SIMCO Electronics - Salt Lake City

65 E. Wadsworth Park Drive, Suite 101, 84020, Draper, UT, United States

Visit website
A2LA 1395.20ISO/IEC 17025Register refreshed 2026-11-20

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free, no obligation.
Scope lines
23
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 23 scope lines.

A2LA certificate 1395.20 · 23 lines

LineMethodMatrixRange
1ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994-See scope document
2--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 2 of 22 Parameter/Equipment Range CMC2, 4 (±) Comments Height Masters & Height Gages Up to 24 in (30 + 1.3L) µin Gage blocks, amplifier, surface plate End Standards, Micrometer Standards, End Rods Up to 12 in (5.9 + 1.8L) µin Pratt & Whitney LabMaster TM, gage blocks Cylindrical Plug Gages & Master Discs Up to 12 in (3.9 + 1.8L) µin Pratt & Whitney LabMaster TM, gage blocks Precision Balls & Spheres (0.01 to 3) in (5.9 + 1.8L) µin Pratt & Whitney LabMaster TM, gage blocks Micrometers – Flatness Parallelism Up to 40 in Up to 1000 mm Up to 1 in Diameter Up to 0.001 in (10 + 14L) µin + 0.6R (0.23 + 14L) µm + 0.6R 7.3 µin 16 µin Gage blocks Optical flat Optical parallel Optical Flats Up to 3 in Diameter 4.7 µin 3 in Van Keuren master optical flat Thread Plugs – UN (60° threads) UN National Pipe (NPT) Pitch Diameter: Up to 12 in, Pitch: (4 to 80) TPI Pitch Diameter: (1 to 300) mm, Pitch: (0.3 to 4) mm Major Diameter: Up to 12 in Pitch Diameter: Up to 3 in Diameter, Pitch: (8 to 27) TPI Notch Height: Up to 1.5 in (70 + 4.2L) µin (1.8 + 0.043L) µm (27 + 4.2L) µin (99 + 3.6L) µin 160 µin Pratt & Whitney SupermicrometerTM with: thread wires (english) Thread wires (metric) Sine fixture & thread wires (english) Surface plate, gage blocks, electronic indicator
to 24 in
3Federal 832 (moody method)-to 40 in
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 4 of 22 II. Electrical – DC/Low Frequency Parameter/Range Frequency CMC2, 7 (±) Comments AC Current3 – Generate (22 to 220) µA (0.22 to 2.2) mA (2.2 to 22) mA (22 to 220) mA (0.22 to 2.2) A (2.2 to 11) A (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 20 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 0.026 % + 16 nA 0.017 % + 10 nA 0.013 % + 8 nA 0.033 % + 12 nA 0.11 % + 8.4 µA 0.026 % + 54 nA 0.017 % + 51 nA 0.013 % + 51 nA 0.021 % + 0.12 µA 0.11 % + 0.65 µA 0.028 % + 0.40 µA 0.017 % + 0.35 µA 0.013 % + 0.35 µA 0.021 % + 0.55 µA 0.11 % + 5.0 µA 0.028 % + 4.0 µA 0.017 % + 3.5 µA 0.013 % + 2.5 µA 0.022 % + 3.5 µA 0.11 % + 10 µA 0.027 % + 35 µA 0.045 % + 80 µA 0.70 % + 0.16 mA 0.047 % + 0.17 mA 0.095 % + 0.38 mA 0.36 % + 0.75 mA Fluke 57XXA/EP calibrator w/5725A
5

