| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994 | - | to 24 in |
| 2 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 2 of 13
Parameter/Equipment Range CMC2, 4 (±) Comments
Levels –
Tubular
Cylindrical
Up to 10ʺ
Up to 60ʺ
5ʺ
5ʺ
Gage blocks & surface
plates
Length Standards
Up to 20 in
(20 to 80) in
(5.0 + 3L) µin
(10 + 5L) µin
Master length standards/
gage blocks/comparator
Micrometers3 –
Precision
Outside Diameter
Bore
Depth
Up to 2 in
Up to 24 in
(24 to 42) in
Up to 4 in
(4 to 8) in
Up to 12 in
16 µin
(85 + 6L) µin
(160 + 3L) µin
(100 + L) µin
(200 + 2L) µin
(100 + L) µin
UMM & gage blocks
Gage block
Length standards
Master ring gage
Master ring gage
Gage blocks
Micrometers, Inside
Diameter
Up to 12 in
Up to 24 in
Up to 36 in
Up to 48 in
Up to 80 in
150 µin
170 µin
190 µin
200 µin
250 µin
UMM,
gage blocks/length
standards
Optical Comparators3 –
Linearity/Squareness
Up to 20 in
(100 + 2L) µin
Gage blocks/length
standards
Pipe Thread Plug –
Pitch Diameter:
(4 to 80) TPI
Major Diameter
Length to Notch
(0.0625 to 10) in
(0.0625 to 10) in
(0.0625 to 10) in
110 µin
100 µin
60 µin
Three wire method in
fixture
Comparator
Gage blocks | to 10 in |
| 3 | ASME/ANSI B1.2-1983; ASME/ANSI B1.3-2007 | - | See scope document |
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 4 of 13
Parameter/Equipment
Range9 CMC2, 4 (±) Comments
Thread Ring Gages (cont.) –
Solid Gages:
Pitch Diameter
Minor Diameter
(0.25 to 8) in
(0.25 to 8) in
(100 + 3D) µin
(100 + 3D) µin
UMM
Video Measurement System3 –
X, Y Axis
Z Axis
Up to 48 in
Up to 12 in
(70 + 5L) µin
(70 + 5L) µin
Verification grids
Gage blocks with
length standards
Vision Measurement System3 –
–
X, Y Axis
Z Axis
Up to 60 in
Up to 18 in
(20 + 5L) µin
(60 + 6L) µin
Verification grids
Gage blocks with
length standards
Height Gages3
Up to 36 in
(25 + 4L) μin
Gage blocks/standard
reference bar
Height Masters
Up to 24 in
(25 + 4L) μin
Gage blocks/standard
reference bar
Protractor –
Digital / Analog
(0 to 360) °
1´
Sine plate, gage blocks,
surface plate, indicator
II. Dimensional Testing/Calibration
1
Parameter/Equipment
Range CMC2, 4 (±) Comments
Length – 2D7
Up to 24 in
(100 + 3L) µin
Vision system | ||
| 5 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 5 of 13
Parameter/Equipment
Range CMC2, 4 (±) Comments
Length – 1D7
Up to 40 in
(41 to 80) in
(5.0 + 3L) µin
(10 + 5L) µin
Linear comparator, or
surface plate metrology
III. Electrical – DC/Low Frequency
Parameter/Equipment
Range
CMC2, 5, 8 (±)
Comments
DC Voltage3 – Generate
(0 to 200) mV
200 mV to 2 V
(2 to 20) V
(20 to 200) V
(200 to 1050) V
5.0 μV/V + 0.1 μV
3.6 μV/V + 0.4 μV
3.6 μV/V + 4.0 μV
5.6 μV/V + 40 μV
5.6 μV/V + 500 μV
Fluke 8508A
DC Voltage3 – Measure
(0 to 200) mV
200 mV to 2 V
(2 to 20) V
(20 to 200) V
(200 to 1050) V
(1 to 10) kV
(10 to 100) kV
5.0 μV/V + 0.1 μV
3.6 μV/V + 0.4 μV
3.6 μV/V + 4.0 μV
5.6 μV/V + 40 μV
5.6 μV/V + 500 μV
0.042 % + 0.9 V
0.025 % + 42 V
Fluke 8508A
High voltage DVR-150
& Fluke 8508A
DC Current3 – Generate
(0 to 200) μA
(200 to 2000) μA
