Registers refreshed 2026-11-20.See refresh dates
United States

Anritsu AUS, California

490 Jarvis Drive, 95037, Morgan Hill, CA, United States

Visit website
A2LA 2160.01ISO/IEC 17025Register refreshed 2026-11-20

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free, no obligation.
Scope lines
10
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 10 scope lines.

A2LA certificate 2160.01 · 10 lines

LineMethodMatrixRange
1ISO/IEC 17025:2017 ANSI/NCSL Z540-1-1994; IEC 17025; ANSI/NCSL Z540-1-1994; ANSI/NCSLI Z540Air; airSee scope document
2

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 2 of 9 II. Electrical – RF/Microwave Parameter/Range Frequency CMC2, 4 (±) Comments S Parameters3 – Measure Transmission & Phase for S12, S21 (-80 to 10) dB, (0 to 180) ° 9 kHz to 4 GHz (4 to 10) GHz (10 to 18) GHz 70 kHz to 20 GHz (20 to 30) GHz (0.04 to 1000) MHz (1 to 20) GHz (20 to 40) GHz (40 to 45) GHz 70 kHz to 20 GHz (20 to 40) GHz (40 to 67) GHz (67 to 70) GHz 70 kHz to 20 GHz (20 to 40) GHz (40 to 50) GHz (50 to 60) GHz (60 to 70) GHz (70 to 80) GHz (80 to 90) GHz (90 to 100) GHz (100 to 110) GHz (110 to 116.5) GHz (40 to 170) GHz 0.015 dB / 0.15° 0.020 dB / 0.25° 0.027 dB / 0.35° 0.01 dB / 1.0° 0.02 dB / 2.0° 0.037 dB / 0.4° 0.022 dB / 0.5° 0.032 dB / 0.8° 0.046 dB / 1.1° 0.035 dB / 0.3° 0.035 dB / 0.6° 0.045 dB / 0.9° 0.05 dB / 2.4° 0.02 dB / 0.5° 0.04 dB / 2.5° 0.05 dB / 2.8° 0.07dB / 3.2° 0.08 dB / 3.9° 0.1 dB / 4.2° 0.12 dB / 4.6° 0.13 dB / 5.4° 0.14 dB / 6.6° 0.16 dB / 6.6° 0.14 dB / 15° VNA with N cal kits VNA & 3.5mm cal kit VNA with K cal kit VNA with V cal kit VNA with 1 mm cal kit VNA with Waveguide cal kit
3-Coaxial Components
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 3 of 9 Parameter/Range Frequency CMC2, 4 (±) Comments Reflection3 S11, S22 – Measure Coaxial Components (0.000 01 to 1.0) lin, (0 to 180) ° 9 kHz to 300 MHz (0.3 to 2) GHz (2 to 6) GHz (6 to 14) GHz (14 to 18) GHz 70 kHz to 8 GHz (8 to 20) GHz (20 to 26) GHz (26 to 30) GHz (0.04 to 1000) MHz (1 to 20) GHz (20 to 40) GHz (40 to 45) GHz 100 kHz to 10 GHz (10 to 20) GHz (20 to 40) GHz (40 to 67) GHz (67 to 69.5) GHz 70 kHz to 20 GHz (20 to 40) GHz (40 to 50) GHz (50 to 60) GHz (60 to 70) GHz (70 to 80) GHz (80 to 90) GHz (90 to 100) GHz (100 to 110) GHz (110 to 116.5) GHz 70 kHz to 145 GHz (40 to 170) GHz 0.002 lin, 0.2° 0.003 lin, 0.3° 0.004 lin, 0.4° 0.006 lin, 0.6° 0.007 lin, 0.8° 0.005 lin, 0.5° 0.008 lin, 1.0° 0.012 lin, 1.5° 0.020 lin, 2.0° 0.003 lin, 0.4° 0.0025 lin, 0.6° 0.0035 lin, 1.0° 0.0055 lin, 1.1° 0.005 lin, 1.5° 0.005 lin, 1.5° 0.007 lin, 3° 0.008 lin, 3.5° 0.008 lin, 4° 0.005 lin / 0.5° 0.009 lin / 0.7° 0.012 lin / 1.0° 0.016 lin / 1.1° 0.019 lin / 1.2° 0.023 lin / 1.2° 0.027 lin / 1.3° 0.030 lin 1.6° 0.033 lin 1.7° 0.037 lin 1.8° 0.006 lin / 0.6° 0.032 lin / 15° VNA with N cal kit VNA with SMA cal kit VNA with K cal kit VNA with V cal kit VNA with 1 mm cal kit VNA with 0.8 mm cal kit VNA with Waveguide cal kit
See scope document
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 4 of 9 Parameter/Range Frequency CMC2, 4 (±) Comments Reflection3 S11, S22 – Measure, Directivity & Test Port Match 9 kHz to 18 GHz 70 kHz to 45 GHz 70 kHz to 70 GHz 125 kHz to 116.5 GHz 0.32 dB 0.1 dB 0.2 dB 0.4 dB VNA, Airline, Autocal Power Level – Absolute6 Measure 0 dBm (-20 to 20) dBm (-40 to -20) dBm (-60 to -40) dBm (-90 to -60) dBm 0 dBm (-20 to 20) dBm (-60 to -20) dBm (-90 to -60) dBm (0.01 to 5) GHz (7 to 15) GHz (15 to 18) GHz (0.01 to 2) GHz (2 to 8) GHz (8 to 12) GHz (12 to 18) GHz (0.01 to 2) GHz (2 to 8) GHz (8 to 12) GHz (12 to 18) GHz (0.01 to 18) GHz (0.01 to 18) GHz (10 to 100) MHz (0.1 to 2) GHz (2 to 8) GHZ (8 to 12) GHz (12 to 43.5) GHz (0.05 to 2) GHz (2 to 8) GHz (8 to 12) GHz (12 to 43.5) GHz (0.05 to 2) GHz (2 to 12) GHz (12 to 33) GHz (33 to 43.5) GHz (0.05 to 18) GHz 0.025 dB 0.030 dB 0.034 dB 0.035 dB 0.042 dB 0.048 dB 0.065 dB 0.062 dB 0.074 dB 0.083 dB 0.1 dB 0.12 dB 0.17 dB 0.027 dB 0.037 dB 0.044 dB 0.056 dB 0.062 dB 0.052 dB 0.061 dB 0.07 dB 0.11 dB 0.094 dB 0.11 dB 0.12 dB 0.15 dB 0.18 dB Type N power sensor Spectrum analyzer power sensor with attenuator Type K power sensor Spectrum analyzer, power sensor with attenuator
5

