Registers refreshed 2026-11-20.See refresh dates
United States

Micro Precision Calibration

9080 Activity Road, Suite C, 92126, San Diego, CA, United States

Visit website
A2LA 935.10ISO/IEC 17025Register refreshed 2026-11-20

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free, no obligation.
Scope lines
25
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We found no published test standard numbers in this lab’s scope. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 25 scope lines.

A2LA certificate 935.10 · 25 lines

LineMethodMatrixRange
1ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994; ANSI/NCSL Z540.3-2006Conductivity; pH-
2-Height Gages; Gage blocks, surface plate; Micrometers; Flatness; Parallelism; Optical flat; Optical parallels; Micrometer Standards; Universal length measuring machine, gage blocks; Indicators – Dial & Digital Indicators, LVDTs; Rigid Rules & Tape Measures – Length; Ruler calibrator; Surface Plate – Local Flatness; Repeat-O-meter; Gage Blocks – Length; Gage block comparator, universal length measuring machine, gage blocks; Plain Cylindrical Ring Gages – Internal Diameter; Cylindrical Pin – Plug Gages; Thread Wires
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 2 of 24 Parameter/Equipment Range CMC2, 8 (±) Comments Height Gages 3 Up to 40 in (62 + 1.7L) µin Gage blocks, surface plate Micrometers 3 – Resolution: 10 µin 50 µin 100 µin 0.001 in Flatness Parallelism Up to 1 in Up to 4 in (4 to 24) in (24 to 40) in 50 µin 80 µin (8.4 + 3.6L) µin (29 + 2.3L) µin (53 + 1.2L) µin 580 µin 8.0 µin 12 µin Gage blocks Optical flat Optical parallels Micrometer Standards (1 to 20) in (26 + 0.40L) µin Universal length measuring machine, gage blocks Indicators – Dial & Digital Indicators, LVDTs3 Up to 6 in (3.0 + 3.0L) µin Universal length measuring machine, gage blocks Rigid Rules & Tape Measures3 – Length Up to 96 in 0.0023 in Ruler calibrator Surface Plate 3 – Local Flatness (-1000 to 1000) µin 21 µin Repeat-O-meter Gage Blocks – Length (0.05 to 4) in (>4 to 20) in (4.3 + 0.40L) µin (0.50 + 1.3 L) µin Gage block comparator, universal length measuring machine, gage blocks Plain Cylindrical Ring Gages – Internal Diameter (0.05 to 12) in (10 + 1.2L) µin Universal length measuring machine, gage blocks Cylindrical Pin – Plug Gages Up to 12 in (5.7 + 1.1L) µin Universal length measuring machine, gage blocks Thread Wires, (80 to 5) TPI (0.007 to 0.1155) µin (13 + 7.7L) µin Universal length measuring machine, gage blocks
See scope document
3-Thread Plug Gages; Thread Rings Gages; Universal Length Measuring Machine; Length Standards; Optical Comparators; Vision Systems & Measuring Microscopes; Fixture Gages
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 3 of 24 Parameter/Equipment Range CMC2, 8 (±) Comments Thread Plug Gages – Major Diameter Simple Pitch Diameter (0.04 to 12) in (0.04 to 12) in (8.2 + 1.8L) µin (30 + 2.8L) µin Universal length measuring machine, gage blocks, master plug, thread measuring wires Thread Rings Gages – Minor Diameter Simple Pitch Diameter (0.1 to 4) in (0.1 to 4) in (21 + 0.60L) µin (22 + 7.0L) µin Universal length measuring machine, gage blocks, master ring, T- square probe Universal Length Measuring Machine3 Up to 4 in (>4 to 20) in (5.0 + 0.30L) µin (1.0 + 1.2L) µin Gage blocks Length Standards (Bars, Rods, Micrometers Setting Standards, & Fixtures Used as Length Standards) Up to 4 in (>4 to 20) in (8.0 + 0.80L) µin (1.0 + 7.3L) µin Gage blocks ULMM Optical Comparators 3 – Vision Systems & Measuring Microscopes X & Y Axis – Error of Indication Angle Up to 1 in (>1 to 8) in Up to 90° (56 + 27L) µin (81 + 2.3L) µin 4.3 arcsecond Gage blocks Glass scale Fixture Gages – Non- Contact Measurements – Length Angle Up to 7 in Up to 90° (97 + 3.3L) µin 4.3 arcsecond Optical comparator
See scope document
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 4 of 24 III. Electrical – DC/Low Frequency Parameter/Equipment Range CMC2, 4 (±) Comments DC Voltage 3 – Generate (0 to 100) mV >100 mV to 1 V (>1 to 10) V (>10 to 100) V (>100 to 1100) V 2.8 µV/V + 360 nV 3.0 µV/V + 620 nV 1.0 µV/V + 3.2 µV 4.2 µV/V + 18 µV 5.2 µV/V + 22 µV Multiproduct calibrator DC Voltage3 – Measure (0 to 100) mV >100 mV to 1 V (>1 to 10) V (>10 to 100) V (>100 to 1000) V (>1000 to 10 000) V (>10 000 to 100 000) V 600 nV/V + 260 nV 2.8 µV/V + 44 nV 3.5 µV/V – 560 nV 3.0 µV/V + 10 µV 8.7 µV/V – 500 µV 10 µV/V + 4.0 µV 25 µV/V + 1.0 µV Multimeter Multimeter / high voltage probe DC Current 3 – Generate (0 to 100) µA >100 µA to 1 mA (>1 to 10) mA (>10 to 100) mA >100 mA to 1 A (>1 to 11) A 1.0 µA/A + 3.0 nA 33 µA/A + 0.11 nA 57 µA/A – 24 nA 210 µA/A – 1.5 µA 100 µA/A + 9.0 µA 950 µA/A – 870 nA Multi-function calibrator DC Current 3 – Measure Up to 100 µA (>0.1 to 1) mA (>1 to 10) mA (>10 to 100) mA (>0.1 to 1) A (>1 to 11) A (>11 to 600) A 44 µA/A 36 µA/A + 0.60 nA 36 µA/A + 1.0 nA 120 µA/A – 820 nA 120 µA/A – 1.0 µA 310 nA/A + 5.0 µA 75 µA/A – 650 µA Multimeter Multimeter with shunt DC Current3 – Toroidal Clamp Meter Other Type Clamp Meter (9.5 to 550) A (9.5 to 550) A 3.0 mA/A + 4.0 mA 6.0 mA/A + 18 mA Multi-product calibrator W/50 turn coil DC Power3 – Generate (1 to 10) W (>10 to 100) W (>100 to 1000) W (>1000 to 5000) W (>5000 to 10 000) W 17 µW/W + 930 µW 95 µW/W 54 µW/W + 4.0 mW 5.0 µW/W + 53 mW 100 µW/W – 430 mW Multi-product calibrator
5-AC Voltage
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 5 of 24 Parameter/Range Frequency CMC2, 4 (±) Comments AC Voltage 3 – Generate Up to 0.022 V (>0.022 to 0.220) V (>0.220 to 2.2) V (>2.2 to 22) V (>22 to 220) V (>220 to 1100) V Up to 20 Hz (>20 to 40) Hz >40 Hz to 20 kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz (>300 to 500) kHz >500 kHz to 1 MHz Up to 20 Hz (>20 to 40) Hz >40 Hz to 20 kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz (>300 to 500) kHz >500 kHz to 1 MHz Up to 20 Hz (>20 to 40) Hz >40 Hz to 20 kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz (>300 to 500) kHz >500 kHz to 1 MHz Up to 20 Hz (>20 to 40) Hz >40 Hz to 20 kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz (>300 to 500) kHz >500 kHz to 1 MHz Up to 20 Hz (>20 to 40) Hz >40 Hz to 20 kHz (>20 to 50) kHz (>50 to 100) kHz ≤50 Hz (>50 to 300) Hz >300 Hz to 1 kHz 320 µV/V + 1.3 µV 200 µV/V + 1.0 µV 44 µV/V + 1.3 µV 360 µV/V + 1.0 µV 630 µV/V + 4.3 µV 1.3 mV/V + 11 µV 1.7 mV/V + 14 µV 2 mV/V + 12 µV 250 µV/V + 1.4 µV 110 µV/V + 1.1 µV 48 µV/V + 1.3 µV 130 µV/V + 1.2 µV 170 µ V/V + 5.3 µV 540 µV/V + 13 µV 820 µV/V + 15 µV 1.2 mV/V + 13 µV 240 µV/V + 1.4 µV 74 µV/V + 1.2 µV 34 µV/V + 1.3 µV 59 µV/V + 1.4 µV 67 µV/V + 5.5 µV 220 µV/V + 14 µV 690 µV/V + 16 µV 840 µV/V + 13 µV 250 µV/V + 1.4 µV 75 µV/V + 1.2 µV 44 µV/V + 1.3 µV 55 µV/V + 1.4 µV 70 µV/V + 5.5 µV 180 µV/V + 14 µV 600 µV/V + 16 µV 700 µV/V + 14 µV 250 µ V/V + 1.4 µV 90 µV/V + 1.1 µV 44 µV/V + 1.3 µV 28 µV/V + 1.4 µV 120 µV/V + 5.4 µV 55 µV/V + 2.0 µV 58 µV/V + 1.0 µV 61 µV/V + 1.1 µV Multi-function calibrator
-
6-AC Voltage
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 6 of 24 Parameter/Range Frequency CMC2, 4 (±) Comments AC Voltage 3 – Measure Up to 0.01 V (>0.01 to 0.1) V (>0.1 to 1) V (>1 to 10) V (>10 to 100) V (>100 to 1000) V Up to 40 Hz >40 Hz to 1 kHz (>1 to 20) kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz >300 kHz to 1 MHz (>1 to 2) MHz Up to 40 Hz >40 Hz to 1 kHz (>1 to 20) kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz >300 kHz to 1 MHz (>1 to 2) MHz Up to 40 Hz >40 Hz to 1 kHz (>1 to 20) kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz >300 kHz to 1 MHz (>1 to 2) MHz Up to 40 Hz >40 Hz to 1 kHz (>1 to 20) kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz >300 kHz to 1 MHz (>1 to 2) MHz Up to 40 Hz >40 Hz to 1 kHz (>1 to 20) kHz (>20 to 50) kHz (>50 to 100) kHz (>100 to 300) kHz >300 kHz to 1 MHz Up to 40 Hz 330 µV/V + 4.0 µV 170 µV/V + 2.0 µV 390 µV/V + 2.0 µV 1.2 mV/V + 2.0 µV 3.2 mV/V + 3.0 µV 13 mV/V + 25 µV 13 mV/V + 16 µV 4.4 mV/V + 170 µV 94 µV/V + 5.0 µV 97 µV/V + 2.0 µV 180 µV/V + 2.0 µV 360 µV/V + 2.0 µV 980 µV/V + 2.0 µV 3.4 mV/V + 23 µV 12 mV/V + 11 µV 16 mV/V + 66 µV 83 µV/V + 47 µV 87 µV/V + 23 µV 170 µV/V + 23µV 350 µV/V + 23 µV 930 µV/V + 27 µV 3.5 mV/V + 120 µV 12 mV/V + 120 µV 17 mV/V + 210 µV 84 µV/V + 460 µV 62 µV/V + 480 µV 170 µV/V + 220 µV 350 µV/V + 230 µV 920 µV/V + 280 µV 3.5 mV/V + 1.1 mV 12 mV/V + 1.0 mV 18 mV/V + 170 µV 230 µV/V + 4.5 mV 240 µV/V + 2.2 mV 250 µV/V + 2.1 mV 410 µV/V + 2.3 mV 1.4 mV/V + 2.2 mV 4.6 mV/V + 12 mV 17 mV/V + 22 mV 480 µV/V + 45 mV Multimeter, synchronous sub- sampled mode
-
7

