Registers refreshed 2026-11-20.See refresh dates
United States

Trescal, Inc. - Long Island, NY

80 Orville Drive, Suite 115, 11716, Bohemia, NY, United States

Visit website
A2LA 3471.01ISO/IEC 17025Register refreshed 2026-11-20

We check which accredited labs can do it, starting with this one, and send you a short comparison of time and rough cost within three working days.

  • Free, no obligation.
Scope lines
32
A2LA certificate
1
Site in our directory
1
A2LA accredited scope

What this lab is accredited for

We found no published test standard numbers in this lab’s scope. The full scope as published is below.

Only this lab. Searches the method, matrix and range of each of the 32 scope lines.

A2LA certificate 3471.01 · 32 lines

LineMethodMatrixRange
1ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994-to 12 in
2

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 2 of 31 II. Electrical – DC/Low Frequency Parameter/Equipment Range CMC2, 4, 5 (±) Comments DC Voltage3 – Generate Up to 220 mV 220 mV to 2.2 V (2.2 to 11) V (11 to 22) V (22 to 220) V (220 to 1000) V (0 to 120) mV (0 to 1.2) V (0 to 12) V (0 to 120) V (0 to 1020) V 0.49 µV + 5.7 nV/mV 0.8 µV + 3.5 µV/V 2.9 µV + 2.5 µV/V 4.3 µV + 2.5 µV/V 43 µV + 3.5 µV/V 0.42 mV + 4.5 µV/V 0.8 µV + 9.3 µV/V 1.9 µV + 5.8 µV/V 10 µV + 6.2 µV/V 0.1 mV + 8.5 µV/V 1 mV + 8.5 µV/V Fluke 5720A Fluke 5560A DC Voltage3 – Generate, Fixed Points 1.018 V 10 V 2.6 µV 23 µV Fluke 732B DC Voltage3 – Measure Up to 200 mV 200 mV to 2 V (2 to 20) V (20 to 200) V (200 to 1000) V (1000 to 10 000) V (1 to 45) kV (45 to 160) kV 0.24 µV + 5 nV/mV 1.1 µV + 3.7 µV/V 10 µV + 3.5 µV/V 0.88 mV + 3.2 µV/V 1.4 mV + 5.2 µV/V 0.018 % 0.13 % 0.13 % Fluke 8508A w/ Fluke 80E-10 Ross Divider, DMM DC Current3 – Generate Up to 220 µA 220 µA to 2.2 mA (2.2 to 22) mA (22 to 220) mA 220 mA to 2.2 A (2.2 to 11) A (11 to 20.5) A (2 to 20) A (20 to 100) A 6 nA + 35 pA/µA 7.1 nA + 30 nA/mA 41 nA + 30 nA/mA 0.73 µA + 50 nA/mA 12 µA + 0.11 mA/A 0.49 mA + 0.34 mA/A 9.4 mA + 0.8 mA/A 13 mA + 0.13 mA/A 38 mA + 0.39 mA/A Fluke 5720A w/ Fluke 5725A Fluke 5520A Valhalla 2555A
3

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 3 of 31 Parameter/Equipment Range CMC2, 4 (±) Comments DC Current3 – Generate (cont) (0 to 120) µA (0 to 1.2) mA (0 to 12) mA (0 to 120) mA (0 to 1.2) A (0 to 3.1) A (0 to 12) A (0 to 30.2) A 6 nA + 97 µA/A 15 nA + 78 µA/A 82 nA + 78 µA/A 0.82 µA + 78 µA/A 10 µA + 0.12 mA/A 0.15 mA + 0.23 mA/A 0.25 mA + 0.23 mA/A 0.5 mA + 0.78 mA/A Fluke 5560A DC Current3 – Measure Up to 220 µA 200 µA to 2 mA (2 to 20) mA (20 to 200) mA 200 mA to 2 A 2.4 nA + 7 pA/µA 7 nA + 11 nA/mA 53 nA + 14 nA/mA 1.5 nA + 47 nA/mA 26 µA + 0.18 mA/A Fluke 8508A Parameter/Range Frequency CMC2, 4, 5 (±) Comments DC Current3 – Measure Up to 15 A (15 to 100) A DC to 60 Hz DC to 60 Hz 0.0034 % 0.0037 % Leeds & Northrup 4360 Leeds & Northrup 4361 Clamp Meters3, AC Current – Generate Toroidal: (10 to 16.5) A (16.5 to 150) A (150 to 1025) A (10 to 16.5) A (16.5 to 150) A (150 to 1025) A Non-Toroidal: (10 to 16.5) A (16.5 to 150) A (150 to 1025) A (10 to 16.5) A (16.5 to 150) A (150 to 1025) A (45 to 65) Hz (65 to 440) Hz (45 to 65) Hz (65 to 440) Hz 26 mA + 2.1 mA/A 50 mA + 1.9 mA/A 0.34 A + 1.9 mA/A 60 mA + 6 mA/A 0.11 A + 5.3 mA/A 0.86 A + 5.3 mA/A 60 mA/A + 3.8 mA/A 0.23 A + 3.7 mA/A 1.2 A + 3.7 mA/A 90 mA + 7.2 mA/A 0.28 A + 6.7 mA/A 1.6 A + 6.7 mA/A Fluke 5520A w/ coil
4

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 4 of 31 Parameter/Equipment Range CMC2, 4 (±) Comments Clamp Meters, DC Current – Generate (10 to 16.5) A (16.5 to 150) A (150 to 1025) A 50 mA + 6.5 mA/A 0.18 A + 3.4 mA/A 0.83 A + 3.3 mA/A Fluke 5520A w/ coil Resistance3 – Generate Fixed Points 0 Ω 1 Ω 1.9 Ω 10 Ω 19 Ω 100 Ω 190 Ω 1 kΩ 1.9 kΩ 10 kΩ 19 kΩ 100 kΩ 190 kΩ 1 MΩ 1.9 MΩ 10 MΩ 19 MΩ 100 MΩ 1 Ω 10 kΩ (0 to 10.9999) Ω (11 to 32.9999) Ω (33 to 109.9999) Ω (110 to 329.9999) Ω (0.33 to 1.099 999) kΩ (1.1 to 3.299 999) kΩ (3.3 to 10.999 99) kΩ (11 to 32.999 99) kΩ (33 to 109.9999) kΩ (110 to 329.9999) kΩ (0.33 to 1.099 999) MΩ (1.1 to 3.299 999) MΩ (3.3 to 10.999 99) MΩ (11 to 32.999 99) MΩ (33 to 109.9999) MΩ (110 to 329.9999) MΩ (330 to 1100) MΩ 50 µΩ 95 µΩ 0.18 mΩ 0.26 mΩ 0.48 mΩ 1.1 mΩ 2.1 mΩ 9 mΩ 17 mΩ 90 mΩ 0.17 Ω 1.1 Ω 2.1 Ω 18 Ω 36 Ω 0.36 kΩ 0.89 kΩ 10 kΩ 12 µΩ 50 mΩ 1.2 mΩ + 24 µΩ/Ω 1.5 mΩ + 24 µΩ/Ω 1.9 mΩ + 22 µΩ/Ω 4.1 mΩ + 22 µΩ/Ω 9.1 mΩ + 22 µΩ/Ω 41 mΩ + 22 µΩ/Ω 92 mΩ + 22 µΩ/Ω 0.41 Ω + 22 µΩ/Ω 0.9 Ω + 22 µΩ/Ω 8.4 Ω + 26 µΩ/Ω 14 Ω + 26 µΩ/Ω 93 Ω + 48 µΩ/Ω 0.4 kΩ + 0.1 mΩ/Ω 4.4 kΩ + 0.2 mΩ/Ω 16 kΩ + 0.4 mΩ/Ω 0.35 MΩ + 2.4 mΩ/Ω 4.4 MΩ + 12 mΩ/Ω Fluke 5720A Fluke 742A-1 Fluke 742A-10K Fluke 5520A
5

