JM Test Systems, LLC.
738 S. Main Street, 77531, Clute, TX, United States
Visit website- Scope lines
- 19
- A2LA certificate
- 1
- Sites in our directory
- 5
What this lab is accredited for
We could not match the method references in this lab’s scope to published test standard numbers. The full scope as published is below.
Only this lab. Searches the method, matrix and range of each of the 19 scope lines.
A2LA certificate 1995.03 · 19 lines
| Line | Method | Matrix | Range |
|---|---|---|---|
| 1 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994 | - | to 4 in |
| 2 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 2 of 18
II. Electrical – DC/Low Frequency
Parameter/Equipment
Range
CMC2, 4, 8 (±)
Comments
DC Voltage
3 – Generate
(0 to 220) mV
(0.22 to 2.2) V
(2.2 to 11) V
(11 to 22) V
(22 to 220) V
(220 to 1100) V
(0 to 330) mV
(0.33 to 33) V
(3.3 to 33) V
(33 to 330) V
(330 to 1020) V
8.6 µV/V + 0.6 µV
7.3 µV/V + 1 µV
7.3 µV/V + 3.5 µV
7.2 µV/V + 6.5 µV
8.4 µV/V + 80 µV
10 µV/V + 500 µV
24 µV/V + 1 µV
13 µV/V + 2 µV
15 µV/V + 20 µV
21 µV/V + 0.15 mV
21 µV/V + 1.5 mV
Fluke 5700A
Fluke 5522A
DC Current3 – Generate
(0 to 220) μA
(0.22 to 22) mA
(2.2 to 22) mA
(22 to 220) mA
(0.22 to 2.2) A
Up to 330 µA
(0.33 to 3.3) mA
(3.3 to 33) mA
(33 to 330) mA
(0.33 to 1.1) A
(1.1 to 3) A
(3 to 11) A
(11 to 20.5) A
(10 to 16.5) A
(16.5 to 150) A
(150 to 1025) A
52 µA/A + 8 nA
52 µA/A + 8 nA
52 µA/A + 80 nA
65 µA/A + 0.8 µA
84 µA/A + 25 µA
0.017 % + 0.02 µA
0.012 % + 0.05 µA
0.012 % + 0.25 µA
0.012 % + 2.5 µA
0.023 % + 40 µA
0.044 % + 40 µA
0.059 % + 0.5 mA
0.12 % + 0.75 mA
0.25 % + 0.002 A
0.26 % + 0.015 A
0.28 % + 0.05 A
Fluke 5700A
Fluke 5522A
Fluke 5522A, 50
turn coil | ||
| 3 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 3 of 18
Parameter/Equipment
Range
CMC2, 4, 8 (±)
Comments
Resistance – Generate
Fixed Points
Fixed points for
Insulation testers
3
0 Ω
1 Ω
1.9 Ω
10 Ω
19 Ω
100 Ω
190 Ω
1 kΩ
1.9 kΩ
10 kΩ
19 kΩ
100 kΩ
190 kΩ
1 MΩ
1.9 MΩ
10 MΩ
19 MΩ
100 MΩ
1 MΩ
10 MΩ
100 MΩ
1 GΩ
10 GΩ
100 GΩ
41 µΩ
98 µΩ/Ω
98 µΩ/Ω
30 µΩ/Ω
28 µΩ/Ω
20 µΩ/Ω
18 µΩ/Ω
17 µΩ/Ω
14 µΩ/Ω
16 µΩ/Ω
14 µΩ/Ω
20 µΩ/Ω
16 µΩ/Ω
24 µΩ/Ω
23 µΩ/Ω
46 µΩ/Ω
51 µΩ/Ω
110 µΩ/Ω
0.012 MΩ
0.12 MΩ
1.2 MΩ
0.012 GΩ
0.24 GΩ
5.8 GΩ
Fluke 5700A
Fixed Resistors | ||
| 4 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 4 of 18
Parameter/Equipment
Range
CMC2, 8 (±)
Comments
Fixed Points for Ground
Testers & Other Low
Resistance Ohmmeters
3
Resistance
3 – Generate
0.1 mΩ
0.5 mΩ
1 mΩ
2 mΩ
5 mΩ
10 mΩ
20 mΩ
50 mΩ
0.1 Ω
0.2 Ω
0.5 Ω
1 Ω