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 7 of 24 Parameter/Range Frequency CMC2, 4 (±) Comments AC Voltage 3 – Measure (cont) (>100 to 1000) V (1000 to 30 000) kV (1000 to 30 000) kV >40 Hz to 1 kHz (>1 to 10) kHz (>10 to 100) kHz 60 Hz 1 kHz 470 µV/V + 22 mV 700 µV/V + 23 mV 1.4 mV/V + 20 mV 6.0 µV/V + 3.0 µV 4.0 µV/V - 38 µV Multimeter / high voltage divider, synchronous sub- sampled mode AC Current 3 – Generate Up to 2.2 mA (>2.2 to 22) mA (>22 to 220) mA Up to 2.2 mA (>2.2 to 22) mA (>22 to 220) mA >220 mA to 2.2 A Up to 2.2 mA (>2.2 to 22) mA (>22 to 220) mA >220 mA to 2.2 A Up to 2.2 mA (>2.2 to 22) mA (>22 to 220) mA >220 mA to 2.2 A Up to 2.2 mA (>2.2 to 22) mA (>22 to 220) mA >220 mA to 2.2 A (2.2 to 11) A (10 to 20) Hz (>20 to 40) Hz >40 Hz to 1 kHz (>1 to 5) kHz (>5 to 10) kHz Up to 65 Hz (>65 to 500) Hz >500 Hz to 1 kHz 160 µA/A + 48 nA 200 µA/A + 11 nA 76 mA/A + 1.2 mA 39 µA/A + 26 nA 120 µA/A - 50 nA 86 µA/A + 230 nA 71 µA/A + 1.8 µA 42 µA/A + 23 nA 110 µA/A - 50 nA 86 µA/A + 230 nA 71 µA/A + 1.8 µA 12 µA/A + 66 nA 110 µA/A – 36 nA 67 µA/A + 420 nA 61 µA/A + 1.1 µA 160 µA/A + 84 nA 140 µA/A + 100 nA 55 µA/A + 940 nA 420 µA/A + 35 µA 300 µA/A + 67 µA 420 µA/A + 11 µ A 370 mA/A + 400 µA Multi-function calibrator
8-AC Current
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 8 of 24 Parameter/Range Frequency CMC2, 4 (±) Comments AC Current 3 – Measure Up to 100 µA >100 µA to 1 mA (>1 to 100) mA >100 mA to 1 A Up to 100 µA >100 µA to 1 mA (>1 to 100) mA >100 to 1 A Up to 100 µA >100 µA to 1 mA (>1 to 10) mA (>10 to 100) mA >100 to 1 A Up to 100 µA >100 µA to 1 mA (>1 to 10) mA (>10 to 100) mA >100 to 1 A >100 µA to 1 mA (>1 to 10) mA (>10 to 100) mA >100 mA to 1 A >100 µA to 1 mA (>1 to 10) mA (>10 to 100) mA >100 mA to 1 A >100 µA to 1 mA (>1 to 10) mA (>10 to 100) mA (>1 to 10) A (>10 to 600) A Up to 20 Hz (>20 to 45) Hz (>45 to 100) Hz >100 Hz to 5 kHz (>5 to 20) kHz (>20 to 50) kHz (>50 to 100) kHz Up to 1 kHz (>1 to 50) kHz Up to 50 kHz 4.6 mA/A + 34 nA 4.8 mA/A + 22 nA 4.8 mA/A 4.9 mA/A – 11 1.8 mA/A + 34 nA 2.0 mA/A + 11 nA 2.0 mA/A 2.0 mA/A – 11 µA 640 µA/A + 36 nA 920 µA/A + 8.0 nA 930 µA/A 930 µA/A + 1.0 nA 1.2 mA/A – 30 µA 640 µA/A + 36 nA 530 µ A/A + 47nA 580 µA/A 580 µA/A + 1.0 nA 1.5 mA/A - 91 µA 700 µA/A + 230 nA 930 µA/A 930 µA/A + 3.0 nA 4 mA/A – 310 µA 4.6 mA/A + 460 nA 5.1 mA/A – 1.0 nA 5.1 mA/A + 15 nA 13 mA/A – 760 µA 6.3 mA/A + 1.8 µA 8.1 mA/A – 2.0 nA 8.1 mA/A + 24 nA 930 µA/A + 11 µA 950 µA/A 4.9 mA/A + 2.8 mA Multimeter Multimeter with shunt
-
9