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 5 of 31 Parameter/Equipment Range CMC2, 4 (±) Comments Resistance3 – Generate Fixed Points (cont) (0 to 12) Ω (12 to 120) Ω 120 Ω to 1.2 kΩ (1.2 to 12) kΩ (12 to 120) kΩ 120 kΩ to 1.2 MΩ (1.2 to 12) MΩ (12 to 120) MΩ (120 to 1200) MΩ 1 mΩ + 19 µΩ/Ω 1.2 mΩ + 19 µΩ/Ω 4.4 mΩ + 19 µΩ/Ω 44 mΩ + 19 µΩ/Ω 0.44 Ω + 19 µΩ/Ω 4.4 Ω + 19 µΩ/Ω 63 Ω + 27 µΩ/Ω 6.5 kΩ + 0.33 mΩ/Ω 0.47 MΩ + 3.1 mΩ/Ω Fluke 5560A Resistance3 – Measure (0 to 2) Ω (2 to 20) Ω (20 to 200) Ω 200 Ω to 2 kΩ (2 to 20) kΩ (20 to 200) kΩ 200 kΩ to 2 MΩ (2 to 20) MΩ (20 to 200) MΩ 0.36 mΩ + 1.2 µΩ/Ω 58 µΩ + 8.4 µΩ/Ω 0.18 mΩ + 7.6 µΩ/Ω 1.4 mΩ + 7.8 µΩ/Ω 14 mΩ + 7.8 µΩ/Ω 0.13 Ω + 8 µΩ/Ω 1.9 Ω + 10 µΩ/Ω 0.12 kΩ + 21 µΩ/Ω 12 kΩ + 0.12 mΩ/Ω Fluke 8508A Parameter/Range Frequency CMC2, 4 (±) Comments Capacitance3 – Generate, Fixed Points 1 pF 10 pF 100 pF 1 kHz 1 MHz 2 MHz 3 MHz 4 MHz 5 MHz 10 MHz 13 MHz 1 kHz to 5 MHz 10 MHz 13 MHz 1 kHz to 5 MHz 10 MHz 13 MHz 0.35 fF 0.36 fF 0.42 fF 0.54 fF 0.72 fF 0.95 fF 2.5 fF 3.7 fF 3.5 fF 3.7 fF 3.8 fF 35 fF 48 fF 60 fF Agilent 16380 series capacitors
6

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 6 of 31 Parameter/Range Frequency CMC2, 4 (±) Comments Capacitance3 – Generate, Fixed Points (cont) 1000 pF 10 000 pF 1 µF 1 kHz 1 MHz 2 MHz 3 MHz 4 MHz 5 MHz 10 MHz 13 MHz 120 Hz to 100 kHz 120 Hz to 100 kHz 0.35 pF 0.35 pF 0.38 pF 0.45 pF 0.56 pF 0.71 pF 1.9 pF 2.8 pF 3.5 pF 0.4 nF Agilent 16380 series capacitors Parameter/Equipment Range CMC2, 4, 5 (±) Comments Capacitance3 – Measure 12 Hz to 100 kHz (1 to 10) pF (10 to 100) pF (100 to 1000) pF (1 to 10) nF (10 to 100) nF (100 to 1000) nF (1 to 10) µF (10 to 100) µF 0.048 pF + 0.4 pF 0.22 pF 1.1 pF 40 pF 0.11 nF 1 nF 9.2 nF 0.14 µF GenRad 1689 M CMC is only valid at 1 kHz Capacitance3 – Measure 1 nF to 10 mF 100 Hz to 100 kHz See matrix below QuadTech 7400 ≤ 100 mV Frequency Range 100 Hz 120 Hz 1 kHz 10 kHz 100 kHz 10 nF 3.9 % 2.8 % 0.28 % 0.13 % 0.13 % 100 nF 0.96 % 0.75 % 0.16 % 0.12 % 0.13 % 1 µF 0.51 % 0.43 % 0.15 % 0.12 % 0.17 % 10 µF 0.46 % 0.4 % 0.15 % 0.15 % 0.42 % 100 µF 0.46 % 0.4 % 0.18 % 0.35 % 3 % 1 mF 0.6 % 0.54 % 0.46 % 2.5 % ---- 10 mF 1.6 % 1.6 % 3.4 % ---- ----
7

