SIMCO Electronics - Chicago
1701 Quincy Avenue, Suite 32, 60540, Naperville, IL, United States
Visit website- Scope lines
- 10
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We found no published test standard numbers in this lab’s scope. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 10 scope lines.
A2LA certificate 1395.07 · 10 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994 | - | See scope document |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 2 of 9
Parameter/Equipment
Range CMC2, 4, 5, 9 (±) Comments
DC Current3 – Measure
(0.1 to 1) µA
(1 to 10) µA
(10 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
39 µA/A + 64 pA
29 µA/A + 130 pA
28 µA/A + 1.1 nA
28 µA/A + 7.5 nA
28 µA/A + 75 nA
44 µA/A + 0.75 µA
0.013 % + 16 µA
HP 3458A, opt 2
DC Current – Clamp-On
Meters
(10 to 25) A
(25 to 150) A
(150 to 1025) A
0.62 % + 0.02 A
0.62 % + 0.14 A
0.63 % + 0.5 A
Fluke 5520A/SC1100,
Fluke 5500A/coil
Resistance3 – Generate
(0 to 11) Ω
(11 to 33) Ω
(33 to 110) Ω
(110 to 330) Ω
(0.33 to 1.1) kΩ
(1.1 to 3.3) kΩ
(3.3 to 11) kΩ
(11 to 33) kΩ
(33 to 110) kΩ
(110 to 330) kΩ
(0.33 to 1.1) MΩ
(1.1 to 3.3) MΩ
(3.3 to 11) MΩ
(11 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(0.33 to 1.1) GΩ
40 µΩ/Ω + 1 mΩ
24 µΩ/Ω + 1.5 mΩ
22 µΩ/Ω + 1.4 mΩ
29 µΩ/Ω + 2 mΩ
22 µΩ/Ω + 2 mΩ
52 µΩ/Ω + 0.02 Ω
22 µΩ/Ω + 0.02 Ω
25 µΩ/Ω + 0.2 Ω
22 µΩ/Ω + 0.2 Ω
44 µΩ/Ω + 2 Ω
28 µΩ/Ω + 2 Ω
52 µΩ/Ω + 30 Ω
0.011 % + 50 Ω
0.028 % + 2.5 kΩ
0.042 % + 3 kΩ
0.24 % + 80 kΩ
1.2 % + 400 kΩ
Fluke 5520A/SC1100
Resistance3 – Measure
(0 to 10) Ω
(10 to 100) Ω
(0.1 to 1) kΩ
(1 to 10) kΩ
(10 to 100) kΩ
(0.1 to 1) MΩ
(1 to 10) MΩ
(10 to 100) MΩ
(100 to 1000) MΩ
19 µΩ/Ω + 90 µΩ
16 µΩ/Ω + 0.9 mΩ
14 µΩ/Ω + 1 mΩ
14 µΩ/Ω + 10 mΩ
14 µΩ/Ω + 90 mΩ
20 µΩ/Ω + 3 Ω
62 µΩ/Ω + 120 Ω
0.059 % + 4 kΩ
0.58 % + 260 kΩ
HP 3458A, opt 2 | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 3 of 9
Parameter/Equipment
Range CMC2, 4, 9 (±) Comments
Capacitance – Generate
(0.19 to 0.3999) nF
(0.4 to 1.0999) nF
(1.1 to 3.2999) nF
(3.3 to 10.9999) nF
(11 to 109.999 99) nF
(110 to 329.999) nF
(0.33 to 1.9999) µF
(2 to 3.2999) µF
(3.3 to 10.9999) µF
(11 to 32.9999) µF
(33 to 109.9999) µF
(110 to 329.9999) µF
(0.33 to 1.099 99) mF
(1.1 to 3.2999) mF
(3.3 to 10.9999) mF
(11 to 32.9999) mF
(33 to 110) mF
0.44 % + 8.1 pF
0.44 % + 8.1 pF
0.42 % + 8.1 pF
0.24 % + 8.1 pF
0.23 % + 85 pF
0.23 % + 0.24 nF
0.23 % + 0.8 nF
0.23 % + 2.3 nF
0.24 % + 7.8 nF
0.34 % + 23 nF
0.38 % + 78 nF
0.38 % + 230 nF
0.37 % + 0.8 µF
0.37 % + 2.3 µF
0.37 % + 7.8 µF
0.59 % + 23 µF
0.86 % + 78 µF
Fluke 5520A
Oscilloscopes3 –
Square Wave Signal
10 Hz to 10 kHz
50 Ω Impedance
1 MΩ Impedance
DC Signal
10 Hz to 10 kHz
50 Ω Impedance
1 MΩ Impedance
Level Sine Amp – 50 Ω
50 kHz Ref.