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 9 of 24 Parameter/Range Frequency CMC2, 4 (±) Comments AC Current 3 – Measure (cont) Toroidal Clamp Meter: (9.5 to 550) A Other Type Clamp Meter: (9.5 to 550) A Up to 65 Hz (>65 to 440) Hz Up to 65 Hz (>65 to 440) Hz 3.0 mA/A + 3.0 mA 10 mA/A – 6.0 mA 8.0 mA/A + 18 mA 14 mA/A + 7.0 mA Multi-product calibrator W/50 turn coil AC Power 3 – Generate – PF = (0 to 1) (1 to 10 000) W 45 Hz to 1 kHz 57 µW/W + 2.0 mW Multi-product calibrator Parameter/Equipment Range CMC2, 4 (±) Comments Resistance 3 – Measure (0 to 9.99) Ω (10 to 99.99) Ω 100 Ω to 0.999 kΩ (1 to 9.99) kΩ (10 to 99.99) kΩ 100 kΩ to 0.99 MΩ (1 to 9.99) MΩ (10 to 99.99) MΩ 100 MΩ to 1 GΩ 57 µΩ/Ω + 28 µΩ 38 µΩ/Ω + 220 µΩ 24 µΩ/Ω + 1.6 mΩ 29 µΩ/Ω - 3.0 mΩ 18 µΩ/Ω + 110 mΩ 12 µΩ/Ω + 780 mΩ 170 µΩ/Ω – 150 Ω 220 µΩ/Ω – 670 Ω 6.4 mΩ/Ω – 620 kΩ Multimeter Resistance 3 – Generate (0 to 10.99) Ω (11 to 32.99) Ω (33 to 109.99) Ω (110 to 329.99) Ω 330 Ω to 1.099 kΩ (1.1 to 3.299) kΩ (3.3 to 10.99) kΩ (11 to 32.99) kΩ (33 to 109.99) kΩ (110 to 329.99) kΩ 330 kΩ to 1.099 MΩ (1.1 to 3.299) MΩ (3.3 to 10.99) MΩ 26 µΩ/Ω + 150 µΩ 31 µΩ/Ω + 2.6 mΩ 4.0 µΩ/Ω + 2.7 mΩ 1.0 µΩ/Ω + 1.7 mΩ 45 µΩ/Ω – 7.5 mΩ 62 µΩ/Ω – 25 mΩ 90 µΩ/Ω – 110 mΩ 1.6 µΩ/Ω + 420 mΩ 48 µΩ/Ω - 1.0 Ω 19 µΩ/Ω + 2.2 Ω 150 µΩ/Ω – 38 Ω 63 µΩ/Ω + 62 Ω 890 µΩ/Ω - 2.4 kΩ Multi-product calibrator
10

