Registers refreshed 2026-11-20.See refresh dates
United States

Test & Measurement

7931 Bond Street, 66214, Lenexa, KS, United States

Visit website
A2LA 6940.01ISO/IEC 17025Register refreshed 2026-11-20

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free, no obligation.
Scope lines
18
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 18 scope lines.

A2LA certificate 6940.01 · 18 lines

LineMethodMatrixRange
1ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994-0.01 to 12 in
2

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 2 of 17 Parameter Range CMC2, 7 (±) Comments Length Standards3 (1 to 12) in (12-36)” (6.9 + 0.22L) µin (10 + 0.25L) µin Labmaster & master gage blocks Gage blocks & indicator Micrometers3 Up to 24 in (Resolution = 50 µin) Up to 24 in (Resolution = 100 µin) Up to 24 in (Resolution = .001 in) 30 µin 58 µin 580 µin Gage blocks Calipers3 Up to 36 in (290 + 4L) µin Gage blocks Height Gages3 Up to 24 in 61 µin Gage blocks Thread Plugs – Major Diameter Pitch Diameter (0.060 to 4.0) in (0.060 to 4.0) in (17 + 2.7L) µin ± 72 µin Labmaster & master gage blocks Labmaster, thread wires, & master gage blocks Thread Rings, Pitch Diameter (0.060 to 4.0) in ± 72 µin Set thread plugs Thread Wires (4 to 80) TPI ± 7.4 µin Labmaster, master gage blocks Digital Protractors Bevel Protractors (0 to 180) ° (0 to 180) ° 0.27 ° 0.64 ° Plate & gage blocks Angle Measurements (0 to 45) ° 5.3 Arc Seconds Sine plate, gage blocks & indicator
3

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 3 of 17 Parameter Range CMC2,7 (±) Comments Surface Plates Repeat Flatness ± 0.001 Up to 161 in Diagonal ± 28 µin ± (25 + 0.5√D 8) µin Repeat-o-meter Auto collimator Optical Comparators3 Magnification Linearity Angle Up to 12 in ±71 µin ±66 µin ±1.1 ° Glass scale Glass scale Angle blocks Tape Measures Up to 100’ 58 µin Comparison to gage blocks II. Electrical – DC/Low Frequency Parameter Range CMC2, 5 (±) Comments DC Voltage – Generate3 Up to 220 mV 220 mV to 2.2 V (2.2 to 11) V (11 to 22) V (22 to 220) V (220 to 1100) V 6.3 μV/V + 0.61 μV 4.3 μV/V + 1.7 μV 2.6 μV/V + 9.3 μV 2.5 μV/V + 21 μV 4.4 μV/V + 0.13 mV 5.5 μV/V + 1.4 mV Fluke 5730A DC Voltage – Measure3 Up to 120 mV (0.12 to 1.2) V (1.2 to 12) V (12 to 120) V (120 to 1050) V (0.700 to 150) kV 9 μV/V + 0.31 μV 8 μV/V + 0.45 μV 8 μV/V + 2 μV 10 μV/V + 50 μV 10 μV/V + 200 μV 5 mV/V + 3.6 V Agilent 3458A High Voltage Inc. DVR- 150 with Agilent 3458A
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 4 of 17 Parameter Range CMC2, 5 (±) Comments DC Current – Generate3 Up to 220 μA (220 to 2.2) mA (2.2 to 22) mA (22 to 100) mA (100 to 220) mA 220 mA to 2.2 A (2.2 to 11) A (11 to 20.5) A (20 to 1000) A 38 μA/A + 5.9 nA 27 μA/A + 17 nA 26 μA/A + 0.18 μA 34 μA/A + 1 μA 26 μA/A + 2.6 μA 52 μA/A + 37 μA 0.19 mA/A + 0.66 mA 0.15 mA/A + 35 mA 5.5 mA/A + 0.25 A Fluke 5730A with 5725A Fluke 5520A Fluke 5520A with 50 turn coil DC Current – Measure3 DC Apparent Power – Generate3 PF=1 Up to 12 μA (12 to 120) μA 120 μA to 1.2 mA (1.2 to 12) mA (12 to 120) mA 120 mA to 1.05 A (1.05 to 20) A (20 to 100) A (100 to 1000) A (2 to 3059) W (3.06 to 20.91) kW 35 μA/A + 5.8 nA 6.7 μA/A + 5.8 nA 17 μA/A + 9.4 nA 17 μA/A + 86 nA 33 μA/A + 0.77 μA 0.11 mA/A + 12 μA ± 0.10 % of rdg ± 0.13 % of rdg ± 0.51 % of rdg ± 0.017 % of rdg ± 0.048 % of rdg Agilent 3458A Valhalla 2575A & 3458A Passive shunts & 34401A Fluke 5520A
5--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 5 of 17 Parameter/Range Frequency CMC2, 5 (±) Comments AC Current – Generate3 Up to 220 μA 220 μA to 2.2 mA (2.2 to 22) mA (22 to 220) mA 220 mA to 2.2 A (2.2 to 11) A (20 to 1000) A (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 20 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 45 Hz to 65 Hz 100 Hz to 440 Hz 0.29 mA/A + 23 nA 0.19 mA/A + 15 nA 0.11 mA/A + 14 nA 0.3 mA/A + 31 nA 1.2 mA/A + 0.12 μA 0.3 mA/A + 97 nA 0.19 mA/A + 76 nA 0.11 mA/A + 76 nA 0.23 mA/A + 0.19 μA 1.3 mA/A + 0.90 μA 0.28 mA/A + 1.0 μA 0.19 mA/A + 0.87 μA 0.11 mA/A + 0.88 μA 0.2 mA/A + 1.8 μA 1.3 mA/A + 8.5 μA 0.3 mA/A + 13 μA 0.18 mA/A + 9.6 μA 0.1 mA/A + 9.1 μA 0.19 mA/A + 17 μA 1.2 mA/A + 62 μA 0.26 mA/A + 0.16 mA 0.43 mA/A + 0.30 mA 7.6 mA/A + 1.2 mA 0.4 mA/A + 0.93 mA 0.87 mA/A + 1.4 mA 3.6 mA/A + 1.5 mA 3.2mA/A + 87 mA 8 mA/A + 0.19 A Fluke 5730A Fluke 5730A w/ Fluke 5725A amplifier Fluke 5520A with 50 turn coil
to 1000 A
6

