Process Instruments
703 Rodi Road, suite 100, 15235, Pittsburgh, PA, United States
Visit website- Scope lines
- 29
- A2LA certificate
- 1
- Site in our directory
- 1
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 29 scope lines.
A2LA certificate 1894.01 · 29 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994 | - | See scope document |
| 2 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 2 of 28
Parameter/Equipment
Range CMC2, 5 (±) Comments
Calipers3
(OD, ID, Depth, Step)
Up to 4 in
(4 to 24) in
(24 to 60) in
(290 + 0.1L) µin
(290 + 0.9L) µin
(310 + 2.5L) µin
Gage blocks
Dial Bore Gages3
(Bore Gages
w/indicators)
Up to 4 in
(4 to 8) in
(66 + 2.0L) µin
(70 + 8.0L) µin
Gage Blocks, rings,
indicator calibrator
Electronic Levels
Angular +/- 990 arcsec
Low Mag: 1000
High Mag: 200
1.2 arcsec
1.1 arcsec
Gage blocks, sine
plate, surface plate
Gage Blocks –
Parallelism (Variation
in Length)
(0.005 to 4) in
(> 4 to 20) in
(0.005 to 4) in
(> 4 to 20) in
(3.1 + 1.1L) µin
(2 + 3.5L) µin
2.3 µin
4.2 µin
Gage block
comparator, ULM
& master gauge
blocks
Height Gages3
Up to 40 in
(190 + 1.8L) µin
Gage blocks
Length Indicators3
(Dial, Digital, Test)
Up to 6 in
(6 + 2.9L) µin
Universal measuring
machine, gage blocks,
indicator calibrator
Thread Plugs –
Pitch Diameter
Major Diameter
Up to 6 in
Up to 6 in
80 µin
(16 + 1.8D) µin
Universal measuring
machine with thread
wires & plain plug
gages | to 40 in |
| 3 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 3 of 28
Parameter/Equipment
Range CMC2, 5 (±) Comments
Tapered Thread Plugs –
Pitch Diameter
Major Diameter
Notch Height
Up to 3 in
Up to 3 in
Up to 1 in
100 µin
(75 + 0.7D) µin
130 µin
Universal measuring
machine with thread
wires, tapered sine
block, gage block,
surface plate & dial
indicator
Plain Plugs, Disks
Up to 8 in
(16 + 1.8D) µin
Universal measuring
machine
Thread Measuring Wires
(2 to 120) TPI
(9 + 2.6L) µin
Universal measuring
machine, gage blocks,
master wires
Length Standards
Up to 8 in
(8 to 24) in
(22 to 40) in
(4 + 3.2L) µin
(2 + 3.5L) µin
(7 + 3.7L) µin
Universal measuring
machine, gage blocks,
indicator & surface plate
Steel Rules3
Up to 40 in
(40 to 72) in
0.0024 in
0.0033 in
Glass scale, master rule
Surface Plates3 –
Overall Flatness
Local Variation in
Flatness
Up to 225 in DL
Up to 225 in DL
12√DL µin
31 µin
Precision level system
Repeat-o-meter
II. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC
2, 9 (±) Comments
DC Voltage – Generate3
(0 to 220) mV
220 mV to 2.2 V
(2.2 to 11) V
(11 to 22) V
(22 to 220) V
(220 to 1100) V
6 µV/V + 0.4 µV
3.5 µV/V + 0.7 µV
2.5 µV/V + 2.5 µV
2.5 µV/V + 4 µV
3.5 µV/V + 40 µV
4.5 µV/V + 0.4 mV
Fluke 5700A/EP | to 3 in |
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 4 of 28
Parameter/Equipment Range CMC2, 4, 9 (±) Comments
DC Voltage – Generate
Fixed Points
10 V
±100 mV
±1 V
±10 V
±100 V
±1000 V
0.5 µV/V
1.5 µV/V
0.5 µV/V
0.5 µV/V
0.6 µV/V
0.7 µV/V
Fluke 732A
Fluke 732A/752A
DC Voltage – Measure3
(0 to 200) mV
(0 to 2) V
(0 to 20) V
(0 to 200) V
(0 to 1050) V
(1 to 10) kV
(10 to 35) kV
5.2 µV/V + 0.1 µV
3.5 µV/V + 0.4 µV
3.5 µV/V + 4 µV
5.5 µV/V + 40 µV
5.5 µV/V + 0.5 mV
0.034 % + 0.034 V
0.035 % + 0.07 V
Fluke 8508A
Vitrek 4700/HVP-35
DC Voltage – Measure
Fixed Points
10 V
±100 mV
±1 V
±10 V
±100 V
±1000 V
0.3 µV/V
1.5 µV/V
0.5 µV/V