(2 to 20) mA
(20 to 200) mA
(200 to 2000) mA
(2 to 20) A
Up to 200 A
(200 to 1000) A
7.0 μA/A + 0.41 nA
6.9 μA/A + 4.1 nA
8.0 μA/A + 41 nA
33 μA/A + 0.8 μA
0.017 % + 16 μA
0.038 % + 0.4 mA
0.25 % + 5 mA
0.25 % + 10 mA
Fluke 8508A
Transmille 4010/coil
DC Current3 – Measure
(0 to 200) μA
(200 to 2000) μA
(2 to 20) mA
(20 to 200) mA
(200 to 2000) mA
(2 to 20) A
(20 to 100) A
7.0 μA/A + 0.41 nA
6.9 μA/A + 4.1 nA
8.0 μA/A + 41 nA
33 μA/A + 0.8 μA
0.017 % + 16 μA
0.038 % + 0.4 mA
0.12 % + 5 mA
Fluke 8508A
Empro 249004 | to 40 in |
| 6 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 6 of 13
Parameter/Equipment
Range
CMC2, 5 (±)
Comments
Resistance3 – Generate
(0 to 11) Ω
(11 to 33) Ω
(33 to 110) Ω
(110 to 330) Ω
(330 to 1100) Ω
(1.1 to 3.3) kΩ
(3.3 to 11) kΩ
(11 to 33) kΩ
(33 to 110) kΩ
(110 to 330) kΩ
(330 to 1100) kΩ
(1.1 to 3.3) MΩ
(3.3 to 11) MΩ
(11 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(0.33 to 1) GΩ
0.012 % + 6.0 mΩ
0.012 % + 10 mΩ
90 µΩ/Ω + 10 mΩ
90 µΩ/Ω + 10 mΩ
90 µΩ/Ω + 60 mΩ
90 µΩ/Ω + 60 mΩ
90 µΩ/Ω + 600 mΩ
90 µΩ/Ω + 600 mΩ
0.011 % + 6.0 Ω
0.011 % + 6.0 Ω
0.015 % + 55 Ω
0.015 % + 55 Ω
0.06 % + 550 Ω
0.10 % + 550 Ω
0.50 % + 5.5 k Ω
0.50 % + 17 kΩ
2 % + 100 kΩ
Fluke 5500A
Transmille 4010
Resistance3 – Measure
(0 to 2) Ω
(2 to 20) Ω
(20 to 200) Ω
(200 to 2000) Ω
(2 to 20) kΩ
(20 to 200) kΩ
(200 to 2000) kΩ
(2 to 20) MΩ
(20 to 200) MΩ
(200 to 2000) MΩ
(2 to 20) GΩ
(20 to 200) GΩ
(200 to 2000) GΩ
15 µΩ/Ω + 4.0 µΩ
9.0 µΩ/Ω + 14 µΩ
7.5 µΩ/Ω + 50 µΩ
7.5 µΩ/Ω + 0.5 mΩ
7.5 µΩ/Ω + 5.0 mΩ
7.5 µΩ/Ω + 50 mΩ
8.5 µΩ/Ω + 1.0 Ω
15 µΩ/Ω + 10 Ω
60 µΩ/Ω + 1.0 kΩ
0.015 % + 110 kΩ
0.050 % + 10 MΩ
75 MΩ
12 GΩ
Fluke 8508A
Fluke 5500A &
Fluke 8508A
Capacitance3 – Generate
(0.33 to 0.5) nF
(0.5 to 1.1) nF
(1.1 to 3.3) nF
(3.3 to 11) nF
(11 to 33) nF
(33 to 110) nF
(110 to 330) nF
(330 to 1100) nF
(1.1 to 3.3) μF
(3.3 to 11) μF
(11 to 33) μF
(33 to 110) μF
(110 to 330) μF
(330 to 1100) μF
(1.1 to 10) mF
(10 to 100) mF
20 pF
0.23 % + 15 pF
0.29 % + 17 pF
0.50 % + 10 pF
0.25 % + 100 pF
0.25 % + 100 pF
0.25 % + 300 pF
0.25 % + 1.0 nF
0.35 % + 3.0 nF
0.35 % + 10 nF
0.40 % + 31 nF
0.50 % + 100 nF
0.70 % + 300 nF
1.0 % + 300 nF
0.5 % + 5 μF
0.5 % + 7 μF
Fluke 5500A
Transmille 4010 | ||
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 7 of 13
Parameter/Equipment
Range
CMC2, 5 (±)
Comments
Oscilloscope3 –
Bandwidth
Vertical Deflection:
DC Level:
50 Ω
1 MΩ
1 MΩ
1 MΩ
Square Wave:
50 Ω / 1 MΩ
100 mHz to 100 MHz
(100 to 550) MHz
550 MHz to 1.1 GHz
± 5.56 V
± 100 mV
± 5 V
± 222 V
(35.52 to 1000) µVpp
(1 to 21) mVpp
(21 to 556) mVpp
556 mV to 10 V
(10 to 210) V
1.5 %
3.0 %
4.6 %
0.14 %
0.25 % + 20 µV
0.095 % + 20 µV
0.030 % + 20 mV
1.2 % + 10 µV
0.044 % + 49 µV
0.045 % + 0.45 mV
0.012 % + 8.3 mV
0.061 % + 0.1 V
Wavetek 9500
Time Marker:
Square Wave Pulse /
Triangle
9 ns to 55 s
11 µs/s + 45 ps
Wavetek 9500
Electrical Calibration of
Thermocouple Indicators
& Indicating Systems3 –
Type B
Type C
Type E
Type J
Type K
(600 to 800) °C
(800 to 1820) °C
(0 to 1000) °C
(1000 to 1800) °C