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 7 of 31 Parameter/Equipment Range CMC2, 4, 5 (±) Comments Capacitance3 – Measure (cont) 1 nF to 10 mF 100 Hz to 100 kHz See matrix below QuadTech 7400 > 100 mV Frequency Range 100 Hz 120 Hz 1 kHz 10 kHz 100 kHz 10 nF 0.95 % 0.71 % 0.087 % 0.058 % 0.059 % 100 nF 0.24 % 0.2 % 0.066 % 0.058 % 0.059 % 1 µF 0.16 % 0.14 % 0.062 % 0.058 % 0.068 % 10 µF 0.15 % 0.14 % 0.062 % 0.061 % 0.15 % 100 µF 0.15 % 0.14 % 0.072 % 0.13 % 0.96 % 1 mF 0.19 % 0.17 % 0.16 % 0.8 % ---- 10 mF 0.52 % 0.51 % 1.1 ---- ---- Capacitance3 – Generate (0.19 to 0.3999) nF (0.4 to 1.0999) nF (1.1 to 3.2999) nF (3.3 to 10.9999) nF (11 to 32.9999) nF (33 to 109.999) nF (110 to 329.999) nF (0.33 to 1.099 99) µF (1.1 to 3.2999) µF (3.3 to 10.9999) µF (11 to 32.9999) µF (33 to 109.999) µF (110 to 329.999) µF (0.33 to 1.099 99) mF (1.1 to 3.2999) mF (3.3 to 10.9999) mF (11 to 32.9999) mF (33 to 110) mF 200 pF to 1.2 nF (1.2 to 12) nF (> 3 to 12) nF (12 to 120) nF (> 30 to 120) nF (0.12 to 1.2) µF (1.2 to 12) µF (12 to 120) µF (0.12 to 1.2) mF (1.2 to 12) mF (12 to 120) mF 10 Hz to 10 kHz 10 Hz to 10 kHz 10 Hz to 10 kHz (10 to 1000) Hz (10 to 1000) Hz (10 to 1000) Hz (10 to 1000) Hz (10 to 600) Hz (10 to 300) Hz (10 to 150) Hz (10 to 120) Hz (10 to 80) Hz Up to 50 Hz Up to 20 Hz Up to 6 Hz Up to 2 Hz Up to 0.6 Hz Up to 0.2 Hz 100 Hz to 10 kHz 150 Hz to 5 kHz (10 to 150) Hz 200 Hz to 1.3 kHz (10 to 200) Hz (2 to 310) Hz (0.5 to 110) Hz (0.5 to 40) Hz (0.1 to 11) Hz (0.03 to 4) Hz (0.01 to 1.3) Hz 8.7 pF + 4 pF/nF 9.7 pF + 4 pF/nF 13 pF + 4 pF/nF 15 pF + 2 pF/nF 0.1 nF + 2 pF/nF 0.15 nF + 2 pF/nF 0.46 nF + 2 pF/nF 1.5 nF + 2 nF/µF 4.6 nF + 2 nF/µF 15 nF + 2 nF/µF 59 nF + 3.2 nF/µF 0.2 µF + 3.6 nF/µF 0.64 µF + 3.6 nF/µF 2 µF + 3.6 µF/mF 7 µF + 3.5 µF/mF 38 µF + 2.5 µF/mF 90 µF + 6 µF/mF 0.37 mF + 8.8 µF/mF 2.4 pF + 0.86 mF/F 6.2 pF + 0.9 mF/F 7.8 pF + 0.9 mF/F 43 pF + 1 mF/F 61 pF + 1 mF/F 0.42 nF + 1 mF/F 4.2 nF + 1 mF/F 40 nF + 1.2 mF/F 0.48 µF + 1.9 mF/F 5.3 µF + 1.9 mF/F 77 µF + 3.9 mF/F Fluke 5522A Fluke 5560A
8--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 8 of 31 Parameter/Equipment Range CMC2, 4 (±) Comments Inductance3 – Measure, 12 Hz to 100 kHz 0.01 µH to 1 mH (1 to 100) mH 100 mH to 1 H (1 to 10) H 1.4 µH 0.1 mH 1.4 mH 11 mH GenRad 1689 M CMC is only valid at 1 kHz Inductance3 – Generate, Fixed Points, 100 Hz to 1 kHz 100 µH 1 mH 10 mH 100 mH 1 H 6.1 nH 1 µH 9.8 µH 54 µH 0.57 mH GenRad 1482 Series Inductors Parameter/Range Frequency CMC2, 5 (±) Comments Inductance – Generate3 (13 to 120) µH (0.120 01 to 1.2) mH (1.2001 to 3.3) mH (3.3 to 12) mH (12 to 83) mH (490 to 550) Hz (550 to 999) Hz 1 kHz (1.001 to 13) kHz (13 to 17) kHz (260 to 330) Hz (330 to 999) Hz 1 kHz (1.001 to 1.6) kHz (1.6 to 2.5) kHz (0.5 to 999) Hz 110 Hz (111 to 800) Hz (800 to 980) Hz (0.5 to 109) Hz 110 Hz (111 to 1000) Hz (1 to 1.4) kHz (0.1 to 99) Hz 100 Hz (101 to 180) Hz (180 to 230) Hz 1.5 % 0.67 % 0.38 % 0.67 % 1.5 % 1.4 % 0.49 % 0.2 % 0.49 % 1.4 % 0.74 % 0.46 % 0.74 % 1.6 % 0.49 % 0.2 % 0.49 % 1.4 % 0.53 % 0.25 % 0.53 % 1.4 % Fluke 5560A
12 to 100000 Hz
9

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 9 of 31 Parameter/Range Frequency CMC2, 5 (±) Comments Inductance – Generate3 (cont) (83 to 120) mH (0.120 01 to 0.65) H (0.65 to 1.2) H (1.2001 to 5.5) H (5.5 to 12) H (12.001 to 30) H (30 to 120) H (0.1 to 99) Hz 100 Hz (101 to 320) Hz (320 to 1000) Hz (0.05 to 9.9) Hz 10 Hz (10.1 to 30) Hz (30 to 50) Hz (0.05 to 9.9) Hz 10 Hz (10.1 to 100) Hz (100 to 170) Hz (0.01 to 2.99) Hz 3 Hz (3.01 to 8) Hz (8 to 16) Hz (0.01 to 2.99) Hz 3 Hz (3.01 to 19) Hz (19 to 37) Hz (0.005 to 1.99) Hz 2 Hz (2.01 to 4) Hz (4 to 16) Hz (0.005 to 1.99) Hz 2 Hz (2.01 to 7) Hz (7 to 14) Hz 0.49 % 0.2 % 0.49 % 1.4 % 0.61 % 0.32 % 0.6 % 1.5 % 0.52 % 0.24 % 0.52 % 1.4 % 0.68 % 0.4 % 0.68 % 1.6 % 0.57 % 0.28 % 0.57 % 1.4 % 0.89 % 0.61 % 0.89 % 1.8 % 0.61 % 0.32 % 0.6 % 1.5 % Fluke 5560A
10

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 10 of 31 Parameter/Range Frequency CMC2, 4 (±) Comments AC Voltage3 – Generate 220 µV to 2.2 mV (2.2 to 22) mV (22 to 220) mV (0.22 to 2.2) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz 4 µV + 0.22 µV/mV 4 µV + 85 nV/mV 4 µV + 75 nV/mV 4 µV + 0.18 µV/mV 5 µV + 0.46 µV/mV 10 µV + 0.9 µV/mV 20 µV + 1.2 µV/mV 20 µV + 2.5 µV/mV 4.5 µV + 0.22 µV/mV 4.2 µV + 85 nV/mV 4.2 µV + 75 nV/mV 4.4µV + 0.18 µV/mV 6.1 µV + 0.46 µV/mV 12 µV + 0.9 µV/mV 23 µV + 1.2 µV/mV 26 µV + 2.5 µV/mV 17 µV + 0.22 µV/mV 9 µV + 85 nV/mV 8.7 µV + 75 nV/mV 11 µV + 0.18 µV/mV 16 µV + 0.46 µV/mV 31 µV + 0.9 µV/mV 48 µV + 1.2 µV/mV 78 µV + 2.5 µV/mV 90 µV + 0.22 mV/V 34 µV + 79 µV/V 18 µV + 39 µV/V 26 µV + 70 µV/V 54 µV + 0.11 mV/V 0.16 mV + 0.34 mV/V 0.56 mV + 0.84 mV/V 0.64 mV + 1.5 mV/V Fluke 5720A
11

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 11 of 31 Parameter/Range Frequency CMC2, 4 (±) Comments AC Voltage3 – Generate (cont) (0.22 to 2.2) V (2.2 to 22) V (22 to 220) V* (220 to 1100) V (220 to 750) V (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz 500 kHz to 1 MHz (15 to 50) Hz ** 50 Hz to 1 kHz 40 Hz to 1 kHz (1 to 20) kHz (20 to 30) kHz (30 to 50) kHz (50 to 100) kHz 90 µV + 0.22 mV/V 34 µV + 79 µV/V 18 µV + 39 µV/V 26 µV + 70 µV/V 54 µV + 0.11 mV/V 0.16 mV + 0.34 mV/V 0.56 mV + 0.84 mV/V 0.64 mV + 1.5 mV/V 0.89 mV + 0.22 mV/V 0.33 mV + 80 µV/V 0.15 mV + 40 µV/V 0.26 mV + 70 µV/V 0.41 mV + 95 µV/V 1.2 mV + 0.26 mV/V 4 mV + 0.9 mV/V 6.1 mV + 1.3 mV/V 8.9 mV + 0.22 mV/V 3.4 mV + 80 µV/V 1.7 mV + 47 µV/V 3.5 mV + 72 µV/V 5.9 mV + 0.13 mV/V 34 mV + 0.8 mV/V 0.13 V + 4.2 mV/V 0.24 V + 7 mV/V 74 mV + 0.26 mV/V 17 mV + 60 µV/V 22 mV + 80 µV/V 34 mV + 0.13 mV/V 93 mV + 0.36 mV/V 93 mV + 0.36 mV/V 0.34 V + 1.3 mV/V Fluke 5720A * 220 V range subject to 2.2E7 V-Hz limitation ** Max output 250 V from 15 Hz to 50 Hz w/ 5725A
12