0.5 µV/V
0.6 µV/V
0.7 µV/V
Fluke 732A
Fluke 732A/752A/
5720A
DC Current – Generate3
(0 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
(0 to 220) µA
(0.22 to 2.2) mA
(2.2 to 22) mA
(22 to 220) mA
(0.22 to 2.2) A
(2.2 to 11) A
(11 to 20.5) A
(20 to 1000) A
5.9 µA/A + 1 nA
6.1 µA/A + 5.3 nA
6.1 µA/A + 50 nA
8.8 µA/A + 0.3 µA
14 µA/A + 7 µA
35 µA/A + 6 nA
30 µA/A + 7 nA
30 µA/A + 40 nA
40 µA/A + 0.7 µA
60 µA/A + 12 µA
0.034 % + 0.48 mA
0.1 % + 750 µA
0.55 %
Fluke 5700A/EP
characterized with
HP 3458A &
standard resistors
Fluke 5700A/EP
Fluke 5520A
Fluke 5520A with
5500/coil | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 5 of 28
Parameter/Equipment Range CMC2, 9 (±) Comments
DC Current – Measure3
(0 to 200) µA
(0 to 2) mA
(0 to 20) mA
(0 to 200) mA
(0 to 2) A
(0 to 20) A
12 µA/A + 0.4 nA
12 µA/A + 4 nA
14 µA/A + 40 nA
48 µA/A + 0.8 µA
0.019 % + 16 µA
0.04 % + 0.4 mA
Fluke 8508A
Resistance – Generate3
Fixed Points
0 Ω
1 Ω
1.9 Ω
10 Ω
19 Ω
100 Ω
190 Ω
1.9 kΩ
19 kΩ
100 kΩ
190 kΩ
1 MΩ
1.9 MΩ
10 MΩ
19 MΩ
100 MΩ
40 µΩ
80 µΩ/Ω
80 µΩ/Ω
21 µΩ/Ω
21 µΩ/Ω
7.5 µΩ/Ω
7.5 µΩ/Ω
7.5 µΩ/Ω
7.5 µΩ/Ω
9 µΩ/Ω
9 µΩ/Ω
15 µΩ/Ω
16 µΩ/Ω
31 µΩ/Ω
39 µΩ/Ω
95 µΩ/Ω
Fluke 5700A/EP
Resistance – Generate3
(0 to 11) Ω
(11 to 33) Ω
(33 to 110) Ω
110 Ω to 1.1 kΩ
(1.1 to 11) kΩ
(11 to 110) kΩ
110 kΩ to 1.1 MΩ
(1.1 to 3.3) MΩ
(3.3 to 11) MΩ
(11 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(330 to 1100) MΩ
40 µΩ/Ω + 1 mΩ
30 µΩ/Ω + 1.5 mΩ
28 µΩ/Ω + 1.4 mΩ
28 µΩ/Ω + 2 mΩ
28 µΩ/Ω + 20 mΩ
28 µΩ/Ω + 200 mΩ
32 µΩ/Ω + 2 Ω
60 µΩ/Ω + 30 Ω
0.013 % + 50 Ω
0.025 % + 2.5 kΩ
0.05 % + 3 kΩ
0.3 % + 100 kΩ
1.5 % + 500 kΩ
Fluke 5520A | ||
| 6 | MIL 6010B/6011A or 6000B | - | See scope document |
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 7 of 28
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Calibration of
RTD Indicators & Indicating
Systems3 – (cont)
Pt 3926, 100 Ω
Pt 3916, 100 Ω
Pt 385, 200 Ω
Pt 385, 500 Ω
Pt 385, 1000 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 630) °C
(-200 to -190) °C
(-190 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
0.05 °C
0.05 °C
0.07 °C
0.09 °C
0.1 °C
0.12 °C
0.25 °C
0.04 °C
0.05 °C
0.06 °C
0.07 °C
0.08 °C
0.09 °C
0.1 °C
0.23 °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
0.12 °C
0.13 °C
0.14 °C
0.16 °C
0.04 °C
0.05 °C
0.05 °C
0.06 °C
0.08 °C
0.08 °C
0.09 °C
0.11 °C
0.03 °C
0.03 °C
0.04 °C
0.05 °C
0.06 °C
0.07 °C
0.07 °C
0.23 °C
Fluke 5520A/8508A | ||
| 8 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 8 of 28
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Calibration of
RTD Indicators & Indicating
Systems
3 – (cont)
PtNi 385, 120 Ω
(Ni 120)
Cu 427, 10 Ω
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(-100 to 260) °C
0.08 °C
0.08 °C
0.14 °C
0.3 °C
Fluke 5520A/8508A
Electrical Calibration of
Thermocouple
Indicators & Indicating
Systems
3 –
Type B
Type C
Type E
Type J
Type K
(600 to 800) °C
(800 to 1000) °C
(1000 to 1550) °C
(1550 to 1820) °C
(0 to 150) °C
(150 to 650) °C
(650 to 1000) °C
(1000 to 1800) °C
(1800 to 2316) °C
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
0.44 °C
0.34 °C
0.3 °C
0.33 °C
0.3 °C
0.26 °C
0.31 °C
0.5 °C
0.84 °C
0.5 °C
0.16 °C
0.14 °C
0.16 °C
0.21 °C
0.27 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
0.33 °C
0.18 °C
0.16 °C
0.26 °C
0.4 °C
Fluke 5520A | ||