(1800 to 2316) °C
(-250 to -100) °C
(-100 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
0.44 °C
0.34 °C
0.31 °C
0.50 °C
0.84 °C
0.51 °C
0.18 °C
0.21 °C
0.27 °C
0.17 °C
0.23 °C
0.34 °C
0.18 °C
0.26 °C
0.40 °C
Fluke 5500A | ||
| 8 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 8 of 13
Parameter/Equipment
Range
CMC2 (±)
Comments
Electrical Calibration of
Thermocouple Indicators
& Indicating Systems3 –
(cont.)
Type N
Type R
Type S
Type T
(-200 to -100) °C
(-100 to 410) °C
(410 to 1300) °C
(0 to 250) °C
(250 to 1000) °C
(1000 to 1767) °C
(0 to 250) °C
(250 to 1400) °C
(1400 to 1767) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 400) °C
0.40 °C
0.22 °C
0.27 °C
0.57 °C
0.35 °C
0.40 °C
0.47 °C
0.37 °C
0.46 °C
0.63 °C
0.24 °C
0.16 °C
Fluke 5500A
Electrical Calibration of
RTD Indicators &
Indicating Systems
3 –
Pt 385, 100 Ω
Pt 385, 100 Ω
(-200 to 200) °C
(200 to 600) °C
(600 to 800) °C
(-200 to 0) °C
(0 to 400) °C
(400 to 800) °C
0.10 °C
0.20 °C
0.30 °C
0.051 °C
0.10 °C
0.14 °C
Beamex MC2
Fluke 5500A | ||
| 9 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 9 of 13
Parameter/Range
Frequency
CMC2, 5, 8, 12 (±)
Comments
AC Voltage3 – Measure
(0 to 200) mV
200 mV to 2 V
(2 to 20) V
(20 to 200) V
(200 to 1050) V
(1 to 10) kV
(10 to 100) kV
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
60 Hz
60 Hz
0.010 % + 4.0 µV
0.010 % + 2.0 µV
80 μV/V + 20 µV
60 μV/V + 20 µV
80 μV/V + 0.2 mV
60 μV/V + 0.2 mV
80 μV/V + 2.0 mV
60 μV/V + 2.0 mV
90 μV/V + 27 mV
0.50 % + 0.3 V
0.57 %
Fluke 8508A
High voltage DVR-
150 & Fluke 8508A
AC Voltage3 – Generate
(0 to 200) mV
200 mV to 2 V
(2 to 20) V
(20 to 200) V
(200 to 1050) V
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
(50 to 100) Hz
(100 to 2000) Hz
(50 to 2000) Hz
0.010 % + 4.0 µV
0.010 % + 2.0 µV
80 μV/V + 20 µV
60 μV/V + 20 µV
80 μV/V + 0.2 mV
60 μV/V + 0.2 mV
80 μV/V + 2.0 mV
60 μV/V + 2.0 mV
90 μV/V + 27 mV
Fluke 8508A
AC Current3 – Generate
(0 to 200) µA
(200 to 2000) μA
(2 to 20) mA
(20 to 200) mA
(200 to 2000) mA
(2 to 20) A
Up to 200 A
(200 to 1000) A
Up to 200 A
(200 to 1000) A
(10 to 10 000) Hz
(10 to 10 000) Hz
(10 to 10 000) Hz
(10 to 10 000) Hz
(10 to 2000) Hz
(10 to 2000) Hz
(45 to 65) Hz
(45 to 65) Hz
(65 to 440) Hz
(65 to 440) Hz
0.025 % + 20 nA
0.025 % + 0.2 μA
0.025 % + 2.0 μA
0.025 % + 20 μA
0.060 % + 200 μA
0.080 % + 2.0 mA
0.17 %
0.11 % + 0.54 A
0.48 %
0.33 % + 0.37 A
Fluke 8508A
Transmille 4010/coil
Transmille 4010/coil | ||
| 10 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 10 of 13
Parameter/Range
Frequency
CMC2, 5, 8, 12 (±)
Comments
AC Current3 – Measure
(0 to 200) µA
(200 to 2000) μA
(2 to 20) mA
(20 to 200) mA
(200 to 2000) mA
(2 to 20) A
(20 to 100) A
(10 to 10 000) Hz
(10 to 10 000) Hz
(10 to 10 000) Hz
(10 to 10 000) Hz
(10 to 2000) Hz
(10 to 2000) Hz
(40 to 100) Hz
0.025 % + 20 nA
0.025 % + 0.2 μA
0.025 % + 2.0 μA
0.025 % + 20 μA
0.060 % + 200 μA
0.080 % + 2.0 mA
0.28 % + 4.0 mA
Fluke 8508A
Empro 249004
IV. Mechanical
Parameter/Equipment Range CMC
2 (±) Comments
