Aldinger Company
Dallas, TX, United States
Visit website- Scope lines
- 38
- A2LA certificate
- 1
- Sites in our directory
- 2
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 38 scope lines.
A2LA certificate 1509.01 · 38 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; IEC 17025; ANSI/NCSL Z540-1-1994 ALDINGER COMPANY 1440 | - | See scope document |
| 2 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 2 of 37
II. Dimensional
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Calipers3
Up to 24 in
(24 to 48) in
Up to 48 in
(48 to 96) in
550 µin
1100 µin
77 µin
1500 µin
Gage blocks (field)
Gage blocks (lab)
Dial & Test Indicators3
Up to 1 in
Up to 3 in
Up to 3 in
20 µin
59 µin
90 µin
Gage blocks (lab)
Gage blocks, ULM,
SuperMicrometer10 (lab)
Gage blocks (field)
Gage Blocks
Up to 1 in
(2 to 4) in
(4 to 20) in
(3.3 + 1.5L) µin
(7.7 + 1.5L) µin
(7 + 2L) µin
Master gage block set,
Federal 130B-24 comparator
Maser gage block set, ULM
Angle Blocks
Up to 90°
0.005°
Sine bar & gage blocks
Squares
Up to 18 in
99 µin
Master square, amplified gage
head
Height Gages3
Up to 48 in
280 µin
89 µin
Gage blocks (field)
Gage blocks (lab)
Length Standards
Up to 18 in
Up to 48 in
(37 + 1.5L) µin
(65 + 1.5L) µin
ULM/SuperMicrometer9,
gage blocks
Levels3
Up to 12 in
(40 + 14L) µin
Surface plate, granite square
Micrometers3
Up to 4 in
(4 to 24) in
(24 to 48) in
Up to 4 in
(4 to 24) in
(24 to 96) in
92 µin
550 µin
1100 µin
15 µin
60 µin
1500 µin
Gage blocks (field)
Gage blocks (lab) | to 96 in |
| 3 | ANSI B1 | - | See scope document |
| 4 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 4 of 37
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Surface Plates – Flatness
& Repeatability3
Indicator Stands
Up to 107 in DL
(-0.002 to 0.002) in
(-0.002 to 0.002) in
(-0.01 to 0.01) in
(48 + 0.41DL) µin
31 µin
45 µin
580 µin
Leveling system (DL is
diagonal of the plate)
Repeat reading gage
Repeat reading gage,
triangular gage block base
Indicator, triangular gage
block base
Optical Comparators3
X-Axis
Y-Axis
Angle
Up to 12 in
Up to 12 in
Up to 90°
(250 + 91L) µin
(250 + 91L) µin
0.029°
Glass scale
Angle blocks
Rulers3
Up to 72 in
0.013 in
Standard ruler
Radius Gages3
Up to 0.75 in
1300 µin
Optical comparator & overlay
Angle Gages3
Up to 180°
0.044°
Optical comparator & vernier
Torque Arms
Up to 24 in
Up to 48 in
0.0047 in
0.003 in
Height gage, test indicator
Gage blocks
Coating/Electromagnetic
Thickness3
Up to 120 mils
0.25 mils
Shims
Tape Measures3
Up to 40 in
Up to 100 ft
0.013 in
0.062 in
Standard ruler
Standard tape
PI Tapes
Up to 72 in
910 µin
Gage blocks | to 107 in |
| 5 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 5 of 37
III. Electrical – DC/Low Frequency
Parameter/Range
Frequency
CMC2, 6, 12 (±)
Comments
AC Current – Measure3
Up to 1 A
(1 to 3) A
60 Hz to 1 kHz
0.16 % + 0.4 mA
0.14 % + 1.8 mA
HP 34401A
AC Current – Generate3
(29 to 330) µA
330 µA to 3.3 mA
(3.3 to 33) mA
(33 to 330) mA
(0.33 to 3) A
(3 to 10) A
(10 to 20) A
(16 to 150) A
(150 to 1000) A
45 Hz to 1 kHz
(45 to 440) Hz
0.13 % + 0.1 µA
0.10 % + 0.15 µA
0.07 % + 2 µA
0.083 % + 20 µA
0.10 % + 100 µA
0.21 % + 2 mA
0.21 % + 5 mA
1 % + 0.25 A
1 % + 0.9 A
Fluke 5522A
Using 50 turn coil
AC Voltage – Generate3
(1 to 330) mV
(0.330 to 3.3) V
(3.3 to 33) V
(33 to 330) V
(330 to 1020) V
45 Hz to 10 kHz
45 Hz to 10 kHz
45 Hz to 10 kHz
45 Hz to 1 kHz
(1 to 10) kHz
45 Hz to 1 kHz
(1 to 5) kHz
0.018 % + 8 µV
0.018 % + 60 µV
0.017 % + 0.6 mV
0.023 % + 2 mV
0.023 % + 6 mV
0.046 % + 10 mV
0.040 % + 10 mV
Fluke 5522A
AC Voltage – Measure3
Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 750) V
Up to 10 kV
Up to 50 kV
60 Hz to 3 kHz
0.063 % + 40 µV
0.063 % + 300 µV
0.062 % + 3 mV
0.064 % + 30 mV
0.075 % + 230 mV
0.12 % + 100 mV
0.1 % + 400 mV
HP 34401A
Vitrek 4700 w/ HVL-70 | to 1020 V |
| 6 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 6 of 37
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
Capacitance – Generate3
Low
High
(0.22 to 3.3) nF
(3.3 to 33) nF
(33 to 330) nF
(330 to 3300) nF
(3.3 to 33) µF
(33 to 330) µF
(330 to 3300) µF
(3.3 to 110) mF
0.5 % + 10 pF
0.26 % + 100 pF
0.26 % + 300 pF
0.26 % + 3 nF
0.4 % + 30 nF
0.45 % + 300 nF
0.45 % + 3 µF
1.1 % + 10 µF
Fluke 5522A
DC Current – Generate3
(0 to 330) µA
Up to 3.3 mA
Up to 33 mA
Up to 330 mA
Up to 3 A
(1.1 to 11) A
(11 to 20.5) A
(10 to 150) A
(150 to 1000) A
0.015 % + 20 nA
0.010 % + 50 nA
0.014 % + 250 nA
0.010 % + 2.5 µA
0.038 % + 40 µA
0.050 % + 500 µA
0.1 % + 750 µA
0.5 % + 0.14 A
0.5 % + 0.5 A
Fluke 5522A
Using 50 turn coil
DC Current – Measure3
Up to 10 mA
(10 to 100) mA