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 5 of 22 Parameter/Range Frequency CMC2, 7 (±) Comments AC Current3 – Generate (cont) (11 to 20.5) A (45 to 100) Hz 100 Hz to 1 kHz (1 to 5) kHz 0.11 % + 1.6 mA 0.11 % + 3.9 mA 2.3 % + 3.9 mA Fluke 552XA calibrator, LComp on AC Current3 – Measure (5 to 100) µA (0.1 to 1) mA (1 to 10) mA (10 to 100) mA (0.1 to 1) A (1 to 3) A (3 to 30) A (30 to 300) A 45 Hz to 1 kHz (45 to 100) Hz 100 Hz to 5 kHz (45 to 100) Hz 100 Hz to 5 kHz (45 to 100) Hz 100 Hz to 5 kHz (45 to 100) Hz 100 Hz to 5 kHz 45 Hz to 2 kHz (2 to 10) kHz (45 to 400) Hz (45 to 400) Hz 0.072 % + 0.035 µA 0.070 % + 0.23 µA 0.037 % + 0.23 µA 0.070 % + 2.3 µA 0.037 % + 2.3 µA 0.070 % + 23 µA 0.037 % + 23 µA 0.093 % + 0.23 mA 0.12 % + 0.23 mA 1.0 % + 3 mA 2.0 % + 6 mA 1.0 % + 60 mA 1.0 % + 0.6 A HP 3458A multimeter GW Instek PCS-1000
6

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 6 of 22 Parameter/Range Frequency CMC2, 7 (±) Comments AC Voltage3 – Generate (0.22 to 2.2) mV (2.2 to 22) mV (22 to 220) mV (0.22 to 2.2) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz 0.04 % + 4.0 µV 0.081 % + 4.0 µV 0.032 % + 4.0 µV 0.033 % + 4.0 µV 0.06 % + 5.0 µV 0.13 % + 10 µV 0.22 % + 20 µV 0.37 % + 20 µV 0.027 % + 4.0 µV 0.013 % + 4.0 µV 0.010 % + 4.0 µV 0.021 % + 4.0 µV 0.051 % + 5.0 µV 0.11 % + 10 µV 0.14 % + 20 µV 0.30 % + 20 µV 0.024 % + 12 µV 97 µV/V + 7.5 µV 82 µV/V + 7.5 µV 0.020 % + 7.5 µV 0.047 % + 17 µV 0.091 % + 20 µV 0.14 % + 25 µV 0.28 % + 45 µV 0.025 % + 0.26 mV 91 µV/V + 0.26 mV 46 µV/V + 0.26 mV 77 µV/V + 0.26 mV 0.011 % + 0.26 mV 0.043 % + 0.27 mV 0.10 % + 0.33 mV 0.18 % + 0.40 mV Fluke 57XXA/EP
7

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 7 of 22 Parameter/Range Frequency CMC2, 7 (±) Comments AC Voltage3 (cont) – Generate (2.2 to 22) V (22 to 220) V (220 to 1100) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (15 to 50) Hz 50 Hz to 1 kHz (1 to 20) kHz (20 to 30) kHz (30 to 50) kHz* (50 to 100) kHz * 0.024 % + 0.40 mV 92 µV/V + 0.15 mV 47 µV/V + 62 µV 76 µV/V + 0.11 mV 0.010 % + 0.20 mV 0.029 % + 0.60 mV 0.10 % + 2.0 mV 0.17 % + 3.2 mV 0.024 % + 4.8 mV 94 µV/V + 3.0 mV 56 µV/V + 2.6 mV 82 µV/V + 2.8 mV 0.016 % + 3.6 mV 0.090 % + 16 mV 0.44 % + 40 mV 0.80 % + 80 mV 0.030 % + 16 mV 74 µV/V + 4.3 mV 0.014 % + 6.0 mV 0.061 % + 6.0 mV 0.071 % + 11 mV 0.23 % + 45 mV Fluke 57XXA/EP w/5725A *750V Maximum AC Voltage3 – Measure 50 µV to 1 mV* (1 to 10) mV (10 to 100) mV (0.3 to 3) MHz* (3 to 10) MHz* (10 to 20) MHz* (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 3) MHz* (3 to 10) MHz* (10 to 20) MHz* (20 to 30) MHz* (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (1 to 3) MHz* (3 to 10) MHz* (10 to 20) MHz* (20 to 30) MHz* 3.0 % + 8.0 µV 8.0 % + 10 µV 20 % + 10 µV 0.035 % + 3.5 µV 0.023 % + 1.3 µV 0.035 % + 1.3 µV 0.012 % + 1.3 µV 0.6 % + 1.3 µV 4.6 % + 2.3 µV 1.5 % + 8.0 µV 2.5 % + 10 µV 6 % + 10 µV 13 % + 14µV 89 µV/V + 4.6 µV 89 µV/V + 2.3 µV 0.017 % + 2.3 µV 0.035 % + 2.3 µV 0.093 % + 2.3 µV 0.35 % + 12 µV 1.2 % + 12 µV 1.5 % + 40 µV 2.5 % + 50 µV 6.0 % + 50 µV 13 % + 64 V *R+S URE3 HP 3458A
8