Balances & Scales3
Up to 50g
Up to 250 g
Up to 500 g
Up to 2000 g
Up to 6000 g
Up to 15 000 g
Up to 20 000 g
Up to 40 000 g
Up to 30 lb
Up to 70 lb
Up to 100 lb
Up to 150 lb
Up to 330lb
Up to 660 lb
Up to 1000 lb
0.13 mg
0.64 mg
2.0 mg
0.015 g
0.020 g
0.16 g
0.22 g
2.3 g
0.0033 lb
0.0072 lb
0.017 lb
0.019 lb
0.13 lb
0.13 lb
0.25 lb
Class 1 & 3 weights
Class 6 weights
Force3 – Measure,
Tension & Compression
Up to 4 lbf
Up to 50 lbf
Up to 100 lbf
Up to 250 lbf
Up to 500 lbf
Up to 1000 lbf
(1000 to 10 000) lbf
(10 000 to 30 000) lbf
0.0018 lbf
0.016 lbf
0.018 lbf
0.033 lbf
0.070 lbf
0.13 lbf
0.5 %
0.5 %
Class 1, 3, & 6 weights
Load cells | ||
| 11 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 11 of 13
Parameter/Equipment
Range CMC2, 8, 10, 11, 14 (±) Comments
Torque Wrenches
Up to 100 oz·in
5 lb-in to 250 lb·ft
250 lb·ft to 2000 lb·ft
1.0 %
0.5 %
1.0 %
Torque calibrator
Torque Calibrators
Up to 10 ozf·in
Up to 100 ozf·in
Up to 50 lbf·in
Up to 400 lbf·in
Up to 1000 lbf·in
Up to 250 lbf·ft
Up to 600 lbf·ft
Up to 2000 lbf·ft
0.025 ozf·in
0.15 ozf·in
0.015 lbf·in
0.091 lbf·in
0.22 lbf·in
0.067 lbf·ft
0.21 lbf·ft
0.32 lbf·ft
Precision weights,
arms, wheels, & trays
Vacuum3 – Negative
Pressure – Measure
(-14.5 to 0) psi
0.030 % + 0.0025 psi
Beamex EXT-1C
Pressure3 – Measure
Pneumatic
Hydraulic
(0 to 12) inH
2O
Up to 40 inH2O
Up to 15 psi
Up to 30 psi
Up to 90 psi
Up to 300 psi
Up to 900 psi
Up to 500 psig
Up to 8000 psig
Up to 16 000 psig
0.0024 inH2O
0.025 % + 0.011 inH2O
0.025 % + 0.0023 psi
0.025 % + 0.003 psi
0.025 % + 0.01 psi
0.025 % + 0.03 psi
0.025 % + 0.09 psi
0.013 % + 0.012 psi
0.011 % + 0.7 psi
0.011 % + 0.8 psi
Dwyer 1425
Beamex EXT100m
Beamex EXT1C
Beamex EXT2C
Beamex EXT6C
Beamex EXT20C
Beamex EXT60
Pressurements 2800/3
V. Thermodynamics
Parameter/Equipment
Range
CMC2, 8, 10 (±)
Comments
Relative Humidity3 –
Measure
(10 to 90) % RH
(90 to 97) % RH
0.35 % + 1.1 % RH
2.0 % RH
Vaisala MI70, HMP76 | ||
| 12 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 12 of 13
Parameter/Equipment
Range
CMC2, 8, 10 (±)
Comments
Relative Humidity3 –
Measuring Equipment
11 % RH
23 % RH
33 % RH
43 % RH
75 % RH
85 % RH
95 % RH
1.1 % RH
1.2 % RH
1.2 % RH
1.3 % RH
1.4 % RH
1.4 % RH
2.0 % RH
Vaisala MI70,
HMP76, salts
Temperature – Measure3
(-200 to 660) °C
0.0035 % + 0.013 °C
Fluke 5609, 1502A
PRT
Temperature – Measuring
Equipment3
(-5 to 126) °C
(126 to 660) °C
0.039 °C
0.0022 % + 0.036 °C
Hart 7102 w/PRT
Fluke 9173 w/PRT
Infrared Temperature –
Measuring Equipment
(35 to 100) °C
(100 to 200) °C
(200 to 350) °C
(350 to 500) °C
0.62 °C
0.84 °C
1.4 °C
1.9 °C
Fluke 4181 IR
blackbody
VI. Time & Frequency
Parameter/Equipment
Range
CMC2, 10 (±)
Comments
Timers & Clocks
1 ms to 1 s
(1 to 100) s
(100 to 3600) s
220 ns
45 µs/s + 350 ns
41 ms
Agilent 53132A
Frequency3 – Measuring
Equipment
Sine Wave
Square Wave
Triangular Wave
100 μHz to 15 MHz
100 μHz to 15 MHz
100 μHz to 100 kHz
21 µHz/Hz + 5 mHz
20 µHz/Hz + 5 mHz
21 µHz/Hz + 0.5 mHz
Agilent 33120A
Frequency3 – Measure
(0.1 to 1000) Hz