Level Sine Flatness – 50 Ω
5 mV to 5.5 V Relative to
50 kHz Ref.
Time Marker – 50 Ω
Rise Time
1 kHz to 10 MHz
1 mV to 6.6 V
pk – pk
1 mV to 130 Vpk - pk
1 mV to 6.6 V
1 mV to 130 V
10 mV to 5 V
(p-p)
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
2 ns to 20 ms
(50 to 5) s
(250 to 350) ps
0.26 % + 100 µV
0.13 % + 100 µV
0.25 % + 40 µV
0.052 % + 40 µV
2 % + 300 µV
1.8 % + 100 µV
1.8 % + 100 µV
3.2 % + 100 µV
2.5 parts in 10
6 s
(25 + 1000·t) µs/s
160 ps
Fluke
5520A/SC1100
t = time in seconds | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 4 of 9
Parameter/Equipment
Range CMC2 (±) Comments
Electrical Simulation of
Thermocouple Indicators &
Indicating Systems3 –
Type E
Type J
Type K
Type T
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 120) °C
(120 to 400) °C
0.5 °C
0.16 °C
0.14 °C
0.16 °C
0.21 °C
0.27 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
0.33 °C
0.18 °C
0.16 °C
0.26 °C
0.4 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
Fluke 5520A | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 5 of 9
Parameter/Range
Frequency CMC2, 4, 9 (±) Comments
AC Voltage3 – Generate
(1 to 33) mV
(33 to 330) mV
(0.33 to 3.3) V
(3.3 to 33) V
(33 to 330) V
(330 to 1020) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
45 Hz to 1 kHz
(1 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.078 % + 6 µV
0.02 % + 6 µV
0.05 % + 6 µV
0.1 % + 6 µV
0.28 % + 12 µV
0.66 % + 50 µV
0.027 % + 8 µV
0.018 % + 8 µV
0.019 % + 8 µV
0.032 % + 8 µV
0.064 % + 32 µV
0.16 % + 70 µV
0.027 % + 40 µV
0.018 % + 50 µV
0.020 % + 50 µV
0.027 % + 40 µV
0.056 % + 100 µV
0.19 % + 70 µV
0.027 % + 0.65 mV
0.018 % + 0.6 mV
0.023 % + 0.6 mV
0.03 % + 0.6 mV
0.071 % + 1.6 mV
0.02 % + 2 mV
0.021 % + 6 mV
0.024 % + 6 mV
0.024 % + 6 mV
0.16 % + 50 mV
0.024 % + 10 mV
0.02 % + 10 mV
0.024 % + 10 mV
Fluke 5520A/SC1100 | ||
| 6 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 6 of 9
Parameter/Range
Frequency CMC2, 5, 9 (±) Comments
AC Voltage3 – Measure
(1 to 10) mV
(10 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.04 % + 3.5 µV
0.025 % + 1.3 µV
0.035 % + 1.3 µV
0.14 % + 1.3 µV
0.58 % + 1.3 µV
4.6 % + 2.5 µV
90 µV/V + 5 µV
90 µV/V + 2.5 µV
0.017 % + 2.5 µV
0.036 % + 2.5 µV
0.093 % + 2.5 µV
0.35 % + 15 µV
1.2 % + 15 µV
90 µV/V + 50 µV
90 µV/V + 25 µV
0.017 % + 25 µV
0.035 % + 25 µV
0.092 % + 25 µV
0.35 % + 130 µV
1.2 % + 130 µV
90 µV/V + 0.5 mV
90 µV/V + 0.25 mV
0.017 % + 0.25 mV
0.035 % + 0.25 mV
0.093 % + 0.25 mV
0.35 % + 1.5 mV
1.2 % + 1.5 mV
0.03 % + 5 mV
0.03 % + 2.5 mV
0.024 % + 2.5 mV
0.041 % + 2.5 mV
0.14 % + 2.5 mV
HP 3458A, opt 2 | ||
| 7 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 7 of 9
Parameter/Range
Frequency CMC2, 4, 5, 9 (±) Comments
AC Voltage – Measure3
(cont)
(100 to 1000) V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.05 % + 50 mV
0.05 % + 25 mV
0.07 % + 25 mV
0.14 % + 25 mV
0.35 % + 25 mV
HP 3458A, opt 2
AC Current – Generate3
(30 to 330) µA
(0.33 to 3.3) mA
(3.3 to 33) mA
(33 to 330) mA
(0.33 to 1.1) A
(1.1 to 3) A
(3 to 11) A
(11 to 20.5) A
45 Hz to 1 kHz
100 Hz to 1 kHz
0.1 % + 0.1 µA
0.078 % + 0.15 µA
0.032 % + 2 µA