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 8 of 22 Parameter/Range Frequency CMC2, 7 (±) Comments AC Voltage3 – Measure (cont) (0.1 to 1) V (1 to 10) V (10 to 100) V (100 to 700) V (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (1 to 3) MHz* (3 to 10) MHz* (10 to 20) MHz* (20 to 30) MHz* (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (1 to 3) MHz* (3 to 10) MHz* (10 to 20) MHz* (20 to 30) MHz* (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz (20 to 50) kHz (50 to 100) kHz 89 µV/V + 46 µV 89 µV/V + 23 µV 0.017 % + 23 µV 0.035 % + 23 µV 0.093 % + 23 µV 0.35 % + 0.12 mV 1.2 % + 0.12 mV 1.5 % + 0.60 mV 2.5 % + 0.60 mV 6.0 % + 0.90 mV 13 % + 1.2 mV 89 µV/V + 0.47 mV 89 µV/V + 0.23 mV 0.017 % + 0.23 mV 0.035 % + 0.23 mV 0.093 % + 0.23 mV 0.35 % + 1.2 mV 1.2 % + 1.2 mV 1.5 % + 3.0 mV 2.5 % + 7.0 mV 6.0 % + 9.0 mV 13 % + 9.0 mV 0.023 % + 4.6 mV 0.023 % + 2.3 mV 0.023 % + 2.3 mV 0.041 % + 2.3 mV 0.14 % + 2.3 mV 0.046 % + 47 mV 0.046 % + 24 mV 0.069 % + 23 mV 0.14 % + 23 mV 0.35 % + 23 mV HP 3458A *R+S URE3 AC Current – Clamp Meters3 (0 to 500) A (500 to 1000) A (45 to 400) Hz (45 to 400) Hz 0.37 % + 0.097 A 0.38 % + 0.2 A Fluke 552XA & 5500A/COIL
9

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 9 of 22 Parameter/Equipment Range CMC2, 7 (±) Comments DC Current3 – Generate 0 µA (0 to 220) µA (0.22 to 2.2) mA (2.2 to 22) mA (22 to 220) mA (0.22 to 2.2) A (2.2 to 11) A (11 to 20.5) A 0.4 nA 42 µA/A + 6.2 nA 37 µA/A + 7.2 nA 37 µA/A + 40 nA 47 µA/A + 0.75 µA 84 µA/A + 12 µA* 0.036 % + 0.48mA 0.079 % + 0.65 mA Reference open Fluke 57XXA/EP w/ 5725A *add (±200 × I²) µA/A for (>100 to 220) mA & (±10 × I²) µA/A for (>1 to 2.2) A Fluke 552XA DC Current – Clamp Meters 3 (0 to 1000) A 0.37 % + 0.065 A Fluke 552XA & 5500A/COIL DC Current3 – Measure 0 A (0 to 100) µA (0.1 to 1) mA (1 to 10) mA (10 to 100) mA (0.1 to 1) A (1 to 3) A* (3 to 30) A* (30 to 300) A* 0.4 nA 26 µA/A + 1.1 nA 26 µA/A + 7.5 nA 27 µA/A + 75 nA 44 µA/A + 0.75 µA 0.013 % + 16 µA 0.030 % + 0.30 mA 0.030 % + 3.0 mA 0.060 % + 30 mA Reference open HP 3458A opt 002 *GW Instek PCS-1000I DC Voltage3 – Generate 0 V (0 to 220) mV (0.22 to 2.2) V (2.2 to 11) V (11 to 22) V (22 to 220) V (220 to 1000) V 27 nV 8.6 µV/V + 1.3 µV 5.2 µV/V + 1.4 µV 3.7 µV/V + 3.1 µV 3.7 µV/V + 4.9 µV 5.2 µV/V + 47 µV 6.9 µV/V + 0.43 mV Reference short Fluke 57XXA/EP DC Voltage3 – Measure 0 V (0 to 100) mV (0.1 to 1) V (1 to 10) V (10 to 100) V (100 to 1000) V 27 nV 6.9 µV/V + 0.45 µV 5.4 µV/V + 0.46 µV 5.4 µV/V + 0.94 µV 8 µV/V + 45 µV 8 µV/V + 0.23 mV* Reference short HP 3458A opt 002 *Add 12 µV/V (V in/1000)2 for inputs >100 V
10