100 mA to 1 A
(1 to 3) A
0.075 % + 6 µA
0.051 % + 5 µA
0.11 % + 100 µA
0.13 % + 600 µA
HP 34401A
DC Voltage – Generate3
(0 to 330) mV
Up to 3.3 V
Up to 33 V
(30 to 330) V
(100 to 1000) V
0.002 % + 1 µV
0.0012 % + 2 µV
0.0013 % + 20 µV
0.0018 % + 0.15 mV
0.0019 % + 1.5 mV
Fluke 5522A
DC Voltage – Measure3
Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
Up to 10 kV
Up to 70 kV
0.0054 % + 3.5 µV
0.0042 % + 7 µV
0.0037 % + 50 µV
0.0048 % + 600 µV
0.0049 % + 10 mV
0.041 % + 30 mV
0.049 % + 200 mV
HP 34401A
Vitrek 4700 w/ HVL-70 | ||
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 7 of 37
Parameter/Equipment
Range
CMC2, 12 (±)
Comments
Oscilloscopes3
Square Wave Amplitude:
50 Ω at 1 kHz
1 MΩ at 1 kHz
DC Voltage Amplitude:
50 Ω Load
1 MΩ Load
Level Sine Wave:
Frequency
Amplitude
Flatness
Time Markers:
50 Ω Load
Rise Time – Generate
1 mV to 6.6 V pk-pk
1 mv to 130 V pk-pk
Up to ± 6.6 V
Up to ± 130 V
Up to 1100 MHz
50 kHz Reference
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
5 s to 50 ms
20 ms to 1 ns
1 kHz to 2 MHz
(200 to 300) ps
(2 to 10) MHz
(200 to 300) ps
0.34 % + 40 µV
0.25 % + 40 µV
0.25 % + 40 µV
0.05 % + 40 µV
2.5 parts × 10
6
2.8 % + 300 µV
4.2 % + 300 µV
4.6 % + 300 µV
6.4 % + 300 µV
7.4 % + 300 µV
2.7 % + 100 µV
3.0 % + 100 µV
4.6 % + 100 µV
5.5 % + 100 µV
28 + 1000t µs/s
3.2 µs/s
120 ps
120 ps
Fluke 5522A/
SC1100
t = time in seconds
Electrical Calibration of
Temperature Controllers
3
Thermocouple Types J, K,
T & E
PT385, 100 Ω
(-200 to < 0) °C
(0 to 1370) °C
(-200 to 800) °C
1.2 °C
0.85 °C
0.66 °C
Fluke 725 | ||
| 8 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 8 of 37
Parameter/Equipment
Range
CMC2, 6. 12 (±)
Comments
Resistance – Generate3
Up to 330 Ω
(330 to 3300) Ω
(3300 to 33 000) Ω
(33 to 330) kΩ
(330 to 3300) kΩ
(3.3 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(330 to 1100) MΩ
0.065 % + 20 mΩ
0.029 % + 100 mΩ
0.029 % + 1.0 Ω
0.0037 % + 10 Ω
0.0063 % + 150 Ω
0.026 % + 2500 Ω
0.063 % + 3.0 kΩ
0.14 % + 100 kΩ
0.14 % + 500 kΩ
Fluke 5522A
Resistance – Measure3
Up to 100 Ω
100 Ω to 1 kΩ
(1 to 10) kΩ
(10 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
0.066 % + 4 mΩ
0.031 % + 10 mΩ
0.031 % + 100 mΩ
0.011 % + 1 Ω
0.012 % + 10 Ω
0.048 % + 100 Ω
0.8 % + 10 kΩ
HP 34401A
Electrical Simulation of
RTD
3,
PT385, 100 Ω
(-200 to 0) °C
(0 to 100) °C
(100 to 400) °C
(400 to 630) °C
(630 to 800) °C
0.058 °C
0.17 °C
0.12 °C
0.14 °C
0.28 °C
Fluke 5522A | to 1100 MΩ |
| 9 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 9 of 37
Parameter/Equipment
Range
CMC2 (±)
Comments
Electrical Simulation of
Thermocouples &
Indicators3
Type E
Type J
Type K
Type T
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
(-250 to -150) °C
(-150 to 0.0) °C
(0.0 to 120) °C
(120 to 400) °C
0.51 °C
0.16 °C
0.19 °C
0.20 °C
0.24 °C
0.28 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
0.33 °C
0.18 °C
0.20 °C
0.24 °C
0.40 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
Fluke 5522A
Welding Devices3
(0 to 350) Amps DC
(0 to 100) Volts DC
(100 to 700) Feed Rate IPM
0.87 ADC
0.012 VDC
3.6 IPM
Load bank, current
shunt, DMM
IV. Fluid Quantities
Parameter/Equipment
Range
CMC2 (±)
Comments
POVA (Piston/Plunger
Operated Volumetric
Apparatus)3 –
Including But Not
Limited To: Pipettes,
Syringes, Dispensers, &
Burettes
(0.1 to 50) µL
(50 to 100) µL
(100 to 500) µL
(500 to 1000) µL
(1000 to 5000) µL
(5000 to 10 000) µL
(10 000 to 20 000) µL
(20 000 to 50 000) µL
0.31 µL
0.31 µL
0.56 µL
0.72 µL
5.4 µL
18 µL
30 µL
72 µL
Gravimetric method | ||
| 10 | ison Balances3 | - | See scope document |
| 11 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 11 of 37
Parameter/Equipment
Range7
CMC2 (±)
Comments
Mass, Fixed Points
1 mg
2 mg
5 mg
10 mg
20 mg
30 mg
50 mg
100 mg
200 mg
300 mg
500 mg
1 g
2 g
5 g
10 g
20 g
50 g
100 g
200 g
500 g
1 kg
2 kg
5 kg
10 kg
20 kg
25 kg
0.8859 g (
1/32 oz)
1.772 g (1/16 oz)
3.544 g (1/8 oz)
7.087 g (1/4 oz)
14.17 g (1/2 oz)
28.35 g (1 oz)
56.7 g (2 oz)
113.4 g (4 oz)
226.8 g (8 oz)
0.4536 g (0.001 lb)
0.9072 g (0.002 lb)
2.27 g (0.005 lb)
4.54 g (0.01 lb)
9.07 g (0.02 lb)
22.68 g (0.05 lb)
45.36 g (0.1 lb)
90.72 g (0.2 lb)
2.2 µg
2.1 µg
2.9 µg
1.4 µg
1.5 µg
2.2 µg
1.6 µg
1.7 µg
1.6 µg
1.6 µg
2.1 µg
11 µg
3.3 µg
5.7 µg
7.2 µg
13 µg
11 µg
39 µg
140 µg
320 µg
390 µg
1.7 mg
2.6 mg
93 mg
94 mg
140 mg
12 µg
94 µg
240 µg
170 µg
170 µg
370 µg
150 µg
740 µg
5.1 mg
130 µg
56 µg
26 µg
16 µg
56 µg
79 µg
37 µg
87 µg
Modified double
substitution
Direct comparison | ||
| 12 | ison Optical Tachometer3 Up to 200 FPM Up to 3000 FPM Up to 29 999 FPM 0 | - | to 20000 lb |