| 9 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 9 of 28
Parameter/Equipment Range CMC2, 6 (±) Comments
Electrical Calibration of
Thermocouple
Indicators & Indicating
Systems3 (cont) –
Type L
Type N
Type R
Type S
Type T
Type U
(-200 to -100) °C
(-100 to 800) °C
(800 to 900) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
120 to 410) °C
(410 to 1300) °C
(0 to 250) °C
(250 to 400) °C
(400 to 1000) °C
(1000 to 1767) °C
(0 to 250) °C
(250 to 1000) °C
(1000 to 1400) °C
(1400 to 1767) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 120) °C
(120 to 400) °C
(-200 to 0) °C
(0 to 600) °C
0.37 °C
0.26 °C
0.17 °C
0.4 °C
0.22 °C
0.19 °C
0.18 °C
0.27 °C
0.57 °C
0.35 °C
0.33 °C
0.4 °C
0.47 °C
0.36 °C
0.37 °C
0.46 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
0.56 °C
0.27 °C
Fluke 5520A | ||
| 10 | - | DC PowerShow page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 10 of 28
Voltage Current (A)
Output Power
(W)
(±) CMC2, 6, 7, 8
DC Power
–
Generate
33 mV
0.33 mA 11 µW 0.03 %
3.299 99 mA 109 µW 0.01 %
3.3 mA 109 µW 0.02 %
32.9999 mA 1.1 mW 0.01 %
33 mA 1.1 mW 0.02 %
329.999 mA 11 mW 0.01 %
0.33 A 11 mW 0.03 %
1.099 99 A 36 mW 0.02 %
1.1 A 36 mW 0.04 %
2.999 99 A 99 mW 0.04 %
3 A 99 mW 0.07 %
10.9999 A 363 mW 0.05 %
11 A 363 mW 0.11 %
20.5 A 677 mW 0.10 %
329.9999 mV
0.33 mA 109 µW 0.03 %
3.299 99 mA 1.1 mW 0.01 %
3.3 mA 1.1 mW 0.02 %
32.9999 mA 11 mW 0.01 %
33 mA 11 mW 0.02 %
329.999 mA 109 mW 0.01 %
0.33 A 109 mW 0.03 %
1.099 99 A 363 mW 0.02 %
1.1 A 363 mW 0.04 %
2.999 99 A 1 W 0.04 %
3 A 1 W 0.07 %
10.9999 A 3.6 W 0.05 %
11 A 3.6 W 0.11 %
20.5 A 6.8 W 0.10 %
0.33 V
0.33 mA 109 mW 0.03 %
3.299 99 mA 1.1 mW 0.01 %
3.3 mA 1.1 mW 0.02 %
32.9999 mA 11 mW 0.01 %
33 mA 11 mW 0.02 %
329.999 mA 109 mW 0.01 %
0.33 A 109 mW 0.03 %
1.099 99 A 363 mW 0.02 %
1.1 A 363 mW 0.04 %
2.999 99 A 1 W 0.04 %
3 A 1 W 0.07 %
10.9999 A 3.6 W 0.05 %
11 A 3.6 W 0.11 %
20.5 A 6.8 W 0.10 %
Comments Fluke 5520A | See scope document |
| 11 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 11 of 28
Voltage Current (A) Output Power
(W)
(±) CMC2, 6, 7, 8
DC Power
–
Generate
(cont)
3.299 999 V
0.33 mA 1.1 mW 0.03 %
3.299 99 mA 10.9 mW 0.01 %
3.3 mA 10.9 mW 0.02 %
32.9999 mA 109 mW 0.01 %
33 mA 109 mW 0.02 %
329.999 mA 1.1 W 0.01 %
0.33 A 1.1 W 0.03 %
1.099 99 A 3.6 W 0.02 %
1.1 A 3.6 W 0.04 %
2.999 99 A 9.9 W 0.04 %
3 A 9.9 W 0.07 %
10.9999 A 36 W 0.05 %
11 A 36 W 0.11 %
20.5 A 68 W 0.10 %
3.3 V
0.33 mA 1.1 mW 0.03 %
3.299 99 mA 11 mW 0.01 %
3.3 mA 10.9 mW 0.02 %
32.9999 mA 109 mW 0.01 %
33 mA 109 mW 0.02 %
329.999 mA 1.1 W 0.01 %
0.33 A 1.1 W 0.03 %
1.099 99 A 3.6 W 0.02 %
1.1 A 3.6 W 0.04 %
2.999 99 A 9.9 W 0.04 %
3 A 9.9 W 0.07 %
10.9999 A 36 W 0.05 %
11 A 36 W 0.11 %
20.5 A 68 W 0.10 %
32.999 99 V
0.33 mA 11 mW 0.03 %
3.299 99 mA 0.1 W 0.01 %
3.3 mA 109 mW 0.02 %
32.9999 mA 1.1 W 0.01 %
33 mA 1.1 W 0.02 %
329.999 mA 10.9 W 0.01 %
0.33 A 10.9 W 0.03 %
1.099 99 A 36.3 W 0.02 %
1.1 A 36.3 W 0.04 %
2.999 99 A 99 W 0.04 %
3 A 99 W 0.07 %
10.9999 A 363 W 0.05 %
11 A 363 W 0.11 %
20.5 A 676 W 0.10 %
Comments Fluke 5520A | ||
| 12 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 12 of 28
Voltage Current (A) Output Power
(W)
(±) CMC2, 6, 7, 8
DC Power
–
Generate
(cont)
33 V
0.33 mA 11 mW 0.03 %
3.299 99 mA 0.1 W 0.01 %
3.3 mA 109 mW 0.02 %
32.9999 mA 1.1 W 0.01 %
33 mA 1.1 W 0.02 %