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 12 of 31 Parameter/Range Frequency CMC2, 5 (±) Comments AC Voltage – Generate3 (1 to 12) mV (12 to 120) mV (0.12 to 1.2) V (1.2 to 12) V (12 to 120) V (0.01 to 3) Hz (3 to 5) Hz (5 to 10) Hz 10 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.01 to 3) Hz (3 to 5) Hz (5 to 10) Hz 10 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.01 to 3) Hz (3 to 5) Hz (5 to 10) Hz (10 to 40) Hz 40.01 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.01 to 3) Hz (3 to 5) Hz (5 to 10) Hz (10 to 40) Hz 40.01 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.01 to 3) Hz (3 to 5) Hz (5 to 10) Hz (10 to 40) Hz 40.01 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz * 0.11 mV + 50 mV/V 8.9 µV + 1.9 mV/V 7.7 µV + 0.68 mV/V 6.1 µV + 0.12 mV/V 6.3 µV + 0.29 mV/V 16 µV + 1.2 mV/V 36 µV + 6.2 mV/V 36 µV + 6.2 mV/V 1.2 mV + 50 mV/V 30 µV + 1.9 mV/V 15 µV + 0.68 mV/V 7.4 µV + 0.11 mV/V 11 µV + 0.27 mV/V 28 µV + 0.62 mV/V 49 µV + 1.6 mV/V 49 µV + 1.6 mV/V 12 mV + 50 mV/V 0.3 mV + 1.9 mV/V 0.15 mV + 0.68 mV/V 74 µV + 0.11 mV/V 21 µV + 0.11 mV/V 42 µV + 0.23 mV/V 0.11 mV + 0.54 mV/V 0.26 mV + 1.5 mV/V 0.26 mV + 1.5 mV/V 0.12 V + 50 mV/V 3 mV + 1.9 mV/V 1.6 mV + 0.68 mV/V 0.49 mV + 0.11 mV/V 0.18 mV + 0.11 mV/V 0.33 mV + 0.23 mV/V 0.77 mV + 0.54 mV/V 2.5 mV + 1.6 mV/V 2.5 mV + 1.6 mV/V 1.2 V + 50 mV/V 30 mV + 1.9 mV/V 16 mV + 0.68 mV/V 4.8 mV + 0.11 mV/V 1.8 mV + 0.11 mV/V 3.3 mV + 0.23 mV/V 7.7 mV + 0.54 mV/V 39 mV + 1.6 mV/V Fluke 5560A * Max output 70 V
13

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 13 of 31 Parameter/Range Frequency CMC2, 5 (±) Comments AC Voltage – Generate3 (cont) (120 to 330) V (330 to 1020) V (3 to 5) Hz (5 to 10) Hz 10 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (3 to 5) Hz (5 to 10) Hz 10 Hz to 10 kHz 0.3 V + 1.9 mV/V 0.16 V + 0.68 mV/V 21 mV + 0.11 mV/V 36 mV + 0.23 mV/V 0.16 V + 1.2 mV/V 0.7 V + 1.9 mV/V 0.3 V + 0.68 mV/V 0.12 V + 0.11 mV/V Fluke 5560A AC Voltage3 – Measure (20 to 200) mV (0.2 to 2) V (2 to 20) V (20 to 200) V (200 to 1000) V (1 to 10) Hz (10 to 40) Hz (40 to 100) Hz 100 Hz to 2 kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz (10 to 40) Hz (40 to 100) Hz 100 Hz to 2 kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz (10 to 40) Hz (40 to 100) Hz 100 Hz to 2 kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz (10 to 40) Hz (40 to 100) Hz 100 Hz to 2 kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz (10 to 40) Hz 40 Hz to 1 kHz 14 µV + 0.16 µV/mV 4.7 µV + 0.12 µV/mV 4.9 µV + 0.1 µV/mV 2.8 µV + 0.1 µV/mV 5.9 µV + 0.13 µV/mV 10 µV + 0.31 µV/mV 25 µV + 0.66 µV/mV 0.15 mV + 0.15 mV/V 38 µV + 0.11 mV/V 36 µV + 88 µV/V 34 µV + 73 µV/V 46 µV + 0.11 mV/V 0.28 mV + 0.14 mV/V 0.3 mV + 0.51 mV/V 0.29 mV + 0.16 mV/V 0.15 mV + 0.12 mV/V 0.31 mV + 80 µV/V 0.11 mV + 74 µV/V 0.2 mV + 0.11 mV/V 0.26 mV + 0.21 mV/V 0.86 mV + 0.51 mV/V 12 mV + 0.16 mV/V 2.7 mV + 0.12 mV/V 2.7 mV + 90 µV/V 2.7 mV + 75 µV/V 3.5 mV + 0.11 mV/V 5.9 mV + 0.21 mV/V 25 mV + 0.51 mV/V 72 mV + 0.15 mV/V 29 mV + 0.11 mV/V 33 mV + 0.11 mV/V Fluke 8508A
14

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 14 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments AC Voltage3 – Measure (1 to 30) kV (30 to 100) kV 0.59 % 0.59 % Ross Divider, DMM AC Current3 – Measure (20 to 200) µA (0.2 to 2) mA (2 to 20) mA (20 to 200) mA (0.2 to 2) A (2 to 20) A (1 to 10) Hz 10 Hz to 10 kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz 10 Hz to 10 kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz 10 Hz to 10 kHz (10 to 30) kHz (30 to 100) kHz (1 to 10) Hz 10 Hz to 10 kHz (10 to 30) kHz (1 to 2) Hz 2 Hz to 10 kHz (10 to 30) kHz 10 Hz to 2 kHz (2 to 10) kHz 0.23 µA + 0.14 nA/µA 20 nA + 0.48 nA/µA 21 nA + 0.62 nA/µA 25 nA + 3.1 nA/µA 0.42 µA + 0.25 µA/mA 0.57 µA + 0.19 µA/mA 0.27 µA + 0.62 µA/mA 0.56 µA + 3.1 µA/mA 48 µA + 33 nA/mA 2.1 µA + 0.29 µA/mA 2.1 µA + 0.62 µA/mA 2.8 µA + 3.1 µA/mA 25 µA + 0.29 µA/mA 21 µA + 0.28 µA/mA 21 µA + 0.58 µA/mA 0.41 mA + 0.05 mA/A 0.28 mA + 0.67 mA/A 0.45 mA + 2.5 mA/A 2.3 mA + 0.72 mA/A 2.5 mA + 1.9 mA/A Fluke 8508A AC Current3 – Generate (0 to 220) µA (29 to 329.99) µA 220 µA to 2.2 mA (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 63 nA + 0.13 nA/µA 12 nA + 0.14 nA/µA 9.2 nA + 0.11 nA/µA 15 nA + 0.25 nA/µA 75 nA + 0.9 nA/µA 0.69 µA + 13 nA/µA 93 nA + 0.23 µA/mA 67 nA + 0.14 µA/mA 60 nA + 0.11 µA/mA 0.15 µA + 0.18 µA/mA 0.85 µA + 0.9 µA/mA Fluke 5720A 5520A Fluke 5720A
15