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 10 of 24 Parameter/Equipment Range CMC2, 4 (±) Comments Resistance 3 – Generate (cont) Fixed Points (11 to 32.99) MΩ (33 to 109.99) MΩ (110 to 330) MΩ 0 Ω 1 Ω 1.9 Ω 10 Ω 19 Ω 100 Ω 190 Ω 1 kΩ 1.9 kΩ 10 kΩ 19 kΩ 100 kΩ 190 kΩ 1 MΩ 1.9 MΩ 10 MΩ 19 MΩ 100 MΩ 1 GΩ 10 GΩ 100 GΩ 1 TΩ 10 TΩ 610 µΩ/Ω + 620 Ω 570 µΩ/Ω + 1.9 kΩ 3.0 mΩ/Ω – 0.29 MΩ 20 µΩ 110 µΩ 210 µΩ 140 µΩ 580 µΩ 1.5 mΩ 2.2 mΩ 10 mΩ 19 mΩ 100 mΩ 190 mΩ 1.3 Ω 3.0 Ω 7.0 Ω 91 Ω 320 Ω 6.8 kΩ 33 kΩ 12 MΩ 580 MΩ 5.8 GΩ 67 GΩ 1.3 TΩ Multi-product calibrator Resistance box Inductance 3 – Measuring Equipment (100 to 1000) µH (>1 to 10) mH (>10 to 100) mH (>100 to 1000) mH 1.0 % IV + 1.6 µH 1.8 % IV mH 1.8 % IV – 0.090 mH 0.92 % IV + 0.060 mH Decade inductor IV= indicated value Parameter/Range Frequency CMC2, 4 (±) Comments Capacitance 3 – Measuring Equipment Up to 3.29 nF (3.29 to 32.9) nF (>32.9 to 329.9) nF 50 Hz to 1 kHz 0.27 % IV + 0.0090 nF 0.30 % IV – 0.014 nF 0.14 % IV + 0.11 nF Multi-product calibrator IV= indicated value
11-capacitance; thermocouple indicators
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 11 of 24 Parameter/Range Frequency CMC2, 4 (±) Comments Capacitance3 – Measuring Equipment (cont) (>329.9 to 1.09) µF (>1.09 to 3.29) µF (>3.29 to 32.9) µF (>32.9 to 329.9) µF (>0.33 to 1.1) mF (100 to 1000) pF (>1000 to 10 000) pF (>0.01 to 0.1) µF (>0.1 to 1) µF 50 Hz to 1 kHz 0.15 % IV – 0.30 pF 0.37 % IV – 2.4 nF 0.33 % IV – 10 nF 0.54 % IV – 0.32 µF 0.27 % IV + 0.50 µF 0.020 % IV + 0.11 pF 0.061 % IV - 0.31 pF 0.054 % IV + 4.0 pF 0.085 % IV - 30 pF Multi-product calibrator Decade capacitor IV= indicated value Parameter/Equipment Range CMC2, 4 (±) Comments Electrical Calibration of Thermocouple Indicators 3 Type B Type C Type E Type J Type K Type N Type R Type S Type T (600 to 800) °C (800 to 1000) °C (1000 to 1820) °C (0 to 1800) °C (1800 to 2310) °C (-250 to 1000) °C (-210 to 1200) °C (-200 to 1372) °C (-200 to -1300) °C (0 to 250) °C (250 to 400) °C (400 to 1767) °C (0 to 250) °C (250 to 1760) °C (-250 to -150) °C (-150 to 400) °C 0.35 °C 0.29 °C 0.23 °C 0.20 °C 0.24 °C 0.070 °C 0.080 °C 0.10 °C 0.10 °C 0.31 °C 0.25 °C 0.21 °C 0.33 °C 0.24 °C 0.12 °C 0.078 °C Multi-product calibrator
-
12-Thermocouple; RTD Indicating Systems
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 12 of 24 Parameter/Equipment Range CMC2, 4 (±) Comments Electrical Calibration of Thermocouple – Measure 3 Type B Type C Type E Type J Type K Type N Type R Type S Type T (600 to 1000) °C (1000 to 1820) °C (0 to 150) °C (150 to 1000) °C (1000 to 1800) °C (1800 to 2310) °C (-250 to 350) °C (350 to 1000) °C (-210 to -100) °C (-100 to 1200) °C (-200 to 1372) °C (-200 to -1300) °C (0 to 400) °C (400 to 1000) °C (1000 to 1767) °C (0 to 1760) °C (-250 to -150) °C (-150 to 0) °C (0 to 120) °C (120 to 400) °C 0.35 °C 0.26 °C 0.19 °C 0.15 °C 0.18 °C 0.24 °C 0.070 °C 0.080 °C 0.090 °C 0.080 °C 0.10 °C 0.094 °C 0.31 °C 0.26 °C 0.21 °C 0.24 °C 0.095 °C 0.082 °C 0.090 °C 0.082 °C Multimeter Electrical Calibration of RTD Indicating Systems 3 – Pt 385, 100 Ω Pt 3926, 100 Ω (-200 to 0) °C (0 to 100) °C (100 to 300) °C (300 to 400) °C (400 to 630) °C (630 to 800) °C (-200 to 0) °C (0 to 100) °C (100 to 300) °C 0.058 °C 0.081 °C 0.10 °C 0.12 °C 0.14 °C 0.27 °C 0.058 °C 0.081 °C 0.10 °C Multi-product calibrator
See scope document
13-RTD Indicating Systems; Oscilloscope
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 13 of 24 Parameter/Equipment Range CMC2 (±) Comments Electrical Calibration of RTD Indicating Systems 3 – (cont) Pt 3926, 100 Ω Pt 3916, 100 Ω Pt 385, 200 Ω Pt 385, 500 Ω Pt 385, 1 kΩ PtNi 385, 120 Ω Cu 427, 10 Ω (300 to 400) °C (400 to 630) °C (-200 to -190) °C (-190 to -80) °C (-80 to 0) °C (0 to 100) °C (100 to 260) °C (260 to 300) °C (300 to 400) °C (400 to 600) °C (600 to 630) °C (-200 to 100) °C (100 to 260) °C (260 to 600) °C (600 to 630) °C (-200 to -80) °C (-80 to 100) °C (100 to 260) °C (260 to 400) °C (400 to 600) °C (600 to 630) °C (-200 to 0) °C (0 to 100) °C (100 to 260) °C (260 to 300) °C (300 to 600) °C (600 to 630) °C (-80 to 100) °C (1000 to 260) °C (-100 to 260) °C 0.12 °C 0.14 °C 0.29 °C 0.046 °C 0.058 °C 0.070 °C 0.081 °C 0.093 °C 0.10 °C 0.12 °C 0.27 °C 0.047 °C 0.058 °C 0.16 °C 0.18 °C 0.047 °C 0.058 °C 0.070 °C 0.093 °C 0.010 °C 0.13 °C 0.035 °C 0.047 °C 0.058 °C 0.070 °C 0.081 °C 0.27 °C 0.093 °C 0.16 °C 0.36 °C Multi-product calibrator Oscilloscope3 – DC Vertical Deflection Accuracy: 50 Ω Load (1 to 10) mV (>10 to 100) mV (>100 to 1000) mV (>1 to 5) V 4.0 μV 45 μV 500 μV 3.0 mV Scope calibrator w/active probe