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 10 of 22 Parameter/Equipment Range CMC2, 4, 7 (±) Comments Resistance3 – Generate Fixed Values Variable 0 Ω 1.0 Ω 1.9 Ω 10 Ω 19 Ω 100 Ω 190 Ω 1.0 kΩ 1.9 kΩ 10 kΩ 19 kΩ 100 kΩ 190 kΩ 1.0 MΩ 1.9 MΩ 100 MΩ 19 MΩ 100 MΩ (0 to 11) Ω (11 to 33) Ω (33 to 110) Ω (110 to 330) Ω (0.33 to 1.1) kΩ (1.1 to 3.3) kΩ (3.3 to 11) kΩ (11 to 33) kΩ (33 to 110) kΩ (110 to 330) kΩ (0.33 to 1.1) MΩ (1.1 to 3.3) MΩ (3.3 to 11) MΩ (11 to 33) MΩ (33 to 110) MΩ (110 to 330) MΩ (0.33 to 1.1) GΩ 1.2 µΩ + 0.6R 98 µΩ/Ω 96 µΩ/Ω 26 µΩ/Ω 27 µΩ/Ω 11 µΩ/Ω 13 µΩ/Ω 9.2 µΩ/Ω 9.8 µΩ/Ω 9.0 µΩ/Ω 9.5 µΩ/Ω 12 µΩ/Ω 13 µΩ/Ω 21 µΩ/Ω 23 µΩ/Ω 45 µΩ/Ω 52 µΩ/Ω 0.011 % 32 µΩ/Ω + 0.78 mΩ 24 µΩ/Ω + 1.2 mΩ 22 µΩ/Ω + 1.1 mΩ 22 µΩ/Ω + 1.6 mΩ 22 µΩ/Ω + 1.6 mΩ 22 µΩ/Ω + 16 mΩ 22 µΩ/Ω + 16 mΩ 22 µΩ/Ω + 0.16 Ω 22 µΩ/Ω + 0.16 Ω 28 µΩ/Ω + 1.6 Ω 28 µΩ/Ω + 1.6 Ω 52 µΩ/Ω + 24 Ω 0.011 % + 40 Ω 0.021 % + 2 kΩ 0.041 % + 2.7 kΩ 0.24 % + 80 kΩ 1.2 % + 0.39 MΩ Reference short Fluke 57XXA/EP Fluke 552XA
11