(1 to 1000) kHz
(1 to 225) MHz
0.38 mHz
44 nHz/Hz + 4.4 mHz
47 nHz/Hz + 0.15 Hz
Agilent 53132A
_______________________________________________ | ||
| 13 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1579.01) 10/06/2025 Page 13 of 13
1 This laboratory offers commercial and field calibration and dimensional testing services.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects su ch as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches;
D is the numerical value of the nominal diameter of the device measured in inches.
5 The measurands stated are generated with the Fluke 5500A , Fluke 5700A, Fluke 8508A, Fluke 89 and
Transmille 4010 series of instruments. This capability is suitable for the calibration of the devices
intended to measure the stated measurand in the ranges indicated. CMCs are expressed as either a specific
value that covers the full range or as a fraction or percentage of the reading plus a fixed floor specification.
6 The measurands stated are generated with the Wavetek 9500. This capability is suitable for the calibration
of the devices intended to measure the stated measurand in the ranges indicated. CMCs are expressed as
either a specific value that covers the full range or as a percentage of output plus one -year floor
specifications where defined.
7 This laboratory meets R205 – Specific Requirements: Calibration Laboratory Accreditation Program for
the types of dimensional tests listed above and is considered equivalent to that of a calibration.
8 In the statement of CMC, percentages are to be read as percentage of reading, unless otherwise noted.
9 Calibration of adjustable thread ring gages above 2.625 inches is performed with a customer supplied
compatible Class X thread setting plug gage - subject to calibration by HJM Precision.
10 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
11 In the statement of CMC, psi is to be read as pounds per square inch.
12 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured
value in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range
or as a percent or fraction of the reading plus a fixed floor specification.
13 The scope meets A2LA’s P112 Flexible Scope Policy.
14 In the statement of CMC, percentages are to be read as percentage of the absolute value of the reading,
unless otherwise noted. | ||
| 14 | ISO/IEC 17025:20 17; ANSI/NCSL Z540-1-1994; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Labs with overlapping accredited scope in United States
- Quality PlusChicopee, MA · ~110 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Fischer Technology, Inc. USAWindsor, CT · ~130 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Technical Maintenance, Inc. - WINDSORWindsor, CT · ~130 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Industronics Service CompanySouth Windsor, CT · ~136 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- CONNECTICUT CALIBRATION TECHNOLOGIES, INC., dba CONNECTICUT CALIBRATIONNewtown, CT · ~150 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.