University of Wisconsin Radiation
Room B1002, WIMR, 1111 Highland Avenue, 53705-2275, Madison, WI, United States
- Scope lines
- 5
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 5 scope lines.
A2LA certificate 1664.01 · 5 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1664.01) 01/20/2025 Page 1 of 4
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY
Room B1002, WIMR
1111 Highland Avenue
Madison, WI 53705-2275
Larry A. DeWerd, Ph.D. Phone: 608 262 0378
CALIBRATION
Valid To: January 31, 2027 Certificate Number: 1664.01
In recognition of the successful completion of the A2LA evaluation process, (including an assessment of
the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is
granted to this laboratory to perform the following calibrations1, 5:
I. Ionizing Radiation and Dosimetry
Parameter/Equipment
Range
CMC2, 3 (±)
Comments
Ionization Chambers/
Therapy Class –
60Co Air Kerma
60Co Dose to Water
137Cs Air Kerma
X-Rays
Up to 150 cGy/min
Up to 150 cGy/min
Up to 20 cGy/min
(20 to 250) kVp
1.5 %
1.4 %
1.6 %
1.0 %
NIST calibrated reference class
ionization chambers
Ionization Chambers/
Diagnostic Class –
X-Rays:
Tungsten Anode
Molybdenum Anode
(20 to 250) kVp
(23 to 35) kVp
1.9 %
1.9 %
NIST calibrated reference class
ionization chambers
Ionization Chambers/
Health Physics Radiation
Protection –
60Co Air Kerma
137Cs Air Kerma
X-Rays
Up to 150 cGy/min
Up to 20 cGy/min
(20 to 250) kVp
1.5 %
1.6 %
1.9 %
NIST calibrated reference class
ionization chambers | to 150 cGy/min |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1664.01) 01/20/2025 Page 2 of 4
Parameter/Equipment
Range
CMC2, 3 (±)
Comments
Well-Type
Brachytherapy Ionization
Chambers –
HDR: 192Ir
LDR
6: 192Ir, 125I, 103Pd,
137Cs, 131Cs
(1 U = 1 µGym2/hr)
Up to 50 kU
Up to 500 U
2.8 %
2.7 %
NIST calibrated
137Cs
brachytherapy sources;
reference class well-type
ionization chambers calibrated
against NIST calibrated
sources for short-lived
isotopes such as
192Ir, 125I,
131Cs, and 103Pd
Radioactive
Brachytherapy
Sources –
LDR6: 137Cs, 131Cs,
192Ir, 125I, 103Pd
Up to 500 U
1.9 %
NIST calibrated Cs
137
brachytherapy sources,
reference class well-type
ionization chambers calibrated
against NIST calibrated
sources for short-lived
isotopes such as
192Ir, 125I,
131Cs, and 103Pd
Well-Type Intra-
Vascular Ionization
Chambers –
90Sr Absorbed Dose
Up to 200 mGy/s
15 %
NIST traceable absorbed dose
to water source calibrations
Electrometers –
Charge (Coulomb)
Current (Ampere)
25 pC to 10 000 nC
1 pA to 1000 nA
0.21 %
0.21 %
NIST traceable time, voltage,
and resistance standards | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1664.01) 01/20/2025 Page 3 of 4
Parameter/Equipment
Range
CMC2, 3 (±)
Comments
Non-Invasive kVp
Meters –
Radiographic
Mammographic
(50 to 150) kVp
(25 to 35) kVp
1.0 %
1.0 %
NIST traceable multi-nuclide
source and calibrated x-ray
units
Radiation Protection
Survey Instruments –
137Cs Exposure
0.02 mR/hr to 20 R/hr
6.5 %
NIST calibrated reference class
ionization chamber used to
calibrate radiation beam air
kerma or exposure rate
TLD4, Radiochromic
Film and Artifact
Irradiation –
60Co
137Cs
X-Ray
High Energy Photons
(6 to 18) MV and
Electrons (4 to 16)
MeV
Up to 150 cGy/min
Up to 20 cGy/min
(20 to 250) kVp
Up to 2400 cGy/min
3.6 %
3.6 %
3.6 %
3.6 %
NIST calibrated reference class
ionization chambers calibrated
in reproduced beams using
replacement technique
TLD4 Services –
Dose Verification
Dose Determination
On-Patient, or Custom
Geometry
3.6 %
5.0 %
NIST calibrated reference class
ionization chamber for
calibration of radiation beam
air kerma, dose, or exposure
Well-Type Ionization
Chambers/Electronic
Brachytherapy Sources
Up to 5 mGy/s
4.0 %
NIST traceable air kerma rate
for electronic brachytherapy
sources
_____________________
1 This laboratory offers commercial calibration service. | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1664.01) 01/20/2025 Page 4 of 4
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavio r of the customer’s device and to influences from the
circumstances of the specific calibration.
3 In the statement of CMC, percentages are percentage of reading, unless otherwise indicated.
4 TLD is Thermoluminescent Dosimeter.
5 This scope meets A2LA’s P112 Flexible Scope Policy.
6 Derived from uncertainty budget for 103Pd, 125I, and 131Cs, which are the most common LDR isotopes.
Source directionality may increase the uncertainty. | ||
| 5 | ISO/IEC 17025:2017; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.