-
14-Oscilloscope
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 14 of 24 Parameter/Equipment Range CMC2 (±) Comments Oscilloscope3 – (cont) DC Vertical Deflection Accuracy: 1 MΩ Load Vertical Deflection Accuracy: 50 Ω Load Square Wave Signal <10kHz Vertical Deflection Accuracy: 1 MΩ Load Square Wave Signal <10kHz Horizontal Cursor Accuracy (Time Base) (1 to 10) mV (>10 to 100) mV (>100 to 1000) mV (>1 to 10) V (>10 to 100) V (>100 to 220) V (1 to 10) mV (>10 to 100) mV (>100 to 1000) mV (>1 to 5) V (1 to 10) mV (>10 to 100) mV (>100 to 1000) mV (>1 to 10) V (>10 to 100) V (>100 to 220) V 5 s 2 s 1 s 500 ms 200 ms 100 ms 50 ms 20 ms 10 ms 5 ms 2 ms 1 ms 500 μs 200 μs 100 μs 50 μs 20 μs 10 μs 5 μs 2 μs 1 μs 500 ns 200 ns 100 ns 4.0 μV 54 μV 390 μV 5.0 mV 33 mV 350 mV 4.0 μV 38 μV 450 μV 3.0 mV 5.0 μV 40 μV 400 μV 4.0 mV 39 mV 490 mV 4.0 ms 2.0 ms 2.0 ms 0.39 ms 0.35 ms 0.28 ms 46 µs 39 µs 35 µs 4.0 µs 4.0 µs 2.0 µs 0.41 µs 0.34 µs 0.22 µs 62 ns 49 ns 47 ns 4.0 ns 3.0 ns 2.0 ns 0.62 ns 0.37 ns 0.33 ns Scope calibrator w/active probe
See scope document
15-Oscilloscope; RF Tuned Power; RF Absolute Power
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 15 of 24 Parameter/Range Frequency CMC2, 5 (±) Comments Oscilloscope3 – (cont) Horizontal Cursor Accuracy (Time Base) Bandwidth 50 Ω Bandwidth 50 Ω, Relative to 50 kHz Rise / Fall Time 3.4 ns 14 ns 50 ns 20 ns 10 ns 5 ns 2 ns 50 kHz 50 kHz to 100 MHz (>100 to 300) MHz (>300 to 600) MHz 600 MHz to 26.5 GHz (>26.5 to 50) GHz 50 kHz to 100 MHz (>100 to 300) MHz (>300 to 600) MHz 600 MHz to 26.5 GHz (>26.5 to 50) GHz (10 to 90) % (10 to 90) % 54 ps 33 ps 27 ps 4.0 ps 3.0 ps 0.44 % 0.71 % 0.76 % 0.81 % 1.1 % 1.9 % 0.38 % 0.78 % 0.83 % 1.1 % 1.9 % 0.39 ns 0.43 ns Scope calibrator w/active probe Multi-product calibrator power splitter, power sensors power meter IV. Electrical – RF/Microwave Parameter/Range Frequency CMC2, 7 (±) Comments RF Tuned Power 3 – Measure (20 to -30) dB (-30 to -130) dB 150 kHz to 26.5 GHz 0.15 dB 0.19 dB Measurement receiver RF Absolute Power 3 – Measure (-20 to 20) dB (-20 to 20) dB (-20 to 20) dB (-70 to -20) dB (-70 to -20) dB (-70 to -20) dB 9 kHz to 18 GHz (>18 to 40) GHz (>40 to 50) GHz (0.01 to 18) GHz (>18 to 40) GHz (>40 to 50) GHz 0.060 dB 0.090 dB 0.16 dB 0.075 dB 0.14 dB 0.23 dB Coaxial thermistor & thermocouple power sensors
-
16-Power Meter; Power Sensor
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 16 of 24 Parameter/Range Frequency CMC2, 5, 7 (±) Comments Power Meter 3 – Measurement Accuracy: (-30 to -130) dB 50 MHz Power Reference 100 mW 30 mW 10 mW 3.0 mW 1 mW 300 µW 100 µW 30 µW 10 µW 3 µW 0 mW 1 mW 0.32 mW 93 µW 32 µW 9.3 µW 3.2 µW 0.93 µW 0.32 µW 93 nW 32 nW 9.3 nW 0.72 nW 7.9 µW Range calibrator, multimeter, coaxial thermistor & thermocouple power sensors Power meter & power sensor Power Sensor Calibration Factor 3 – (-20 to 20) dB (-70 to -20) dB 50 MHz 9 kHz to 0.001 GHz (>0.001 to 0.01) GHz (>0.01 to 2) GHz (>2 to 4.2) GHz (>4.2 to 12.4) GHz (>12.4 to 18) GHz (>18 to 22) GHz (>22 to 26.5) GHz (>26.5 to 40) GHz (>40 to 50) GHz (0.01 to 0.03) GHz (>0.03 to 4) GHz (>4 to 18) GHz (>18 to 34) GHz (>34 to 40) GHz (>40 to 50) GHz 1.1 % 1.5 % 1.2 % 1.1 % 1.5 % 1.5 % 1.7 % 2.2 % 2.3 % 2.8 % 4.6 % 1.8 % 1.8 % 2.1 % 3.4 % 4.2 % 6.9 % Coaxial thermistor & thermocouple power sensors Frequency Modulation Rejection 3 – Equipment Rate: 400 HZ to 1 kHz Dev.: ≤5 kHz peak Dev.: ≤50 kHz peak Residuals AM3 250 kHz to 10 MHz 10 MHz to 26.5 GHz 150 kHz to 26.5GHz 0.12 % AM 0.12 % AM 0.0073 % (rms) Measuring receiver
-
17