0.032 % + 20 µA
0.04 % + 0.1 mA
0.05 % + 0.1 mA
0.081 % + 2 mA
0.12 % + 5 mA
Fluke 5520A/SC1100;
LCOMP off
AC Current – Measure3
(5 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
45 Hz to 1 kHz
(45 to 100) Hz
100 Hz to 5 kHz
(45 to 100) Hz
100 Hz to 5 kHz
(45 to 100) Hz
100 Hz to 5 kHz
(45 to 100) Hz
100 Hz to 5 kHz
0.075 % + 0.04 µA
0.072 % + 0.25 µA
0.041 % + 0.25 µA
0.072 % + 2.5 µA
0.041 % + 2.5 µA
0.072 % + 25 µA
0.041 % + 25 µA
0.10 % + 0.25 mA
0.12 % + 0.25 mA
HP 3458A, opt 2 | 100 to 1000 V |
| 8 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 8 of 9
Parameter/Range
Frequency CMC2, 4, 9 (±) Comments
AC Current – Clamp on
Meter
(10 to 25) A
(25 to 150) A
(150 to 1000) A
(45 to 60) Hz
(64 to 440) Hz
(45 to 60) Hz
(64 to 440) Hz
(45 to 60) Hz
(64 to 440) Hz
0.71 % + 0.03 A
1.3 % + 0.03 A
0.71 % + 0.25 A
1.7 % + 0.25 A
0.74 % + 0.9 A
1.6 % + 0.9 A
Fluke 5520A/SC1100,
Fluke 5500A/coil
II. Mechanical
Parameter/Equipment
Range
CMC2, 6, 8 (±)
Comments
Torque Wrenches –
Measure
(20 to 200) ozf∙in
(5 to 50) lbf∙in
(40 to 400) lbf∙in
(100 to 1000) lbf∙in
(25 to 250) lbf∙ft
0.61 %
0.96 %
0.62 %
0.82 %
0.60 %
CDI torque
transducers
III. Time & Frequency
Parameter/Equipment
Range
CMC2, 8 (±)
Comments
Frequency3 –
Measuring Equipment
10 MHz
0.01 Hz to 2 MHz
0.85 mHz
1.9 µHz/Hz + 5 µHz
HP 58503A GPS
Fluke 5520A
___________________________________________________________
1 This laboratory offers commercial and field calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration. | ||
| 9 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1395.07) Revised 04/16/2025 Page 9 of 9
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects su ch as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 Based on using the standard at the temperature the Fluke 5520A with SC1100 was calibrated (tcal ± 5 °C)
and assuming the instrument is zeroed at least every seven days or when the ambient temperature changes
more than 5 °C. For resistance, a zero calibration is performed at least every 12 hours within ± 1 oC of
use.
5 Based on using the standard at the temperature the HP 3458A was calibrated (tcal ± 5 °C) and an auto-
calibration (ACAL) was performed within the previous 24 hours (± 1 °C of ambient).
6 In the statement of CMC, the value is defined as the percentage of reading unless otherwise noted.
7 This scope meets A2LA’s P112 Flexible Scope Policy.
8 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
9 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification. | ||
| 10 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Labs with overlapping accredited scope in United States
- Mitutoyo America CorporationAurora, IL · ~14 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Florida Metrology, LLC d/b/a Acme MetrologyVilla Park, IL · ~19 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- DicksonAddison, IL · ~22 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- GraftelRoselle, IL · ~24 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- J. H. MetrologyRolling Meadows, IL · ~36 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.