329.999 mA 10.9 W 0.01 %
0.33 A 10.9 W 0.03 %
1.099 99 A 36 W 0.02 %
1.1 A 36 W 0.04 %
2.999 99 A 99 W 0.04 %
3 A 99 W 0.07 %
10.9999 A 363 W 0.05 %
11 A 363 W 0.11 %
20.5 A 677 W 0.10 %
329.9999 V
0.33 mA 0.1 W 0.03 %
3.299 99 mA 1.1 W 0.01 %
3.3 mA 1.1 W 0.02 %
32.9999 mA 10.9 W 0.01 %
33 mA 10.9 W 0.02 %
329.999 mA 109 W 0.01 %
0.33 A 109 W 0.03 %
1.099 99 A 363 W 0.02 %
1.1 A 363 W 0.04 %
2.999 99 A 990 W 0.04 %
3 A 990 W 0.07 %
10.9999 A 3.6 kW 0.05 %
11 A 3.6 kW 0.11 %
20.5 A 6.8 kW 0.10 %
330 V
0.33 mA 0.1 W 0.03 %
3.299 99 mA 1.1 W 0.01 %
3.3 mA 1.1 W 0.02 %
32.9999 mA 10.9 W 0.01 %
33 mA 10.9 W 0.02 %
329.999 mA 109 W 0.01 %
0.33 A 109 W 0.03 %
1.099 99 A 363 W 0.02 %
1.1 A 363 W 0.04 %
2.999 99 A 990 W 0.04 %
3 A 990 W 0.07 %
10.9999 A 3.6 kW 0.05 %
11 A 3.6 kW 0.11 %
20.5 A 6.8 kW 0.10 %
Comments Fluke 5520A | ||
| 13 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 13 of 28
Voltage Current (A) Output Power
(W)
(±) CMC2, 6, 7, 8
DC Power
–
Generate
(cont)
1000 V
0.33 mA 0.3 W 0.03 %
3.299 99 mA 3 W 0.01 %
3.3 mA 3.3 W 0.02 %
32.9999 mA 33 W 0.01 %
33 mA 33 W 0.02 %
329.999 mA 330 W 0.01 %
0.33 A 330 W 0.03 %
1.099 99 A 1.1 kW 0.02 %
1.1 A 1.1 kW 0.04 %
2.999 99 A 3 kW 0.04 %
3 A 3 kW 0.07 %
10.9999 A 11 kW 0.05 %
11 A 11 kW 0.11 %
20.5 A 20.5 kW 0.10 %
Comments Fluke 5520A
AC Power – Generate
Frequency Voltage Current (A) Output Power
(W)
(±) CMC2, 6, 7, 8
(45 to 65)
Hz
PF = 1
33 mV
3.3 mA 109 µW 0.11 %
32.999 mA 1.1 mW 0.06 %
33 mA 1.1 mW 0.11 %
329.99 mA 11 mW 0.06 %
0.33 A 11 mW 0.09 %
1.099 99 A 36 mW 0.07 %
1.1 A 36 mW 0.08 %
2.999 99 A 99 mW 0.07 %
3 A 99 mW 0.13 %
10.9999 A 363 mW 0.09 %
11 A 363 mW 0.17 %
20.5 A 677 mW 0.15 %
329.999 mV
3.3 mA 1.1 mW 0.10 %
32.999 mA 11 mW 0.05 %
33 mA 11 mW 0.10 %
329.99 mA 109 mW 0.05 %
0.33 A 109 mW 0.08 %
1.099 99 A 363 mW 0.06 %
1.1 A 363 mW 0.07 %
2.999 99 A 1 W 0.07 %
3 A 1 W 0.13 %
10.9999 A 3.6 W 0.08 %
11 A 3.6 W 0.17 %
20.5 A 6.8 W 0.15 %
Comments Fluke 5520A | ||
| 14 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 14 of 28
AC Power – Generate
Frequency Voltage Current (A) Output Power
(W)
(±) CMC2, 6, 7, 8
(45 to 65)
Hz
PF = 1
0.33 V
3.3 mA 1.1 mW 0.11 %
32.999 mA 11 mW 0.06 %
33 mA 11 mW 0.11 %
329.99 mA 109 mW 0.06 %
0.33 A 109 mW 0.09 %
1.099 99 A 363 mW 0.07 %
1.1 A 363 mW 0.08 %
2.999 99 A 1 W 0.07 %
3 A 1 W 0.13 %
10.9999 A 3.6 W 0.08 %
11 A 3.6 W 0.17 %
20.5 A 6.8 W 0.15 %
3.2999 V
3.3 mA 10.9 mW 0.10 %
32.999 mA 109 mW 0.05 %
33 mA 109 mW 0.10 %
329.99 mA 1.1 W 0.05 %
0.33 A 1.1 W 0.08 %
1.099 99 A 3.6 W 0.06 %
1.1 A 3.6 W 0.07 %
2.999 99 A 9.9 W 0.07 %
3 A 9.9 W 0.13 %
10.9999 A 36 W 0.08 %
11 A 36 W 0.17 %
20.5 A 68 W 0.15 %
3.3 V
3.3 mA 10.9 mW 0.11 %
32.999 mA 109 mW 0.06 %
33 mA 109 mW 0.11 %
329.99 mA 1.1 W 0.06 %
0.33 A 1.1 W 0.09 %
1.099 99 A 3.6 W 0.07 %
1.1 A 3.6 W 0.08 %
2.999 99 A 9.9 W 0.07 %
3 A 9.9 W 0.13 %
10.9999 A 36 W 0.08 %
11 A 36 W 0.17 %
20.5 A 68 W 0.15 %
32.9999 V
3.3 mA 109 mW 0.10 %
32.999 mA 1.1 W 0.05 %
33 mA 1.1 W 0.10 %
329.99 mA 10.9 W 0.05 %
0.33 A 10.9 W 0.08 %
1.099 99 A 36.3 W 0.06 %
1.1 A 36.3 W 0.07 %
2.999 99 A 99 W 0.07 %
3 A 99 W 0.13 %
10.9999 A 363 W 0.08 %
11 A 363 W 0.17 %
20.5 A 676 W 0.15 %
Comments Fluke 5520A | 45 to 65 Hz |
| 15 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 15 of 28
AC Power – Generate