| 13 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 13 of 37
Parameter/Equipment
Range
CMC2, 8, 11 (±)
Comments
Torque Wrenches –
Measure3
(1 to 10) lbf⋅in
(10 to 50) lbf⋅in
(30 to 400) lbf⋅in
(80 to 1000) lbf⋅in
(240 to 3000) lbf⋅in
(1200 to 6000) lbf⋅in
(2400 to 12 000) lbf⋅in
0.33 lbf⋅in
0.17 lbf⋅in
1.8 lbf⋅in
5.1 lbf⋅in
21 lbf⋅in
23 lbf⋅in
58 lbf⋅in
CDI/Norbar torque
calibrator
Norbar torque
calibrator
Torque Measuring
Equipment – Transducers
Up to 200 lbf⋅in
Up to 1200 lbf⋅in
Up to 12 000 lbf⋅in
0.10 %
0.091 %
0.10 %
Torque arm &
weights
Vacuum Measuring
Equipment – Gauges3
Up to 13 psig
0.033 psig
Druck DPI 610
Rotations Speed –
Measurement3
Contact:
Up to 200 RPM
(200 to 3000) RPM
(3000 to 20 000) RPM
Optical:
Up to 3000 RPM
(3000 to 30 000) RPM
0.67 RPM
13 RPM
16 RPM
0.5 RPM
5.0 RPM
Monarch PLT200
Indirect Verification of
Rockwell Hardness
Testers
3
HRB:
Low
Medium
High
HRC:
Low
Medium
High
1.0 HRB
0.65 HRB
0.49 HRB
0.45 HRC
0.36 HRC
0.37 HRC
Indirect verification
Indirect Verification of
Brinell Hardness Testers
3
HBW 10/500:
Low
HBW 10/3000:
High
4.0 HBW
4.8 HBW
Indirect verification | ||
| 14 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 14 of 37
VI. Thermodynamics
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Temperature – Measuring
Equipment3
(-100 to 300) °C
(-148 to 572) °F
0.09 °C
0.16 °F
Hart 1502A/5614 PRT or
Fluke 1523
Temperature – Measuring
Equipment
(-40 to 200) °C
(-40 to 392) °F
0.061 °C
0.11 °F
Hart 1502A/5614 PRT in
wet bath
Temperature –Measure3
(-100 to 600) °C
(-148 to 1112) °F
(> 600 to 1200) °C
(> 1112 to 2192) °F
1.5 °C
2.7 °F
10 °C
18 °F
Fluke 51 or 725 with
thermocouple
IR Thermometer3
(-20 to 150) °C
(Amb + 10 to 400) °C
1.1 °C
2.1 °C
Mikron M340 (lab only)
Omega BB703
Emissivity ≈ 0.95
Wavelength (8 to 14) µm
Humidity – Measure &
Measuring Equipment3
(20 to 90) % RH
3.4 % RH
SensoScientific B13-200
VII. Time & Frequency
Parameter/Equipment
Range
CMC2, 6, 11 (±)
Comments
Frequency – Measuring
Equipment3
Up to 1100 MHz
2.5 parts × 106
Fluke 5522A/SC1100
Frequency – Measure3
40 Hz to 300 kHz
0.014 % Indication
HP 34401
Timers & Stopwatches3
Up to 1 hr
(> 1 to 24) hr
0.14 s
0.59 s
Reference stopwatch
Stroboscopes3
Up to 3600 FPM
Up to 30 000 FPM
0.11 FPM
0.72 FPM
Universal counter | ||
| 15 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 15 of 37
SATELLITE
ALDINGER COMPANY
3130 Rogerdale Rd. STE 186
Houston, TX 77042
Dustin Detten Phone: 214 638 1808
CALIBRATION
I. Chemical
Parameter/Equipment
Range
CMC2 (±)
Comments
Conductivity Meters3
1 µS/cm
10 µS/cm
100 µS/cm
1000 µS/cm
10 000 µS/cm
100 000 µS/cm
0.55 µS/cm
0.41 µS/cm
2.8 µS/cm
20 µS/cm
190 µS/cm
1900 µS/cm
Reference conductivity
standards
pH Meters3
Electrical Calibration
Probe/Meter
Standardization
(0 to 14) pH
4 pH
7 pH
10 pH
0.015 pH
0.030 pH
0.030 pH
0.030 pH
mV/pH calibrator
Reference buffer
standards
II. Dimensional
Parameter/Equipment
Range
CMC2 (±)
Comments
Calipers3
Up to 24 in
(24 to 48) in
Up to 48 in
(48 to 96) in
550 µin
1100 µin
77 µin
1500 µin
Gage blocks (field)
Gage blocks (lab) | ||
| 16 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 16 of 37
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Dial & Test Indicators3
Up to 1 in
Up to 3 in
Up to 3 in
20 µin
59 µin
90 µin
Gage blocks (lab)
Gage blocks, ULM,
SuperMicrometer
10 (lab)
Gage blocks (field)
Height Gages3
Up to 48 in
280 µin
89 µin
Gage blocks (field)
Gage blocks (lab)
Levels3
Up to 12 in
(40 + 14L) µin
Surface plate, granite square
Micrometers3
Up to 4 in
(4 to 24) in
(24 to 48) in
Up to 4 in
(4 to 24) in
(24 to 96) in
92 µin
550 µin
1100 µin
15 µin
60 µin
1500 µin
Gage blocks (field)
Gage blocks (lab)
Protractors3
Up to 90°
0.049°
Surface plate, angle blocks,
sine bar w/ gage blocks
Surface Plates – Flatness
& Repeatability3
Indicator Stands
Up to 107 in DL
(-0.002 to 0.002) in
(-0.002 to 0.002) in
(-0.01 to 0.01) in
(48 + 0.41DL) µin
31 µin
45 µin
580 µin
Leveling system (DL is
diagonal of the plate)
Repeat reading gage
Repeat reading gage,
triangular gage block base
Indicator, triangular gage
block base
Optical Comparators3
X-Axis
Y-Axis
Angle
Up to 12 in
Up to 12 in
Up to 90°
(250 + 91L) µin
(250 + 91L) µin
0.029°
Glass scale
Angle blocks
Rulers3
Up to 72 in
0.013 in
Standard ruler | to 107 in |
| 17 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 17 of 37
Parameter/Equipment
Range
CMC2 (±)
Comments
Radius Gages3
Up to 0.75 in
1300 µin
Optical comparator & overlay
Angle Gages3
Up to 180°
0.044°
Optical comparator & vernier
Coating/Electromagnetic
Thickness3
Up to 120 mils
0.25 mils
Shims
Tape Measures3
Up to 40 in
Up to 100 ft