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 5 of 9 Parameter/Range Frequency CMC2, 4 (±) Comments Power Level – Absolute 6 Measure: Type N, Type K, Type V, & W Connector 0 dBm (-20 to 20) dBm (-40 to -20) dBm (-60 to -40) dBm (-90 to -60) dBm (0 to 100) MHz (0.1 to 2) GHz (2 to 8) GHz (8 to 12) GHz (12 to 18) GHz (18 to 40) GHz (40 to 49) GHz (49 to 67) GHz (67 to 70) GHz (0.05 to 2) GHz (2 to 8) GHz (8 to 12) GHz (12 to 40) GHz (40 to 50) GHz (50 to 70) GHz (0.05 to 8) GHz (8 to 12) GHz (12 to 40) GHz (40 to 50) GHz (0.05 to 8) GHz (8 to 12) GHz (12 to 26) GHz (26 to 33) GHz (33 to 40) GHz (40 to 50) GHz (0.05 to 8) GHz (8 to 12) GHz (12 to 26) GHz (26 to 33) GHz (33 to 40) GHz (40 to 50) GHz 0.024 dB 0.036 dB 0.042 dB 0.054 dB 0.07 dB 0.06 dB 0.10 dB 0.15 dB 0.18 dB 0.044 dB 0.051 dB 0.059 dB 0.09 dB 0.13 dB 0.21 dB 0.08 dB 0.09 dB 0.13 dB 0.17 dB 0.09 dB 0.10 dB 0.12 dB 0.13 dB 0.14 dB 0.17 dB 0.18 dB 0.2 dB 0.25 dB 0.26 dB 0.28 dB 0.35 dB Type V power sensor Spectrum analyzer, power sensor, attenuator
6