Up to 11 Ω
(11 to 33) Ω
(33 to 110) Ω
(110 to 330) Ω
(0.33 to 1.1) kΩ
(1.1 to 3.3) kΩ
(3.3 to 11) kΩ
(11 to 33) kΩ
(33 to 110) kΩ
(110 to 330) kΩ
(0.33 to 1.1) MΩ
(1.1 to 33) MΩ
(3.3 to 11) MΩ
(11 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
(330 to 1100) MΩ
0.39 µΩ
1.6 µΩ
3.2 µΩ
6.3 µ
Ω
15 µΩ
29 µΩ
28 µΩ
66 µΩ
0.12 mΩ
0.25 mΩ
0.60 mΩ
1.2 mΩ
51 µΩ/Ω + 0.001 Ω
35 µΩ/Ω + 0.0015 Ω
33 µΩ/Ω + 0.0014 Ω
33 µΩ/Ω + 0.002 Ω
33 µΩ/Ω + 0.002 Ω
33 µΩ/Ω + 0.02 Ω
33 µΩ/Ω + 0.02 Ω
33 µΩ/Ω + 0.2 Ω
33 µΩ/Ω + 0.2 Ω
37 µΩ/Ω + 2 Ω
38 µΩ/Ω + 2
Ω
70 µΩ/Ω + 30 Ω
0.015 % + 50 Ω
0.029 % + 2.5 kΩ
0.060 % + 3 kΩ
0.35 % + 0.1 MΩ
1.7 % + 0.5 MΩ
Current shunts
Fluke 5522A
DC Power3 – Generate
Up to 109 µW
(0.109 to 1.09) mW
(1.09 to 10.9) mW
(10.9 to 109) mW
(0.109 to 1.09) W
(1.09 to 10.9) W
(10.9 to 109) W
(109 to 337) W
(337 to 990) W
(0.99 to 3.06) kW
(3.06 to 11.2) kW
11.2 to 20.9) kW
5.5 W to 16.8 kW
(16.8 to 153) kW
(153 to 1045) kW
0.023 %
0.014 %
0.013 %
0.013 %
0.013 %
0.013 %
0.013 %
0.013 %
0.027 %
0.045 %
0.064 %
0.12 %
0.19 %
0.20 %
0.21 %
Fluke 5522A
Fluke 5522A + 50
turn coil | ||
| 5 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 5 of 18
Parameter/Range
Frequency
CMC2, 8 (±)
Comments
AC Voltage
3 – Generate
Up to 33 mV
(33 to 330) mV
(0.33 to 3.3) V
(3.3 to 33) V
(33 to 330) V
(330 to 1020) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
45 Hz to 1 kHz
(1 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.095 % + 6 µV
0.019 % + 6 µV
0.024 % + 6 µV
0.12 % + 6 µV
0.4 % + 12 µV
0.93 % + 50 µV
0.035 % + 8 µV
0.017 % + 8 µV
0.019 % + 8 µV
0.041 % + 8 µV
0.093 % + 32 µV
0.23 % + 70 µV
0.035 % + 50 µV
0.017 % + 60 µV
0.022 % + 60 µV
0.035 % + 50 µV
0.081 % + 0.13 mV
0.28 % + 0.6 mV
0.035 % + 0.65 mV
0.017 % + 0.6 mV
0.028 % + 0.6 mV
0.04 % + 0.6 mV
0.1 % + 1.6 mV
0.022 % + 2 mV
0.023 % + 6 mV
0.029 % + 6 mV
0.035 % + 6 mV
0.23 % + 50 mV
0.035 % + 10 mV
0.029 % + 10 mV
0.035 % + 10 mV
Fluke 5522A | ||
| 6 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 6 of 18
Parameter/Range
Frequency
CMC2, 4, 8 (±)
Comments
AC Voltage
3 – Generate
Up to 2.2 mV
(2.2 to 22) mV
(22 to 220) mV
(0.22 to 2.2) V
(2.2 to 22) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
0.2 % + 4.5 µV
0.11 % + 4.5 µV
0.083 % + 4.5 µV
0.14 % + 4.5 µV
0.2 % + 7 µV
0.34 % + 13 µV
0.44 % + 25 µV
0.57 % + 25 µV
0.064 % + 5 µV
0.029 % + 5 µV
0.017 % + 5 µV
0.045 % + 5 µV
0.094 % + 7 µV
0.14 % + 12 µV
0.2 % + 25 µV
0.38 % + 25 µV
0.06 % + 13 µV
0.023 % + 8 µV
0.011 % + 8 µV
0.034 % + 8 µV
0.088 % + 25 µV