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 17 of 24 Parameter/Range Frequency CMC2 (±) Comments Frequency Modulation Distortion 3 – Bandwidth: 20 Hz to 250 kHz 150 kHz to 26.5 GHz 0.12 % IV Measuring receiver IV = indicated value Frequency Modulation3 – Magnitude Deviation Rate: 20 Hz to 10 kHz Dev: ≤ 40 kHz Rate: 50 Hz to 200 kHz Dev: ≤ 400 kHz Rate: 50 Hz to 200 kHz Dev: ≤ 400 kHz Rate: 50 Hz to 200 kHz Dev: ≤ 400 kHz 250 kHz to 10 MHz 10 MHz to 6.6 GHz (6.6 to 13.2) GHz (13.2 to 26.5) GHz 0.10 Hz 0.20 Hz 0.40 Hz 0.50 Hz Measuring receiver Amplitude Modulation 3 – Rejection Rate: 400 HZ to 1 kHz Depth.: ≤50 % 150 kHz to 6.6 GHz (6.6 to 13.2) GHz (13.2 to 26.5) GHz 17 Hz 27 Hz 50 Hz Measuring receiver Amplitude Modulation3 – Depth Rate: 50 Hz to 10 kHz Depth: (5 to 99) % Rate: 50 Hz to 100 kHz Depth: (20 to 99) % Rate: 50 Hz to 100 kHz Depth: (5 to 20) % Rate: 50 Hz to 100 kHz Depth: (20 to 99) % Rate: 50 Hz to 100 kHz Depth: (5 to 20) % 150 kHz to 10 MHz 10 MHz to 3 GHz 10 MHz to 3 GHz (3 to 26.5) GHz (3 to 26.5) GHz 0.20 % IV 0.30 % IV 0.30 % IV 0.90 % IV 0.50 % IV Measuring receiver IV = indicated value
18

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 18 of 24 Parameter/Range Frequency CMC2 (±) Comments Amplitude Modulation 3 – Flatness Rate: (9 Hz to 10 kHz) Depth: (5 to 99) % 10 MHz to 26.5 GHz 0.15 % IV Measuring receiver IV = indicated value Residuals FM3 – Bandwidth: 50 Hz to 3 kHz 150 kHz to 10 MHz (3 to 6.6) GHz (6.6 to 13.2) GHz (13.2 to 26.5) GHz 1.3 Hz 1.9 Hz 2.7 Hz 3.8 Hz Measuring receiver Amplitude Modulation3 – Distortion Rate: 20 Hz to 250 kHz BW (0.01 to 100) % 150 kHz to 26.5 GHz 0.14 % IV Measuring receiver IV = indicated value Phase Modulation3 – Accuracy – Rate: 20 Hz to 200 kHz Deviation: >0.7 Rad Deviation: >0.3 Rad Rate: 20 Hz to 200 kHz Deviation: >2 Rad Deviation: >0.6 Rad Rate: 20 Hz to 200 kHz Deviation: >4 Rad Deviation: >1.2 Rad 100 kHz to 6.6 GHz 100 kHz to 6.6 GHz (6.6 to 13.2) GHz (6.6 to 13.2) GHz (13.2 to 26.5) GHz (13.2 to 26.5) GHz 0.10 % IV 0.20 % IV 0.10 % IV 0.20 % IV 0.10 % IV 0.20 % IV Measuring receiver IV = indicated value Phase Residuals 3 Rate: 50 Hz to 3 kHz 100 kHz to 26.5 GHz 0.010 rad Measuring receiver
19