Frequency Voltage Current (A) Output Power
(W)
(±) CMC2, 6, 7, 8
(45 to 65)
Hz
PF = 1
33 V
3.3 mA 109 mW 0.10 %
32.999 mA 1.1 W 0.05 %
33 mA 1.1 W 0.10 %
329.99 mA 10.9 W 0.05 %
0.33 A 10.9 W 0.08 %
1.099 99 A 36 W 0.06 %
1.1 A 36 W 0.07 %
2.999 99 A 99 W 0.07 %
3 A 99 W 0.13 %
10.9999 A 363 W 0.08 %
11 A 363 W 0.17 %
20.5 A 677 W 0.15 %
329.999 V
3.3 mA 1.1 W 0.10 %
32.999 mA 10.9 W 0.05 %
33 mA 10.9 W 0.10 %
329.99 mA 109 W 0.05 %
0.33 A 109 W 0.08 %
1.099 99 A 363 W 0.06 %
1.1 A 363 W 0.07 %
2.999 99 A 990 W 0.07 %
3 A 990 W 0.13 %
10.9999 A 3.6 kW 0.08 %
11 A 3.6 kW 0.17 %
20.5 A 6.8 kW 0.15 %
330 V
3.3 mA 1.1 W 0.11 %
32.999 mA 10.9 W 0.06 %
33 mA 10.9 W 0.11 %
329.99 mA 109 W 0.06 %
0.33 A 109 W 0.09 %
1.099 99 A 363 W 0.07 %
1.1 A 363 W 0.08 %
2.999 99 A 990 W 0.07 %
3 A 990 W 0.13 %
10.9999 A 3.6 kW 0.08 %
11 A 3.6 kW 0.17 %
20.5 A 6.8 kW 0.15 %
1000 V
3.3 mA 3.3 W 0.11 %
32.999 mA 33 W 0.06 %
33 mA 33 W 0.11 %
329.99 mA 330 W 0.06 %
0.33 A 330 W 0.09 %
1.099 99 A 1.1 kW 0.07 %
1.1 A 1.1 kW 0.08 %
2.999 99 A 3 kW 0.07 %
3 A 3 kW 0.13 %
10.9999 A 11 kW 0.08 %
11 A 11 kW 0.17 %
20.5 A 20.5 kW 0.15 %
Comments Fluke 5520A | 45 to 65 Hz |
| 16 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 16 of 28
Parameter/Range Frequency CMC2 (±) Comments
Capacitance – Generate3
(0.19 to 0.4) nF
(0.4 to 1.1) nF
(1.1 to 3.3) nF
(3.3 to 11) nF
(11 to 33) nF
(33 to 110) nF
(110 to 330) nF
330 nF to 1.1 µF
(1.1 to 3.3) µF
(3.3 to 11) µF
(11 to 33) µF
(33 to 110) µF
(110 to 330) µF
330 µF to 1.1 mF
(1.1 to 3.3) mF
(3.3 to 11) mF
(11 to 33) mF
(33 to 110) mF
10 Hz to 10 kHz
10 Hz to 10 kHz
10 Hz to 3 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
(10 to 600) Hz
(10 to 300) Hz
(10 to 150) Hz
(10 to 120) Hz
(10 to 80) Hz
(0 to 50) Hz
(0 to 20) Hz
(0 to 6) Hz
(0 to 2) Hz
(0 to 0.6) Hz
(0 to 0.2) Hz
0.5 % + 0.01 nF
0.5 % + 0.01 nF
0.5 % + 0.01 nF
0.25 % + 0.01 nF
0.25 % + 0.1 nF
0.25 % + 0.1 nF
0.25 % + 0.3 nF
0.25 % + 1 nF
0.25 % + 3 nF
0.25 % + 10 nF
0.4 % + 30 nF
0.45 % + 100 nF
0.45 % + 300 nF
0.45 % + 1 µF
0.45 % + 3 µF
0.45 % + 10 µF
0.75 % + 30 µF
1.1 % + 100 µF
Fluke 5520A
AC Voltage – Generate3
Up to 2.2 mV
(2.2 to 22) mV
(22 to 220) mV
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
0.022 % + 4 µV
85 µV/V + 4 µV
75 µV/V + 4 µV
0.018 % + 4 µV
0.046 % + 5 µV
0.09 % + 10 µV
0.12 % + 20 µV
0.25 % + 20 µV
0.022 % + 4 µV
85 µV/V + 4 µV
75 µV/V + 4 µV
0.018 % + 4 µV
0.046 % + 5 µV
0.09 % + 10 µV
0.12 % + 20 µV
0.25 % + 20 µV
0.022 % + 12 µV
85 µV/V + 7 µV
75 µV/V + 7 µV
0.018 % + 7 µV
0.042 % + 17 µV
0.075 % + 20 µV
Fluke 5700A/EP | ||
| 17 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 17 of 28
Parameter/Range
Frequency CMC2, 9 (±) Comments
AC Voltage – Generate3 (cont)
(22 to 220) mV
(0.22 to 2.2) V
(2.2 to 22) V
(22 to 220) V
(220 to 250) V
(220 to 1100) V
(220 to 750) V
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(15 to 40) Hz
(40 to 50) Hz
(0.05 to 1) kHz
(1 to 20) kHz
(20 to 30) kHz
(30 to 50) kHz
(50 to 100) kHz
0.12 % + 25 µV
0.25 % + 45 µV
0.022 % + 40 µV
80 µV/V + 15 µV
40 µV/V + 8 µV
70 µV/V + 10 µV
0.011 % + 30 µV
0.034 % + 80 µV
0.09 % + 200 µV
0.15 % + 300 µV
0.022 % + 400 µV
80 µV/V + 150 µV
40 µV/V + 50 µV
70 µV/V + 100 µV
95 µV/V + 200 µV
0.026 % + 600 µV
0.09 % + 2 mV
0.13 % + 3.2 mV
0.022 % + 4 mV
80 µV/V + 1.5 mV
47 µV/V + 0.6 mV