0.12 % + 25 µV
0.18 % + 35 µV
0.36 % + 80 µV
0.055 % + 80 µV
0.018 % + 25 µV
80 µV/V + 6 µV
0.013 % + 16 µV
0.027 % + 70 µV
0.048 % + 130 µV
0.11 % + 350 µV
0.24 % + 0.85 mV
0.055 % + 0.8 mV
0.018 % + 250 µV
81 µV/V + 60 µV
0.013 % + 160 µV
0.027 % + 350 µV
0.055 % + 1.5 mV
0.13 % + 4.3 mV
0.3 % + 8.5 mV
Fluke 5700A | ||
| 7 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 7 of 18
Parameter/Range
Frequency
CMC2, 4, 8 (±)
Comments
AC Voltage
3 – Generate
(cont)
(22 to 220) V
(220 to 250) V
(220 to 1100) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
500 kHz to 1 MHz
(15 to 50) Hz
50 Hz to 1 kHz
0.055 % + 8 mV
0.018 % + 2.5 mV
87 µV/V + 0.8 mV
0.023 % + 3.5 mV
0.052 % + 8 mV
0.15 % + 90 mV
0.48 % + 90 mV
1.2 % + 190 mV
0.041 % + 16 mV
91 µV/V + 3.5 mV
Fluke 5700A
Note: Voltage
output above 100
kHz exhibits a
linear drop off of
-2.2 mV per Hz
above 100 kHz.
AC Current
3 – Generate
Up to 330 µA
(0.33 to 3.3) mA
(3.3 to 33) mA
(0.33 to 1.1) A
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
0.23 % + 0.1 µA
0.17 % + 0.1 µA
0.14 % + 0.1 µA
0.35 % + 0.15 µA
0.92 % + 0.2 µA
1.8 % + 0.4 µA
0.23 % + 0.15 µA
0.14 % + 0.15 µA
0.12 % + 0.15 µA
0.23 % + 0.2 µA
0.58 % + 0.3 µA
1.2 % + 0.6 µA
0.21 % + 2 µA
0.1 % + 2 µA
0.046 % + 2 µA
0.092 % + 2 µA
0.23 % + 3 µA
0.46 % + 4 µA
0.21 % + 20 µA
0.1 % + 20 µA
0.046 % + 20 µA
0.12 % + 50 µA
0.23 % + 0.1 mA
0.46 % + 0.2 mA
Fluke 5522A | ||
| 8 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 8 of 18
Parameter/Range
Frequency
CMC2, 4, 8 (±)
Comments
AC Current3 – Generate
(10 to 220) µA
220 µA to 2.2 mA
(2.2 to 22) mA
(22 to 220) mA
220 mA to 2.2 A
(1.1 to 3) A
(3 to 11) A
(11 to 20.5) A
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
20 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 45) Hz
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(45 to 100) Hz
100 Hz to 1 kHz
(1 to 5) kHz
(45 to 100) Hz
100 Hz to 1 kHz
(1 to 5) kHz
0.076 % + 25 nA
0.039 % + 20 nA
0.02 % + 16 nA
0.063 % + 40 nA
0.17 % + 80 nA
0.076 % + 40 nA
0.039 % + 35 nA
0.02 % + 35 nA
0.063 % + 0.4 µA
0.17 % + 0.8 µA
0.073 % + 0.4 µA
0.038 % + 0.35 µA
0.02 % + 0.35 µA
0.063 % + 4 µA
0.16 % + 8 µA
0.073 % + 4 µA
0.038 % + 3.5 µA
0.02 % + 3.5 µA
0.063 % + 40 µA
0.16 % + 80 µA
0.068 % + 35 µA
0.078 % + 80 µA
0.87 % + 160 µA
0.21 % + 0.1 mA
0.069 % + 0.1 mA
0.69 % + 1 mA
2.9 % + 5 mA
0.069 % + 2 mA
0.12 % + 2 mA
3.5 % + 2 mA
0.14 % + 5 mA
0.17 % + 5 mA
3.5 % + 5 mA
Fluke 5700A
Fluke 5522A | 10 to 20.5 A |
| 9 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 9 of 18
