FindTestingLabs
Registers refreshed 2026-10-25.See refresh dates
United States

Eustis Co., Inc./Pyrocom Calibration

12407-B Mukilteo Speedway, #200, 98087, Lynnwood, WA, United States

Visit website
A2LA 2496.01ISO/IEC 17025Register refreshed 2026-10-23

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free. No lab gets your details until you choose.
Scope lines
5
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 5 scope lines.

A2LA certificate 2496.01 · 5 lines

LineMethodMatrixRange
1ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994-See scope document
2

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2496.01) 08/15/2024 Page 2 of 4 Parameter/Equipment Range CMC2, 3 (±) Comments Temperature – Measuring Equipment (cont) Calibration of Thermocouples: Type J Type B Type R Type S Type E Type J Type K Type N Type T Type B Type R Type S Type E Type J Type K Type N Type T Type T Type B Type E Type J Type K Type N Type R Type S Type B Type E Type J Type K Type N Type R Type S (0 to 100) °C (25 to 100) °C (101 to 125) °C (101 to 300) °C (126 to 300) °C (301 to 400) °C (301 to 600) °C (601 to 660) °C 0.31 °C 0.67 °C 0.59 °C 0.52 °C 0.27 °C 0.31 °C 0.40 °C 0.33 °C 0.27 °C 0.67 °C 0.60 °C 0.52 °C 0.29 °C 0.33 °C 0.40 °C 0.33 °C 0.29 °C 0.35 °C 0.68 °C 0.30 °C 0.33 °C 0.41 °C 0.33 °C 0.60 °C 0.52 °C 0.85 °C 0.60 °C 0.62 °C 0.66 °C 0.62 °C 0.79 °C 0.73 °C 5628 PRT
3

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2496.01) 08/15/2024 Page 3 of 4 Parameter/Equipment Range CMC2, 3 (±) Comments Temperature – Measuring Equipment (cont) Calibration of Thermocouples: Type B Type E Type J Type K Type N Type R Type S Type E Type B Type J Type K Type N Type R Type S Type B Type K Type N Type R Type S Type B Type K Type N Type R Type S (661 to 700) °C (701 to 800) °C (701 to 1000) °C (1001 to 1100) °C (1101 to 1210) °C 1.1 °C 0.88 °C 0.89 °C 0.92 °C 0.89 °C 1.1 °C 0.98 °C 0.90 °C 1.1 °C 0.91 °C 0.94 °C 0.91 °C 1.1 °C 0.99 °C 1.3 °C 1.2 °C 1.2 °C 1.3 °C 1.2 °C 2.8 °C 2.7 °C 2.7 °C 2.7 °C 2.7 °C Type S thermocouple
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 2496.01) 08/15/2024 Page 4 of 4 Parameter/Equipment Range CMC2, 3 (±) Comments Temperature – Measuring Equipment Calibration of RTDs: Pt 385, 100 Ω and Pt 392, 100 Ω -196 °C (-196 to -81) °C (-80 to -1) °C (0 to 100) °C (101 to 300) °C (301 to 600) °C (601 to 660) °C 0.12 °C 0.16 °C 0.12 °C 0.13 °C 0.14 °C 0.15 °C 0.54 °C 5628 PRT ___________________________________________ 1 This laboratory offers commercial calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. 3 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter. 4 This scope meets A2LA’s P112 Flexible Scope Policy.
5ISO/IEC 17025:2017; R205--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

12407-B Mukilteo Speedway, #200, Lynnwood, WA, 98087, US

47.82787, -122.30539

Open in Maps

Accreditation data refreshed from the A2LA register on .

This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.