0.013 in
0.062 in
Standard ruler
Standard tape
III. Electrical – DC/Low Frequency
Parameter/Range
Frequency
CMC2, 6, 12 (±)
Comments
AC Current – Measure3
Up to 1 A
(1 to 3) A
60 Hz to 1 kHz
0.16 % + 0.4 mA
0.14 % + 1.8 mA
HP 34401A
AC Current – Generate3
(29 to 330) µA
330 µA to 3.3 mA
(3.3 to 33) mA
(33 to 330) mA
(0.33 to 3) A
(3 to 10) A
(10 to 20) A
(16 to 150) A
(150 to 1000) A
45 Hz to 1 kHz
(45 to 440) Hz
0.13 % + 0.1 µA
0.10 % + 0.15 µA
0.07 % + 2 µA
0.083 % + 20 µA
0.10 % + 100 µA
0.21 % + 2 mA
0.21 % + 5 mA
1 % + 0.25 A
1 % + 0.9 A
Fluke 5522A
Using 50 turn coil
AC Voltage – Generate3
(1 to 330) mV
(0.330 to 3.3) V
(3.3 to 33) V
(33 to 330) V
(330 to 1020) V
45 Hz to 10 kHz
45 Hz to 10 kHz
45 Hz to 10 kHz
45 Hz to 1 kHz
(1 to 10) kHz
45 Hz to 1 kHz
(1 to 5) kHz
0.018 % + 8 µV
0.018 % + 60 µV
0.017 % + 0.6 mV
0.023 % + 2 mV
0.023 % + 6 mV
0.046 % + 10 mV
0.040 % + 10 mV
Fluke 5522A | ||
| 18 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 18 of 37
Parameter/Range
Frequency
CMC2, 6, 12 (±)
Comments
AC Voltage – Measure3
Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 750) V
Up to 10 kV
Up to 50 kV
60 Hz to 3 kHz
0.063 % + 40 µV
0.063 % + 300 µV
0.062 % + 3 mV
0.064 % + 30 mV
0.075 % + 230 mV
0.12 % + 100 mV
0.1 % + 400 mV
HP 34401A
Vitrek 4700 w/ HVL-70
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
Capacitance – Generate3
Low
High
(0.22 to 3.3) nF
(3.3 to 33) nF
(33 to 330) nF
(330 to 3300) nF
(3.3 to 33) µF
(33 to 330) µF
(330 to 3300) µF
(3.3 to 110) mF
0.5 % + 10 pF
0.26 % + 100 pF
0.26 % + 300 pF
0.26 % + 3 nF
0.4 % + 30 nF
0.45 % + 300 nF
0.45 % + 3 µF
1.1 % + 10 µF
Fluke 5522A
DC Current – Generate3
(0 to 330) µA
Up to 3.3 mA
Up to 33 mA
Up to 330 mA
Up to 3 A
(1.1 to 11) A
(11 to 20.5) A
(10 to 150) A
(150 to 1000) A
0.015 % + 20 nA
0.010 % + 50 nA
0.014 % + 250 nA
0.010 % + 2.5 µA
0.038 % + 40 µA
0.050 % + 500 µA
0.1 % + 750 µA
0.5 % + 0.14 A
0.5 % + 0.5 A
Fluke 5522A
Using 50 turn coil
DC Current – Measure3
Up to 10 mA
(10 to 100) mA
100 mA to 1 A
(1 to 3) A
0.075 % + 6 µA
0.051 % + 5 µA
0.11 % + 100 µA
0.13 % + 600 µA
HP 34401A
DC Voltage – Generate3
(0 to 330) mV
Up to 3.3 V
Up to 33 V
(30 to 330) V
(100 to 1000) V
0.002 % + 1 µV
0.0012 % + 2 µV
0.0013 % + 20 µV
0.0018 % + 0.15 mV
0.0019 % + 1.5 mV
Fluke 5522A | to 750 V |
| 19 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 19 of 37
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
DC Voltage – Measure3
Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
Up to 10 kV
Up to 70 kV
0.0054 % + 3.5 µV
0.0042 % + 7 µV
0.0037 % + 50 µV
0.0048 % + 600 µV
0.0049 % + 10 mV
0.041 % + 30 mV
0.049 % + 200 mV
HP 34401A
Vitrek 4700 w/ HVL-
70
Oscilloscopes3
Square Wave Amplitude:
50 Ω at 1 kHz
1 MΩ at 1 kHz
DC Voltage Amplitude:
50 Ω Load
1 MΩ Load
Level Sine Wave:
Frequency
Amplitude
Flatness
Time Markers:
50 Ω Load
Rise Time – Generate
1 mV to 6.6 V pk-pk
1 mv to 130 V pk-pk
Up to ± 6.6 V
Up to ± 130 V
Up to 1100 MHz
50 kHz Reference
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
5 s to 50 ms
20 ms to 1 ns
1 kHz to 2 MHz
(200 to 300) ps
(2 to 10) MHz
(200 to 300) ps
0.34 % + 40 µV
0.25 % + 40 µV
0.25 % + 40 µV
0.05 % + 40 µV
2.5 parts × 10
6
2.8 % + 300 µV
4.2 % + 300 µV
4.6 % + 300 µV
6.4 % + 300 µV
7.4 % + 300 µV
2.7 % + 100 µV
3.0 % + 100 µV
4.6 % + 100 µV
5.5 % + 100 µV
28 + 1000t µs/s
3.2 µs/s
120 ps
120 ps
Fluke 5522A/
SC1100
t = time in seconds | to 70000 V |
| 20 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 20 of 37
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
Electrical Calibration of
Temperature Controllers3
Thermocouple types J, K, T
& E
PT385, 100 Ω
(-200 to < 0) °C
(0 to 1370) °C
(-200 to 800) °C
1.2 °C
0.85 °C
0.66 °C
Fluke 725
Resistance – Generate3
Up to 330 Ω
(330 to 3300) Ω
(3300 to 33 000) Ω
(33 to 330) kΩ
(330 to 3300) kΩ
(3.3 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(330 to 1100) MΩ
0.065 % + 20 mΩ
0.029 % + 100 mΩ
0.029 % + 1.0 Ω
0.0037 % + 10 Ω
0.0063 % + 150 Ω
0.026 % + 2500 Ω
0.063 % + 3.0 kΩ
0.14 % + 100 kΩ
0.14 % + 500 kΩ
Fluke 5522A
Resistance – Measure3
Up to 100 Ω
100 Ω to 1 kΩ
(1 to 10) kΩ
(10 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
0.066 % + 4 mΩ
0.031 % + 10 mΩ
0.031 % + 100 mΩ
0.011 % + 1 Ω
0.012 % + 10 Ω
0.048 % + 100 Ω
0.8 % + 10 kΩ
HP 34401A
Electrical Simulation of
RTD
3
PT385, 100 Ω
(-200 to 0) °C
(0 to 100) °C
(100 to 400) °C
(400 to 630) °C
(630 to 800) °C
0.058 °C
0.17 °C
0.12 °C
0.14 °C
0.28 °C
Fluke 5522A | ||
| 21 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 21 of 37
Parameter/Equipment
Range
CMC2 (±)
Comments
Electrical Simulation of