Parameter/Range
Frequency
CMC2, 8 (±)
Comments
AC Current
3 – Generate
(cont)
Up to 330 µA
(0.33 to 3.3) mA
(3.3 to 33) mA
(33 to 330) mA
(0.33 to 3) A
(3 to 20.5) A
(10 to 16.5) A
(16.5 to 150) A
(150 to 1025) A
(10 to 100) Hz
100 Hz to 1 kHz
(10 to 100) Hz
100 Hz to 1 kHz
(10 to 100) Hz
100 Hz to 1 kHz
(10 to 100) Hz
100 Hz to 1 kHz
(10 to 100) Hz
(100 to 400) Hz
(10 to 100) Hz
(100 to 400) Hz
(45 to 65) Hz
(65 to 440) Hz
(45 to 65) Hz
(65 to 440) Hz
(45 to 65) Hz
(65 to 440) Hz
0.29 % + 0.2 µA
0.69 % + 0.5 µA
0.29 % + 0.3 µA
0.69 % + 0.8 µA
0.092 % + 4 µA
0.23 % + 10 µA
0.092 % + 40 µA
0.23 % + 0.1 mA
0.14 % + 0.2 µA
0.35 % + 1 mA
0.14 % + 2 mA
1.2 % + 5 mA
0.029 % + 0.003 A
0.083 % + 0.003 A
0.32 % + 0.025 A
0.87 % + 0.027 A
0.32 % + 0.09 A
1.5 % + 0.1 A
Fluke 5522A with
LCOMP On
Fluke 5522A, 50
turn coil
AC Power
3 – Generate
(45 to 65 Hz, PF=1)
Up to 10.9 µW
(10.9 to 109 µW
(0.109 to 1.09) mW
(1.09 to 10.9) mW
(10.9 to 109) mW
(0.109 to 1.09) W
(1.09 to 10.9) W
(10.9 to 37) W
(37 to 337) W
(0.337 to 1.12) kW
(1.12 to 3.06) kW
(3.06 to 11.2) kW
(11.2 to 20.9) kW
0.55 W to 16.8 kW
(16.8 to 1045) kW
0.17 %
0.13 %
0.06 %
0.056 %
0.055 %
0.055 %
0.055 %
0.058 %
0.063 %
0.076 %
0.081 %
0.09 %
0.17 %
0.23 %
0.25 %
Fluke 5522A
Fluke 5522A + 50
turn coil | to 1025 A |
| 10 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 10 of 18
Parameter/Equipment
Range
CMC2, 8 (±)
Comments
Capacitance
3 – Generate
(220 to 400) pF
(0.4 to 1.1) nF
(1.1 to 3.3) nF
(3.3 to 11) nF
(11 to 33) nF
(33 to 110) nF
(110 to 330) nF
(0.33 to 1.1) µF
(1.1 to 3.3) µF
(3.3 to 11) µF
(11 to 33) µF
(33 to 110) µF
(110 to 330) µF
(0.33 to 1.1) mF
(1.1 to 3.3) mF
(3.3 to 11) mF
(11 to 33) mF
(33 to 110) mF
0.58 % + 0.01 nF
0.58 % + 0.01 nF
0.58 % + 0.01 nF
0.29 % + 0.01 nF
0.29 % + 0.1 nF
0.29 % + 0.1 nF
0.29 % + 0.3 nF
0.29 % + 1 nF
0.29 % + 3 nF
0.29 % + 10 nF
0.46 % + 30 nF
0.52 % + 0.1 µF
0.52 % + 0.3 µF
0.52 % + 1 µF
0.52 % + 3 µF
0.52 % + 10 µF
0.87 % + 30 µF
1.2 % + 100 µF
Fluke 5522A
Electrical Calibration of
Thermocouple Indicators3 –
Generate & Measure
Type E
Type J
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1200) °C
0.39 °C
0.13 °C
0.12 °C
0.13 °C
0.17 °C
0.21 °C
0.13 °C
0.12 °C
0.14 °C
0.19 °C
Fluke 5522A | ||
| 11 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 11 of 18
Parameter/Equipment
Range
CMC2, 8 (±)
Comments
Electrical Calibration of
Thermocouple Indicators
3 –
Generate & Measure (cont)
Type K
Type N
Type T
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1000) °C
