Technical Maintenance, Inc. - WHITE
3660 Highway 411 N.E., 30184, White, GA, United States
Visit website- Scope lines
- 6
- A2LA certificate
- 1
- Sites in our directory
- 2
What this lab is accredited for
We found no published test standard numbers in this lab’s scope. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 6 scope lines.
A2LA certificate 2348.10 · 6 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994; ANSI/NCSL Z540.3-2006 | - | to 300 mm |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. Page 2 of 5
(A2LA Cert. No. 2348.10) 06/25/2026
II. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC2, 3 (±) Comment
DC Voltage – Generate
(0 to 329.999) mV
330 mV to 3.299 99 V
(3.3 to 32.999 9) V
(33 to 329.999) V
(330 to 1020) V
0.028 % of rdg + 10 µV
0.013 % of rdg + 15 µV
0.016 % of rdg + 150 µV
0.015 % of rdg + 1.5 mV
0.015 % of rdg + 5.5 mV
Fluke 5080A
DC Voltage – Measure
(0 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 1000) V
0.0076 % of rdg + 3.5 µV
0.0058 % of rdg + 7 µV
0.0051 % of rdg + 50 µV
0.0064 % of rdg + 600 µV
0.0056 % of rdg + 10 mV
Fluke 8845A
DC Current – Generate
(0 to 329.99) µA
(0.330 to 3.2999) mA
(3.3 to 32.999) mA
(33 to 329.999) mA
330 mA to 1.0999 A
(1.1 to 2.999 9) A
(3 to 10.999) A
(11 to 20.5) A
(20.5 to 150) A
(150 to 1050) A
0.1 % of rdg + 0.1 µA
0.095 % of rdg + 0.25 µA
0.083 % of rdg + 1.25 µA
0.1 % of rdg + 16.5 µA
0.18 % of rdg + 220 µA
0.19 % of rdg + 220 µA
0.29 % of rdg + 2.5 mA
0.58 % of rdg + 3.8 mA
0.44 % of rdg + 0.14 A
0.55 % of rdg + 0.14 A
Fluke 5080A
Fluke 5080A,
5500A/coil x50
Resistance – Generate
0 Ω
100 Ω
1000 Ω
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
1000 MΩ
0.014 Ω
0.049 % of rdg
0.029 % of rdg
0.029 % of rdg
0.044 % of rdg
0.11 % of rdg
0.12 % of rdg
1 % of rdg
1 % of rdg
Fluke 5080A
Tsuruga 5804
resistor
AC Voltage – Generate
45 Hz to 1 kHz
(1 to 32.99) mV
(33 to 329.99) mV
330 mV to 3.299 9 V
(3.3 to 32.999) V
(33 to 101.99) V
(102 to 329.99) V
(330 to 1020) V
0.42 % of rdg + 60 µV
0.2 % of rdg + 60 µV
0.15 % of rdg + 180 µV
0.16 % of rdg + 1.8 mV
0.19 % of rdg + 18 mV
0.19 % of rdg + 18 mV
0.2 % of rdg + 180 mV
Fluke 5080A | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. Page 3 of 5
(A2LA Cert. No. 2348.10) 06/25/2026
Parameter/Range
Frequency
CMC2, 3 (±) Comment
AC Voltage – Measure
(0 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 750) V
10 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
10 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
10 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
10 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
10 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
0.12 % of rdg + 0.04 mV
0.19 % of rdg + 0.05 mV
0.79 % of rdg + 0.08 mV
4.8 % of rdg + 0.5 mV
0.13 % of rdg + 0.3 mV
0.18 % of rdg + 0.5 mV
0.74 % of rdg + 0.8 mV
4.7 % of rdg + 5 mV
0.12 % of rdg + 3 mV
0.18 % of rdg + 5 mV
0.74 % of rdg + 8 mV
4.7 % of rdg + 50 mV
0.12 % of rdg + 30 mV
0.18 % of rdg + 50 mV
0.74 % of rdg + 80 mV
4.7 % of rdg + 500 mV
0.12 % of rdg + 0.3 V
0.18 % of rdg + 0.5 V