Thermocouples &
Indicators3
Type E
Type J
Type K
Type T
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
(-250 to -150) °C
(-150 to 0.0) °C
(0.0 to 120) °C
(120 to 400) °C
0.51 °C
0.16 °C
0.19 °C
0.20 °C
0.24 °C
0.28 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
0.33 °C
0.18 °C
0.20 °C
0.24 °C
0.40 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
Fluke 5522A
Welding Devices3
(0 to 350) Amps DC
(0 to 100) Volts DC
(100 to 700) Feed Rate IPM
0.87 ADC
0.012 VDC
3.6 IPM
Load bank, current
shunt, DMM
IV. Fluid Quantities
Parameter/Equipment
Range
CMC2 (±)
Comments
POVA (Piston/Plunger
Operated Volumetric
Apparatus)3 –
Including But Not
Limited To: Pipettes,
Syringes, Dispensers, &
Burettes
(.1 to 50) µL
(50 to 100) µL
(100 to 500) µL
(500 to 1000) µL
(1000 to 5000) µL
(5000 to 10 000) µL
(10 000 to 20 000) µL
(20 000 to 50 000) µL
0.31 µL
0.31 µL
0.56 µL
0.72 µL
5.4 µL
18 µL
30 µL
72 µL
Gravimetric method | ||
| 22 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 22 of 37
V. Mechanical
Parameter/Equipment
Range
CMC2, 7, 11 (±)
Comments
Balances3
Up to 20 g
Up to 200 g
Up to 200 g
Up to 1000 g
Up to 5000 g
Up to 20 g
Up to 200 g
Up to 1000 g
Up to 5000 g
Up to 10 000 g
Up to 20 000 g
0.11 mg
0.69 mg
1.3 mg
6.6 mg
34 mg
0.81 mg
4.8 mg
24 mg
130 mg
370 mg
460 mg
Class 1 weights
Class 2 weights
Class 4 weights
Force – Measuring
Equipment3
Tension/Compression
Compression
Tension
Up to 100 lbf
(100 to 1000) lbf
(1000 to 10 000) lbf
(10 000 to 50 000) lbf
(10 000 to 100 000) lbf
(50 000 to 500 000) lbf
(100 to 1000) lbf
(1000 to 10 000) lbf
(10 000 to 50 000) lbf
(10 000 to 100 000) lbf
0.0064 lbf
0.26 % Indication
0.24 % Indication
0.33 % Indication
0.20 % Indication
0.4 % Indication
0.23 % Indication
0.23 % Indication
0.29 % Indication
0.20 % Indication
Class F weights w/
load cells | to 20000 g |
| 23 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 23 of 37
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Optical Tachometer3
Up to 200 FPM
Up to 3000 FPM
Up to 29 999 FPM
0.12 FPM
0.39 FPM
3.9 FPM
Calibrated strobe
(FPM = Flashes per
minute)
Pressure Measuring
Equipment – Gages,
Transducers &
Transmitters3
Up to 300 psig
Up to 3000 psig
Up to 5000 psig
Up to 10 000 psig
Up to 30 000 psig
Up to 2 in H
2O
Up to 300 psig
Up to 3000 psig
Up to 10 000 psig
(10 to 17) psia
0.18 psig
3.8 psig
3.9 psig
12 psig
43 psig
0.0026 in H
2O
0.051 psig
0.4 psig
1.3 psig
0.000 78 psia
Digital pressure
gage
Dwyer hook gage
GE/Pace 1003
Barometric
reference
Scales3
Up to 10 lb
Up to 20 lb
Up to 50 lb
Up to 100 lb
Up to 200 lb
Up to 500 lb
Up to 1000 lb
Up to 5000 lb
Up to 20 000 lb
0.000 25 lb
0.000 25 lb
0.0063 lb
0.0063 lb
0.0064 lb
0.0064 lb
0.18 lb
0.19 lb
0.21 lb
Class 4 weights
Class F weights | ||
| 24 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 24 of 37
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Torque Wrenches –
Measure3
(1 to 10) lbf⋅in
(10 to 50) lbf⋅in
(30 to 400) lbf⋅in
(80 to 1000) lbf⋅in
(240 to 3000) lbf⋅in
(1200 to 6000) lbf⋅in
(2400 to 12 000) lbf⋅in
0.33 lbf⋅in
0.17 lbf⋅in
1.8 lbf⋅in
5.1 lbf⋅in
21 lbf⋅in
23 lbf⋅in
58 lbf⋅in
CDI/Norbar torque
calibrator
Norbar torque
calibrator
Vacuum Measuring
Equipment – Gauges
3
Up to 13 psig
0.033 psig
Druck DPI 610
Rotations Speed –
Measurement3
Contact:
Up to 200 RPM
(200 to 3000) RPM
(3000 to 20 000) RPM
Optical:
Up to 3000 RPM
(3000 to 30 000) RPM
0.67 RPM
13 RPM
16 RPM
0.5 RPM
5.0 RPM
Monarch PLT200
Indirect Verification of
Rockwell Hardness
Testers3
HRB:
Low
Medium
High
HRC:
Low
Medium
High
1.0 HRB
0.65 HRB
0.49 HRB
0.45 HRC
0.36 HRC
0.37 HRC
Indirect verification
Indirect Verification of
Brinell Hardness Testers3
HBW 10/500:
Low
HBW 10/3000:
High
4.0 HBW
4.8 HBW
Indirect verification | ||
| 25 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 25 of 37
VI. Thermodynamic
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Temperature – Measuring
Equipment3
(-100 to 300) °C
(-148 to 572) °F
0.09 °C
0.16 °F
Hart 1502A/5614 PRT or
Fluke 1523
Temperature –Measure3
(-100 to 600) °C
(-148 to 1112) °F
(> 600 to 1200) °C
(> 1112 to 2192) °F
1.5 °C
2.7 °F
10 °C
18 °F
Fluke 51 or 725 with
thermocouple
IR Thermometer3
(-20 to 150) °C
(Amb + 10 to 400) °C
1.1 °C
2.1 °C
Mikron M340 (lab only)
Omega BB703
Emissivity ≈ 0.95
Wavelength (8 to 14) µm
Humidity – Measure &
Measuring Equipment3
(20 to 90) % RH
3.4 % RH
SensoScientific B13-200
VII. Time & Frequency
Parameter/Equipment
Range
CMC2, 6, 11 (±)
Comments
Frequency – Measuring
Equipment3
Up to 1100 MHz
2.5 parts × 106
Fluke 5522A/SC1100
Frequency – Measure3
40 Hz to 300 kHz