(1000 to 1372) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 410) °C
(410 to 1300) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 120) °C
(120 to 400) °C
0.26 °C
0.15 °C
0.13 °C
0.20 °C
0.31 °C
0.31 °C
0.18 °C
0.16 °C
0.15 °C
0.21 °C
0.49 °C
0.20 °C
0.13 °C
0.12 °C
Fluke 5522A
Electrical Calibration of
RTD Indicators
3 – Generate
Pt 385, 100 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 630) °C
(630 to 800) °C
0.04 °C
0.04 °C
0.06 °C
0.07 °C
0.08 °C
0.09 °C
0.19 °C
Fluke 5522A
Electrical Calibration of
Phase Indicators3 –
Generate
(10 to 65) Hz
(65 to 500) Hz
500 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(10 to 30) kHz
0.12 °phase
0.3 °phase
0.6 °phase
2.9 °phase
6 °phase
12 °phase
Fluke 5522A | ||
| 12 | - | -Show page textExtracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 12 of 18
Parameter/Equipment
Range
CMC2, 5, 8 (±)
Comments
DC Voltage3 – Measure
(0 to 100) mV
(0.10 to 1.0) V
(1.0 to 10) V
(10 to 100) V
(100 to 1000) V
11 µV/V + 0.3 µV
10 µV/V + 0.3 µV
10 µV/V + 0.5 µV
13 µV/V + 30 µV
14 µV/V + 100 µV
Agilent 3458A
DC High Voltage3 –
Generate & Measure
(1 to 100) kV
1.2 %
Phenix KVM-(series)
Resistance3 – Measure
Up to 10 Ω
(10 to 100) Ω
(0.1 to 1) kΩ
(1 to 10) kΩ
(10 to 100) kΩ
(0.1 to 1) MΩ
(1 to 10) MΩ
(10 to 100) MΩ
10 MΩ to 1 GΩ
(1 to 10) GΩ
(10 to 100) GΩ
21 µΩ/Ω + 50 µΩ
18 µΩ/Ω + 0.5 mΩ
17 µΩ/Ω + 0.5 mΩ
16 µΩ/Ω + 5 mΩ
19 µΩ/Ω + 50 mΩ
24 µΩ/Ω + 2 Ω
70 µΩ/Ω + 100 Ω
0.060 % + 1 kΩ
77 μΩ/Ω
0.54 %
0.70 %
Agilent 3458A
Voltage source & µA
range of 3458A
DC Current3 – Measure
Up to 100 µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
(1 to 10) A
(10 to 50) A
(50 to 200) A
(200 to 1000) A
35 µA/A + 0.8 nA
35 µA/A + 5 nA
40 µA/A + 50 nA
53 µA/A + 0.5 µA
0.018 % + 10 µA
0.12 %
0.14 %
0.32 %
0.39 %
Agilent 3458A
Current shunts,
DMM | to 1000 A |
| 13 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 13 of 18
Parameter/Range
Frequency
CMC2, 5, 8 (±)
Comments
AC Voltage3 – Measure
(1 to 10) mV
(10 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 700) V
(10 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(10 to 40) Hz
40 Hz to 1 kHz
(1to20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(10 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(10 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(10 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(10 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
0.051 % + 3 µV
0.056 % + 1.1 µV
0.083 % + 1.1 µV
0.46 % + 1.1 µV
0.78 % + 1.1 µV
4.7 % + 2 µV
0.013 % + 4 µV