0.74 % of rdg + 0.8 V
4.7 % of rdg + 5 V
Fluke 8845A
AC Current – Generate
45 Hz to 1 kHz
(45 to 65) Hz
(65 to 440) Hz
(29 to 329.9) µA
330 µA to 3.299 9 mA
(3.3 to 32.999) mA
(33 to 329.99) mA
330 mA to 1.099 9 A
(1.1 to 2.999 9) A
(3 to 10.999) A
(11 to 20.5) A
(20.5 to 1000) A
0.31 % of rdg + 0.75 µA
0.27 % of rdg + 0.9 µA
0.23 % of rdg + 12 µA
0.32 % of rdg + 120 µA
0.28 % of rdg + 1.2 mA
0.33 % of rdg + 1.5 mA
0.47 % of rdg + 6 mA
0.8 % of rdg + 15 mA
0.57 % of rdg + 1 A
0.86 % of rdg + 0.98 A
Fluke 5080A
Fluke 5080A,
5500A/coil x50 | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. Page 4 of 5
(A2LA Cert. No. 2348.10) 06/25/2026
Parameter/Equipment
Range CMC2 (±) Comment
Electrical Simulation of
Thermocouple Indicators
–
Type J
(-200 to 0) °C
(0 to 1200) °C
1.2 °C
0.83 °C
Fluke 725
Electrical Simulation of
RTD Indicators –
Pt100-385
Pt100-392
(0 to 200) °C
(0 to 200) °C
0.39 °C
0.36 °C
Fluke 725
III. Mechanical
Parameter/Equipment Range CMC2, 4 (±) Comment
Scales & Balances
(0.1 to 250) kg
9 g + 0.6R
Class F weights
Pressure
Up to 100 psi
Up to 689 kPa
Up to 300 psi
Up to 2068 kPa
Up to 1000 psi
Up to 6895 kPa
Up to 10 000 psi
Up to 68 948 kPa
0.09 psi
0.62 kPa
0.22 psi
1.5 kPa
10 psi
69 kPa
10 psi
69 kPa
Fluke 700P06
Fluke 700P27
Fluke 700P08
Fluke 700P31 | ||
| 5 | - | RTD; Thermocouple; FrequencyShow page textExtracted from the scope document; tables may not line up. View scope for the original. Page 5 of 5
(A2LA Cert. No. 2348.10) 06/25/2026
IV. Thermodynamics
Parameter Range CMC2, 6 (±) Comment
RTD – Measure
Pt100-385
Pt100-3926
(0 to 200) °C
(0 to 200) °C
0.39 °C
0.36 °C
Fluke 725
Thermocouple – Measure
Type J
(50 to 105) °C
1.2 °C
Fluke 725, Fluke 9140
dry block
V. Time & Frequency
Parameter Range CMC2 (±) Comment
Frequency
(45 to 119.99) Hz
(120 to 1000) Hz
0.017 % of rdg + 2 mHz
0.16 % of rdg + 2 mHz
Fluke 5080A
____________________________________________________________
1 This laboratory does not offer commercial calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.
4 In the statement of CMC, R is the numerical value of the resolution of the device under test in microinches.
5 This scope meets A2LA’s P112 Flexible Scope Policy.
6 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter. | See scope document |
| 6 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994; ANSI/NCSL Z540.3-2006; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Other Locations
1 other site in this organization
Labs with overlapping accredited scope in United States
- Hottinger Bruel & KjaerDuluth, GA · ~78 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Trescal, Inc. - Atlanta, GAAlpharetta, GA · ~79 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Constellation Powerlabs LLC - Piedmont, SCPiedmont, SC · ~119 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Ludlum Measurements Inc. Knoxville Calibration DivisionKnoxville, TN · ~148 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Yokogawa Corporation of AmericaNewnan, GA · ~169 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.