wrench
(0.25 to 250) lbf.in
2.5 %
Torque analyzer
IV. Time & Frequency
Parameter/Equipment
Range
CMC2, 6 (±)
Comments
Frequency –
Measuring Equipment
10 MHz reference
1 Hz to 70 GHz4
9 x 10-11 Hz/Hz
1.4 x 10-9 Hz/Hz
GPS receiver
Synthesizer w/ GPS
disciplined oscillator
____________________________________________
1 This laboratory offers commercial calibration and field calibration services.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavio r of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the Calibration and
Measurement Capability Uncertainty (CMC) found on the A2LA Scope. Allowance must be made for
aspects such as the environment at the place of calibration and for other possible adverse effects such as
those caused by transportation of the calibration equipment. The usual allowance for the uncertainty
introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own,
could result in the actual measurement uncertainty achievable on a customer’s site being larger than the
CMC
4 This scope meets A2LA’s P112 Flexible Scope Policy.
5 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification.
6 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter. | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 2160.05) 07/31/2024 Page 5 of 5
7 In the statement of CMC, percentages are percentage of reading, unless otherwise indicated.
8 Enlisted values represent absolute power level uncertainty, including test port match. | ||
| 6 | ISO/IEC 17025:2017; ISO-ILAC-IAF Communiqué dated April 2017; IEC 17025; ANSI/NCSL Z540-1-1994 and R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.