0.016 % + 2 µV
0.027 % + 2 µV
0.12 % + 2 µV
0.23 % + 2 µV
0.41 % + 10 µV
1.2 % + 10 µV
0.011 % + 40 µV
0.016 % + 20 µV
0.024 % + 20 µV
0.041 % + 20 µV
0.11 % + 20 µV
0.36 % + 100 µV
1.2 % + 100 µV
0.029 % + 0.4 mV
0.012 % + 0.2 mV
0.020 % + 0.2 mV
0.037 % + 0.2 mV
0.095 % + 0.2 mV
0.36 % + 1 mV
1.2 % + 1 mV
0.025 % + 4 mV
0.025 % + 2 mV
0.026 % + 2 mV
0.048 % + 2 mV
0.15 % + 2 mV
0.47 % + 10 mV
1.7 % + 10 mV
0.048 % + 40 mV
0.048 % + 20 mV
0.070 % + 20 mV
0.14 % + 20 mV
0.35 % + 20 mV
Agilent 3458A | ||
| 14 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 14 of 18
Parameter/Range
Frequency
CMC2, 5, 8 (±)
Comments
AC High Voltage –
Generate & Measure
3
(0.7 to 6) kV
(6 to 80 kV)
(DC to 500) Hz
60 Hz
1.5 %
1.6 %
Fluke 80K-6, DMM
Phenix KVM-(series)
AC Current3 – Measure
Up to 100 μA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
(1 to 10) A
(10 to 50) A
(50 to 200) A
(200 to 1000) A
(10 to 20) Hz
(20 to 40) Hz
(40 to 100) Hz
(0.1 to 5) kHz
(10 to 20) Hz
(20 to 40) Hz
(40 to 100) Hz
(0.1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
(40 to 100) Hz
(0.1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
(40 to 100) Hz
(0.1 to 5) kHz
(5 to 10) kHz
(10 to 20) Hz
(20 to 40) Hz
(40 to 100) Hz
(0.1 to 5) kHz
(5 to 10) kHz
(45 to 65) Hz
(45 to 65) Hz
(45 to 65) Hz
(45 to 65) Hz
0.46 % + 0.03 µA
0.18 % + 0.03 µA
0.074 % + 0.03 µA
0.074 % + 0.03 µA
0.46 % + 0.2 µA
0.18 % + 0.2 µA
0.076 % + 0.2 µA
0.047 % + 0.2 µA
0.074 % + 0.2 µA
0.46 % + 2 µA
0.18 % + 2 µA
0.077 % + 2 µA
0.048 % + 2 µA
0.074 % + 2 µA
0.46 % + 20 µA
0.18 % + 20 µA
0.077 % + 20 µA
0.049 % + 20 µA
0.075 % + 20 µA
0.47 % + 200 µA
0.2 % + 200 µA
0.12 % + 200 µA
0.14 % + 200 µA
0.35 % + 200 µA
0.12 %
0.14 %
0.32 %
0.39 %
Agilent 3458A
Current shunts,
DMM | ||
| 15 | Direct cross-float method- Ruska 2400 | - | See scope document |
| 16 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 16 of 18
Parameter/Equipment
Range
CMC2, 6, 9 (±)
Comments
Pneumatic Pressure – Measure
&/or Measuring Equipment
Effective Area Determination
of High Accuracy Piston-
Cylinder Unit (PCU) or Ball-
Nozzle Units (BNU)
Calibration of Transfer
Standards & Pressure
Devices
(0.14 to 30) psi
(0.14 to 300) psi
0.0085 %
0.0081 %
Direct cross-float
method– PK DWT,
Std. Mass
RK-300/1100WC
Torque Devices (Wrenches,
Drivers, Screwdrivers)
3
5 lbfꞏin to 2000 lbfꞏft
1.2 %
Torque Calibrator
IV. Thermodynamics
Parameter/Equipment
Range
CMC
2, 9 (±)
Comments
Temperature – Measure3
(-38 to 0) °C
(0 to 156) °C
(156 to 420) °C
(420 to 660) °C
0.019 °C