75 µV/V + 1 mV
0.013 % + 2.5 mV
0.08 % + 16 mV
0.42 % + 40 mV
0.7 % + 80 mV
0.026 % + 16 mV
80 µV/V + 4 mV
60 µV/V + 3.5 mV
47 µV/V + 6 mV
0.036 % + 11 mV
0.036 % + 11 mV
0.13 % + 45 mV
Fluke 5700A/EP
Fluke 5700A/EP
with 5725A
Fluke 5700A/EP
Fluke 5700A/EP
with 725A
AC Voltage - Generate3
Wideband Absolute
(0.3 to 1.1) mV
(10 to 30) Hz
(0.03 to 120) kHz
(0.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
0.80 % + 0.4 µV
0.60 % + 0.4 µV
0.60 % + 3.4 µV
0.60 % + 3.4 µV
0.70 % + 3.4 µV
0.90 % + 3.4 µV
1.50 % + 3.4 µV
Fluke 5700A/EP/03 | ||
| 18 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 18 of 28
Parameter/Range
Frequency CMC2, 9 (±) Comments
AC Voltage - Generate3
Wideband Absolute (cont)
(1.1 to 3.3) mV
(3.3 to 11) mV
(11 to 33) mV
(33 to 110) mV
(110 to 330) mV
(0.33 to 1.1) V
(1.1 to 3.5) V
(10 to 30) Hz
(0.03 to 120) kHz
(.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
(10 to 30) Hz
(0.03 to 120) kHz
(0.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
(10 to 30) Hz
(0.03 to 120) kHz
(0.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
(10 to 30) Hz
(0.03 to 120) kHz
(0.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
(10 to 30) Hz
(0.03 to 120) kHz
(0.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
(10 to 30) Hz
(0.03 to 120) kHz
(0.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
(10 to 30) Hz
(.03 to 120) kHz
(.12 to 1.2) MHz
(1.2 to 2) MHz
(2 to 10) MHz
(10 to 20) MHz
(20 to 30) MHz
0.75 % + 1.0 µV
0.55 % + 1.0 µV
0.55 % + 4.0 µV
0.55 % + 4.0 µV
0.65 % + 4.0 µV
0.85 % + 4.0 µV
1.45 % + 4.0 µV
0.65 % + 0.4 µV
0.45 % + 0.4 µV
0.45 % + 7.0 µV
0.45 % + 7.0 µV
0.55 % + 7.0 µV
0.75 % + 7.0 µV
1.35 % + 7.0 µV
0.60 % + 10 µV
0.40 % + 10 µV
0.40 % + 13 µV
0.40 % + 13 µV
0.50 % + 13 µV
0.70 % + 13 µV
1.30 % + 13 µV
0.60 % + 60 µV
0.40 % + 40 µV
0.40 % + 40 µV
0.40 % + 43 µV
0.50 % + 43 µV
0.70 % + 43 µV
1.30 % + 43 µV
0.55 % + 100 µV
0.35 % + 100 µV
0.35 % + 103 µV
0.35 % + 103 µV
0.45 % + 103 µV
0.65 % + 103 µV
1.25 % + 103 µV
0.55 % + 400 µV
0.35 % + 400 µV
0.35 % + 403 µV
0.35 % + 403 µV
0.45 % + 403 µV
0.65 % + 403 µV
1.25 % + 403 µV
0.50 % + 500 µV
0.30 % + 500 µV
0.30 % + 503 µV
0.30 % + 503 µV
0.40 % + 503 µV
0.60 % + 503 µV
1.20 % + 503 µV
Fluke 5700A/EP/03 | ||
| 19 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 19 of 28
Parameter/Range
Frequency CMC2, 9 (±) Comments
AC Voltage – Measure3
(0.6 to 2.2) mV
(0.6 to 2.2) mV
(2.2 to 7) mV
(7 to 22) mV
(22 to 70) mV
(70 to 220) mV
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
0.17 % + 1.3 µV
0.074 % + 1.3 µV
0.042 % + 1.3 µV
0.081 % + 2 µV
0.12 % + 2.5 µV
0.23 % + 4 µV
0.24 % + 8 µV
0.35 % + 8 µV
0.085 % + 1.3 µV
0.037 % + 1.3 µV
0.021 % + 1.3 µV
0.04 % + 2 µV
0.06 % + 2.5 µV
0.12 % + 4 µV
0.13 % + 8 µV
0.23 % + 8 µV
0.029 % + 1.3 µV
0.019 % + 1.3 µV
0.011 % + 1.3 µV
0.021 % + 2 µV
0.031 % + 2.5 µV
0.081 % + 4 µV
0.089 % + 8 µV
0.17 % + 8 µV
0.024 % + 1.5 µV
0.012 % + 1.5 µV
65 µV/V + 1.5 µV
0.013 % + 2 µV
0.026 % + 2.5 µV
0.051 % + 4 µV
0.067 % + 8 µV
0.11 % + 8 µV
0.021 % + 1.5 µV
85 µV/V + 1.5 µV
38 µV/V + 1.5 µV
69 µV/V + 2 µV
0.016 % + 2.5 µV
0.025 % + 4 µV
0.038 % + 8 µV
0.1 % + 8 µV
Fluke 5790A | ||
| 20 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 20 of 28
Parameter/Range
Frequency CMC2, 7, 9 (±) Comments
AC Voltage – Measure3