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 6 of 9 Parameter/Range Frequency CMC2, 4 (±) Comments Power Level – Absolute6 Measure: Type N, Type K, Type V, W & Waveguide Connector 0 dBm (-35 to 20) dBm (0 to -60) dBm (-80 to -60) dBm (-100 to -80) dBm (-20 to 0) dBm (-60 to -20) dBm (-90 to -60) dBm (-125 to 0) dBm (10 to 50) MHz (0.05 to 2) GHz (2 to 8) GHz (8 to 12) GHz (12 to 40) GHz (40 to 50) GHz (50 to 75) GHz (75 to 92) GHz (92 to 110) GHz (0.01 to 2) GHz (2 to 12) GHz (12 to 26) GHz (26 to 40) GHz (40 to 50) GHz (50 to 67) GHz (67 to 95) GHz (95 to 110) GHz (0.01 to 20 GHz (20 to 50) GHz (50 to 67) GHz (67 to 70) GHz (0.01 to 50) GHz (0.01 to 50) GHz (40 to 170) GHz (0 to 3) GHz (3 to 40) GHz 0.03 dB 0.04 dB 0.05 dB 0.07 dB 0.09 dB 0.13 dB 0.18 dB 0.20 dB 0.21 dB 0.06 dB 0.08 dB 0.10 dB 0.12 dB 0.17 dB 0.24 dB 0.27 dB 0.30 dB 0.1 dB 0.15 dB 0.18 dB 0.21 dB 0.25 dB 1.3 dB 0.35 dB 0.66 dB 1.5 dB 0.25dB 1.4 dB Type W power sensor Spectrum analyzer Spectrum analyzer with waveguide connector Measuring receiver Receiver with mixer
7--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 7 of 9 Parameter/Range Frequency CMC2, 4 (±) Comments Frequency Modulation – Measure Carrier 1 GHz 5 GHz 10 GHz 20 GHz 30 GHz 40 GHz Modulation 10% 0.02 % 0.08 % 0.15 % 0.30 % 0.45 % 0.60 % Function generator, measuring receiver LO synthesizer Amplitude Modulation – Measure (1 to 20) GHz (20 to 40) GHz (10 to 100) % 2.6 % 3.9 % Function generator & measuring receiver Pulse Modulation – Measure Power Rise, Fall & Compression High Level Noise Magnitude High Level Noise Phase (0.01 to 40) GHz (0.5 to 40) GHz < 500 KHz (0.005 to 2.5) GHz (2.5 to 5) GHz (5 to 20) GHz (20 to 40) GHz (40 to 67) GHz (67 to 70) GHz < 500KHz (0.005 to 2.5) GHz (2.5 to 5) GHz (5 to 20) GHz (20 to 40) GHz (40 to 67) GHz (67 to 70) GHz 0.09 dB 0.011 ns 0.000 59 dB-rms 0.000 33 dB-rms 0.000 25 dB-rms 0.000 28 dB-rms 0.000 32 dB-rms 0.000 23 dB-rms 0.000 22 dB-rms 0.0055 deg-rms 0.0025 deg-rms 0.0029 deg-rms 0.0018 deg-rms 0.0028 deg-rms 0.0024 deg-rms 0.0019 deg-rms Wide band oscilloscope & function generator VNA
to 70 GHz
8

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2160.01) 03/08/2024 Page 8 of 9 Parameter/Range Frequency CMC2, 4, 5 (±) Comments Noise, Spurs & Distortion – Measure Phase Noise (0.01 to 16) GHz Carrier Spurious Signals Dynamic Noise Level >-156 dBm Harmonic Distortion 2 nd Harmonic Offset (0.01 to 10) MHz (0.01 to 45) GHz (0.01 to 70) GHz (0.01 to 1) GHz (1 to 43.5) GHz 3.5 dB 3 dB 0.74 dB 0.56 dB 0.15 dB Spectrum analyzer Spectrum analyzer VNA or SPA Spectrum analyzer III. Mechanical Parameter/Equipment Range CMC2, 5 (±) Comments Torque Devices (0.25 to 250) lbf.in 2.5 % Torque analyzer IV. Optical Quantities Parameter/Equipment Range CMC2, 8 (±) Comments Optical Wavelength & Power – Measure Optical – Electrical Responsivity Transmission & Reflection (850 to 1550) nm (-80 to 0) dBm, (0 to 180) ° 70 kHz to 80 GHz (80 to 90) GHz (90 to 100) GHz (100 to 110) GHz 1 dB 1.5 dB 3 dB 4.5 dB VNA & photodiode
9NIST TMAS-70000 to 170000000000 Hz
10ISO/IEC 17025:2017; ISO-ILAC-IAF Communiqué dated April 2017; IEC 17025; ANSI/NCSL Z540-1-1994 and the requirements of ANSI/NCSL Z540--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

490 Jarvis Drive, Morgan Hill, CA, 95037, US

37.14545, -121.65899

Open in Maps

Accreditation data refreshed from the A2LA register on .

This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.