0.014 % Indication
HP 34401
Timers & Stopwatches3
Up to 1 hr
(> 1 to 24) hr
0.14 s
0.59 s
Reference stopwatch
Stroboscopes3
Up to 3600 FPM
Up to 30 000 FPM
0.11 FPM
0.72 FPM
Universal counter | ||
| 26 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 26 of 37
SATELLITE
ALDINGER COMPANY
8023 Interstate 30
Little Rock, AR 72209
Dustin Detten Phone: 214 638 1808
CALIBRATION
I. Chemical
Parameter/Equipment
Range
CMC2 (±)
Comments
Conductivity Meters3
1 µS/cm
10 µS/cm
100 µS/cm
1000 µS/cm
10 000 µS/cm
100 000 µS/cm
0.55 µS/cm
0.41 µS/cm
2.8 µS/cm
20 µS/cm
190 µS/cm
1900 µS/cm
Reference conductivity
standards
pH Meters3
Electrical Calibration
Probe/Meter
Standardization
(0 to 14) pH
4 pH
7 pH
10 pH
0.015 pH
0.030 pH
0.030 pH
0.030 pH
mV/pH calibrator
Reference buffer
standards
II. Dimensional
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Calipers3
Up to 24 in
(24 to 48) in
Up to 48 in
(48 to 96) in
550 µin
1100 µin
77 µin
1500 µin
Gage blocks (field)
Gage blocks (lab) | ||
| 27 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 27 of 37
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Dial & Test Indicators3
Up to 1 in
Up to 3 in
Up to 3 in
20 µin
59 µin
90 µin
Gage blocks (lab)
Gage blocks, ULM,
SuperMicrometer
10 (lab)
Gage blocks (field)
Height Gages3
Up to 48 in
Up to 48 in
280 µin
89 µin
Gage blocks (field)
Gage blocks (lab)
Levels3
Up to 12 in
(40 + 14L) µin
Surface plate, granite square
Micrometers3
Up to 4 in
(4 to 24) in
(24 to 48) in
Up to 4 in
(4 to 24) in
(24 to 96) in
92 µin
550 µin
1100 µin
15 µin
60 µin
1500 µin
Gage blocks (field)
Gage blocks (lab)
Protractors3
Up to 90°
0.049°
Surface plate, angle blocks,
sine bar w/ gage blocks
Surface Plates – Flatness
& Repeatability3
Indicator Stands
Up to 107 in DL
(-0.002 to 0.002) in
(-0.002 to 0.002) in
(-0.01 to 0.01) in
(48 + 0.41DL) µin
31 µin
45 µin
580 µin
Leveling system (DL is
diagonal of the plate)
Repeat reading gage
Repeat reading gage,
triangular gage block base
Indicator, triangular gage
block base
Optical Comparators3
X-Axis
Y-Axis
Angle
Up to 12 in
Up to 12 in
Up to 90°
(250 + 91L) µin
(250 + 91L) µin
0.029°
Glass scale
Angle blocks
Rulers3
Up to 72 in
0.013 in
Standard ruler | to 107 in |
| 28 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 28 of 37
Parameter/Equipment
Range
CMC2, 4 (±)
Comments
Radius Gages3
Up to 0.75 in
1300 µin
Optical comparator & overlay
Angle Gages3
Up to 180°
0.044°
Optical comparator & vernier
Coating/Electromagnetic
Thickness3
Up to 120 mils
0.25 mils
Shims
Tape Measures3
Up to 40 in
Up to 100 ft
0.013 in
0.062 in
Standard ruler
Standard tape
III. Electrical – DC/Low Frequency
Parameter/Range
Frequency
CMC2, 6, 12 (±)
Comments
AC Current – Measure3
Up to 1 A
(1 to 3) A
60 Hz to 1 kHz
0.16 % + 0.4 mA
0.14 % + 1.8 mA
HP 34401A
AC Current – Generate3
(29 to 330) µA
330 µA to 3.3 mA
(3.3 to 33) mA
(33 to 330) mA
(0.33 to 3) A
(3 to 10) A
(10 to 20) A
(16 to 150) A
(150 to 1000) A
45 Hz to 1 kHz
(45 to 440) Hz
0.13 % + 0.1 µA
0.10 % + 0.15 µA
0.07 % + 2 µA
0.083 % + 20 µA
0.10 % + 100 µA
0.21 % + 2 mA
0.21 % + 5 mA
1 % + 0.25 A
1 % + 0.9 A
Fluke 5522A
Using 50 turn coil
AC Voltage – Generate3
(1 to 330) mV
(0.330 to 3.3) V
(3.3 to 33) V
(33 to 330) V
(330 to 1020) V
45 Hz to 10 kHz
45 Hz to 10 kHz
45 Hz to 10 kHz
45 Hz to 1 kHz
(1 to 10) kHz
45 Hz to 1 kHz
(1 to 5) kHz
0.018 % + 8 µV
0.018 % + 60 µV
0.017 % + 0.6 mV
0.023 % + 2 mV
0.023 % + 6 mV
0.046 % + 10 mV
0.040 % + 10 mV
Fluke 5522A | ||
| 29 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 29 of 37
Parameter/Range
Frequency
CMC2, 6, 12 (±)
Comments
AC Voltage – Measure3
Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 750) V
Up to 10 kV
Up to 50 kV
60 Hz to 3 kHz
0.063 % + 40 µV
0.063 % + 300 µV
0.062 % + 3 mV
0.064 % + 30 mV
0.075 % + 230 mV
0.12 % + 100 mV
0.1 % + 400 mV
HP 34401A
Vitrek 4700 w/ HVL-70
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
Capacitance – Generate3
Low
High
(0.22 to 3.3) nF
(3.3 to 33) nF
(33 to 330) nF
(330 to 3300) nF
(3.3 to 33) µF
(33 to 330) µF
(330 to 3300) µF
(3.3 to 110) mF
0.5 % + 10 pF
0.26 % + 100 pF
0.26 % + 300 pF
0.26 % + 3 nF
0.4 % + 30 nF
0.45 % + 300 nF
0.45 % + 3 µF
1.1 % + 10 µF
Fluke 5522A
DC Current – Generate3
(0 to 330) µA
Up to 3.3 mA
Up to 33 mA
Up to 330 mA
Up to 3 A
(1.1 to 11) A
(11 to 20.5) A
(10 to 150) A
(150 to 1000) A
0.015 % + 20 nA
0.010 % + 50 nA
0.014 % + 250 nA
0.010 % + 2.5 µA
0.038 % + 40 µA
0.050 % + 500 µA
0.1 % + 750 µA
0.5 % + 0.14 A
0.5 % + 0.5 A
Fluke 5522A
Using 50 turn coil
DC Current – Measure3
Up to 10 mA
(10 to 100) mA
100 mA to 1 A
(1 to 3) A
0.075 % + 6 µA