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 15 of 31 Parameter/Range Frequency CMC2, 4 (±) Comments AC Current3 – Generate (cont) (0.33 to 3.2999) mA (2.2 to 22) mA (3.3 to 32.999) mA (22 to 220) mA (33 to 329.99) mA 220 mA to 2.2 A (2.2 to 11) A (11 to 20.5) A (11 to 20) A (11 to 20) A (20 to 100) A (10 to 30) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz 20 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (45 to 100) Hz 100 Hz to 1 kHz (1 to 5) kHz Up to 1 kHz 3.3 µA + 8 µA/mA 0.92 µA + 0.23 µA/mA 0.69 µA + 0.14 µA/mA 0.6 µA + 0.11 µA/mA 0.95 µA + 0.18 µA/mA 7 µA + 0.9 µA/mA 14 µA + 3.2 µA/mA 9.3 µA + 0.23 µA/mA 6.8 µA + 0.14 µA/mA 5 µA + 0.11 µA/mA 7.5 µA + 0.18 µA/mA 30 µA + 0.9 µA/mA .27 mA + 3.2 µA/mA 89 µA + 0.24 mA/A 0.17 mA + 0.39 mA/A 1.5 mA + 6 mA/A 1.1 mA + 0.4 mA/A 2.3 mA + 0.85 mA/A 8.0 mA + 3.3 mA/A 15 mA + 0.96 mA/A 17 mA + 1.2 mA/A 0.27 A + 24 mA/A 42 mA + 1.7 mA/A 0.15 A + 1.7 mA/A 5520A 5720 5520A 5720A 5520A Fluke 5720A w/5725A Fluke 5520A Valhalla 2555A
16--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 16 of 31 Parameter/Range Frequency CMC2, 4 (±) Comments AC Current – Generate3 (12 to 120) µA (0.12 to 1.2) mA (1.2 to 12) mA (12 to 120) mA (0.12 to 1.2) A (1.2 to 3.1) A (3.1 to 12) A (12 to 30) A (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (3 to 45) Hz 45 Hz to 1 kHz (1 to 5) kHz 12 nA + 0.19 mA/A 12 nA + 0.19 mA/A 12 nA + 0.19 mA/A 52 nA + 1.2 mA/A 1 µA + 3.9 mA/A 0.13 µA + 0.19 mA/A 0.13 µA + 0.19 mA/A 0.13 µA + 0.19 mA/A 0.25 µA + 1.2 mA/A 5.5 µA + 3.9 mA/A 1.3 µA + 0.19 mA/A 1.3 µA + 0.19 mA/A 1.3 µA + 0.19 mA/A 2.5 µA + 1.2 mA/A 15 µA + 3.9 mA/A 13 µA + 0.19 mA/A 6.5 µA + 0.12 mA/A 10 µA + 0.19 mA/A 25 µA + 1.2 mA/A 0.15 mA + 3.9 mA/A 0.13 mA + 0.18 mA/A 75 µA + 0.19 mA/A 0.1 mA + 0.19 mA/A 0.53 mA + 1.9 mA/A 0.77 mA + 3.9 mA/A 0.85 mA + 0.29 mA/A 0.58 mA + 0.23 mA/A 0.65 mA + 0.29 mA/A 2.8 mA + 1.9 mA/A 1.9 mA + 0.29 mA/A 1.2 mA + 0.23 mA/A 1.7 mA + 0.29 mA/A 6.9 mA + 1.9 mA/A 19 mA + 0.78 mA/A 15 mA + 0.54 mA/A 55 mA + 3.9 mA/A Fluke 5560A
12 to 30 A
17--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 17 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments Oscilloscopes3 – DC 50 Ω 1 MΩ Square 50 Ω 1 MΩ Fast Edge Leveled Sine Amplitude Leveled Sine Flatness Relative to 50 kHz Relative to 10 MHz Time Marker (0 to 6.6) V (0 to 6.6) V 1 mV to 6.6 Vp-p 1 mV to 130 Vp-p ≤ 150 ps 50 kHz Reference 10 MHz Reference 50 kHz to 100 MHz (100 to 300) MHz (300 to 500) MHz (500 to 600) MHz 600 MHz to 1.1 GHz (1.1 to 1.6) GHz (1.6 to 2.1) GHz 5 s to 50 ms 20 ms to 500 ps 0.84 mV + 2.4 mV/V 0.95 mV + 0.25 mV/V 0.72 mV + 2.4 mV/V 2.2 mV + 0.49 mV/V 29 ps 1.8 mV + 20 mV/V 1.9 mV + 20 mV/V 1 mV + 38 mV/V 0.94 mV + 45 mV/V 0.92 mV + 65 mV/V 0.91 mV + 72 mV/V 0.97 mV + 86 mV/V 1.2 mV + 86 mV/V 1 mV + 0.1 V/V 0.0012 % + 5 ns 0.0003 % Fluke 5820A Distortion3 – Measure (-60 to 0) dB (-100 to 0) dB 20 Hz to 20 kHz (20 to 100) kHz 100 kHz to 2.5 GHz (2.5 to 22) GHz (22 to 26.5) GHz 1.2 dB 2.3 dB 2.1 dB 3.3 dB 4.2 dB 8903B 8563EC
to 6.6 V
18

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 18 of 31 Parameter/Equipment Range CMC2, 4, 5 (±) Comments Duty Cycle3, 8 – Generate (1 to 9.99) % (10 to 49.99) % 50 % (50.01 to 90) % (90.01 to 99) % 0.33 % 0.02 % 0.013 % 0.02 % 0.32 % Fluke 5520A Electrical Calibration of Thermocouple Indicators & Indicating Systems3 – Type E Type J Type K Type R Type S Type T (-250 to -100) °C (-100 to -25) °C (-25 to 350) °C (350 to 650) °C (650 to 1000) °C (-210 to -100) °C (-100 to -30) °C (-30 to 150) °C (150 to 760) °C (760 to 1200) °C (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 1000) °C (1000 to 1372) °C (0 to 250) °C (250 to 400) °C (400 to 1000) °C (1000 to 1750) °C (0 to 250) °C (250 to 1000) °C (1000 to 1400) °C (1400 to 1750) °C (-250 to -150) °C (-150 to 0) °C (0 to 120) °C (120 to 400) °C 0.59 °C 0.21 °C 0.17 °C 0.19 °C 0.29 °C 0.33 °C 0.25 °C 0.19 °C 0.2 °C 0.28 °C 0.39 °C 0.28 °C 0.19 °C 0.32 °C 0.5 °C 0.67 °C 0.41 °C 0.39 °C 0.48 °C + 0.061 m °C/°C 0.65 °C 0.42 °C 0.43 °C 0.55 °C + 0.054 m °C/°C 0.74 °C 0.3 °C 0.21 °C 0.16 °C Martel 3001
19