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 19 of 24 Parameter/Range Frequency CMC2, 7 (±) Comments Phase Distortion 3 – Bandwidth: 20 Hz to 250 kHz 150 kHz to 26.5 GHz 0.14 % IV Measuring receiver IV = indicated value Harmonics3 – Measure 30 Hz to 50 GHz 1.3 dB Spectrum analyzer Reflection Magnitude & Phase S 11/22 – Measure3 5 Hz to 500 MHz 0.064 lin 0.36° Vector network analyzer, calibration kit & network analyze r Transmission Magnitude & Phase S 21/12 – Measure3 5 Hz to 500 MHz 0.014 dB 0.36° Vector network analyzer, calibration kit & network analyze r Reflection Magnitude & Phase S 11/22 – Measure3 30 Hz to 1.3 GHz (>1.3 to 3) GHz (>3 to 6) GHz (>6 to 8) GHz (>8 to 18) GHz (>18 to 26.5) GHz 0.002 lin 0.6° 0.002 lin 1.3° 0.004 lin 1.7° 0.005 lin 1.5° 0.010 lin 1.5° 0.013 lin 1.5° Network analyzer, cal kit Transmission Magnitude & Phase S 21/12 – Measure3 30 Hz to 1.3 GHz (>1.3 to 3) GHz (>3 to 6) GHz (>6 to 8) GHz (>8 to 18) GHz (>18 to 26.5) GHz 0.060 dB 0.4° 0.080 dB 0.8° 0.15 dB 0.8° 0.12 dB 1.8° 0.23 dB 1.8° 0.29 dB 1.8° Network analyzer, cal kit Impedance S 11/22 – Measure3 5 Hz to 500 MHz 30 Hz to 3 GHz (>3 to 6) GHz (>6 to 8) GHz (>8 to 18) GHz (>18 to 26.5) GHz 6.4 % IV 0.20 % IV 0.40 % IV 0.50 % IV 1.0 % IV 1.3 % IV Network analyzer, cal kit IV= indicated value Input Noise Level – Measure 3 30 Hz to 26.5 GHz 0.70 dB Cal kit
20