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 6 of 17 Parameter/Range Frequency CMC2, 5 (±) Comments AC Current – Measure3 (0.6 to 120) μA 120 μA to 1.2 mA (1.2 to 12) mA (12 to 120) mA 120 mA to 1.05 A (1.05 to 3) A (3 to 100) A (10 to 20) Hz (20 to 45) Hz 45 Hz to 1 kHz (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz (10 to 20) Hz (20 to 45) Hz (45 to 100) Hz 100 Hz to 5 kHz 45 Hz to 1 kHz 45 Hz to 1 kHz 4 mA/A + 33 nA 1.5 mA/A + 33 nA 0.59 mA/A + 33 nA 4 mA/A + 0.22 μA 1.5 mA/A + 0.22 μA 0.6 mA/A + 0.22 μA 0.3 mA/A + 0.22 μA 4 mA/A + 2.2 μA 1.5 mA/A + 2.2 μA 0.6 mA/A + 2.2 μA 0.3 mA/A + 2.2 μA 4 mA/A + 22 μA 0.6 mA/A + 22 μA 0.6 mA/A + 22 μA 0.3 mA/A + 22 μA 4 mA/A + 0.23 mA 1.6 mA/A + 0.23 mA 0.78 mA/A + 0.23 mA 0.98 mA/A + 0.23 mA ± 0.33 % of rdg ± 0.22 % of rdg Agilent 3458A Keysight 34401A Valhalla 2575A with 3458A
7--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 7 of 17 Parameter/Range Frequency CMC2, 5 (±) Comments AC Voltage – Generate3 Up to 2.2 mV (2.2 to 22) mV (22 to 220) mV 220 mV to 2.2 V (2.2 to 22) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz (40 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz 0.3 mv/V + 5.1 μV 0.12 mV/V + 5.1 μV 0.10 mV/V + 5.1 μV 0.25 mV/V + 5.1 μV 0.59 mV/V + 6.1 μV 1.3 mV/V + 12 μV 1.7 mV/V + 25 μV 3.3 mV/V + 26 μV 0.3 mV/V + 5.2 μV 0.12 mV/V + 5.2 μV 0.1 mV/V + 5.2 μV 0.25 mV/V + 5.3 μV 0.58 mV/V + 6.7 μV 1.3 mV/V + 14 μV 1.6 mV/V + 28 μV 2.8 mV/V + 48 μV 0.3 mV/V + 17 μV 0.12 mV/V + 9 μV 0.07 mV/V + 9 μV 0.15mV/V + 11 μV 0.4 mV/V + 23 μV 0.73 mV/V + 44 μV 1.6 mV/V + 68 μV 2.7 mV/V + 0.23 mV 0.3 mV/V + 70 μV 0.11 mV/V + 42 μV 0.05 mV/V + 25 μV 0.08 mv/V + 35 μV 0.1 mV/V + 61 μV 0.37 mV/V + 0.20 mV 1.2 mV/V + 0.38 mV 1.8 mV/V + 1.1 mV 0.29 mV/V + 0.85 mV 0.1 mV/V + 0.52 mV 0.04 mV/V + 0.31 mV 0.08 mV/V + 0.36 mV 0.09 mV/V + 0.54 mV 0.26 mV/V + 2.2 mV 1.2 mV/V + 4.7 mV 1.4 mV/V + 17 mV Fluke 5730A
to 2.2 V
8