0.051 % + 5 µA
0.11 % + 100 µA
0.13 % + 600 µA
HP 34401A
DC Voltage – Generate3
(0 to 330) mV
Up to 3.3 V
Up to 33 V
(30 to 330) V
(100 to 1000) V
0.002 % + 1 µV
0.0012 % + 2 µV
0.0013 % + 20 µV
0.0018 % + 0.15 mV
0.0019 % + 1.5 mV
Fluke 5522A | to 750 V |
| 30 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 30 of 37
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
DC Voltage – Measure3
Up to 100 mV
100 mV to 1 V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
Up to 10 kV
Up to 70 kV
0.0054 % + 3.5 µV
0.0042 % + 7 µV
0.0037 % + 50 µV
0.0048 % + 600 µV
0.0049 % + 10 mV
0.041 % + 30 mV
0.049 % + 200 mV
HP 34401A
Vitrek 4700 w/ HVL-
70
Oscilloscopes3
Square Wave Amplitude:
50 Ω at 1 kHz
1 MΩ at 1 kHz
DC Voltage Amplitude:
50 Ω Load
1 MΩ Load
Level Sine Wave:
Frequency
Amplitude
Flatness
Time Markers:
50 Ω Load
Rise Time – Generate
1 mV to 6.6 V pk-pk
1 mv to 130 V pk-pk
Up to ± 6.6 V
Up to ± 130 V
Up to 1100 MHz
50 kHz Reference
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
5 s to 50 ms
20 ms to 1 ns
1 kHz to 2 MHz
(200 to 300) ps
(2 to 10) MHz
(200 to 300) ps
0.34 % + 40 µV
0.25 % + 40 µV
0.25 % + 40 µV
0.05 % + 40 µV
2.5 parts × 10
6
2.8 % + 300 µV
4.2 % + 300 µV
4.6 % + 300 µV
6.4 % + 300 µV
7.4 % + 300 µV
2.7 % + 100 µV
3.0 % + 100 µV
4.6 % + 100 µV
5.5 % + 100 µV
28 + 1000t µs/s
3.2 µs/s
120 ps
120 ps
Fluke 5522A/
SC1100
t = time in seconds | to 70000 V |
| 31 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 31 of 37
Parameter/Equipment
Range
CMC2, 6, 12 (±)
Comments
Electrical Calibration of
Temperature Controllers3
Thermocouple types J, K, T
& E
PT385, 100 Ω
(-200 to < 0) °C
(0 to 1370) °C
(-200 to 800) °C
1.2 °C
0.85 °C
0.66 °C
Fluke 725
Resistance – Generate3
Up to 330 Ω
(330 to 3300) Ω
(3300 to 33 000) Ω
(33 to 330) kΩ
(330 to 3300) kΩ
(3.3 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(330 to 1100) MΩ
0.065 % + 20 mΩ
0.029 % + 100 mΩ
0.029 % + 1.0 Ω
0.0037 % + 10 Ω
0.0063 % + 150 Ω
0.026 % + 2500 Ω
0.063 % + 3.0 kΩ
0.14 % + 100 kΩ
0.14 % + 500 kΩ
Fluke 5522A
Resistance – Measure3
Up to 100 Ω
100 Ω to 1 kΩ
(1 to 10) kΩ
(10 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
0.066 % + 4 mΩ
0.031 % + 10 mΩ
0.031 % + 100 mΩ
0.011 % + 1 Ω
0.012 % + 10 Ω
0.048 % + 100 Ω
0.8 % + 10 kΩ
HP 34401A
Electrical Simulation of
RTD3
PT385, 100 Ω
(-200 to 0) °C
(0 to 100) °C
(100 to 400) °C
(400 to 630) °C
(630 to 800) °C
0.058 °C
0.17 °C
0.12 °C
0.14 °C
0.28 °C
Fluke 5522A | ||
| 32 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 32 of 37
Parameter/Equipment
Range
CMC2 (±)
Comments
Electrical Simulation of
Thermocouples &
Indicators3
Type E
Type J
Type K
Type T
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
(-250 to -150) °C
(-150 to 0.0) °C
(0.0 to 120) °C
(120 to 400) °C
0.51 °C
0.16 °C
0.19 °C
0.20 °C
0.24 °C
0.28 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
0.33 °C
0.18 °C
0.20 °C
0.24 °C
0.40 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
Fluke 5522A
Welding Devices3
(0 to 350) Amps DC
(0 to 100) Volts DC
(100 to 700) Feed Rate IPM
0.87 ADC
0.012 VDC
3.6 IPM
Load bank, current
shunt, DMM
IV. Fluid Quantities
Parameter/Equipment
Range
CMC2 (±)
Comments
POVA (Piston/Plunger
Operated Volumetric
Apparatus)3 –
Including But Not
Limited To: Pipettes,
Syringes, Dispensers, &
Burettes
(0.1 to 50) µL
(50 to 100) µL
(100 to 500) µL
(500 to 1000) µL
(1000 to 5000) µL
(5000 to 10 000) µL
(10 000 to 20 000) µL
(20 000 to 50 000) µL
0.31 µL
0.31 µL
0.56 µL
0.72 µL
5.4 µL
18 µL
30 µL
72 µL
Gravimetric method | ||
| 33 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 33 of 37
V. Mechanical
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Balances3
Up to 20 g
Up to 200 g
Up to 200 g
Up to 1000 g
Up to 5000 g
Up to 20 g
Up to 200 g
Up to 1000 g
Up to 5000 g
Up to 10 000 g
Up to 20 000 g
0.11 mg
0.69 mg
1.3 mg
6.6 mg
34 mg
0.81 mg
4.8 mg
24 mg
130 mg
370 mg
460 mg
Class 1 weights
Class 2 weights
Class 4 weights
Force – Measuring
Equipment3
Tension/Compression
Compression
Tension
Up to 100 lbf
(100 to 1000) lbf
(1000 to 10 000) lbf
(10 000 to 50 000) lbf
(10 000 to 100 000) lbf
(50 000 to 500 000) lbf
(100 to 1000) lbf
(1000 to 10 000) lbf
(10 000 to 50 000) lbf
(10 000 to 100 000) lbf
0.0064 lbf
0.26 % Indication
0.24 % Indication
0.33 % Indication
0.20 % Indication
0.4 % Indication
0.23 % Indication
0.23 % Indication
0.29 % Indication
0.20 % Indication
Class F weights w/
load cells | to 20000 g |
| 34 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 34 of 37
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Optical Tachometer3
Up to 200 FPM
Up to 3000 FPM