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 19 of 31 Parameter/Equipment Range CMC2 (±) Comments Electrical Calibration of RTD Indicators & Indicating Systems3 – Pt 385, 100 Ω (-200 to -80) °C (-80 to 0) °C (0 to 100) °C (100 to 300) °C (300 to 400) °C (400 to 630) °C (630 to 800) °C 0.047 °C 0.058 °C 0.058 °C 0.047 °C 0.048 °C 0.061 °C + 0.016 m °C/°C 0.064 °C + 0.021 m °C/°C Martel 3001 Parameter/Range Frequency CMC2 (±) Comments Phase3 (PF=1) (10 to 65) Hz (65 to 500) Hz 500 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz (3 to 65) Hz (65 to 500) Hz 500 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz Ø 0.1 ° Ø 0.26 ° Ø 0.51 ° Ø 2.6 ° Ø 5.1 ° Ø 10 ° Ø 0.1 ° Ø 0.2 ° Ø 0.4 ° Ø 1.9 ° Ø 3.9 ° Ø 7.8 ° Fluke 5520A Fluke 5560A
20

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 20 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments AC Power3 – Generate See Table below See Table below Fluke 5520A (45 to 65) Hz (3.366 to 33.66) W (33.66 to 336.6) W (0.3366 to 1.122) kW (1.122 to 3.06) kW (3.06 to 11.22) kW (11.22 to 20.9) kW Ø 0 ° PF=1.000 0.052 % 0.068 % 0.076 % 0.083 % 0.17 % Ø 10 ° PF=0.985 0.06 % 0.075 % 0.082 % 0.088 % 0.18 % Ø 20 ° PF=0.940 0.082 % 0.093 % 0.099 % 0.1 % 0.18 % Ø 30 ° PF=0.866 0.11 % 0.12 % 0.13 % 0.2 % Ø 40 ° PF=0.766 0.16 % 0.17 % 0.23 % Ø 50 ° PF=0.643 0.21 % 0.22 % 0.27 % Ø 60 ° PF=0.500 0.31 % 0.35 % Ø 70 ° PF=0.342 0.48 % 0.49 % 0.51 % Ø 80 ° PF=0.174 0.99 % 1 % (65 to 500) Hz (33.66 to 336.6) W (0.3366 to 1.122) kW (1.122 to 3.06) kW (3.06 to 11.22) kW (11.22 to 20.9) kW Ø 0 ° PF=1.000 0.052 % 0.068 % 0.076 % 0.12 % 0.21 % Ø 10 ° PF=0.985 0.094 % 0.1 % 0.11 % 0.14 % 0.22 % Ø 20 ° PF=0.940 0.17 % 0.18 % 0.2 % 0.26 % Ø 30 ° PF=0.866 0.26 % 0.28 % 0.33 % Ø 40 ° PF=0.766 0.37 % 0.39 % 0.42 % Ø 50 ° PF=0.643 0.53 % 0.56 % Ø 60 ° PF=0.500 0.76 % 0.77 % 0.78 % Ø 70 ° PF=0.342 1.2 % Ø 80 ° PF=0.174 2.5 %
21--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 21 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments AC Power – Generate3 (cont) See Table below See Table below Fluke 5560A 500 Hz to 1 kHz (33.66 to 336.6) W (0.3366 to 1.122) kW (1.122 to 3.06) kW (3.06 to 11.22) kW (11.22 to 20.9) kW Ø 0 ° PF=1.000 0.052 % 0.068 % 0.076 % 0.12 % 0.21 % (45 to 65) Hz (1.22 to 12.2) W (12.2 to 122) W (0.122 to 1.22) kW (1.22 to 3.16) kW (3.16 to 12.2) kW (12.2 to 30.8) kW Ø 0 ° PF=1.000 0.038 % 0.029 % 0.034 % 0.048 % 0.04 % 0.11 % Ø 10 ° PF=0.985 0.049 % 0.042 % 0.046 % 0.057 % 0.051 % 0.11 % Ø 20 ° PF=0.940 0.074 % 0.07 % 0.072 % 0.079 % 0.075 % 0.12 % Ø 30 ° PF=0.866 0.11 % 0.15 % Ø 40 ° PF=0.766 0.15 % Ø 50 ° PF=0.643 0.21 % Ø 60 ° PF=0.500 0.3 % Ø 70 ° PF=0.342 0.48 % Ø 80 ° PF=0.174 0.99 % (65 to 500) Hz (12.2 to 122) W (0.122 to 1.22) kW (1.22 to 3.16) kW (3.16 to 12.2) kW (12.2 to 30.8) kW Ø 0 ° PF=1.000 0.029 % 0.034 % 0.048 % 0.04 % 0.11 % Ø 10 ° PF=0.985 0.068 % 0.071 % 0.078 % 0.074 % 0.12 % Ø 20 ° PF=0.940 0.14 % 0.17 % Ø 30 ° PF=0.866 0.2 % 0.21 % 0.23 % Ø 40 ° PF=0.766 0.29 % 0.3 % 0.31 % Ø 50 ° PF=0.643 0.42 % (65 to 500) Hz (12.2 to 122) W (0.122 to 1.22) kW (1.22 to 3.16) kW (3.16 to 12.2) kW (12.2 to 30.8) kW Ø 60 ° PF=0.500 0.61 %
500 to 1000 Hz
22

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 22 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments AC Power – Generate3 (cont) See Table below See Table below Fluke 5560A (65 to 500) Hz (12.2 to 122) W (0.122 to 1.22) kW (1.22 to 3.16) kW (3.16 to 12.2) kW (12.2 to 30.8) kW Ø 70 ° PF=0.342 0.96 % Ø 80 ° PF=0.174 2 % 500 Hz to 1 kHz (12.2 to 122) W (0.122 to 1.22) kW (1.22 to 3.16) kW (3.16 to 12.2) kW (12.2 to 30.8) kW Ø 0 ° PF=1.000 0.029 % 0.034 % 0.048 % 0.04 % 0.11 % Ø 10 ° PF=0.985 0.13 % 0.16 % Ø 20 ° PF=0.940 0.26% 0.28 % Ø 30 ° PF=0.866 0.41% 0.42 % Ø 40 ° PF=0.766 0.59 % 0.6 % Ø 50 ° PF=0.643 0.83 % Ø 60 ° PF=0.500 1.2 % Ø 70 ° PF=0.342 1.9 % Ø 80 ° PF=0.174 4 % (1 to 5) kHz (6 to 60) W (60 to 600) W (0.6 to 1.55) kW Ø 0 ° PF=1.000 0.071 % 0.082 % Ø 10 ° PF=0.985 0.64 % Ø 20 ° PF=0.940 1.3 % Ø 30 ° PF=0.866 2 % Ø 40 ° PF=0.766 2.8 % Ø 50 ° PF=0.643 4 % (5 to 10) kHz (3 to 30) W (30 to 300) W Ø 0 ° PF=1.000 0.28 % 0.34 %
23