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 11 of 22 Parameter/Equipment Range CMC2, 4, 7 (±) Comments Resistance3 – Measure 0 Ω (0 to 10) Ω (10 to 100) Ω (0.1 to 1) kΩ (1 to 10) kΩ (10 to 100) kΩ (0.1 to 1) MΩ (1 to 10) MΩ (10 to 100) MΩ (0.1 to 1.0) GΩ 1.2 µΩ + 0.6R 19 µΩ/Ω + 85 µΩ 16 µΩ/Ω + 0.82 mΩ 14 µΩ/Ω + 0.82 mΩ 14 µΩ/Ω + 8.2 mΩ 14 µΩ/Ω + 83 mΩ 20 µΩ/Ω + 2.7 Ω 62 µΩ/Ω + 0.12 kΩ 0.059 % + 3.7 kΩ 0.6 % + 0.26 MΩ Reference short HP 3458A Capacitance – Generate (0.19 to 0.3999) nF (0.4 to 1.039 99) nF (1.1 to 3.2999) nF (3.3 to 10.9999) nF (11 to 32.9999) nF (33 to 109.999) nF (110 to 329.999) nF (0.33 to 1.099 99) µF (1.1 to 3.299 99) µF (3.3 to 10.9999) µF (11 to 32.9999) µF (33 to 109.999) µF (110 to 329.999) µF (0.33 to 1.099 99) mF (1.1 to 3.2999) mF (3.3 to 10.9999) mF (11 to 32.9999) mF (33 to 110) mF 0.44 % + 8.1 pF 0.39 % + 8.1 pF 0.39 % + 8.1 pF 0.20 % + 8.1 pF 0.20 % + 78 pF 0.20 % + 80 pF 0.20 % + 0.23 nF 0.20 % + 0.78 nF 0.20 % + 2.3 nF 0.20 % + 23 nF 0.31 % + 23 nF 0.35 % + 78 nF 0.35 % + 0.23 µF 0.35 % + 0.78 µF 0.35 % + 2.3 µF 0.35 % + 7.8 µF 0.60 % + 23 µF 0.85 % + 78 µF Fluke 552XA
12

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 12 of 22 Parameter/Equipment Range CMC2 (±) Comments Electrical Simulation of Thermocouples3 – Type E Type J Type K Type N Type R/S Type T (-270 to -200) °C (-200 to 0) °C (0 to 600) °C (600 to 1000) °C (-210 to -100) °C (-100 to 900) °C (900 to 1200) °C (-270 to -200) °C (-200 to -100) °C (-100 to 600) °C (600 to 1000) °C (1000 to 1372) °C (-270 to -200) °C (-200 to 0) °C (0 to 1300) °C (-50 to 0 °C (0 to 500) °C (500 to 1768) °C (-270 to -200) °C (-200 to 0) °C (0 to 400) °C 0.091 °C 0.049 °C 0.041 °C 0.044 °C 0.056 °C 0.050 °C 0.054 °C 0.15 °C 0.068 °C 0.053 °C 0.060 °C 0.069 °C 0.23 °C 0.094 °C 0.070 °C 0.35 °C 0.29 °C 0.28 °C 0.13 °C 0.068 °C 0.049 °C Fluke 552XA Vdc output mode with external thermocouple reference probe
13

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 13 of 22 Parameter/Equipment Range CMC2, 5, 7 (±) Comments Oscilloscopes3 – DC Signal Square Wave Leveled Sine Wave: 5 mV to 5 V 5 mV to 2 V Edge Transition Timing Marker Functions Pulse Width ± 1 mV to ± 200 V dc (into 50 Ω) ± 1 mV to ± 50 Vdc (into 1 MΩ) 40 μV to <1 mV pk-pk 1 mV to 200 Vpk-pk 50 kHz Reference Amplitude 0.1 Hz to 300 MHz (>300 to 550) MHz (>0.55 to 2.5) GHz (>2.5 to 3.0) GHz (>3.0 to 6.5) GHz 500 ps, (10 Hz to 2 MHz rate) 150 ps, (10 Hz to 2 MHz rate) 70 ps, (10 Hz to 1 MHz rate) 9.0091 ns to <83 μs ≥83 μs to 55 s (1 to 100) ns 0.061 % + 25 μV 0.83 % + 10 μV 0.16 % + 10 μV 1.2 % 2.2 % 2.5 % 2.8 % 2.8 % 3.5 % 33 ps 14 ps 8 ps 0.2 μs/s 2.3 μs/s 4.1 % + 200 ps Fluke 9500B Oscilloscope calibrator with 9530/9560 active heads
14