(cont)
(220 to 700) mV
(0.7 to 2.2) V
(2.2 to 7) V
(7 to 22) V
(22 to 70) V
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
0.021 % + 1.5 µV
76 µV/V + 1.5 µV
33 µV/V + 1.5 µV
51 µV/V + 2 µV
79 µV/V + 2.5 µV
0.018 % + 4 µV
0.03 % + 8 µV
0.096 % + 8 µV
0.02 %
66 µV/V
24 µV/V
46 µV/V
71 µV/V
0.016 %
0.026 %
0.09 %
0.02 %
67 µV/V
24 µV/V
48 µV/V
81 µV/V
0.019 %
0.04 %
0.12 %
0.02 %
67 µV/V
27 µV/V
48 µV/V
81 µV/V
0.019 %
0.04 %
0.12 %
0.02 %
68 µV/V
32 µV/V
57 µV/V
94 µV/V
0.02 %
0.041 %
0.12 %
Fluke 5790A | ||
| 21 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 21 of 28
Parameter/Range
Frequency CMC2, 7, 9 (±) Comments
AC Voltage – Measure3
(cont)
(70 to 220) V
(220 to 700) V
(700 to 1000) V
(1 to 10) kV
(10 to 30) kV
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(10 to 20) Hz
(20 to 40) Hz
(0.04 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(30 to 200) Hz
(30 to 200) Hz
0.02 %
68 µV/V
31 µV/V
69 µV/V
98 µV/V
0.021 %
0.05 %
0.02 %
99 µV/V
41 µV/V
0.013 %
0.05 %
0.02 %
99 µV/V
38 µV/V
0.013 %
0.05 %
0.12 % + 0.1 V
0.1 % + 0.2 V
Fluke 5790A
Vitrek 4700/HVP-
35 | ||
| 22 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 22 of 28
Parameter/Range
Frequency CMC2, 9 (±) Comments
AC Current – Generate3
(9 to 220) µA
(0.22 to 2.2) mA
(2.2 to 22) mA
(22 to 220) mA
(0.22 to 2.2) A
(2.2 to 11) A
(11 to 20.5) A
(20 to 1000) A
Clamp Meters
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
20 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(45 to 100) Hz
100 Hz to 1 kHz
(1 to 5) kHz
(45 to 65) Hz
(65 to 440) Hz
0.023 % + 16 nA
0.014 % + 10 nA
0.011 % + 8 nA
0.025 % + 0.12 nA
0.09 % + 65 nA
0.023 % + 40 nA
0.014 % + 35 nA
0.011 % + 35 nA
0.018 % + 110 nA
0.09 % + 650 nA
0.023 % + 400 nA
0.014 % + 350 nA
0.011 % + 350 nA
0.018 % + 550 nA
0.09 % + 5 µA
0.023 % + 4 µA
0.014 % + 3.5 µA
0.011 % + 2.5 µA
0.018 % + 3.5 µA
0.09 % + 10 µA
0.024 % + 35 µA
0.039 % + 80 µA
0.6 % + 160 µA
0.04 % + 170 µA
0.085 % + 380 µA
0.33 % + 750 µA
0.12 % + 5 mA
0.15 % + 5 mA
3 % + 5 mA
0.3 %
0.81 %
Fluke 5700A/EP
Fluke 5700A/EP
with 5725A
Fluke 5520A
Fluke 5520A with
5500A coil
AC Current – Measure3
(2 to 200) µA
200 µA to 2 mA
(2 to 20) mA
(20 to 200) mA
200 mA to 2 A
(2 to 20) A
10 Hz to 10 kHz
10 Hz to 10 kHz
10 Hz to 10 kHz
10 Hz to 10 kHz
10 Hz to 2 kHz
(2 to 10) kHz
10 Hz to 2 kHz
(2 to 10) kHz
0.05 % + 0.02 µA
0.03 % + 0.2 µA
0.03 % + 2 µA
0.05 % + 20 µA
0.062 % + 200 µA
0.073 % + 200 µA
0.082 % + 2 mA
0.25 % + 2 mA
Fluke 8508A | ||
| 23 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 23 of 28
Parameter/Range Frequency CMC2, 6, 7, 9 (±) Comments
Oscilloscope – Leveled
Sine Wave3
Absolute
5 mV to 5.5 V
Flatness
(Relative to 50 kHz)
5 mV to 5.5 V
Amplitude
DC Signal
50 Ω load
1 MΩ load
Time Marker, 50 Ω
Load
Edge – Rise Time
50 Ω load
≤ 2 MHz
> 2 MHz
50 kHz reference
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
(600 to 1100) MHz
(-6.6 to 6.6) V
(-130 to 130) V
5 s to 50 ms
20 ms to 1 ns
≤ 300 ps
≤ 300 ps
> 350 ps
2 % + 300 µV
3.5 % + 300 µV
4 % + 300 µV
6 % + 300 µV
7 % + 300 µV
1.5 % + 100 µV
2 % + 100 µV
4 % + 100 µV
5 % + 100 µV
0.25 % + 40 µV
0.05 % + 40 µV
(25 + 1000t) μs/s
2.5 μs/s
+0 ps/-100 ps
+0 ps/-100 ps
+0 ps/-100 ps
Fluke 5520A with
SC1100 option.
CMC’s do not
include loading
effect of UUT
Fluke 5520A
w/SC1100 option.