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 8 of 17 Parameter/Range Frequency CMC2, 5 (±) Comments AC Voltage – Generate3 (cont) (22 to 220) V (220 to 750) V (220 to 1100) V (1.2 to 12) mV (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (30 to 50) kHz (50 to 100) kHz (1 to 20) kHz (20 to 30) kHz (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 0.29 mV/V + 0.85 mV 0.10 mV/V + 0.52 mV 0.04 mV/V + 0.31 mV 0.08 mV/V + 0.36 mV 0.09 mV/V + 0.54 mV 0.26 mV/V + 2.2 mV 1.2 mV/V + 4.7 mV 1.4 mV/V + 17 mV 0.59 mV/V + 22 mV 2.3 mV/V + 49 mV 0.15 mV/V + 25 mV 0.59 mV/V + 27 mV 0.28 mV/V + 3.3 μV 0.19 mV/V + 1.3 μV 0.27 mV/V + 1.6 μV 0.97 mV/V + 1.6 μV 5 mV/V + 2 μV 40 mV/V + 3.5 μV Fluke 5730A Agilent 3458A AC Voltage – Measure3 (1.2 to 12) mV (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 0.28 mV/V + 3.3 μV 0.19 mV/V + 1.3 μV 0.27 mV/V + 1.6 μV 0.97 mV/V + 1.6 μV 5 mV/V + 2 μV 40 mV/V + 3.5 μV Agilent 3458A
9