Up to 29 999 FPM
0.12 FPM
0.39 FPM
3.9 FPM
Calibrated strobe
(FPM = Flashes per
minute)
Pressure Measuring
Equipment – Gages,
Transducers &
Transmitters3
Up to 300 psig
Up to 3000 psig
Up to 5000 psig
Up to 10 000 psig
Up to 30 000 psig
Up to 2 in H
2O
Up to 300 psig
Up to 3000 psig
Up to 10 000 psig
(10 to 17) psia
0.18 psig
3.8 psig
3.9 psig
12 psig
43 psig
0.0026 in H
2O
0.051 psig
0.4 psig
1.3 psig
0.000 78 psia
Digital pressure
gage
Dwyer hook gage
GE/Pace 1003
Barometric
reference
Scales3
Up to 10 lb
Up to 20 lb
Up to 50 lb
Up to 100 lb
Up to 200 lb
Up to 500 lb
Up to 1000 lb
Up to 5000 lb
Up to 20 000 lb
0.000 25 lb
0.000 25 lb
0.0063 lb
0.0063 lb
0.0064 lb
0.0064 lb
0.18 lb
0.19 lb
0.21 lb
Class 4 weights
Class F weights | ||
| 35 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 35 of 37
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Torque Wrenches –
Measure3
(1 to 10) lbf⋅in
(10 to 50) lbf⋅in
(30 to 400) lbf⋅in
(80 to 1000) lbf⋅in
(240 to 3000) lbf⋅in
(1200 to 6000) lbf⋅in
(2400 to 12 000) lbf⋅in
0.33 lbf⋅in
0.17 lbf⋅in
1.8 lbf⋅in
5.1 lbf⋅in
21 lbf⋅in
23 lbf⋅in
58 lbf⋅in
CDI/Norbar torque
calibrator
Norbar torque
calibrator
Vacuum Measuring
Equipment – Gauges
3
Up to 13 psig
0.033 psig
Druck DPI 610
Rotations Speed –
Measurement3
Contact:
Up to 200 RPM
(200 to 3000) RPM
(3000 to 20 000) RPM
Optical:
Up to 3000 RPM
(3000 to 30 000) RPM
0.67 RPM
13 RPM
16 RPM
0.5 RPM
5.0 RPM
Monarch PLT200
Indirect Verification of
Rockwell Hardness
Testers3
HRB:
Low
Medium
High
HRC:
Low
Medium
High
1.0 HRB
0.65 HRB
0.49 HRB
0.45 HRC
0.36 HRC
0.37 HRC
Indirect verification
Indirect Verification of
Brinell Hardness Testers3
HBW 10/500:
Low
HBW 10/3000:
High
4.0 HBW
4.8 HBW
Indirect verification | ||
| 36 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 36 of 37
VI. Thermodynamic
Parameter/Equipment
Range
CMC2, 11 (±)
Comments
Temperature – Measuring
Equipment3
(-100 to 300) °C
(-148 to 572) °F
0.09 °C
0.16 °F
Hart 1502A/5614 PRT or
Fluke 1523
Temperature –Measure3
(-100 to 600) °C
(-148 to 1112) °F
(> 600 to 1200) °C
(> 1112 to 2192) °F
1.5 °C
2.7 °F
10 °C
18 °F
Fluke 51 or 725 with
thermocouple
IR Thermometer3
(-20 to 150) °C
(Amb + 10 to 400) °C
1.1 °C
2.1 °C
Mikron M340 (lab only)
Omega BB703
Emissivity ≈ 0.95
Wavelength (8 to 14) µm
Humidity – Measure &
Measuring Equipment3
(20 to 90) % RH
3.4 % RH
SensoScientific B13-200
VII. Time & Frequency
Parameter/Equipment
Range
CMC2, 6, 11 (±)
Comments
Frequency – Measuring
Equipment3
Up to 1100 MHz
2.5 parts × 106
Fluke 5522A/SC1100
Frequency – Measure3
40 Hz to 300 kHz
0.014 % Indication
HP 34401
Timers & Stopwatches3
Up to 1 hr
(> 1 to 24) hr
0.14 s
0.59 s
Reference stopwatch
Stroboscopes3
Up to 3600 FPM
Up to 30 000 FPM
0.11 FPM
0.72 FPM
Universal counter
__________________________________________ | ||
| 37 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1509.01) Revised 07/09/2025 Page 37 of 37
1 This laboratory offers commercial calibration service and field calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. , resolution,
repeatability) must also be considered and this, on its own, could result in the actual measurement
uncertainty achievable on a customer’s site being larger than the CMC.
4 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches
and R is the numerical value of the resolution of the device in microinches unless otherwise noted.
5 Adjustable thread rings are set to applicable specification using calibrated master set plug gages.
6 CMC for the HP 34401 is based on 1- year specifications within a temperature range of 18 ° C to 28 °C.
Field calibrations will be performed within 18 °C to 28 °C, 30 % to 55 % humidity.
7 Where ranges are not specified, the CMC stated is for the cardinal points only.
8 In the statement of CMC, percentage (%) refers to percent of reading, unless otherwise noted.
9 "Supermicrometer" is a registered trademark with a last listed owner of Pratt & Whitney Measurement
Systems, Inc., Connecticut U.S.A.
10 This scope meets A2LA’s P112 Flexible Scope Policy.
11 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
12 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.
13 Aldinger Company is also doing business as Utah Scale Center for field calibration services, with
administrative offices located at 199 W Cottage Avenue, Sandy, UT 84070. | ||
| 38 | ISO/IEC 17025:2017; ISO-ILAC-IAF Communiqué dated April 2017; IEC 17025; ANSI/NCSL Z540-1-1994 and R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Other Locations
1 other site in this organization
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.