0.023 °C
0.052 °C
0.099 °C
PRT, readout
Temperature – Measuring
Equipment3
(-25 to 0) °C
(0 to 156) °C
(156 to 250) °C
0.013 °C
0.019 °C
0.051 °C
PRT, readout, bath | ||
| 17 | NIST SP 960-12 | - | See scope document |
| 18 | Extracted from the scope document; tables may not line up. View scope for the original. (A2LA Cert. No. 1995.03) 12/09/2025 Page 18 of 18
5 Agilent 3458A CMCs are based upon the temper ature the standard was calibrated (tcal ± 5 oC) and an
auto calibration (ACAL) was performed within the previous 24 hours (± 1 oC of ambient temperature.)
CMCs are also based upon 1-year floor specifications . CMCs are expressed as either a specific value
that covers the full range or as a combination of a percent or the fraction of the reading/output plus a
range specification.
6 In the statement of CMC, percentages are percentages of reading, unless otherwise indicated.
7 Applicable when the internal time base (oscillator) of the device under test is/can be measured directly
by the frequency counter.
8 The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification.
9 The type of instrument or material being calibrate d is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
10 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches.
11 This scope meets A2LA’s P112 Flexible Scope Policy. | ||
| 19 | ISO/IEC 17025:2017; ANSI/NCSL Z540-1-1994 | - | - |
Ranges are reproduced from the accredited scope document as published, and only where we found that range printed in the document for the line. Otherwise the cell reads “See scope document”. An absent range means the register did not state one. Line numbers are ours, for cross-reference; they follow the order of the scope document.
Location
Other Locations
4 other sites in this organization
Labs with overlapping accredited scope in United States
- Endress+HauserPearland, TX · ~61 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- G.T. MICHELLI CO., LLC - HOUSTONHouston, TX · ~82 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Technology & Calibration, Inc. dba Tech CalHouston, TX · ~82 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Troxler Electronic Laboratories, Inc.Houston, TX · ~82 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
- Mensor, LPSan Marcos, TX · ~264 kmA2LAShares 1 accredited test methodANSI/NCSL Z540-1-1994
Accreditation data refreshed from the A2LA register on .
This page mirrors a public register. For a purchasing, audit or regulatory decision, confirm against the A2LA entry linked above.