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 9 of 17 Parameter/Range Frequency CMC2, 5 (±) Comments AC Voltage – Measure3 (cont) (12 to 120) mV 120 mV to 1.2 V (1.2 to 12) V (12 to 120) V (120 to 700) V (0.700 to 150) kV AC Apparent Power – Generate 3 PF=1 2 W to 20.91 kW (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz 300 kHz to 1 MHz (1 to 2) MHz (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (1 to 2) MHz (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (1 to 2) MHz (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz 1 to 40 Hz 40 Hz to 1 kHz 1 to 20 kHz (20 to 50) kHz (50 to 100) kHz 40 Hz to 1 kHz (45 to 65) Hz 65 μV/V + 4.9 μV 66 μV/V + 4.8 μV 0.14 mV/V + 2.3 μV 0.29 mV/V + 3.7 μV 0.77 mV/V + 7 μV 3 mV/V + 13 μV 10 mV/V + 16 μV 15 mV/V + 52 μV 68 μV/V + 43 μV 67 μV/V + 25 μV 0.14 mV/V + 26 μV 0.3 mV/V + 29 μV 0.8 mV/V + 29 μV 3 mV/V + 0.11 mV 10 mV/V + 0.11 mV 15 mV/V + 0.28 mV 64 μV/V + 0.52 mV 64 μV/V + 0.31 mV 0.14 mV/V + 0.2 mV 0.3 mV/V + 0.26 mV 0.8 mV/V + 0.31 mV 3 mV/V + 1.1 mV 10 mV/V + 1.2 mV 15 mV/V + 3.2 mV 0.2 mV/V + 4.3 mV 0.2 mV/V + 2.6 mV 0.2 mV/V + 2.8 mV 0.35 mV/V + 2.8 mV 1.2 mV/V + 3.2 mV 4 mV/V + 11 mV 15 mV/V + 11 mV 0.4 mV/V + 43 mV 0.4 mV/V + 26 mV 0.6 mV/V + 28 mV 1.2 mV/V + 28 mV 3 mV/V + 28 mV 18 mV/V + 3.6 V ± 0.09 % of rdg Agilent 3458A High Voltage Inc. DVR- 150 with Agilent 3458A Fluke 5520A
10--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 10 of 17 Parameter/Equipment Range CMC2, 5 (±) Comments DC Resistance – Generate, Fixed Points3 1 Ω 1.9 Ω 10 Ω 19 Ω 100 Ω 190 Ω 1 kΩ 1.9 kΩ 10 kΩ 19 kΩ 100 kΩ 190 kΩ 1 MΩ 1.9 MΩ 10 MΩ 19 MΩ 100 MΩ 0.11 mΩ 0.21 mΩ 0.28 mΩ 0.52 mΩ 1.3 mΩ 2.4 mΩ 9 mΩ 16 mΩ 80 mΩ 16 Ω 1 Ω 2.3 Ω 19 Ω 42 Ω 0.47 kΩ 1.1 kΩ 12 kΩ Fluke 5730A DC Resistance – Measure 3 Up to 0.1 Ω (0.1 to 12) Ω (12 to 120) Ω 120 Ω to 1.2 kΩ (1.2 to 12) kΩ (12 to 120) kΩ 120 kΩ to 1.2 MΩ (1.2 to 12) MΩ (12 to 120) MΩ 120 MΩ to 1.2 GΩ 0.11 mΩ/Ω + 2 μΩ 15 μΩ/Ω + 54 μΩ 11 μΩ/Ω + 0.57 mΩ 9 μΩ/Ω + 1.7 mΩ 9 μΩ/Ω + 17 mΩ 8.9 μΩ/Ω + 0.19 Ω 13 μΩ/Ω + 4.8 Ω 41 μΩ/Ω + 0.13 kΩ 0.45 mΩ/Ω + 7 kΩ 4.5 mΩ/Ω + 0.6 MΩ Agilent 34420A & Agilent 3458A DC Resistance – Generate 3 Up to 11 Ω (11 to 33) Ω (33 to 110) Ω (110 to 330) Ω 330 Ω to 1.1 kΩ (1.1 to 3.3) kΩ (3.3 to 11) kΩ (11 to 110) kΩ (110 to 330) kΩ 330 kΩ to 1.1 MΩ (1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ 330 MΩ to 1.1 GΩ 41 μΩ/Ω + 1.1 mΩ 32 μΩ/Ω + 1.6 mΩ 28 μΩ/Ω + 1.6 mΩ 29 μΩ/Ω + 2.1 mΩ 26 μΩ/Ω + 6.2 mΩ 29 μΩ/Ω + 21 mΩ 29 μΩ/Ω + 22 mΩ 29 μΩ/Ω + 0.22 Ω 33 μΩ/Ω + 2.1 Ω 33 μΩ/Ω + 2.3 Ω 61 μΩ/Ω + 32 Ω 0.14 mΩ/Ω + 56 Ω 0.24 mΩ/Ω + 3.2 kΩ 0.51 mΩ/Ω + 3.1 kΩ 3.1 mΩ/Ω + 0.11 MΩ 16 mΩ/Ω + 0.55 MΩ Fluke 5520A
to 1100000000 Ω
11--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 11 of 17 Parameter/Equipment Range CMC2, 5 (±) Comments Capacitance – Generate3 10 Hz to 10 kHz Up to 400 pF 400 pF to 1.1 nF (1.1 to 3.3) nF (3.3 to 11) nF (11 to 33) nF (33 to 110) nF (110 to 330) nF 330 nF to 1.1 μF (1.1 to 3.3) μF (3.3 to 11) μF (11 to 33) μF (33 to 110) μF (110 to 330) μF 330 μF to 1.1 mF (1.1 to 3.3) mF (3.3 to 11) mF (11 to 33) mF (33 to 110) mF 0.44 % of rdg + 12 pF 0.45 % of rdg + 12 pF 0.48 % of rdg + 12 pF 0.25 % of rdg + 12 pF 0.26 % of rdg + 0.1 nF 0.25 % of rdg + 0.11 nF 0.26 % of rdg + 0.3 nF 0.26 % of rdg + 1.1 nF 0.26 % of rdg + 3 nF 0.26 % of rdg + 10 nF 0.41 % of rdg + 30 nF 0.45 % of rdg + 0.13 μF 0.47 % of rdg + 0.3 μF 0.45 % of rdg + 1.1 μF 0.45 % of rdg + 3.1 μF 0.46 % of rdg + 10 μF 0.74 % of rdg + 35 μF 1.1 % of rdg + 0.1 mF Fluke 5520A Inductance – Generate Fixed Value3 100 Hz to 10 kHz 100 mH 200 mH ± 0.14 mH ± 0.33 mH General Radio 1482-L & 1482-M (standard inductors)
10 to 10000 Hz
12