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 20 of 24 Parameter/Range Frequency CMC2, 7 (±) Comments Trace Noise Magnitude & Phase 3 – Measure 30 Hz to 3 GHz (>3 to 20) GHz (>20 to 26.5) GHz 0.070 dB 1.0° 0.20 dB 2.3° 0.30 dB 3.2° Cal kit Insertion Delay 3 – Group Delay Flatness F0= +/- 21 MHz 220 MHz to 6 GHz 220 MHz to 6 GHz 0.40 ns 0.40 ns Network analyzer, calibration unit V. Mechanical Parameter/Equipment Range CMC 2, 7 (±) Comments Pressure 3 – Measure & Measuring Equipment (-25 to 25) in H2O (-15 to 0) psi (0 to 15) psi (>15 to 300) psi (>300 to 500) psi (>500 to 15 000) psi 0.0030 in H 2O 0.0019 psi 0.0018 psi 0.037 psi 0.012 % IV + 0.000 43 psi 0.014 % IV + 1.2 psi Pressure calibrator IV= indicated value Dead weight tester Torque Tools 3 – Measure (0.1 to 199.9) ozfꞏin (0.15 to 9) lbfꞏin (10 to 100) lbfꞏin (0.1 to 200) lbf∙ft (>200 to 1000) lbf∙ft 1.2 % IV + 0.05 ozfꞏin 0.030 % IV + 0.05 lbfꞏin 0.50 % IV + 0.1 lbfꞏin 0.60 % IV + 0.04 lbf∙ft 1.3 % IV – 0.9 lbf∙ft Torque calibrator IV= indicated value Torque Analyzers & Transducers 3 – Measure Up to 100 lbfꞏin 0.020 % IV + 0.060 lbfꞏin Torque wheel/arm, weights Mass (1 to 100) mg 200 mg 500 mg 3.2 µg 6.0 µg 6.1 µg Ultra Class, class 0 & 1 weights with balance
21-weights; scales; balances; pipettes; syringes; burettes; dispensers; liquid handlers; titrators; dilutors; accelerometers
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 21 of 24 Parameter/Equipment Range CMC2, 7 (±) Comments Mass (cont) 1 g 2 g 5 g 10 g 20 g 50 g 100 g 200 g 500 g 1 kg 2 kg 3 kg 5 kg 10 kg 15 kg 20 kg 25 kg 8.4 µg 8.6 µg 10 µg 12 µg 16 µg 30 µg 52 µg 100 µg 0.60 mg 790 µg 2.1 mg 1.6 mg 2.6 mg 5.3 mg 5.9 mg 6.4 mg 6.7 mg Ultra Class, class 0 & 1 weights with balance Class 1, 2, & 3 weights with balance Scales & Balances3 Up to 250 g >250 g to 25 kg (>25 to 200) kg 1.5 µg/g + 7.0 µg 0.30 µg/g + 57 µg 0.23 µg/g + 2.3 mg Ultra Class, Class 1 & 3 weights, Class F weights POVA (Piston/Plunger Operated Volumetric Apparatus) Pipettes, Syringes, Piston/Displacement Burettes, Dispensers, Liquid Handlers, Titrators, & Dilutors Up to 50 µL (>0.050 to 200) mL 0.12 % IV + 23 nL 0.17 % IV Analytical balance IV = indicated value Accelerometers – Frequency Response 10 Hz to 2 kHz (>2 to 10) kHz (1 to 5) g 1.6 % IV 2.2 % IV Vibration calibration system, standard accelerometers IV = indicated value
-
22-Force; Humidity; Temperature
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 22 of 24 Parameter/Equipment Range CMC2, 5, 7 (±) Comments Force 3 – Measure & Measuring Equipment Tension Compression Up to 500 lbf Up to 25 000 lbf (>25 000 to 50 000) lbf (>50 000 to 100 000) lbf 0.0060 % + 0.0060 lbf 0.0010 % + 0.50 lbf 0.0010 % + 0.70 lbf 0.0010 % + 0.70 lbf Load cell & press VI. Thermodynamics Parameter/Equipment Range CMC 2, 7 (±) Comments Humidity 8 – Measuring Equipment (11 to 75) % RH (>75 to 99) % RH 0.010 % RH / % RH + 0.50 % RH 0.080 % RH / % RH + 0.50 % RH Humidity salts Humidity 3, 8 – Measure (0 to 11) % RH (>11 to 33) % RH (>33 to 100) % RH 0.022 % RH / % RH + 0.51 % RH 0.0010 % RH / % RH + 0.75 % RH 0.0070 % RH / % RH + 0.55 % RH Humidity meter Temperature3 – Measuring Equipment (Industrial/Process) (-80 to 420) °C (>420 to 450) °C 0.030 °C 0.070 °C Bath, drywell calibrator temperatur e meter, SPRT ITS-90 SPRT, PRT, RTD Probes Resistance Values Temperature Coefficients (-80 to <0.01) °C a 4 b 4 (≥0.01 to 420) °C a 8 b 8 1.8 mΩ - 0.22µΩ/Ω 0.094 % 0.046 % 1.8 mΩ + 0.25µΩ/Ω 0.095 % 0.026 % Bath, drywell calibrator, meter, SPRT
-
23-Temperature; Frequency; Timers; Photo Tachometer
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 23 of 24 Parameter/Equipment Range CMC2 (±) Comments Temperature 3 – Infrared Thermometers (10 to 300) °C 0.030 % IV + 1.2 °C Black body calibrator Ɛ = 0.99, λ = (8 to 14) µm IV = indicated value VII. Time & Frequency Parameter/Equipment Range CMC 2, 7 (±) Comments Frequency 3 – Measure Measuring Equipment Up to Hz to 8.7 MHz (>8.7 to 87) MHz (>87 to 250) MHz (>250 to 870) MHz >870 MHz to 40 GHz 1 Hz to 50 GHz (6.8 x 10 -11 Hz x f ) + 1.1 µHz (6.8 x 10-10 Hz x f ) Hz (3.3 x 10-11 Hz x f ) - 23 mHz (6.8 x 10-11 Hz x f ) - 160 mHz (6.8 x 10-11 Hz x f ) + 590 mHz 5.9 x 10 -12 Hz x f GPS receiver, counter GPS receiver, signal generator f = frequency in Hz Timers 3 – Measure Up to 9:59:59.999 hr 5.6 x 10-7/sec + 0.07 ms Function generator, counter Photo Tachometer3 – Measure Up to 100 RPM (>100 to 1000) RPM (>1000 to 10 000) RPM (>10 000 to 100 000) RPM (>100 000 to 200 000) RPM (0.0010 + 48 x 10 -6/R) RPM (0.012 + 1.1 x 10-6/R) RPM (0.045 + 2.1 x 10-6/R) RPM (0.45 + 2.1 x 10-6/R) RPM (0.51 + 1.5 x 10-6/R) RPM Function generator with lamp R = Reading in RPM
-
24

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 0935.10) Revised 03/24/2026 Page 24 of 24 Parameter/Equipment Range CMC2, 7 (±) Comments Rotational Speed 3 – Measure (Up to 100) RPM (>100 to 1000) RPM (>1000 to 10 000) RPM (>10 000 to 100 000) RPM (>100 000 to 200 000) RPM (0.0010 + 120 x 10 -6/R) RPM (0.0040 + 120 x 10-6/R) RPM (0.022 + 130 x 10-6/R) RPM (0.33 + 120 x 10-6/R) RPM (1.0 + 110 x 10-6/R) RPM Photo tachometer R = Reading in RPM ___________________________________________ 1 This laboratory offers commercial dimensional testing/calibration service and field calibration services. 2 Calibration and Measurement Capability (CMC) is th e smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded uncerta inties expressed at approximately the 95 % confidence level, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. 3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC. 4 The stated measured values are determined usi ng the indicated instrument (see Comments). This capability is suitable for the calibration of the devices intended to measure or generate the measured value in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a fraction/percentage of the reading plus a fixed floor specification. 5 In the statement of CMC, the value is defined as the percentage of reading unless otherwise noted. 6 This scope meets A2LA’s P112 Flexible Scope Policy. 7 The type of instrument or material being calibra ted is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter. 8 In the statement of CMC, L is the numerical value of the nominal length in the indicated units. 9 This laboratory meets R205 – Specific Requirements: Calibration Laboratory Accreditation Program for the types of dimensional tests listed above and is considered equivalent to that of a calibration certificate.
25ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994; ANSI/NCSL Z540.3-2006; R205--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

9080 Activity Road, Suite C, San Diego, CA, 92126, US

32.71571, -117.16383

Open in Maps

Accreditation data refreshed from the A2LA register on .

This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.