CMC’s do not
include loading
effect of UUT
t is the time in
seconds
Markers in a 5-2-1
sequence
III. Fluid Quantities
Parameter/Equipment Range CMC2, 7, 10 (±) Comments
Gas Flow3 – Measuring
Equipment
Up to 100 slm
0.2 %
DHI
molbox1/molblocs | ||
| 24 | NIST Class F | Oil | See scope document |
| 25 | ASTM Class 2; ASTM Class 3; ASTM Class F | Standards/Artifacts; Scales / Balances; Force Gages; Tachometer | See scope document |
| 26 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 26 of 28
V. Thermodynamics
Parameter/Equipment Range CMC2, 7, 10 (±) Comments
Infrared Pyrometry3 –
Infrared Thermometers
& Calibrators
(-30 to 140) °C
(35 to 100) °C
(100 to 200) °C
(200 to 350) °C
(350 to 500) °C
(150 to 1200) °C
(600 to 2300) °C
(-15 to 1200) °C
(600 to 1500) °C
(1500 to 2300) °C
0.2 % + 1 °C
0.55 °C
0.79 °C
1.4 °C
1.8 °C
0.2 % + 2 °C
0.25 % + 2 °C
0.16 % + 0.3 °C
0.16 % + 0.3 °C
0.25 %
Black body calibrator
Pyrometer standard
Optical Pyrometry
(800 to 1100) °C
(1100 to 1500) °C
(1500 to 1900) °C
(1900 to 2300) °C
1.8 °C
1.9 °C
2 °C
3 °C
Tungsten ribbon filament
lamp, current shunt, DMM
Humidity³ –
Measuring Equipment
Measure3
(10 to 95) % RH
(10 to 80) % RH
0.5 % RH
1.1 % RH
Thunder 1200
Vaisala HMI41/HMP46
Temperature –
Measuring Equipment &
Measure
3
(-30 to 140) °C
(140 to 420) °C
(420 to 660) °C
(660 to 1200) °C
(-80 to 0) °C
(0 to 300) °C
(300 to 500) °C
(0 to 1250) °C
0.14 °C
0.4 °C
0.6 °C
1.6 °C
0.009 °C
0.006 °C
0.011 °C
0.8 °C
Dry block calibrator
SPRT, Precision bath,
Current comparator
resistance bridge.
Comparison to noble metal
thermocouple | ||
| 27 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 27 of 28
Parameter/Equipment Range CMC2, 7, 10 (±) Comments
Temperature – Measuring
Equipment & Measure3 (cont)
Fixed Points
156 °C
231 °C
420 °C
660 °C
-196 °C
-80 °C
-38 °C
0.01 °C
0.006 °C
0.006 °C
0.011 °C
0.012 °C
0.004 °C
0.009 °C
0.009 °C
0.006 °C
SPRT, Precision bath,
Current comparator
resistance bridge.
SPRT, Aluminum cell,
Current comparator
resistance bridge
LN2, SPRT, Current
comparator resistance
bridge
SPRT, Current comparator
resistance bridge
SPRT, Triple point of
water cell, Current
comparator resistance
bridge
VI. Time & Frequency
Parameter/Equipment Range CMC2, 10 (±) Comments
Frequency – Counters,
Generators
Fixed Points
Measuring Equipment
Measure
Timers
Measure
3
Measuring Equipment
3
Time Intervals, Timers
3
(5 & 10) MHz
0.01 Hz to 20 MHz
0.01 Hz to 225 MHz
6 ns to 1 x 10
6 s
(0.01 to 225) MHz
50 kHz to 1100 MHz
6 ns to 1 x 10
6 s
2 parts in 1012 Hz
3 parts in 1011 Hz
3 parts in 1011 Hz
5.8 parts in 10
7 s
1 part in 10
7 Hz
2.5 parts in 106 Hz
6 parts in 107 s
Fluke 910
HP 3325B locked w
Fluke 910
Fluke PM6680
locked w Fluke 910
Fluke PM6680B
locked w/ Fluke 910
Fluke PM6680B
Fluke
5520A/SC1100
_____________________________ | ||
| 28 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1894.01) Revised 07/07/2025 Page 28 of 28
1 This laboratory offers commercial calibration service.
2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3 Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4 Based on using the standard at the temperature the HP 3458A was calibrated (tcal ± 5 oC) and an auto-
calibration (ACAL) was performed within the previous 24 hours (± 1 oC of ambient temperature). CMC
is based upon 1-year specifications and is read as parts in 10 6 or percent output plus floor specification
or as parts in 106 or percent of reading plus parts in 106 or percent of range.
5 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches.
In the statement of CMC, D is the numerical value of the nominal diameter of the device measured in
inches. In the statement of CMC, DL is the diagonal length of the surface plate measured in inches.
6 Based on using the standard at the temperature the Fluke 5520A was calibrated (tcal ± 5 oC) and assuming
the instrument is zeroed at least every seven days or when the ambient temperature changes more than 5
oC. For resistance, a zero calibration is performed at least every 12 hours within ± 1 oC of use. For AC
Current, CMCs are determined with LCOMP Off. CMC is based upon 1- year specifications and is read
as ppm or percent output plus floor specification. When “ppm” is used on the scope, it refers to parts in
106.
7 In the statement of CMC, the value is defined as the percentage of reading.
8 The CMC values shown are for the voltage and current values applied at that wattage level. They are not
to be considered fixed points and may vary. The actual measurement uncertainty will be calculated using
the voltage and current level applied with consideration of the device under test.
9 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.
10 The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
11 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 29 | ISO/IEC 17025:20 17; ANSI/NCSL Z540-1-1994; R205 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Labs with overlapping accredited scope in United States
- BHD Test & MeasurementThornton, CO · ~2,117 kmA2LAShares 2 accredited methods and materialsANSI/NCSL Z540-1-1994Oil
- Fluke Calibration, Phoenix - Primary Pressure and FlowPhoenix, AZ · ~2,936 kmA2LAShares 2 accredited methods and materialsANSI/NCSL Z540-1-1994Oil
- SIMCO Electronics - Los AngelesPanorama City, CA · ~3,440 kmA2LAShares 2 accredited methods and materialsANSI/NCSL Z540-1-1994Oil
- Ohm-LabsPittsburgh, PA · under 1 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Instrument Calibration SolutionsAliquippa, PA · ~29 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.