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 12 of 17 Parameter/Equipment Range CMC2 (±) Comments Electrical Simulation of Thermocouples3 – Type E Type J Type K Type N Type R Type S Type T (-250 to -100) °C (-100 to -25) °C (-25 to 350) °C (350 to 650) °C (650 to 1000) °C (-210 to -100) °C (-100 to -30) °C (-30 to 150) °C (150 to 760) °C (760 to 1000) °C (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 1000) °C (1000 to 1372) °C (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 410) °C (410 to 1300) °C Up to -250 °C (250 to 400) °C (400 to 1000) °C (1000 to 1767) °C Up to 250 °C (250 to 1000) °C (1000 to 1400) °C (1400 to 1767) °C (-250 to -150) °C (-150 °C to 0) °C Up to 120 °C (120 to 400) °C 0.51 °C 0.17 °C 0.15 °C 0.17 °C 0.22 °C 0.28 °C 0.17 °C 0.15 °C 0.18 °C 0.24 °C 0.34 °C 0.19 °C 0.17 °C 0.27 °C 0.41 °C 0.41 °C 0.23 °C 0.20 °C 0.19 °C 0.28 °C 0.58 °C 0.36 °C 0.34 °C 0.41 °C 0.48 °C 0.37 °C 0.38 °C 0.47 °C 0.64 °C 0.25 °C 0.17 °C 0.15 °C Fluke 5520A
13--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 13 of 17 Parameter/Equipment Range CMC2 (±) Comments Oscilloscopes3 – Amplitude Into 50 Ω DC Amplitude Into 50 Ω at 10 Hz to 10 kHz Amplitude Into 1 MΩ at DC Amplitude Into 1 MΩ at 10 Hz to 1 kHz Amplitude Into 1 MΩ at (1 to 10) kHz Leveled Sine Wave: 50 kHz & 100 MHz 50 kHz to 100 MHz (100 to 300) MHz (300 to 500) MHz (500 to 600) MHz Up to 110 mV 110 mV to 2.2 V (2.2 to 6.6) V Up to 25 mV (25 to 110) mV 110 mV to 2.2 V (2.2 to 6.6) V Up to 110 mV 110 mV to 2.2 V (2.2 to 11) V (11 to 130) V Up to 25 mV (25 to 110) mV 110 mV to 2.2 V (2.2 to 11) V (11 to 130) V Up to 25 mV (25 to 110) mV 110 mV to 2.2 V (2.2 to 11) V (11 to 130) V (5 to 100) mV (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 1 V (1 to 5) V 2.5 mV/V + 41 μV 2.5 mV/V + 92 μV 2.4 mV/V + 0.59 mV 2.4 mV/V + 47 μV 2 mV/V + 0.12 mV 1.7 mV/V + 2.2 mV 1.7 mV/V + 6.6 mV 0.25 mV/V + 41 μV 0.23 mV/V + 92 μV 0.21 mV/V + 0.59 mV 0.21 mV/V + 5.9 mV 0.21 mV/V + 26 μV 0.14 mV/V + 0.11 mV 0.12 mV/V + 2.2 mV 0.12 mV/V + 11 mV 0.12 mV/V + 0.13 V 0.44 mV/V + 47 μV 0.26 mV/V + 0.12 mV 0.13 mV/V + 2.2 mV 0.12 mV/V + 11 mV 0.12 mV/V + 0.13 V 20 mV/V + 0.31 mV 20 mV/V + 0.31 mV 19 mV/V + 0.87 mV 20 mV/V + 0.91 mV 35 mV/V + 0.31 mV 35 mV/V + 0.87 mV 35 mV/V + 0.91 mV 40 mV/V + 0.31 mV 39 mV/V + 0.87 mV 40 mV/V + 0.91 mV 55 mV/V + 0.31 mV 55 mV/V + 0.87 mV 55 mV/V + 0.91 mV 60 mV/V + 0.31 mV 59 mV/V + 0.87 mV 60 mV/V + 0.91 mV Fluke 5800A
to 600 MHz
14