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 14 of 22 III. Electrical – RF/Microwave Parameter/Range Frequency CMC2, 5, 7 (±) Comments Amplitude Modulation – Measure Rate: 50 Hz to 10 kHz Depth: (5 to 99) % Rate: 20 Hz to 10 kHz Depth: to 99 % Rate: 50 Hz to 50 kHz Depth: (5 to 99) % Rate: 20 Hz to 100 kHz Depth: to 99 % Rate: 50 Hz to 50 kHz Depth: (5 to 99) % Rate: 20 Hz to 100 kHz Depth: to 99 % 150 kHz to 10 MHz 150 kHz to 10 MHz >10 MHz to 1.3 GHz >10 MHz to 1.3 GHz (>1.3 to 26.5) GHz (>1.3 to 26.5) GHz 2.1 % + 1 digit 3 % + 1 digit 1.1 % + 1 digit 3 % + 1 digit 1.6 % + 1 digit 3 % + 1 digit HP 8902A measuring receiver with 11722/92A sensors & 11793A mixer Frequency Modulation – Measure Rate: 20 Hz to 10 kHz Dev: <40 kHz/pk Rate: 50 Hz to 100 kHz Dev: <400 kHz/pk Rate: 20 Hz to 200 kHz Dev: <400 kHz/pk 250 kHz, 10 MHz 10 MHz to 26.5 GHz 10 MHz to 26.5 GHz 2 % + 1 digit 1.1 % + 1 digit 5 % + 1 digit HP 8902A measuring receiver Frequency Response (Flatness) (0.1 to 20) MHz @ 1 Vrms (>20 to 50) MHz @ 1 Vrms (>50 to 100) MHz @ 1 Vrms (0.1 to 20) MHz @ 3 Vrms (>20 to 50) MHz @ 3 Vrms (>50 to 100) MHz @ 3 Vrms 0.16 % 0.61 % 1.8 % 0.48 % 0.75 % 1.4 % Thermal converters & DMM
15

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 15 of 22 Parameter/Range Frequency CMC2, 5, 7 (±) Comments RF Attenuation – Measure (0 to 10) dB (>10 to 20) dB (>20 to 30) dB (>30 to 40) dB (>40 to 50) dB (>50 to 60) dB (>60 to 70) dB (>70 to 80) dB (>80 to 90) dB (>90 to 100) dB (>100 to 110) dB (0 to 10) dB (>10 to 20) dB (>20 to 30) dB (>30 to 40) dB (>40 to 50) dB (>50 to 60) dB (>60 to 70) dB (>70 to 80) dB (>80 to 90) dB (>90 to 100) dB (>100 to 110) dB (0 to 10) dB (>10 to 20) dB (>20 to 30) dB (>30 to 40) dB (>40 to 50) dB (>50 to 60) dB (>60 to 70) dB (>70 to 80) dB (>80 to 90) dB (>90 to 100) dB (>100 to 110) dB 0.1 MHz to 1.3 GHz (>1.3 to 18) GHz (>18 to 26.5) GHz 0.064 dB 0.068 dB 0.079 dB 0.079 dB 0.095 dB 0.12 dB 0.14 dB 0.16 dB 0.20 dB 0.22 dB 0.33 dB 0.078 dB 0.083 dB 0.093 dB 0.1 dB 0.13 dB 0.15 dB 0.17 dB 0.2 dB 0.23 dB 0.33 dB 0.34 dB 0.095 dB 0.097 dB 0.1 dB 0.12 dB 0.14 dB 0.16 dB 0.17 dB 0.21 dB 0.23 dB 0.32 dB 0.34 dB HP 8902A measuring receiver with 11722A sensor HP 8902A measuring receiver with 11792A sensor
16