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 23 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments AC Power – Generate3 (cont) See Table below See Table below Fluke 5560A (5 to 10) kHz (3 to 30) W (30 to 300) W Ø 10 ° PF=0.985 1.4 % Ø 20 ° PF=0.940 2.7 % Ø 30 ° PF=0.866 4.2 % (10 to 30) kHz (3 to 300) W Ø 0 ° PF=1.000 1.1 % Ø 10 ° PF=0.985 3.4 % Ø 20 ° PF=0.940 5.9% Parameter/Frequency Range CMC2, 4, 5 (±) Comments DC Power3 – Generate Up to 330 µA Up to 3.3 mA Up to 33 mA Up to 330 mA Up to 1.1 A Up to 3 A Up to 11 A Up to 20.5 A Up to 120 µA Up to 1.2 mA Up to 12 mA Up to 120 mA Up to 1.2 A Up to 3.1 A Up to 12 A Up to 30.2 A Up to 1020 V Up to 337 mW (0.337 to 3.366) W (3.366 to 33.66) W (33.66 to 336.6) W (0.3366 to 1.122) kW (1.122 to 3.06) kW (3.06 to 11.22) kW (11.22 to 20.91) kW Up to 122 mW (0.12 to 1.224) W (1.225 to 12.24) W (12.24 to 122.4) W (0.1224 to 1.224) kW (1.224 to 3.162) kW (3.162 to 12.24) kW (12.24 to 30.8) kW 0.017 % 0.021 % 0.0093 % 0.0093 % 0.019 % 0.043 % 0.044 % 0.13 % 0.015 % 0.0092 % 0.0087 % 0.0087 % 0.013 % 0.028 % 0.025 % 0.08 % Fluke 5520A Fluke 5560A
24

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 24 of 31 III. RF/Microwave Parameter/Range Frequency CMC2, 4, 5 (±) Comments Power Meter3 Power Reference @ 1 mW Power Accuracy 50 MHz 3 µW to 100 mW 0.60 % 0.29 % Agilent 432A, 8478B, Agilent 3458A Range calibrator Relative Power3 – Measure (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB (-70 to -80) dB (-80 to -90) dB (-90 to -100) dB (-100 to -110) dB (-110 to -120) dB (-120 to -127) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB (-70 to -80) dB (-80 to -90) dB (-90 to -100) dB (-100 to -110) dB (-110 to -120) dB (-120 to -127) dB 2.5 MHz to 1.3 GHz (1.3 to 26.5) GHz 0.047 dB 0.06 dB 0.074 dB 0.087 dB 0.1 dB 0.1 dB 0.11 dB 0.12 dB 0.13 dB 0.15 dB 0.15 dB 0.28 dB 0.29 dB 0.066 dB 0.06 dB 0.074 dB 0.087 dB 0.1 dB 0.1 dB 0.11 dB 0.12 dB 0.13 dB 0.15 dB 0.15 dB 0.28 dB 0.29 dB 8902A
25

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 25 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments Absolute Power3 – Measure (0 to +20) dBm (-10 to 0) dBm (-60 to -10) dBm (+10 to +20) dBm (-30 to +10) dBm 9 kHz to 6 GHz 100 kHz to 4.2 GHz (4.2 to 26.5) GHz 100 kHz to 4.2 GHz (4.2 to 26.5) GHz 2.9 % 3.3 % 3.9 % 1.1 % 3.2 % 3.6 % 4.7 % Agilent E4418B w/ Agilent E9304A Agilent 8482A Agilent 8485A Agilent 8482A Agilent 8485A Amplitude Modulation3 – Measure Rate: 50 Hz to 10 kHz Depths: (5 to 99) % Rate: 20 Hz to 10 kHz Depths: to 99 % Rate: 50 Hz to 50 kHz Depths: (5 to 99) % Rate: 20 Hz to 100 kHz Depths: to 99 % Rate: 50 Hz to 50 kHz Depths: (5 to 99) % Rate: 20 Hz to 100 kHz Depths: to 99 % 150 kHz to 10 MHz 150 kHz to 10 MHz (0.01 to 1.3) GHz (0.01 to 1.3) GHz (1.3 to 26.5) GHz (1.3 to 26.5) GHz 0.14 % depth + 0.024 % depth/% depth 0.11 % depth + 0.023 % depth/% depth 0.15 % depth + 0.01 % depth/% depth 0.11 % depth + 0.023 % depth/% depth 0.33 % depth + 0.008 % depth/% depth 0.31 % depth + 0.021 % depth/% depth 8902A Frequency Modulation3 – Measure Rate: 20 Hz to 10 kHz Dev.: ≤ 40 kHz Peak Rate: 50 Hz to 100 kHz Dev.: ≤ 400 kHz Peak Rate: 20 Hz to 200 kHz Dev.: ≤ 400 kHz Peak 250 kHz to 10 MHz (0.01 to 1.3) GHz (0.01 to 1.3) GHz 3.5 Hz + 16 Hz/kHz 3.5 Hz + 7.8 Hz/kHz 6.8 Hz + 39 Hz/kHz 8902A
26

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 26 of 31 Parameter/Range Frequency CMC2, 4, 5 (±) Comments Frequency Modulation3 – Measure (cont) Rate: 50 Hz to 100 kHz Dev.: ≤ 400 kHz Peak Rate: 20 Hz to 200 kHz Dev.: ≤ 400 kHz Peak (1.3 to 26.5) GHz (1.3 to 26.5) GHz 6.8 Hz + 7.8 Hz/kHz 6.8 Hz + 39 Hz/kHz 8902A Phase Modulation3 – Measure Rate: 200 Hz to 10 kHz (0 to 100) rad Rate: 200 Hz to 20 kHz (0 to 100) rad Rate: 200 Hz to 20 kHz (0 to 100) rad 150 kHz to 10MHz (0.01 to 1.3) GHz (1.3 to 26.5) GHz 36 mrad + 37 mrad/rad 36 mrad + 37 mrad/rad 36 mrad + 37 mrad/rad 8902A Power Sensor Calibration Factor 9 kHz to 10 MHz 10 MHz to 2 GHz (2 to 3.6) GHz (3.6 to 4.6) GHz (4.6 to 9) GHz (9 to 11.8) GHz (11.8 to 16.25) GHz (16.25 to 18) GHz 1.7 % 0.57 % 0.59 % 0.61 % 0.65 % 0.74 % 0.78 % 0.94 % Power sensor calibration system w/ Tegam 1805B power meter, Agilent E9304A, F1109 power reference Displayed Average Noise Level3, 8 DC to 2 GHz (> 2 to ≤ 3) GHz (> 3 to ≤ 6) GHz 0.39 dB 0.16 dB 0.3 dB 50 Ω termination
27--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 27 of 31 Parameter/Range Frequency CMC2 (±) Comments S-Parameters (Reflection & Transmission) 300 kHz to 6 GHz See tables below (Interpolate CMC’s between cardinal points listed) Network analyzers with calibration/ verification kits Type N: Reflection S11/S22 Linear Magnitude (ρ) E5071C with 85032B rho 300 kHz to 10 MHz 10 MHz to 3 GHz (3 to 6) GHz 0 0.0039 0.0052 0.011 0.1 0.0046 0.0061 0.012 0.2 0.0054 0.0073 0.014 0.3 0.0064 0.0086 0.017 0.4 0.0074 0.01 0.02 0.5 0.0086 0.012 0.024 0.6 0.01 0.015 0.028 0.7 0.012 0.017 0.033 0.8 0.013 0.02 0.038 0.9 0.015 0.023 0.044 1.0 0.017 0.026 0.05 Type N: Reflection S11/S22 Phase (0 to 180)° E5071C with 85032B rho 300 kHz to 10 MHz 10 MHz to 3 GHz (3 to 6) GHz 0 ---- ---- ---- 0.1 2.6 3.4 7 0.2 1.5 2.1 4.1 0.3 1.2 1.6 3.2 0.4 1.1 1.5 2.9 0.5 0.99 1.4 2.7 0.6 0.95 1.4 2.7 0.7 0.95 1.4 2.7 0.8 0.95 1.4 2.7 0.9 0.96 1.5 2.8 1.0 0.98 1.5 2.9 Type N: Transmission S12/S21 Linear Magnitude (dB) E5071C with 85032B dB 300 kHz to 10 MHz 10 MHz to 3 GHz (3 to 6) GHz -90 8 2.6 2.7 -80 3.4 1 1 -70 1.3 0.42 0.47 -60 0.45 0.23 0.28 -50 0.19 0.16 0.21 -40 0.12 0.12 0.18 -30 0.088 0.097 0.15 -20 0.067 0.076 0.13
300000 to 6000000000 Hz
28