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 14 of 17 Parameter/Equipment Range CMC2 (±) Comments Oscilloscopes3 – (cont) Leveled Sine Wave: 600 MHz to 1.1 GHz (1.1 to 1.6) GHz (1.6 to 2.1) GHz Leveled Sine Wave Flatness: 50 kHz to 100 MHz (Relative to 50 kHz) (100 to 300) MHz (300 to 500) MHz (500 to 600) MHz 600 MHz to 1.6 GHz (Relative to 10 MHz) (1.6 to 2.1) GHz Time Marker Edge Rise Time (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 1 V (1 to 5) V (5 to 100) mV 100 mV to 5 V (5 to 100) mV 100 mV to 5 V (5 to 100) mV 100 mV to 5 V (5 to 100) mV 100 mV to 5 V (5 to 100) mV 100 mV to 5 V (5 to 100) mV 100 mV to 5 V (2 to 20) ms (50 to 5) ms 300 ps 70 mV/V + 0.31 mV 69 mV/V + 0.87 mV 70 mV/V + 0.91 mV 70 mV/V + 0.31 mV 69 mV/V + 0.87 mV 70 mV/V + 0.91 mV 80 mV/V + 0.31 mV 79 mV/V + 0.87 mV 80 mV/V + 0.91 mV 15 mV/V + 0.12 mV 15 mV/V + 0.86 mV 20 mV/V + 0.12 mV 20 mV/V + 0.86 mV 35 mV/V + 0.12 mV 35 mV/V + 0.86 mV 40 mV/V + 0.12 mV 40 mV/V + 0.86 mV 50 mV/V + 0.12 mV 50 mV/V + 0.86 mV 60 mV/V + 0.12 mV 60 mV/V + 0.86 mV 1 μs/s 3 μs/s + 6 μs 100 ps Fluke 5800A
15NIST handbook105 Class F-See scope document
16

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 16 of 17 IV. Thermodynamics Parameter/Equipment Range CMC2, 4 (±) Comments Humidity – Measuring Equipment 11.30 % RH 22.51 % RH 52.89 % RH 75.29 % RH 84.34 % RH 93.58 % RH 2.8 % RH 2.8 % RH 2.8 % RH 2.8 % RH 2.8 % RH 3.3 % RH Vaisala HMI41/HMP46 w/ saturated salt solutions Temperature – Generate & Measure 3 (-80 to 25) °C (25 to 400) °C (400 to 700) °C 0.028 °C 0.000 06 °C/°C + 0.017 °C 2.1 °C Fluke 1502A w/ SPRT & Isotech drywell Fluke 5520A w/ characterized Type S thermocouple & Isotech drywell 2 V. Time & Frequency Parameter/Equipment Range CMC2, 4 (±) Comments Stopwatches, Timers & Clocks3 ≤ 1 Hour (1 to 48 Hours ± 0.85 µs/s + 0.59 ms ± 0.35 s Fluke 5800A with HP 5335B, totalize method Frequency – Measuring Equipment3 ≤ 2 MHz ± 1.9 µHz + 5 µHz Fluke 5520A Frequency – Measure3 0.01 Hz to 200 MHz ± 0.58 nHz/Hz + 5.8 pHz Arbiter 1206B & frequency counter Period – Measure3 100 µs to 72 Hours ± 0.57 ns/s + 150 ns Arbiter 1206B & frequency counter _________________________________________ 1 This laboratory is available for commercial calibration and field calibration service.
17

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 6940.01) 7/31/2024 Page 17 of 17 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. 3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC. 4 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter. 5 The stated measured values are determined using the indicated instrument (see Comments). This capability is suitable for the calibration of the devices intended to measure or generate the measured value in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percent or fraction of the reading plus a fixed floor specification. 6 This scope meets A2LA’s P112 Flexible Scope Policy. 7 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches, R is the resolution of the instrument, and D is the numerical value of the nominal length of the diagonal.
18ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994; R205--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

7931 Bond Street, Lenexa, KS, 66214, US

38.96975, -94.78458

Open in Maps

Accreditation data refreshed from the A2LA register on .

This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.