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 16 of 22 Parameter/Range Frequency CMC2, 5, 7 (±) Comments Absolute RF Power – Measure 1 µW to 100 mW* (-30 to +20 dBm) 100 pW to 1.0 µW (-70 to -30 dBm) 0 dBm at 50 MHz (100 to 300) kHz >300 kHz to 1 MHz >1 MHz to 2 GHz (>2 to 12.4) GHz (>12.4 to 18) GHz (>18 to 26.5) GHz (>26.5 to 40) GHz (>40 to 50) GHz (0.05 to 0.1) GHz (0.1 to 2) GHz (>2 to 12.4) GHz (>12.4 to 18) GHz (>18 to 34) GHz (>34 to 40) GHz (>40 to 50) GHz 0.6 % 1.7 % + 140 nW 1.2 % + 140 nW 1.2 % + 140 nW 1.6 % + 140 nW 2.3 % + 140 nW 2.4 % + 140 nW 2.8 % + 140 nW 4.1 % + 140 nW 2.5 % + 57 pW 1.9 % + 57 pW 1.8 % + 57 pW 2.1 % + 57 pW 2.9 % + 57 pW 3.4 % + 57 pW 5.4 % + 57 pW Power meter & thermistor with DC substitution 8480-series power sensors & EPM-series meter, *add 1.9 % to CMC for sensor linearity at levels greater than 10 mW (+10 dBm) 8487D power sensor & EPM-series meter
17

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 17 of 22 Parameter/Range Frequency CMC2, 5, 7 (±) Comments Tuned RF Power – Measure 100 kHz to 1.3 GHz (>1.3 to 18) GHz (>18 to 26.5) GHz (+10 to 0) dBm (<0 to -10) dBm (<-10 to -20) dBm (<-20 to -30) dBm (<-30 to -40) dBm (<-40 to -50) dBm (<-50 to -60) dBm (<-60 to -70) dBm (<-70 to -80) dBm (<-80 to -90) dBm (<-90 to -100) dBm (<-100 to -110) dBm (<-110 to -120) dBm (+10 to 0) dBm (<0 to -10) dBm (<-10 to -20) dBm (<-20 to -30) dBm (<-30 to -40) dBm (<-40 to -50) dBm (<-50 to -60) dBm (<-60 to -70) dBm (<-70 to -80) dBm (<-80 to -90) dBm (<-90 to -100) dBm (<-100 to -110) dBm (<-110 to -120) dBm (+10 to 0) dBm (<0 to -10) dBm (<-10 to -20) dBm (<-20 to -30) dBm (<-30 to -40) dBm (<-40 to -50) dBm (<-50 to -60) dBm (<-60 to -70) dBm (<-70 to -80) dBm (<-80 to -90) dBm (<-90 to -100) dBm (<-100 to -110) dBm (<-110 to -120) dBm 2.5 % 2.5 % 2.7 % 3.2 % 3.5 % 3.8 % 4.2 % 5 % 5.4 % 5.9 % 6.4 % 6.9 % 13 % 2.5 % 2.6 % 2.7 % 3 % 3.4 % 3.8 % 4.2 % 5 % 5.4 % 5.9 % 6.4 % 6.9 % 13 % 2.5 % 2.5 % 2.7 % 2.9 % 3.4 % 3.8 % 4.2 % 5 % 5.4 % 5.9 % 6.3 % 6.9 % 13 % HP 8902A measuring receiver with 11722A sensor HP 8902A measuring receiver with 11792A sensor
18

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 18 of 22 Parameter/Range Frequency CMC2, 5, 7 (±) Comments Phase Modulation – Measure Rate: 200 Hz to 10 kHz Rate: 20 Hz to 200 kHz 150 kHz to 10 MHz 10 MHz to 26.5 GHz 4 % + 0.015 rad 5 % + 0.019 rad HP 8902A measuring receiver w/ HP 11792A sensor Harmonic Distortion – Measure (-80 to 0) dB (-65 to 0) dB (-117 to 0) dB 20 Hz to 20 kHz (20 to 100) kHz 9 kHz to 1.3 GHz (1.3 to 22) GHz (>22 to 50) GHz 1.0 dB 2.0 dB 3.1 dB 3.5 dB 3.9 dB HP 8903A audio analyzer HP 8565E spectrum analyzer IV. Mechanical Parameter/Equipment Range CMC2, 4 (±) Comments Durometers – Indenter Geometry: Diameter Angle8 Extension Force Calibration: Type A, B, E, O Type C, D, DO Up to 0.010 in (0.010 to 0.50) in (29 to 36) degrees Up to 0.1 in (56 to 821) gf (0 to 4.6) kgf (150 + 21L) µin (110 + 21L) µin 290 arc sec (140 + 30L) µin 1.9 gf 11 gf Starrett optical comparator Mettler-Toledo precision balance
19