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 28 of 31 Parameter/Range Frequency CMC2, 4 (±) Comments S-Parameters (Reflection & Transmission) 300 kHz to 6 GHz See tables below (Interpolate CMC’s between cardinal points listed) Network analyzers with calibration/ verification kits Type N: Transmission S12/S21 Linear Magnitude (dB) E5071C with 85032B dB 300 kHz to 10 MHz 10 MHz to 3 GHz (3 to 6) GHz -10 0.056 0.066 0.12 0 0.046 0.056 0.11 10 0.063 0.073 0.13 Type N: Transmission S12/S21 Phase (0 to 180)° E5071C with 85032B dB 300 kHz to 10 MHz 10 MHz to 3 GHz (3 to 6) GHz -90 ---- 21 21 -80 29 7.1 7.4 -70 8.9 2.9 3.2 -60 3 1.5 1.9 -50 1.3 1 1.4 -40 0.79 0.81 1.2 -30 0.59 0.65 1 -20 0.45 0.51 0.88 -10 0.38 0.44 0.82 0 0.32 0.38 0.76 10 0.45 0.52 0.9 Attenuation3 – Generate (1 to 11) dB (12 to 121) dB DC to 8 GHz (8 to 12.4) GHz (12.4 to 18) GHz DC to 8 GHz (8 to 12.4) GHz (12.4 to 18) GHz 0.22 dB 0.29 dB 0.51 dB 0.22 dB + 0.0011 x ΔA Change from 12 dB 0.28 dB + 0.001 x ΔA Change from 12 dB 0.51 dB + 0.0006 x ΔA Change from 12 dB Agilent 8494H Agilent 8494H, 8496H
29--
Show page text

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 29 of 31 IV. Mechanical Parameter/Equipment Range CMC2, 5, 7 (±) Comments Scales3 Up to 400 lbs 0.82 x R Class F weights Balances3 0.0001 g Resolution 0.001 g Resolution 0.01 g Resolution 0.1 g Resolution 1 g & 10 g Resolution (5 to 500) mg > 500 mg to 5 g > 5 g to 6 kg 5 mg to 10 g (> 10 to 50) g > 50 g to 6 kg 5 mg to 300 g > 300 g to 6 kg 5 mg to 2 kg (2 to 6) kg 5 mg to 6 kg 0.82 x R 0.91 x R 1.0 x R 0.82 x R 0.91 x R 1.4 x R 0.82 x R 1.4 x R 0.85 x R 1.0 x R 0.82 x R Class 1, 3 & 4 weights Pressure –Measuring Equipment3 (-1 to 0) psig (0 to 1) psig (0 to 15) psig (15 to 30) psig (30 to 500) psig (500 to 3000) psig (3000 to 5000) psig (0 to 1) inH2O 0.013 psi + 0.6 R 0.012 psi 0.052 psi 0.034 psi 0.44 psi 0.85 psi 3 psi 0.001 inH 2O Pressure modules Druck DPI 610 pressure calibrator Torque Tools – Wrenches, Gages & Drivers3 (0.5 to 5) lbf·in (5 to 50) lbf·in (50 to 400) lbf·in (400 to 1000) lbf·in (83 to 600) lbf·ft 0.58 % 0.58 % 0.29 % 0.3 % 0.29 % Torque testers Torque – Measuring Equipment3 5 ozf·in to 10 lbf·in (10 to 1000) lbf·in 0.05 % 0.036 % Torque arms & Class 4 & Class F weights
to 400 lbs
30

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 30 of 31 V. Thermodynamics Parameter/Equipment Range CMC2, 7 (±) Comments Relative Humidity – Measuring Equipment3 11 %, 33 %, 75 % 1.4 % RH Saturated salts/ Vaisala MI 70, HMP 77B Relative Humidity – Measure3 (20 to 80) % RH 1.4 % RH Vaisala MI 70, HMP 77B Temperature – Measuring Equipment3 Digital & Analog Thermometers (-25 to 140) °C (140 to 660) °C 0.06 °C 0.1 °C Accumac 1751 PRT, 8508A, Calibration Baths & Wells Temperature – Measure 3 Fridges, Freezers, Chambers, Thermal Baths (-196 to 660) °C 0.011 °C + ΔT(Change from -196 °C) x 0.000 003 Accumac AM 1751/ w 8508A Infrared Temperature – Measuring Equipment3 (35 to 500) °C 0.54 °C + ΔT(Change from 35 °C) x 0.0042 Fluke 4181 Ɛ = 0.9 to 1 λ = (8 to 14) µm VI. Time & Frequency Parameter/Equipment Range CMC2, 7 (±) Comments Frequency 10 MHz 12 µHz 58503A GPS Receiver Frequency – Measuring Equipment3 100 mHz to 20 MHz 20 MHz to 50 GHz 55 µHz + 26 pHz/Hz 0.58 mHz + 1 pHz/Hz 3325B function generator E8257D signal generator
31

Extracted from the scope document; tables may not line up. View scope for the original.

(A2LA Cert. No. 3471.01) 07/18/2025 Page 31 of 31 Parameter/Equipment Range CMC2, 7 (±) Comments Frequency – Measure3 0.1 Hz to 3 GHz (3 to 26.5) GHz (26.5 to 50) GHz 5.8 µHz + 59 pHz/Hz 0.11 Hz + 0.36 pHz/Hz 0.12 Hz + 0.53 pz/Hz 53131A frequency counter 5351B microwave counter E4448A spectrum analyzer Stopwatches & Timers ± (0 to 19.99) sec/day 0.037 sec/day Helmut Klein 4500 Timometer _____________________________________________ 1 This laboratory offers commercial calibration service and field calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. 3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g., resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC. 4 The stated measured values are determined using the indicated instrument (see Comments). This capability is suitable for the calibration of the devices intended to measure or generate the measured value in the ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percentage or fraction of the reading plus a fixed floor specification. 5 In the statement of CMC, percentages are to be read as percent of reading unless otherwise noted. In the statement of CMC, R represents the resolution of the unit under test, L is the numerical value of the length of the device measured in inches. 6 This scope meets A2LA’s P112 Flexible Scope Policy. 7 The type of instrument or material being calibrated is defined by the parameter. This indicates the laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated for the listed measurement parameter.
32ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994--

Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.

Location

80 Orville Drive, Suite 115, Bohemia, NY, 11716, US

40.77061, -73.12007

Open in Maps

Accreditation data refreshed from the A2LA register on .

This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.