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 19 of 22 Parameter/Equipment Range CMC2, 4, 5, 10 (±) Comments Mass 1 mg to 40 g (>40 to 200) g >200 g to 6 kg (>6 to 32) kg 0.048 mg 0.16 mg 12 mg 0.12 g Class S weights, analytical balances Scales & Balances3 (1 to 50) mg (>50 to 500) mg >500 mg to 5 g (>5 to 20) g >20 g to 30 kg (5 to 700) lb 0.025 mg + 0.6R 0.033 mg + 0.6R 0.058 mg + 0.6R 0.078 mg + 0.6R 3.1 µg/g + 0.6R 0.01 % + 0.6R Class S weights Class F weights Force – Measuring Equipment Compression & Tension 10 gf to 320 kgf (0.002 to 700) lbf 0.05 % + 0.6R 0.05 % + 0.6R Class F weights Pneumatic Pressure – Measuring Equipment Dry Nitrogen Gas (0 to 90) psi(a) (90 to 300) psi(a) (0 to 90) psi(g) (90 to 300 psi(g) (300 to 1000) psi(a) 0.0021 % + 0.009 psi + 0.6R 0.01 % + 0.000 64 psi + 0.6R 0.0021 % + 0.009 psi + 0.6R 0.01 % + 0.000 64 psi + 0.6R 0.01 % of IV + 0.6R Additel 780S Controller, ADT160A-RD- AP300Q Module Additel 780S Controller, ADT160A-RD- AP1KQ Module
20--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 20 of 22 Parameter/Equipment Range CMC2, 4, 5, 10 (±) Comments Vacuum (-12.5 to 0) psi(g) 0.0021 % + 0.009 psig + 0.6R Additel 780S controller, ADT160A-RD- AP300Q Module Torque Handtools – Measuring Equipment 3 10 ozf·in to 1000 lbf·ft 1.0 % + 0.6R Digital torque load cells Torque Measurement Devices – Generate 10 ozf·in to 1000 lbf·ft 0.06 % Calibration arms & weights V. Thermodynamics Parameter/Equipment Range CMC2, 4, 10 (±) Comments Temperature3 – Measure (-180 to 650) °C (650 to 1000) °C 0.014 °C + 0.6R 0.63 °C + 0.6R SPRT, super thermometer Type S thermocouple Temperature3 – Measuring Equipment (-40 to 120) °C (120 to 650) °C (650 to 1000) °C 0.014 °C + 0.6R 0.22 °C + 0.6R 0.70 °C + 0.6R SPRT super thermometer with: Bath Drywell Type S thermocouple & Fluke Blackstack
-12.5 to 0 psi(g)
21NIST SP 960-12-to 24 hr
22

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 1395.20) Revised 04/16/2025 Page 22 of 22 3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibratio n and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC. 4 Unless otherwise noted, in the statement of CMC Uncertainty, L is the numerical value of the nominal length of the device measured in inches. In the statement of CMC Uncertainty, R is the numerical value of the resolution of the device in microinches. In the statement of CMC Uncertainty, IV represents indicated value. 5 In the statement of CMC, the value is defined as the percentage of reading unless otherwise noted. 6 D = Largest Diameter or Diagonal, as applicable to Surface Plate shape. 7 The stated measured values are determined using the indicated instrument (see Comments). This capability is suitable for the calibration of the devices intended to measure or generate the measured value in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percent or fraction of the reading plus a fixed floor specification. 8 Angular measurements using geometric construction of 2-lines measured by motion. 9 This scope meets A2LA’s P112 Flexible Scope Policy. 10 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter. 11 Repeatability of the Unit Under Test has not been utilized in the calculation of the CMC value for this measurement parameter.
23ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

65 E. Wadsworth Park Drive, Suite 101, Draper, UT, 84020, US

40.52478, -111.86280

Open in Maps

Accreditation data refreshed from the A2LA register on .